The Thermedx 37~5 electrocautery probes are indicated for electrosurgical cutting and coagulation during a variety of urological, gynecological, laparoscopic and general surgical procedures.
Device Story
Thermedx 37-5 Electrocautery Probes are hand-held electrosurgical instruments designed for cutting and coagulation; they integrate suction and irrigation capabilities. The device features various tip configurations (J Hook, L Hook, Spatula, Needle) and connects to a Trumpet valve disposable assembly (model TV0002). It is used in surgical settings by clinicians. The probe connects to an electrosurgical unit via a cable with PVC insulation and copper wires. By delivering electrical energy to the surgical site, the device enables precise tissue cutting and hemostasis, potentially reducing the need for separate instruments and improving surgical efficiency.
Clinical Evidence
Bench testing only. The device was verified to comply with established specifications and validated for safety and effectiveness using electrical safety and biocompatibility standards (IEC 601-2-2 and ISO 10993-1).
Technological Characteristics
Monopolar electrosurgical probe; 33 cm length, 5 mm diameter; ABS coupling; PVC-insulated copper cable. Materials are biocompatible per ISO 10993. Operates via electrosurgical unit energy. Tip configurations: J Hook, L Hook, Spatula, Needle.
Indications for Use
Indicated for patients undergoing urological, gynecological, laparoscopic, or general surgical procedures requiring electrosurgical cutting and coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
K102275
page 1 of 2
## 510(k) Summary
MAR 2 5 2011
## Traditional 510(k) Notification Thermedx 37~5TM Electrode Probes Thermedx, LLC
, ﺗ Applicant Thermedx, LLC 31200 Solon Rd. Unit#1, Solon Oh 44139 440-542-0883 Contact Person: John R. Scoville, Jr., VP Quality and Regulatory Affairs Date Prepared: March 23, 2011
#### Device Name
Proprietary Name: Thermedx 37~5TM Electrocautery Probes Common Name: Electro Cautery Probe Classification Name: Electrosurgical Cutting and Coagulation and Accessories (product code GEI)
#### Indications for Use
The Thermedx 37~5 electrocautery probes are indicated for electrosurgical cutting and coagulation during a variety of urological, gynecological, laparoscopic and general surgical procedures.
#### Device Description
The Thermedx 37~5 electrode probes are electrosurgical devices which enable the user to have the functions of cutting and coagulation along with suction and irrigation in a single hand held device. These probes are intended to be used in conjunction with the Trumpet valve disposable assembly (model number TV0002), The probes come in a variety of tip configurations. Monopolar electrodes include a J Hook. L Hook. Spatula and Needle tip. . All electrodes have working dimensions of 33 cm in length and 5 mm diameter. Electrodes are manufactured of biocompatible materials per ISO 10993 and are safe, effective and durable for their intended use. All electrodes have an ABS coupling to connect to the trumpet valve disposable assembly and are connected to an electrosurgical unit with a cable of PVC insulation and copper wires.
#### Substantial Equivalence
The monopolar electrode probes included in the 37-5 fluid management system are substantially equivalent in terms of safety and effectiveness to the following devices.
- Modulap Disposable Electrosurgical Probes (K994319, KP83623) ATC Technologies part numbers l . 4180, 4181, 4182 and 4183
- 2. Genicon Re-Usable Monopolar Probes (K061417), Genicon, part numbers 707-005-133, 707-005-233 and 707-005-333.
- Stryker Stiletto Electrosurgical Probe (K963765), Stryker Endoscopy 3.
- Conmed Electrosurgical Probe (K012018), Conmed 4.
#### Technological Comparison
This 510(k) Notification demonstrates the substantial equivalence of the Thermedx 37-5 using the substantial equivalence criteria in FDA's K86-3 510(k) Guidance. The new device has the same intended use as the predicate devices, and the minor differences in indications for use statements do not alter the device's intended therapeutic effect. The new device also has similar technological characteristics to those of the predicate devices, and the minor differences are not significant with respect to the safety or effectiveness of the new device. All devices were tested to the same electrical safety and biological testing standards (IEC 601-2-2 and ISO 10993-1) demonstrating equivalence. No animal or clinical test results were required to demonstrate substantial equivalence.
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "K102275" with a line going through it. Below the line, the text "page 2 of 2" is written. The text is written in black ink on a white background.
#### Safety and Effectiveness
The Thermedx 37~5 Electrocautery probes have been developed in accordance with FDA's design controls requirements. The design and development process has ensured that all applicable safety and effectiveness issues are addressed, including but not limited to labeling, shelf life, biocompatibility and electrical safety. The product has been fully verified to comply with its established specifications and applicable standards and validated to ensure the device is safe and effective.
All statements and representations set forth above regarding or related to "substantially equivalent" or "substantial equivalence" are in the limited context of the definition and purpose of substantial equivalence in the Federal Food, Drug, and Cosmetic Act, as amended, and applicable regulations of the Food and Drug Administration, and are not made in the context of, for any purpose related to, or as an admission against interest under, any other laws or regulations, including patent laws (whether in the context of patent infringement or otherwise).
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The text is in all caps and appears to be in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Thermedx, LLC % Mr. John R. Scoville 31200 Solon Road, Unit #1 Solon, Ohio 44139 -
MAR 2 5 2011
Re: K102275
Trade/Device Name: Thermedx 37-5 Electrocautery Probes, Models: MPJ200, MPL210, MPJ220, MPJ230 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 07, 2011 Received: March 08, 2011
Dear Mr. Scoville:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
### Page 2 - Mr. John Scoville
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Henry B. Reite
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text 'K 102275 page 1 of 1' written in black ink on a white background. The text is handwritten and appears to be part of a document or label. The phrase 'page 1 of 1' suggests that this is the only page in the document. The text is slightly slanted and the handwriting is somewhat uneven.
# Indications for Use Form (Text Version)
# Indications for Use
510(k) Number (if known):
Device Name: Thermedx 37~5™ Electrocautery Probes
Indications for Use:
The Thermedx 37~5 electrocautery probes are indicated for electrosurgical cutting and coagulation during a variety of urology, gynecology, laparoscopy and general surgery procedures.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
Page __ of ___
510(k) Number K102275
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.