HALT 2000GI ELECTROSURGICAL RF ABLATION SYSTEM MODEL HALT 2000GI
K094009 · Halt Medical, Inc. · GEI · Jun 14, 2010 · General, Plastic Surgery
Device Facts
Record ID
K094009
Device Name
HALT 2000GI ELECTROSURGICAL RF ABLATION SYSTEM MODEL HALT 2000GI
Applicant
Halt Medical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 14, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic, and intraoperative coagulation and ablation of soft tissue.
Device Story
The Halt 2000GI™ is an electrosurgical radiofrequency (RF) ablation system used by clinicians in percutaneous, laparoscopic, or intraoperative settings. The system consists of an RF generator, a disposable RF probe (Tulip™) with seven deployable needle electrodes, an extension cable, dispersive electrode pads, and a pneumatic foot pedal. The clinician places the probe into target soft tissue and deploys the electrodes. The generator delivers up to 200 W of RF power at 460 kHz. The system operates in temperature control, manual control, or coagulation modes. A GUI displays real-time impedance and temperature feedback from the probe sensors. The system automatically stops energy delivery if parameters exceed safety limits; the user can also manually stop energy via the foot pedal. The device creates zones of tissue necrosis for therapeutic benefit. It is intended for prescription use only.
Clinical Evidence
Bench and animal testing only. Bench testing confirmed ablation capability, producing reproducible, discretely demarcated zones of tissue necrosis. Biocompatibility testing confirmed compliance with ISO 10993-1, ISO 10993-5, and ISO 10993-10. Electrical and mechanical safety verified per EN/IEC 60601-1, 60601-1-2, 60601-2-2, and ANSI/AAMI HF-18.
Technological Characteristics
RF generator delivering 460 kHz energy; max 200 W power. Probe materials: Nitinol, surgical grade stainless steel, PEEK. Features 7 deployable needle electrodes. Feedback: real-time impedance and temperature sensors. Connectivity: pneumatic foot pedal, extension cables. Software: validation testing performed. Standards: EN/IEC 60601 series, ANSI/AAMI HF-18, ISO 10993 series.
Indications for Use
Indicated for percutaneous, laparoscopic, and intraoperative coagulation and ablation of soft tissue in patients requiring such procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K094009" at the top. Below the text is a horizontal line. Under the line is the word "Halt" in a stylized font, with the word "MEDICAL" in smaller letters below it.
JUN 1 4 2010
01 (3)
t Ti
## 510(K) SUMMARY
As Required By 21 CFR 807.92
| Date of Summary Preparation | June 7, 2010 |
|---------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter and Owner's Name<br>and Address | Halt Medical, Inc.<br>131 Sand Creek Road, Suite B<br>Brentwood, CA 94513<br>Main: (925) 634-7943<br>Fax: (925) 634-7841 |
| Contact Person | Clarisa A. Tate<br>Director of RA/QA, Halt Medical, Inc.<br>Office: (925) 271-0626<br>e-mail: ctate@haltmedical.com |
| Trade Name | Halt 2000GI™ Electrosurgical Radiofrequency<br>Ablation System |
| Common/Classification Name | Electrosurgical, cutting and coagulation device<br>and accessories |
| Classification | Class II |
| Product Code | GEI |
| Classification Panel | General and Plastic Surgery |
| Classification Regulation | 21 CFR §878.4400 |
| Legally Marketed Device to<br>which substantial equivalence is<br>claimed | Rita Medical Systems - Electrosurgical RF<br>Generator and accessories:<br>• Rita Model 1500 (K993944)<br>• Rita Model 1500X (K032149) |
| Intended Use | The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System<br>is indicated for use in percutaneous, laparoscopic, and intraoperative<br>coagulation and ablation of soft tissue. |
| Device<br>Description | The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System<br>includes the following system components:<br><br>• Halt 2000GI Radiofrequency Generator (RF Generator): that<br>provides RF energy to the RF Probe |
. .
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the number 2094009 at the top, followed by a horizontal line. Below the line is the word "Halt" in a stylized font, with the word "MEDICAL" in smaller letters below and to the right of "Halt". The word "Halt" is written in a cursive style, with the "H" and "t" connected by a curved line.
Page 2 of 3
- Disposable RF Probe (Tulip™, TU 1000), a hand piece with a . trocar-pointed shaft and 7 deployable RF needle electrodes.
- RF Probe Extension Cable that connects the RF Probe to the . RF Generator.
- Dispersive Electrode Pads (TSP 115) Set that provides the . return path for the RF energy applied by the RF Probe. (Use only Dispersive Electrode Pads provided by Halt Medical, Inc.)
- Dispersive Electrode Pad Set Extension Cable (TSPCBL) that ● connects the Dispersive Electrode Pad Set to the RF Generator.
- Power Cord that is a medical grade power cord providing AC . power to the RF Generator.
- Foot Pedal that is a pneumatic foot pedal with tubing used to � turn RF energy on and off.
The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System is designed to deliver up to 200 W of RF power at 460 kHz in three operational modes: Temperature Control. Manual Control and Coagulation Mode. A touch screen with a graphical user interface (GUI) enables selection of operational parameters such as the mode of operation, the ablation time, the target temperature, and the power delivery level. With the RF probe placed in the tissue to be ablated and its electrodes deployed, RF power can be turned on. The system parameters are continuously monitored and displayed on the RF Generator. If the measured parameters are outside the acceptable limits, the RF energy delivery automatically stops and a message appears on the graphical user interface. RF energy during an ablation or coagulation can also be stopped at any time by the user by pressing the foot pedal.
Technological Characteristics Compared to Predicate Devices
The design features and principal modes of operation of the Halt 2000GI™ System is equivalent to commercially available RF Ablation Systems specifically the predicate devices listed above.
Substantial equivalence is established with respect to similar intended use, principal design, type of energy used or delivered, materials, performance and safety requirements.
The key similarities and differences of the Halt RFA System to the legally marketed RITA RF Ablation Systems are summarized below:
- . Both Utilizes RF energy at equivalent frequencies (approximately 450 kHz), sinusoidal waveforms and
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the number K094009 written in black ink, with a line underneath it. Below the line is the word "Halt" in a stylized font, with the word "Medical" in smaller letters underneath. The word "Halt" has a design element that looks like an arrow going through the "t".
Page 3 of 3
equivalent power levels (150 – 250 Watts)
| Both incorporates the same surgical and technical concepts for the in situ delivery and control of RF energy through deployable array needle electrodes with impedance and temperature feedback under visual control. | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Halt RF Probe (Tulip) features seven deployable needle electrodes, a live trocar tip, and provides real-time simultaneous feedback from all seven temperature sensors while RITA® Probe features nine deployable needle electrodes, a hollow shaft with a needle tip, and provides real-time temperature feedback from five independent temperature sensors within the array. | |
| Halt RF Probe (Tulip) is designed to create single ablations up to 6 cm in size while RITA Probe is up to 5 cm. Both can also be used to ablate the needle track. | |
| Both the RF Probes use similar medical grade metal alloys and plastic polymers (i.e. Nitinol®, surgical grade Stainless Steel and PEEK). | |
| Performance<br>Testing | The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System was subjected to a battery of electrical, mechanical, and software validation testing, as well as applicable safety requirements (EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-2, ANSI/AAMI HF-18). The system passed all testing. |
| | Animal and bench testing also successfully demonstrates that the Halt RF Ablation System performs as intended and per specifications. The ablation capability was confirmed and the radiofrequency ablation provides a reproducible, discretely demarcated zone of tissue necrosis which was surrounded by normal tissue perfused with blood. |
| | A series of biocompatibility testing also demonstrated that the device materials are safe, suitable, and appropriate for their intended use and in compliance with ISO 10993-1, ISO 10993-5, and ISO 10993-10. |
| Conclusion | The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System with its indication for use, is substantially equivalent to the legally marketed medical device as demonstrated by the technological characteristics comparison and performance testing completed for this device. |
ે વ
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Halt Medical. Inc. % Mr. Russ DeLonzor President & COO 131 Sand Creek Road, Suite B Brentwood, California 94513
JUN 14 2010
Re: K094009
Trade/Device Name: Halt 2000GI Electrosurgical Radiofrequency Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI
Dated: April 13, 2010 Received: April 15, 2010
Dear Mr. DeLonzor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
...
{4}------------------------------------------------
Page 2 - Mr. Russ DeLonzor
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucrn1 i 800 .html or the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR, Part 807.97). For questions regarding the reporting of adverse events under the MDR (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Millikin
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Form
510(k) Number (if known): K094009
Device Name: Halt 2000GI™ Electrosurgical Radiofrequency Ablation System
Indications for Use:
The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic, and intraoperative coagulation and ablation of soft tissue.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (
Diagnostic Devices (OVD)
---
on Sign-Off (Divisi Division of Surgical Orthopedic, and Restorative Devices
510(k) Number K094009
Evaluation and Safety
Division Sign-Off
510(k)
Office of In Vitro Diagnostic Device
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.