PAJUNKS DISPOSABLE INSERTS FOR HANDLE INSTRUMENTS AND HF ELECTRODES
K062072 · PAJUNK GmbH Medizintechnologie · GEI · Sep 22, 2006 · General, Plastic Surgery
Device Facts
Record ID
K062072
Device Name
PAJUNKS DISPOSABLE INSERTS FOR HANDLE INSTRUMENTS AND HF ELECTRODES
Applicant
PAJUNK GmbH Medizintechnologie
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 22, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Pajunks disposable inserts are instruments insulated for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures.
Device Story
Modular laparoscopic surgical system comprising reusable handles and disposable instrument inserts; inserts include grasping forceps, scissors, biopsy punches, and clamps/needle holders. System utilizes mechanical locking mechanism to secure inserts into insulated shafts and handles. Handles feature ergonomic design for left/right-handed use, integrated rotation stop, and optional ratchet switch for coagulation control. Operated by surgeons in clinical/laparoscopic settings. Device enables tissue manipulation and monopolar electrosurgical coagulation. Disposable inserts allow for economic replacement of sensitive components, ensuring sharp cutting edges for every intervention.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and intended use comparison to predicate device.
Technological Characteristics
Modular surgical instrument system; materials include glass fiber-reinforced plastic for handles; mechanical locking/latching mechanism for insert-to-shaft-to-handle assembly; monopolar electrosurgical capability; manual operation; sterile, single-use inserts; reusable handles.
Indications for Use
Indicated for use by surgeons to grasp, manipulate, dissect, retrieve, biopsy, cut, or perform optional monopolar coagulation of internal tissue or organs during laparoscopic procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Special Premarket Notification Submission - Djsposable inserts
Pajunk GmbH Karl-Hall-Strasse 01 78187 Geisingen Germany Fon: +49(0)7704-9291-0 Fax: +49(0)7704-9291-610 Establishment Registration Number: 9611612
Submitter Information/ production site:
# Contact:
Christian Quass, Director Regulatory Affairs Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605
E-Mail: christian.quass@pajunk.com
E-Mail: stefan.dayagi@pajunk-usa.com
German American Trade Center
5126 South Royal Atlanta Drive
30084 Tucker, Georgia
Fon: +01(0)770-493-9305
# Contract Sterilizer
SteriPro Lab & EO Facility Dreieichstrasse. 7 64546 Mörfelden Tel +49 6105 23091 or +49 (0) 6105 93470 Fax +49 6105 24760 Germany
# Device Information:
#### Pajunks disposable inserts for Handle instruments and HF Trade Names: Electrodes
# Common Name: Disposable Modular Inserts for Handle instruments
Classification Name: Disposable monopolar surgical instrument
Classification Reference: 21 CFR 876.1500, April 1. 2005
Poposed Classification: Regulatory Class: II
Proposed Product Code: GEI, GCJ
Predicate Devices: Pajunks monopolar electrodes marketed under K033249
Page 1 of 3
# Special 510(k) Premarket Notification Submissio Summary of Safety and Effectiveness
Date of Preparation: August 29th, 2006
USA Contact:
Pajunk USA
USA
Contact
Stefan Dayagi
SEP 2 2 2006
Image /page/0/Picture/22 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters on a black background. Below that, in smaller, white letters, is the word "MEDIZINTECHNOLOGIE".
{1}------------------------------------------------
Special Premarket Notification Submission - Disposable inserts
Image /page/1/Picture/2 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters on a black background. Below that, the word "MEDIZINTECHNOLOGIE" is in smaller, black letters on a white background.
**Device and Technology Description:**
The Pajunk MIS-System contains but is not limited to electrode-inserts for comfortable and safe coagulation. PAJUNK has extended its range of instruments for MIS by a reuseable, modular system with disposeable inserts, which permits the economic replacement of the sensitive inserts.
The instrument inserts consist of four basic types: Grasping Forceps, Scissors, Biopsy Punches and Clamps /Needle Holders. All handles can be combined with all insulated shafts and instrument inserts, regardless of their size and shape.
The shaft screws onto the handle tightly, and it ensures the safe and secure connection of the two elements. Every module of the handle-instrument is kept in its bracing future, secured against distortion. A snap-on mechanism locks the instrument insert in the shaft: This whole element can then in turn be latched onto the handle by a special locking device. This latching of the connecting rod and the shaft minimizes the torsion stress exerted upon the instrument insert; the double security against distortion transfers the torque also to the entire shaft.
For the 1293- and 1292- Combination Handle an ergonomically placed switch allows the ratchet to be easily switched on or off during usage. Jacketed with glass fiber-reinforced plastic, the precise mechanical construction of the PAJUNK handle ensures maximum stability. Designed for left and right handed use, the ergonomically shaped handles are simple and comfortable to operate. The rotary wheel can be adjusted with a little pressure applied by the index finger. A built-in rotation stop secures the desired position when grasping.
And with that, PAJUNK® has created the option to always work and also to coagulate with new, guaranteed sharp scissors at every intervention. Three different sizes are available: micro, medium and large. They are packed individually, sterile, and they can be replaced very easily.
The fast and easy assembly is based on the ingenious locking mechanism. The scissor-insert is clicked into the connecting rod and tightly screwed together with the guidance tube. Please observe, that the replacement can only be carried out if the scissors are closed.Biocompatibility information for Pajunk's adaptable disposable inserts is located in Section 15.0 of this submission.
### Intended Use
Pajunks disposable inserts are instruments insulated for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures.
{2}------------------------------------------------
[062072
Special Premarket Notification Submisston - Disposable inserts
Image /page/2/Picture/2 description: The image shows the logo for PAJUNK, a company specializing in medical technology. The logo consists of the company name in bold, uppercase letters, with the word "MEDIZINTECHNOLOGIE" written in a smaller font size below it. There is a horizontal line below the text.
# Substantial Equivalence
Based on the information within this submission, it is our conclusion that Pajunk's reusable inserts marketed under 510(k) number K033249, and disposable inserts (subject device) are substantially equivalent.
Pajunk's electrodes are as safe and effective as the predicate devices when used by professional surgeons in accordance with the instructions for use supplied with these medical devices.
### Conclusion:
The comparison between the predicate device and the proposed device demonstrates that the proposed Pajunk devices are at least as safe and effective as, as well as substantially equivalent to Pajunks predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged in a circular fashion around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 2 2006
Pajunk GmbH Medizintechnologie % Christian Quass Regulatory Affairs Karl-Hall-Strasse 01 78187 Geisingen, Germany
Re: K062072
Trade/Device Name: Pajunks disposable inserts for Handle instruments and HF Electrodes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: August 29, 2006 Received: September 1, 2006
Dear Christian Quass:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Christian Quass
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Contsumer assistance at its will free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jarbare Buent
N.N. Motes
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Special Premarket Notification Schmission - Disposable inserts
Image /page/5/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in smaller, white letters. The logo appears to be for a medical technology company.
# Indications for use
510(k) Number:
Device Name:
Pajunks disposable inserts for Handle instruments and HF Electrodes
Indications for Use:
Pajunks disposable inserts are instruments insulated for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures.
Image /page/5/Picture/8 description: The image shows the words "Prescription Use" in bold font. Below that, in a smaller font, is the text "(Per 21 CFR 801.109)". The text is left-aligned and appears to be part of a document or label.
Image /page/5/Picture/9 description: The image contains a large, bold letter 'X'. The letter is black and stands out against a white background. There are short, horizontal lines extending from the bottom of each leg of the 'X', giving the impression that it is underlined or emphasized. The overall impression is simple and graphic.
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sunbare Mnchum
(Division Sign-Off) (Division of General, Restorative, and Neurological Devices
**510(k) Number** k060272
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.