KEPLER MEDTEC DISPOSABLE BIPOLAR CABLE

K110462 · Kepler Medtec · GEI · Sep 26, 2011 · General, Plastic Surgery

Device Facts

Record IDK110462
Device NameKEPLER MEDTEC DISPOSABLE BIPOLAR CABLE
ApplicantKepler Medtec
Product CodeGEI · General, Plastic Surgery
Decision DateSep 26, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Kepler MedTec Disposable Bipolar Cable is intended to connect an electrosurgical device to an electrosurgical generator. It is indicated for use with bipolar forceps during general surgical procedures.

Device Story

Disposable bipolar cable; connects electrosurgical generator to bipolar forceps; delivers electrical power for cutting and coagulation. Features male 2-pin banana connector for generator interface; female 2-pin plug connector for forceps. Used in general surgical procedures; operated by surgical staff. Compatible with standard US 5.8 mm pin-spaced forceps and various generators (ValleyLab, Conmed, Medtronic, Bovie). Single-use; provided sterile.

Clinical Evidence

No clinical data. Bench testing only, including compliance with IEC 60601-2-2 for electrical safety and biocompatibility/sterilization standards (ANSI/AAMI 10993-7, ISO 11135-1, ISO 11137-1, ISO 11138-1).

Technological Characteristics

Disposable bipolar cable; 3660 mm length; 2-pin banana (male) and 2-pin plug (female) connectors. Compliant with IEC 60601-2-2. Sterilized via EtO. Biocompatible materials.

Indications for Use

Indicated for use with bipolar forceps during general surgical procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K110462 Pg 1 of 2 #### Traditional 510(k) Summary SEP 2 6 2011 . Kepler MedTec A) SUBMITTED BY : 223 Avenue D Willston, VT 05495 Registration # 3008160705 CONTACT: Sharyn Orton, PhD MEDIcept Inc. 200 Homer Ave Ashland, MA 01721 401-330-8264 508-231-8861 Fax B) DEVICE NAME: Disposable Bipolar Cable COMMON NAME: Disposable Bipolar Cable DEVICE CLASS: 21 CFR 878.4400 Electrosurgical cutting and coagulation device and accessories, Class II PRODUCT CODE: GEI C) PREDICATES: . K970993 DeRoyal Industries, Inc. - Disposable Bipolar Electrosurgical Cable D) DEVICE DESCRIPTION: The Kepler MedTec Disposable Bipolar Cable includes available in multiple colors. The Kepler MedTec Disposable Bipolar Cable is provided sterile. Bipolar forceps (handpiece) are connected to a generator through a bipolar cable. The bipolar cable is flexible and delivers the electrical power from the generator to the binolar forcep. There is a male and female end termination on the cable. The male end termination (2-pin banana connector) is connected to the generator; the female end termination (2-pin plug type connector) to the bipolar forceps. The Kepler MedTec Disposable Bipolar Cable is compatible with: - Bipolar Forceps using standard US connection with 5.8 mm spaced pins . - . VallyLab, Conmed, Medtronic and Bovie generators E) INTENDED USE: The Kepler MedTec Disposable Bipolar Cable is intended to connect an electrosurgical device to an electrosurgical generator. It is indicated for use with bipolar forceps during general surgical procedures. Kepler MedTec Traditional 510(k) Disposable Bipolar Cable September 2011 MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721 {1}------------------------------------------------ K110462 pg 2 of 2 ## F) SUBSTANTIAL EQUIVALENCE COMPARISON AND DISCUSSION | Device Features | Kepler MedTec<br>Disposable Bipolar Cables | DeRoyal<br>Disposable Bipolar cables | |------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | TBD | K970993 | | Intended Use/Indication<br>for Use | Are intended to connect an -<br>electrosurgical device to an<br>electrosurgical generator.<br>Indicated for use with bipolar<br>forceps during general surgical<br>procedures. | Indicated for use during<br>electrosurgical procedures to<br>provide transmission of<br>electrical power from a<br>electrosurgical generator to a<br>bipolar instrument | | Patient Usage | Single use disposable provided<br>sterile | Single use disposable provided<br>sterile | | Cable length | 3660 mm | 366 cm | | Electrical safety/EMC | IEC 60601-2-2 compliant | IEC 60601-2-2 compliant | The Kepler MedTec Disposable Bipolar Cable has the same intended use, target population, clinical setting, and technology as the predicate device. Kepler MedTec believes that the differences between Kepler MedTec Disposable Bipolar Cable and the predicate (i.e. cable color, materials, any slight differences in dimensions; reusability) does not raise new issues of safety or effectiveness. ### G) CONCLUSION The Kepler MedTec Disposable Bipolar Cable has the same intended use, target population, clinical setting, and technology as the predicate devices. Kepler MedTec believes the Kepler MedTec Disposable Bipolar Cable is therefore substantially equivalent to the predicate devices based on intended usage, technology comparison and system performance. #### H) TESTING No clinical testing was submitted with this application. ## I) OTHER The Kepler MedTec Disposable Cable is: - � Compliant with IEC 60601-2-2 (2006): Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment - For EtO sterilization, compliant with ANSVAAMI 10993-7; ISO 11135-1; ISO 11137-1; . ISO 11138-1 - . Biocompatible Kepler MedTec Traditional 510(k) Disposable Bipolar Cable September 2011 MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and three horizontal bars above it. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MI) 20993-0002 Kepler MedTec % MEDIcept. Inc. Sharyn Orton, Ph.D. 200 Homer Avenue Ashland, Massachusetts 01721 SEP 26 2011 · Re: K110462 Frade/Device Name: Kepler MedTec Disposable Bipolar Cable Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 06, 2011 Received: September 08, 2011 Dear Dr. Orton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Sharyn Orton, Ph.D. or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Eric M. Keith - Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K110462 # Indications for Use Form 510(k) Number (if known): . . Device Name: Disposable Bipolar Cable .............................................................................................................................................................................. Indications for Use: The Kepler MedTec Disposable Bipolar Cables are intended to connect an electrosurgical device to an electrosurgical generator. They are indicated for use with bipolar forceps during general surgical procedures. Prescription Use X 21CFR 801, Subpart D OR Over-the-Counter Use 21CFR 801.109 ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) .............................................................................................................................................................................. ## Concurrence of CDRH, Office of Device Evaluation (ODE) Sign Off (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K110462 Kepler MedTec Traditional 510(k) Disposable Bipolar Cable February 10, 2011 MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...