K051956 · Sometech Incorporated · GEI · Sep 8, 2005 · General, Plastic Surgery
Device Facts
Record ID
K051956
Device Name
DR. OPPEL ST-501
Applicant
Sometech Incorporated
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 8, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Dr. OPPEL ST-501 is intended for the removal and destruction of skin legions and the coagulation of tissue. Non-sterile and reusable electrodes are used in conjunction with an electrosurgical handpiece and generator.
Device Story
Dr. OPPEL ST-501 is an electrosurgical unit (ESU) providing high-frequency RF energy for radiosurgery. System components include a generator, handpiece, electrodes (needle, ball, loop, blade, bipolar forceps), patient plate cable, and foot switch. Operator selects power level and time via front panel push buttons; status feedback provided by lamps. Output power is activated via foot or hand switches. Device operates in monopolar or bipolar modes to cut or coagulate tissue. Used in clinical settings by healthcare professionals for dermatological and surgical procedures. Benefits include precise tissue destruction and hemostasis.
Indicated for removal and destruction of skin lesions and coagulation of tissue in patients requiring electrosurgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Aaron 900 High Frequency Desiccator/Handpiece (K000961)
{0}------------------------------------------------
## 510(k) Summary
051956
0009
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: JUNE 7, 2005
l. Submitter:
| | Company |
|---------|---------------------------------------------------------------------------------|
| Name | SOMETECH INCORPORATED |
| Address | Byucksan Digital Valley III 212-13<br>Guro-dong Guro-gu<br>Seoul, 152-050 Korea |
| Phone | +82 2 547-3422 |
| Contact | Young-Soo Seol |
#### 2. Device:
Proprietary Name - Dr. OPPEL ST-501 Common Name - Electrosurgical unit and accessories
Classification Name - Electrosurgical cutting and coagulation device and Accessories
#### 3. Predicate Device:
Aaron 900 High Frequency Desiccator/Handpiece, Aaron Medical Industries, K000961 Aaron Reusable Electrosurgical Electrode, Aaron Medical Industries, K014201 Surgitron 120 IEC, Ellman International, Inc., K013255
## 4. Classifications & Citations:
21CFR 878.4400, GEI Class 2 510(k) Guidance Document for General Surgical Electrosurgical Devices
#### 5. Description:
Dr. OPPEL ST-501 is a electrosurgical unit using high frequency current to cut and/or coagulate tissue. It consists of Electrosurgical Unit generator, Handpiece, Electrodes, a cable for disposable patient plate and foot switch. It is a compact source of high power RF energy to be employed for a variety of radiosurgery procedures. This action is action is achieved by front panel selection of power level and time. All selection is effected through push buttons and lamps that give the operator feedback of status. The final output power control is made through foot and/or hand switches.
Both Monopolar and Bipolar electrodes are provided. Monopolar electrodes have needle, ball, loop and blade types. Electrodes are supplied non-sterile and reusable.
- 6. Indication for use:
Dr. OPPEL ST-501 is intended for the removal and destruction of skin legions and the coagulation of tissue. Non-sterile and reusable electrodes are used in conjunction with an electrosurgical handpiece and generator.
. ﻳ
SOMETECH INCORPORATED
好:
{1}------------------------------------------------
510(k) Submission -- Dr. OPPEL ST-501
# 7. Technological characteristics
| Feature | Dr. OPPEL ST-501 | Aaron 900 High Frequency Desiccator/ Handpiece | Aaron Reusable Electrosurgical Electrode | Surgitron 120 IEC |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | Dr. OPPEL ST-501 is intended for the removal and destruction of skin legions and the coagulation of tissue. Non-sterile and reusable electrodes are used in conjunction with an electrosurgical handpiece and generator. | The Aaron 900 High Frequency Desiccator and Handpiece are intended for the removal and destruction of skin legions and the coagulation of tissue. | Aaron Reusable Electrosurgical Electrodes are non-sterile, the reusable electrosurgical electrodes, used in conjunction with an electrosurgical handpiece and generator to deliver RF energy used to cut and excise tissue or to coagulate blood vessels during surgery. | 1. Cutting - Snoring, Submucosal palatal shrinkage, etc.<br>2. Blended Cutting and Coagulation - Snoring, Submucosal palatal shrinkage, etc.<br>3. Hemostasis - Control of Bleeding, Epilation, etc.<br>4. Fulguration - Basa Cell Carcinoma, Papilloma, etc.<br>5. Bipolar - Pinpoint, Precise Coagulation, etc. |
| Design<br>specification | IEC601-2-2<br>ANSI/AAMI HF-18 | IEC601-2-2<br>UL2601-1 | N/A | IEC601-2-2<br>UL544 |
| Output<br>energy | 100 Watt at 200Ω | 30 Watt | N/A | 120 Watt |
| Output<br>waveform | 4.0 MHz Partially rectified | 550 kHz Partially rectified | N/A | 4.0 MHz Continuous, Fully rectified, Partially rectified, 1.7MHz fulguration |
| Delivery<br>system | Monopolar and Bipolar | Monopolar and Bipolar | N/A | Monopolar and Bipolar |
| Tip<br>configuration | Needle<br>Blade<br>Ball<br>Loop<br>Bipolar Forceps | N/A | Needle<br>Blade<br>Ball<br>Loop<br>Bipolar Forceps | Needle<br>Blade<br>Ball<br>Loop<br>Bipolar Forceps |
| Sterilization<br>method | Non-sterile | N/A | Non-sterile | Non-sterile |
| Standard | IEC601-1<br>IEC601-1-2<br>IEC601-2-2<br>ANSI/AAMI HF18 | UL2601-1<br>EN60601-1-2<br>IEC601-2-2 | N/A | UL2601<br>IEC601-1<br>IEC601-2-2 |
SOMETECH INCORPORATED ":
い
{2}------------------------------------------------
051956
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
- 8. Similarities to, and differences from, the above predicate devices:
、 :
4 :
Dr. OPPEL ST-501 is similar to the above predicate devices based on the intended use, Dr. OPPEL ST-501 1s Sinitial to the "alord" processor "processor of the mance characteristics.
technology used, the claims, the material composition employed and personal wou technology used, the claims, the material composition employ and output waveform.
#### 9. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, FDA's In accordance with the Techar 1000, Brug and Sectrosurgical Devices" and based on the "STU(K) Guldance Document for General Surgical Slouveaught - Previous Chat Dr.
information provided in this premarket notification Sometech Incorporated concludes desiress as information provided in this premarket notheation weeks. Interprises of the predicate devices as described herein.
END
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and a circular border containing the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA". The eagle is positioned in the center of the circle, with its head facing to the right.
SEP - 8 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Sometech Incorporation c/o Mr. Marc M. Mouser Office Coordinator Underwriters Laboratories, Inc. Laboratory and Testing 2600 N.W. Lake Road Camas, Washington 98607-8542
Re: K051956
Trade/Device Name: Dr. OPPEL ST-501 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 22, 2005 Received: August 25, 2005
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
### Page 2- Mr. Marc M. Mouser
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR-Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Neil R. Ogden
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number K _051956
Dr. OPPEL ST-501 Device Name: :::
Indication for use: Dr. OPPEL ST-501 is intended for the removal and destruction of skin legions and the coagulation of tissue. Non-sterile and reusable electrodes are used in conjunction with an electrosurgical handpiece and generator.
Prescription Use AND/OR (Part 21CFR801 Subpart D)
Over-The-Counter Use (21CFR801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
NilRedglen
(Division Sign-Off) Sw.m.416
Division of General, Restorative. and Neurological Devices
510(k) Number K051956
.. •
. ﻳﺘﻢ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ
Indication for use Page # 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.