K101108 · Erbe USA, Inc. · GEI · Jul 16, 2010 · General, Plastic Surgery
Device Facts
Record ID
K101108
Device Name
ERBE ESU MODEL VIO 100 C WITH ACCESSORIES
Applicant
Erbe USA, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 16, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ERBE ESU Model VIO 100 C with Accessories is intended to deliver High Frequency (HF) electrical current for the cutting and/or coagulation of issue.
Device Story
ERBE ESU Model VIO 100 C is a tabletop electrosurgical generator delivering high-frequency (HF) electrical current (380 kHz) for tissue cutting and coagulation. Operated by physicians in clinical settings via footswitches or hand instruments. System features monopolar and bipolar receptacles; programmable modes (Auto Cut, Soft Coag, Forced Coag, Bipolar Soft Coag, Dry Cut); and automatic start functionality. Includes neutral electrode monitoring system for pad connection and skin contact. Provides audio/visual alerts for malfunctions or user errors. Device is non-sterile and reusable. Output power is limited to 100 watts. Simplifies surgical workflows by providing fundamental modes without complex 'Effect' settings found in higher-end models. Benefits include cost-effective, basic electrosurgical capability for standard interventional procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence established through verification and validation of design changes and performance testing.
Technological Characteristics
Tabletop electrosurgical generator; 100W max output; 380 kHz operating frequency. Features monopolar/bipolar receptacles, neutral electrode monitoring system, and footswitch connectivity. Non-sterile, reusable. Programmable user interface for mode selection and power settings. Audio/visual error monitoring system.
Indications for Use
Indicated for cutting and/or coagulation of tissue using high frequency electrical current in interventional applications. For prescription use only.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Kl01108
## ERBE USA Incorporated Special 510(k): ERBE ESU Model VIO 100 C with Accessories
## 510(k) SUMMARY
| Submitted By: | ERBE USA, Inc.<br>2225 Northwest Parkway<br>Marietta, GA 30067<br>Tel: 770-955-4400<br>Fax: 770-955-2577 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------|
| Contact Person: | John Tartal<br>QA/RA Manager |
| Date Prepared: | March 30, 2010 |
| Common Name: | ElectroSurgical Unit (ESU/Generator) System |
| Trade/Proprietary Name: | ERBE ESU Model VIO 100 C with Accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories (21 CFR 878.4400) |
| Product Code: | 79GEI |
| Legally Marketed<br>Predicate Device: | ERBE VIO ESU (Model VIO 200 S), 510(k) Number:<br>K080715 |
#### Device Description:
#### Unit
The ERBE ESU Model VIO 100 C is a Generator that uses High Frequency (HF) electrical current waveforms to cut and/or coagulate tissue. It has a display as well as various cutting and coagulation modes to provide the physician standard modalities in interventional applications (i.e. its ability to generate the HF current). The System has an automatic start feature. The equipment is programmable and various accessories (e.g. footswitches, hand instruments, etc.) as well as modes may be assigned to perform specific functions. When activated, the Device has an audio as well as a visual erroring system (i.e. malfunctions or user errors are detected with medical personnel being alerted visually and/or by sound with, in some cases, no energy being delivered.). The Unit is supplied non-sterile and is reusable.
Note: VIO stands for Variable Cut and Coagulation.
{1}------------------------------------------------
## ERBE USA Incorporated Special 510(k): ERBE ESU Model VIO 100 C with Accessories
## Footswitches (Accessories)
Designated Footswitches have been developed for use with the ERBE ESU Model VIO 100 C. A Footswitch provides the physician a means to activate a mode of the ESU by depressing a foot pedal. There is a choice of two different types of Footswitches as follows:
VIO C One (Single) Pedal Coag Footswitch (Note: Activates coagulation mode.) VIO C Two (Dual/Double) Pedal Footswitch (Note: Activates respectively cut and coagulation modes.)
## Intended Use:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The ERBE ESU Model VIO 100 C is intended to deliver high frequency electrical current for the cutting and/or coagulation of tissue.
# Similarities and Differences of the Modified Device to the Current Device (Predicate Comparison/Substantial Equivalence):
#### Similarities
The modified ESU (ERBE Model VIO 100 C) has the same intended use and uses the same basic accessories (i.e. A/C Cord, Adapters, Connecting Cables, etc.) as the predicate ESU (ERBE Model VIO 200 C). Both Units have a Monopolar, Bipolar, and Neutral Electrode Receptacle. Also, available modes [i.e., Auto Cut, Soft Coaq, Forced Coag, Bipolar Soft Coag] in the modified device are the same Modes that are in the predicate (Note: Modes do not have effect settings and have power limitations with the modified device as compared to the predicate.). Both Generators have user interface displays to select modes, power settings, etc. The modified and predicate devices are programmable and have the Auto Start feature with the Bipolar Soft Coag Mode. Also, each Unit has audio and visual error monitoring. The modified ESU is manufactured by ERBE Elektromedizin GmbH in Germany and like the predicate Generator will be supplied as non-sterile and is reusable. The packaging and labeling (e.g. Outer Package Label, User Manual, etc.) is similar for each device as well.
#### Differences
The ERBE ESU Model VIO 100 C is different than the predicate VIO 200 S Model in that the modified ESU is a basic Model with standard market features. It is lighter, smaller, and designed for use on a tabletop. Compared to the predicate device, the modified Unit has a Neutral Electrode Monitoring System which does not monitor the current density and symmetry of a split return electrode. However, the modified Unit monitors pad connection and with a split return electrode it monitors skin contact. Since the VIO 100 C provides less power than the predicate device, the monitoring system is adequate and according to the industry standard. Also, there are less programming capabilities [i.e., four (4) program possibilities] for the proposed device compared with nine (9) for the predicate ESU which makes the Unit Jess complicated.
{2}------------------------------------------------
## ERBE USA Incorporated Special 510(k): ERBE ESU Model VIQ 100 C with Accessories
The ERBE ESU Model VIO 100 C also is different than the ERBE VIO 200 S ESU in that the available wattage is less (i.e., 100 watts as compared to 200 watts). However, a maximum 100 watt Generator has been found to be adequate (Note: There are many ESUs on the U.S. market with maximum wattage being less than 100 watts.). The New Model also operates at slightly higher frequency (380 kHz) as compared to 350 kHz for the predicate device. Also there are no "Effect" settings for each of the modes with the modified Unit as compared to the predicate. Simply a power setting is used to increase or decrease the available wattage for delivery to target tissue (i.e., with a wattage setting increase, the effect is greater) for the modified ESU which eliminates the need for effect settings.
Modes that are in the predicate and not in the modified device are Endo Cut Q, Endo Cut I, and APC Forced as well as argon-assisted modes. An additional cut mode (Dry Cut) was included in the modified Unit as compared to the predicate (Note: This mode is in other 510(k) cleared ERBE ESU Models.). In summary, the modes in the ERBE Model VIO 100 C are principal ones to perform basic cutting and coagulating activities.
Finally, the ERBE ESU Model VIO 100 C also is different than the ERBE VIO 200 S ESU in that both Units have designated Footswitches. However, both of the ESU's Footswitches are made with similar materials and activate, via a foot pedal, cut as well as coagulation modes.
#### Conclusion:
The ERBE ESU Model VIO 100 C with Accessories has the same intended use, principles of operation, and similar technological characteristics as the predicate ESU in the previously cleared 510(k). The modifications involve having a Generator that is more basic (i.e., less complicated with fundamental modes) with designated Footswitches thus having a more cost effective System. In conclusion, all the changes were verified or validated. As a result, the changes did not raise safety or efficacy concerns nor adversely affect safety or effectiveness.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three distinct, curved lines forming its body and wings. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
ERBE USA, Inc. % Mr. John Tartal 2225 Northwest Parkway Marietta, GA 30067
# JUL 1 5 2010
Re: K101108
Trade/Device Name: ERBE ESU Model VIO 100 C with Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: April 2, 2010 Received: April 20, 2010
Dear Mr. Tartal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. John Tartal
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm117809.html 1 the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR, Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Barbara Bauer up
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: ERBE USA, Inc.'s ERBE ESU Model VIO 100 C with Accessories
Indications For Use:
The ERBE ESU Model VIO 100 C with Accessories is intended to deliver High Frequency (HF) electrical current for the cutting and/or coagulation of issue.
Prescription Use X (Part 21 CFR 801 Subpart D)
· AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101108
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.