K083608 · Erbe USA, Inc. · GEI · Aug 26, 2009 · General, Plastic Surgery
Device Facts
Record ID
K083608
Device Name
ERBE HYBRID KNIFE
Applicant
Erbe USA, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 26, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ERBE Hybrid Knife™ is intended to be used in combination with the Water Jet System/ERBEJET 2 and an ERBE ESU VIO Model to cut and dissect soft tissue in neurosurgery and soft tissue such as the liver, kidney, etc. within the abdomen, including Total Mesorectal Excision (TME) via the Water Jet System as well as to provide monopolar cutting and coagulation of the target tissue by the ESU in open and as well as endoscopic surgery.
Device Story
Hybrid Knife combines water-jet dissection and electrosurgical cutting/coagulation; used with ERBEJET 2 Water Jet System and VIO Electrosurgical Unit (ESU). Input: pressurized sterile saline for water-jet cutting; HF energy for cautery. Device features retractable electrode (up to 5 mm extension) for monopolar cutting/coagulation. Operated by clinician in open or endoscopic surgery (via >2.8 mm working channel). Footswitches control water-jet and ESU activation. Output: physical tissue dissection and thermal coagulation. Benefits: enables precise tissue management in single device; reduces instrument exchanges. Single-use, sterile.
Clinical Evidence
Bench testing only. Biocompatibility testing performed on specific materials. Performance testing conducted to verify functionality of the Hybrid Knife with a specifically designed software version of the ERBEJET 2 Unit.
Technological Characteristics
Single-use, sterile, disposable device. Components: handle, probe tubing, retractable electrode, tubing to Water Jet, cable to ESU. Dimensions: 2 mm outer diameter, 2200 mm length. Energy: HF electrosurgical energy and pressurized saline water-jet. Sterilization: Ethylene Oxide. Connectivity: connects to ERBEJET 2 and VIO ESU.
Indications for Use
Indicated for patients undergoing neurosurgery or abdominal soft tissue surgery (e.g., liver, kidney, TME) requiring tissue dissection and coagulation. Used in open or endoscopic procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
ERBE Monopolar Attachment for Helix Hydro-Jet (K062712)
ERBE ERBEJET 2 System (Includes Applicator) (K072404)
Submission Summary (Full Text)
{0}------------------------------------------------
Ko836.08
ERBE USA Incorporated Traditional 510(k): ERBE Hybrid Knife™
Page 1 of 2
#### 510(k) SUMMARY
AUG 2 6 2009
ERBE USA, Inc. Submitted By: 2225 Northwest Parkway Marietta, GA 30067 Tel: 770-955-4400 Fax: 770-955-2577
John Tartal Contact Person: QA/RA Manager
February 26, 2008 Date Prepared:
Water Cutter/Dissector with Electrosurgical Cutting and Common Name: Coagulation Capabilities
ERBE Hybrid Knife™ Trade/Proprietary Name:
Jet Lavage (21 CFR Part 880.5475) and Electrosurgical Classification Name: cutting and coagulation device and accessories (21 CFR 878.4400)
FQH and GEI Product Code:
Legally Marketed Predicate Device:
ERBE Monopolar Attachment for Helix Hydro-Jet 510(k) Number: K062712 and ERBE ERBEJET 2 System (Includes Applicator) 510(k) Number: K072404
#### Device Description:
The ERBE Hybrid Knife™ is used with the Water Jet Model ERBEJET® 2 and an ERBE ElectroSurgical Unit (ESU), VIO Model. The Water Jet delivers pressurized sterile normal saline through the Hybrid Knife to cut and dissect soft tissue. The ESU supplies High Frequency (HF) energy through a retractable electrode of the Hybrid Knife for the cutting and coaqulation of tissue. The ERBE Hybrid Knife™ consists of tubing to the Water Jet, a cable to the ESU, a handle, probe tubing, and an electrode. Clinicians would assemble the device and attach it respectively to the Water Jet and ESU. For endoscopic procedures, the Hybrid Knife is placed down the channel of an endoscope that has a working channel greater than 2.8 mm. Upon the set up of the Water Jet and ESU, the Hybrid Knife is ready for use. The pedal of the ERBEJET 2 Footswitch activates its water-jet cutting capabilities. To activate cautery, the recessed electrode is extended out its tip (up to 5 mm) and the ESU's Footswitch is depressed. The ERBE Hybrid Knife™ is provided sterile and is single use.
{1}------------------------------------------------
### ERBE USA Incorporated Traditional 510(k): ERBE Hybrid Knife™
#### Intended Use:
The ERBE Hybrid Knife™ is intended to be used in combination with the Water Jet System/ERBEJET 2 and an ERBE ESU VIO Model to cut and dissect soft tissue in neurosurgery and soft tissue such as the liver, kidney, etc. within the abdomen, including Total Mesorectal Excision (TME) via the Water Jet System as well as to provide monopolar cutting and coagulation of the target tissue by the ESU in open and as well as endoscopic surgery.
## Similarities and Differences of the Proposed Device to the Current Devices (Predicate Comparison/Substantial Equivalence):
#### Similarities
The ERBE Hybrid Knife™ has the same basic intended use as the predicate ERBE Monopolar Attachment for Helix Hydro-Jet. That is it combines water-jet cutting and dissection of soft tissue with the ability of providing cautery. It is similar to the other predicate, ERBEJET 2 Applicator, in that it works with the ERBEJET 2 Unit, functions upon the same operating principles as well as is sterilized via Ethylene Oxide, single use, and disposable. The ERBE Hybrid Knife is also manufactured by ERBE Elektromedizin GmbH.
#### Differences
The types of materials used for the ERBE Hybrid Knife™ are similar to the predicates but specific materials are slightly different. Therefore, biocompatibility of the specific materials for the Hybrid Knife was demonstrated. See Biocompatibility Study, III-18 to III-27. One of the other differences is that the electrode is retractable for the Hybrid Knife as opposed to being just at the end of the tip of the predicate ERBE Monopolar Attachment for Helix Hydro-Jet. Structurally and dimensionally the Hybrid Knife is different than the predicate ERBEJET 2 Applicator, in that its tip is internally different and its Outer Diameter is much smaller (2 mm versus 5 mm) with a much longer length 2200 mm versus 336 mm). The size was altered for use in 2.8 mm or larger working channels of endoscopic scopes. With a specifically designed software version of the ERBEJET 2 Unit, performance testing demonstrated the functionality of the Hybrid Knife. See Performance Testing, III-28 to III-31.
#### Conclusion:
The ERBE Hybrid Knife™ has the same basic intended use, principles of operation, and technological characteristics as the predicate devices in the previously cleared 510(k)s. The modifications involve having a device that connects to the Water Jet Model ERBEJET 2 and an ESU VIO Model, has a smaller outer diameter and longer working length for endoscopic use if desired, as well as retractable electrode. In conclusion, all the changes were verified or validated. As a result, the changes did not raise safety or efficacy concerns nor adversely affect safety or effectiveness.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract emblem that resembles a stylized bird or a caduceus, a symbol often associated with medicine and healthcare.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
ERBE USA, Inc. % Mr. John Tartal QA/RA Manager 2225 Northwest Parkway Marietta, Georgia 30067
## AUG 2 6 2009
Re: K083608
Trade/Device Name: ERBE USA, Inc.'s ERBE Hybrid Knife" Regulation Number: 21 CFR 880.5475 Regulation Name: Jet Lavage Regulatory Class: Class II Product Code: FQH and GEI Dated: August 19, 2009 Received: August 20, 2009
Dear Mr. Tartal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to 1 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. ·
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. John Tartal
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Sincerely yours,
for
Mark N. Melkerson
Mark N. Melkerson Director
Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 6083608
## Indications for Use
510(k) Number (if known):
Device Name: ERBE USA, Inc.'s ERBE Hybrid Knife™
Indications For Use:
The ERBE Hybrid Knife™ is intended to be used in combination with the Water Jet System/ERBEJET 2 and an ERBE ESU VIO Model to cut and dissect soft tissue in neurosurgery and soft tissue such as the liver, kidney, etc. within the abdomen, including Total Mesorectal Excision (TME) via the Water Jet System as well as to provide monopolar cutting and coagulation of the target tissue by the ESU in open and as well as endoscopic surgery.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Pth
Jassion Sign-Off) ion of Surgical, Orthopedic, Restorative Devices
510(k) Number K083608
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.