K113265 · Genii, Inc. · GEI · Mar 22, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113265
Device Name
GENII GI 4000 ELECTROSUGICAL GENERATOR
Applicant
Genii, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 22, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The gi 4000 Electrosurgical Generator is intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas in flexible endoscopic applications. The integrated lavage pump is intended to wash tissue or mucosa within the gut when used in conjunction with washing catheters or bipolar endoscopic probes, endoscope water jet channels, or endoscope working channels. The Genii ArConnect™ Conducting Adapter is intended for use with the gi 4000 during argon assisted coagulation.
Device Story
The gi 4000 is an electrosurgical generator for flexible endoscopy; provides monopolar, bipolar, and argon-assisted coagulation outputs; includes an integrated peristaltic lavage pump for gut irrigation. Operated by physician via touch screen and dual-pedal footswitch. Inputs include user-selected power settings and argon gas from a disposable canister; outputs include RF energy and lavage fluid delivered through compatible endoscopic accessories. Argon-assisted coagulation uses 99.999% pure argon gas to create ionized RF arcs. The ArConnect adapter facilitates delivery of RF energy and argon gas to single-use probes. System requires patient return electrode for monopolar/argon modes; bipolar mode is return-electrode-free. Output power is independently controlled via LCD panel. Device assists in tissue management and visualization during endoscopic procedures.
Clinical Evidence
No clinical data. Bench testing only. In vitro tissue study compared lesions created by the gi 4000 against the Erbe ICC 200E with APC 300 system, demonstrating substantial similarity in performance. Electrical safety and EMC testing confirmed compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. Biocompatibility testing (cytotoxicity, irritation, sensitization) confirmed the ArConnect cable is non-toxic.
Technological Characteristics
Electrosurgical generator with 7 monopolar, 1 monopolar argon, and 1 bipolar output. Peristaltic lavage pump. Frequency 315-550 KHz; max power 60-200W. ArConnect adapter materials: PVC, brass, ABS, HDPE, PE, Buna-U. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, CISPR 11, IEC 62304, ISO 14971. Connectivity: touch screen interface. Software: V&V performed.
Indications for Use
Indicated for patients undergoing flexible endoscopic procedures requiring electrosurgical cutting, coagulation, or argon-assisted coagulation, and gut tissue/mucosa irrigation. Contraindications not specified.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Olympus AFU-100 Peristaltic Pump Unit (PPU) (K073207)
Erbe APC 300 Argon Plasma Coagulator and Accessories (K963189)
Erbe ICC 200 Electrosurgical Workstation for Flexible Endoscopy (K933157)
{0}------------------------------------------------
### Section 005 - 510(k) Summary
### Submission Correspondent:
| Name: | Emergo Group, Inc.<br>www.emergogroup.com |
|----------|----------------------------------------------------------|
| Address: | 611 West Fifth Street<br>Third Floor<br>Austin, TX 78701 |
| Phone: | (512) 327-9997 |
| Fax: | (512) 327-9998 |
Contact: Julie Powell Email: jpowell@emergogroup.com
### Submission Sponsor:
Genii, Inc. 2145 Woodlane Drive Suite 101-W St. Paul, MN 55125 Tel. 651-501-4810 Fax. 651-501-4819
| Date Prepared: | 2 November 2011; Revised 23 February 2012; Revised 14 March 2012 |
|-----------------------|------------------------------------------------------------------|
| Trade Name: | Genii gi 4000 Electrosurgical Generator |
| Regulation Number: | 878.4400 |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories. |
| Product Code: | GEI |
| Classification Panel: | General and Plastic Surgery |
| Regulatory Class: | Class II |
#### Device Description:
The gi 4000 has seven monopolar outputs, a monopolar argon assisted coagulation output, and a bipolar output. Users make selections from a touch screen that displays stored default power setting start points, convenience and safety information, as well as indications as to current selection mode and return electrode monitoring status.
The gi 4000 has an integrated peristatic lavage pump designed to provide lavage washing in the gut, either by accessories connected directly to endoscopes which have dedicated washing channels, or by accessories which introduce lavage fluid into the biopsy port of an endoscope or via dedicated washing channels incorporated into a therapeutic accessory such as a bipolar
MAR 2 2 2012
{1}------------------------------------------------
<113265
Genii, Inc. Traditional 510(k) Premarket Submission qi 4000 Electrosurgical Generator
endostasis probe. The system is operated by a dual pedal footswitch which controls the lavage pump and power delivery.
The argon output is designed for an argon assisted coagulation method. This method uses 99.99% pure argon gas to provide ionized RF arcs to the targeted tissue. The monopolar outputs are designed for cutting and coagulation methods. The unit is designed with a standard monopolar active cord receptacle (foot switch activated).
Endoscopic accessories connect to the monopolar receptacle via a widely available detachable active cord accessory. Both the monopolar and argon outputs require the use of a patient return electrode (grounding pad) to return the RF energy from the patient back to the generator. The unit is designed with a standard GI bipolar single jack receptacle. No patient return electrode (grounding pad) is needed for bipolar methods. The output power of the argon, monopolar and bipolar methods are independently controlled through a touch panel LCD display located on the front panel of the generator.
The argon gas for the argon output is exclusively supplied by the disposable Genii Argon Gas Canister. Each canister contains 49 liters of ultra-high purity (99.999% pure) argon gas (certified). When using the Argon Coagulation Method, the ArConnect is intended for single-use only and will be supplied as a non-sterile, disposable item. The ArConnect has been designed to allow the delivery of monopolar high frequency electrosurgical energy and argon gas to a single use argon coagulation probe sold separately. The ArConnect is supplied individually packaged in disposable packaging and is not to be reprocessed or re-used.
### Indications for Use:
The gi 4000 Electrosurgery Generator is intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas in flexible endoscopic applications. The integrated lavage purnp is intended to wash tissue or mucosa within the gut when used in conjunction with washing catheters or bipolar endoscopic probes, endoscope water jet channels, or endoscope working channels. The Genii ArConnect™ Conducting Adapter is intended for use with the gi 4000 during argon assisted coagulation.
#### Predicate Devices:
- 1. Olympus AFU-100 Peristaltic Pump Unit (PPU); K073207
- 2. Erbe APC 300 Argon Plasma Coagulator and Accessories; K963189
- 3. Erbe ICC 200 Electrosurgical Workstation for Flexible Endoscopy; K933157
- Erbe APC Connector Hose and Probes; K013348 বে
### Summary of Technical Comparisons:
The following table compares gi 4000 Electrosurgery Generator to the predicate devices with respect to intended use, target population, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
{2}------------------------------------------------
| Manufacturer | Genii | Olympus | Erbe APC 300 | Erbe ICC 200 |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Gi 4000 | Olympus AFU-100 | Erbe APC 300 | Erbe ICC 200 |
| 510(k) Number | NA | K073207 | K963189 | K933157 |
| Product Code | GEI | FEQ (with GEI) | GEI | GEI |
| Regulation Number | 878.4400 | 876.1500 (with<br>878.4400) | 878.4400 | 878.4400 |
| Regulation Name | Electrosurgical cutting and<br>coagulation device and<br>accessories | Pump, air non<br>manual, for<br>endoscope (with<br>Electrosurgical<br>cutting and<br>coagulation device<br>and accessories) | Electrosurgical<br>cutting and<br>coagulation device<br>and accessories | Electrosurgical cutting<br>and coagulation device<br>and accessories |
| Intended Use | The gi 4000 Electrosurgery<br>Generator is intended to deliver<br>electrosurgical outputs to<br>perform cutting and/or<br>coagulation and coagulation with<br>or without delivery of argon gas<br>in flexible endoscopic<br>applications. The integrated<br>lavage pump is intended to wash<br>tissue or mucosa within the gut<br>when used in conjunction with<br>washing catheters or bipolar<br>endoscopic probes, endoscope<br>water jet channels, or endoscope<br>working channels. | The Olympus AFU-<br>100 peristaltic<br>flushing pump is<br>intended for<br>irrigation of<br>instruments or<br>irrigation/flushing/<br>cleansing of tissue<br>surfaces and<br>wounds<br>supporting<br>endoscopic<br>diagnosis or<br>therapy. | The Erbe APC 300<br>Argon Plasma<br>Coagulator is an<br>argon gas delivery<br>system that is<br>designed for<br>coagulation when<br>used in<br>combination with<br>Erbe ICC<br>electrosurgery<br>Generators and<br>Erbe APC Probes. | The ICC 200 is intended<br>to deliver high frequency<br>electrical current for the<br>cutting/or coagulation of<br>tissue. |
| Pump Type | Peristaltic | Peristaltic | n/a | n/a |
| Flow Rate | 0-735 ml/minute(tubing<br>dependent)200 ml max through<br>leur valves | 0-600ml/minute<br>(tubing<br>dependent) 200 ml<br>max through leur<br>valves | n/a | n/a |
| Dimensions<br>(Width x Height x<br>Depth) | 16" x 6.5" x 16" (19.5 X 7 X 19<br>max overall including pump head,<br>water bottle holder and feet) | 295 X 430 X115<br>mm (note<br>millimeters not<br>inches) | APC 16" x 4" x<br>14.5" (needs cart<br>for gas tanks) | 11" x 6" x 14.5" |
| Max Voltage Peak | 4200 Vp. | n/a | n/a | 4000 Vp |
| Frequency | 315-550 KHz. | n/a | n/a | 330-1,000 KHz (1 MHz). |
| Output Power Limit | 60-200 watts. | n/a | n/a | 120-200 watts |
| Gas Flow Rate | 0.1-2.0 mL/minute. | n/a | 0.1-9.0<br>mL/minute. | n/a |
| Gas Purity | 99.999%. | n/a | 99.998%. | n/a |
| Tank Size | 49 liters. | n/a | 2 X 934 liters. | n/a |
| Electrical Safety/other<br>standards | IEC 60601-1<br>IEC 60601-1-2,<br>IEC 60601-2-2,<br>CISPR 11,<br>IEC 62304,<br>ISO 14971 | IEC 60601-1,<br>UL 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-1-4 | ANSI/AAMI HF-<br>18/1993, IEC<br>60601-1-2 | Not known |
| Wave Form | | n/a | n/a | |
| •Mono Soft Coag | Continuous Sine | | | Continuous Sine |
| •Mono Coag | Modulated | | | Modulated |
| •Mono Blend Cut | Alternating 50%DC | | | n/a |
.
{3}------------------------------------------------
K113265
| Manufacturer | Genii | Olympus | Erbe APC 300 | Erbe ICC 200 |
|------------------------|-----------------------------------------|-----------------|--------------|-------------------|
| Trade Name | Gi 4000 | Olympus AFU-100 | | |
| • Mono Cut | Continuous sine | | | Continuous sine |
| • Bipolar | Continuous sine | | | Continuous sine |
| • Argon | Modulated | | | Modulated |
| Max Voltage Peak | | n/a | n/a | |
| • Mono Soft Coag | 300 | | | 190 |
| • Mono Coag | 2200 | | | 2300-2500 |
| • Mono Blend Cut /PBC | 1625/1050 | | | n/a (EC) |
| • Mono Pulse Cut | 1000 | | | 380-600/190 (EC) |
| • Mono Cut | 780 | | | 600 |
| • Bipolar | 104 | | | 190 |
| • Argon | 4200 | | | 4000 |
| Frequency (KHz) | | n/a | n/a | |
| • Mono Soft Coag | 350KHz | | | 330KHz |
| • Mono Coag | 460-500KHz | | | 1,000KHz |
| • Mono Blend Cut/PBC | 350/350 | | | n/a |
| • Mono Pulse Cut | 350 | | | 330 |
| • Mono Cut | 350 | | | 330 |
| • Bipolar | 350 | | | 350 |
| • Argon | 460-500 | | | 1,000KHz |
| Max Power | | n/a | n/a | |
| • Mono Soft Coag | 120 Watts @ 100 Ω | | | 120 Watts @ 125 Ω |
| • Mono Coag | 120 Watts @ 500 Ω | | | 120 Watts @ 350 Ω |
| • Mono Blend Cut/ PBC | 120 Watts @ 500 Ω/<br>120 Watts @ 500 Ω | | | n/a |
| • Mono Pulse Cut | 120 Watts @ 500 Ω | | | 200 Watts @ 500 Ω |
| • Mono Cut | 200 Watts @ 500 Ω | | | 200 Watts @ 500 Ω |
| • Bipolar | 60 Watts @ 400 Ω | | | 120 Watts @ 125 Ω |
| • Argon | 120 Watts @ 500 Ω | | | 99 Watts @ 350 Ω |
| Duty Cycle | | n/a | n/a | |
| • Mono Soft Coag | 100% | | | 100% |
| • Mono Coag | 6% | | | 4%-6% |
| • Mono Cut | 100% | | | 100% |
| • Bipolar | 100% | | | 100% |
| Crest Factor | | n/a | n/a | |
| • Mono Soft Coag | 1.4 | | | 1.4 |
| • Mono Cut | 1.4 | | | 1.4 |
| • Bipolar | 1.4 | | | 1.4 |
| Power Curve | | n/a | n/a | |
| • Mono Soft Coag | Narrow | | | Narrow |
| • Mono Coag | Wide | | | Wide |
| • Bipolar | Narrow | | | Narrow |
| Pulse Speed | | n/a | n/a | |
| • Mono Pulse Blend Cut | 700ms ON/50ms OFF | | | 750ms ON/50ms OFF |
{4}------------------------------------------------
| Manufacturer | Genii | |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Genii ArConnect | Erbe APC Connector Hose |
| 510(k) Number | NA | K013348 |
| Product Code | GEI | GEI |
| Regulation Number | 878.4400 | 878.4400 |
| Regulation Name | Electrosurgical cutting and coagulation device and<br>accessories | Electrosurgical cutting and coagulation<br>device and accessories |
| Intended Use | The Genii ArConnect conducting adapter is intended for<br>use with the Genii GI4000 during argon assisted<br>coagulation. | The APC Connector Hose and Probes are<br>intended for use in argon plasma<br>coagulation. The device is used to treat<br>many conditions in endoscopy for various<br>surgical procedures. |
| Material | PVC tubing Shore A75+3Clear,<br>Cable 24AWG<br>Terminal: brass with nickel and Gold plating;<br>Insert: ABS PA707 with EX-33084 white;<br>Overmold HDPE LH523, nature; PE white, #PEM864;<br>Gas Tube: ABS PA707, with EX-33084 White;<br>O ring: AS 568-008, Katon P/N: OR-NB50NL-008-BK,<br>ShoreA: 50, BUNA-U;<br>Heat shrink sleeve band: 82*98.5* | Polyamide, PEEK, Thermoflex, silicone. |
| Length | 5 feet. | 6.5 feet. |
| Back Flow | Integrated in the generator gas flow design | A separate back flow filter/valve is sold<br>separately to be used with the hose. |
| How Supplied | Single use; disposable. | Re-useable. |
The following table compares Genii ArConnect accessory to the predicate accessory device.
### Non-clinical Data - Bench Testing:
As part of demonstrating safety and effectiveness of the Genii gi 4000 Electrosurgical Generator and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, the following testing has been performed:
In vitro testing Electrical Safety EMC Biocompatibility Software Testing
In Vitro Testing: An in vitro tissue study was conducted to compare the lesions created by the subject gi 4000 electrosurgical generator and the predicate Erbe ICC 200E with APC 300 system when using equivalent outputs. This in vitro study demonstrated substantial similarity between lesions created by the subject gi 4000 device and the predicate Erbe ICC 200E with APC 300 system. The data demonstrate that the performance of the subject device is equivalent to the predicate devices.
Electrical Safety and EMC: The gi 4000 Electrosurgical Generator has been tested and met the requirements of the Electrical Safety and EMC standards used.
{5}------------------------------------------------
K113265
> Biocompatibility Testing: Cytotoxicity, Irritation and Sensitization testing have been completed on the ArConnect Conducting Cable. The test results demonstrate that the ArConnect Conducting Cable is non-toxic.
Software Verification and Validation: Software verification and validation testing has been performed on the gi 4000 Electrosurgical Generator.
### Safety and Effectiveness:
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the differences between the Genii gi 4000 Electrosurgical Generator and the predicate devices do not raise any questions regarding its safety and effectiveness. The Genii gi 4000 Electrosurgical Generator, as designed and manufactured, therefore is determined to be substantially equivalent to the referenced predicate devices.
{6}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Genii, Incorporated % Emergo Group, Incorporated Ms. Julie Powell 611 West 5th Street, Third Floor Austin, Texas 78701
MAR 2 2 2012
Re: K113265
Trade/Device Name: gi 4000 Electrosurgical Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 14, 2012 Received: March 15, 2012
Dear Ms. Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
## Page 2 – Ms. Julie Powell
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Device Name: gi 4000 Electrosurgical Generator
Indications for Use:
The gi 4000 Electrosurgical Generator is intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas in flexible endoscopic applications. The integrated lavage pump is intended to wash tissue or mucosa within the gut when used in conjunction with washing catheters or bipolar endoscopic probes, endoscope water jet channels, or endoscope working channels. The Genii ArConnect™ Conducting Adapter is intended for use with the gi 4000 during argon assisted coagulation.
Prescription Use __ X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off Office of Device Evaluation
210(k)
(ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113265
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.