K032838 · Celon AG Medical Instruments · GEI · Sep 26, 2003 · General, Plastic Surgery
Device Facts
Record ID
K032838
Device Name
CELON ENT SYSTEM
Applicant
Celon AG Medical Instruments
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 26, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
This application system, comprising the CelonLab ENT power control unit and the CelonProBreath, CelonProSleep, and CelonProSleep PLUS bipolar coagulation electrodes, is indicated for ablation and coagulation of soft tissue in otorhinolaryngology (ENT) surgery including: CelonProBreath: Submucosal tissue shrinkage for nasal airway obstruction by reduction of hypertrophic nasal turbinates; CelonProSleep: Submucosal tissue shrinkage and tissue coagulation in the uvula/soft palate for the treatment of snoring; CelonProSleep PLUS: Submucosal tissue shrinkage and tissue coagulation in the uvula/soft palate for the treatment of snoring. The system is intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
Device Story
Celon ENT System consists of CelonLab ENT power control unit and bipolar coagulation electrodes (CelonProBreath, CelonProSleep, CelonProSleep PLUS). Device delivers bipolar RF current to soft tissue; no neutral electrode required. Operator selects power level via unit display and activates via foot switch. System provides acoustic feedback proportional to tissue impedance; frequency changes indicate coagulation status. Automatic power control monitors impedance; if limit reached (coagulation complete), system emits pulsed tone and ceases output to prevent tissue overdose. Procedure time varies (3 seconds to minutes) based on applicator, tissue, and power. Used in clinical ENT settings by trained medical personnel. Benefits include controlled tissue shrinkage/coagulation for nasal obstruction and snoring treatment.
Clinical Evidence
Bench testing only. Performance validation testing conducted to confirm device functionality and safety. Materials verified for biocompatibility per ISO 10993 and USP 23 Class VI. Electrical safety and electromagnetic compatibility verified per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and ANSI/AAMI HF-18.
Technological Characteristics
Bipolar electrosurgical coagulation system. Components: CelonLab ENT power control unit, bipolar electrodes (CelonProBreath, CelonProSleep, CelonProSleep PLUS), CelonFootSwitch. Materials: Biocompatible per ISO 10993/USP 23 Class VI. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, ANSI/AAMI HF-18. Features: Automatic power control, impedance-based acoustic monitoring, automatic output cessation. Connectivity: Standalone unit.
Indications for Use
Indicated for ENT surgery patients requiring soft tissue ablation and coagulation, specifically for submucosal tissue shrinkage of hypertrophic nasal turbinates (nasal airway obstruction) or uvula/soft palate tissue (snoring treatment). Intended for use by trained medical personnel.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
### SEP 2 6 2003
# K032838 (pg 1 of 4)
### 510(k) Summary of Safety and Effectiveness
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.92
| Submitter: | Celon AG medical instruments<br>Rheinstrasse 8<br>14513 Teltow<br>Germany |
|------------------------|-------------------------------------------------------------------------------------|
| Contact Person: | André Roggan, Ph. D.<br>Director Research and Development |
| | Phone: +49-3328-3519-0<br>Fax: +49-3328-3519-23 |
| United States Agent: | Donald R. Bruce<br>1005 California Ave.<br>San Jose, CA 95125 |
| | Phone: 408-691-6333<br>Fax: 408-294-5258 |
| Date Summary Prepared: | June 27th, 2003 |
| Product Identification | |
| Proprietary Name: | Celon ENT System |
| Classification Name: | Electrosurgical cutting and coagulation<br>device and accessories (21 CFR 878.4400) |
| Common Name: | Electrosurgical Device and Accessories |
| Manufacturer: | Celon AG medical instrument<br>Rheinstrasse 8<br>14513 Teltow<br>Germany |
| Distributor: | pending |
{1}------------------------------------------------
$$\text{אַנון געון געון אַזאַז אַז אַז}$$
#### Legally Marketed Device:
The Celon ENT System is of comparable type and is substantially equivalent to the following currently marketed devices:
| Device | 510(k) Number | Date Cleared |
|-----------------------------------------|---------------|---------------------|
| ArthroCare AccENT Electrosurgery System | K973478 | January 9th, 1998 |
| ArthroCare ENTec Reflex Wand Electrodes | K000036 | February 4th, 2000 |
| ArthroCare ENTec Reflex Wand Electrodes | K000778 | May 3rd, 2000 |
| Somnus Somnoplasty System | K971450 | July 17th, 1997 |
| Somnus Somnoplasty System | K973618 | December 19th, 1997 |
It complies with the same or equivalent standards and has the same intended use.
#### Device Description
The Celon ENT System is comprised of a power control unit (CelonLab EN7) and bipolar coagulation electrodes. The CelonLab ENT power control unit delivers a bipolar output via the bipolar coagulation electrodes of type CelonProBreath, CelonProSleep, and CelonProSleep PLUS. A neutral electrode (return conductor) is not required. The power control unit is user-friendly with a user display showing all the necessary process parameters. After switching on the equipment, the user selects the level of power and activates the radio frequency (RF) current by depressing the foot switch. The power control unit delivers an acoustic signal as an indicator of the coaqulation status. Usually, general electrosurgical units have a fixed activation tone. However, in the case of the CelonLab ENT power control unit, the frequency of the acoustic signal is proportional to the tissue impedance. This permits acoustic monitoring of the coagulation status by the operator, since the latter is directly correlated with the impedance. If the impedance increases above a certain limit, signifying that the coagulation process is complete, this is additionally indicated by a clock pulsed sound and the power output is ceased automatically. Overdose effects in the treated tissue area are excluded by this power control function. The procedure time will be set automatically by the power control unit. The time interval between activation of the power control unit and the end of the coagulation procedure depends on applicator type, tissue properties and power setting, and can vary between 3 seconds and several minutes.
Accessories included with the power control unit include a line power cable and single pedal foot pedal (CelonFootSwitch).
{2}------------------------------------------------
## K032838 (pg 3 of 4)
Materials: Materials and construction of the Celon ENT System are compliant with the international electrical safety standards IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and the American national standard ANSI/AAMI HF-18. Additionally, the materials used in the design and construction of the CelonProBreath, CelonProSleep, and CelonProSleep PLUS bipolar coagulation electrodes, which have direct contact with the patient, are biocompatible according to ISO 10993 or USP 23 Class VI.
Design: The system is designed to be a bipolar electrosurgical coagulation device with an automatic power control, including automatic output power reduction and acoustic monitoring of coagulation status.
#### Intended Use of the Device
This Celon ENT System has the same intended use as the legally marketed devices. This application system, comprising the CelonLab ENT power control unit and the CelonProBreath, CelonProSleep, and CelonProSleep PLUS bipolar coagulation electrodes, is indicated for ablation and coagulation of soft tissue in otorhinolaryngology (ENT) surgery including:
| CelonProBreath: | Submucosal tissue shrinkage for nasal airway obstruction by reduction of hypertrophic nasal turbinates |
|---------------------|----------------------------------------------------------------------------------------------------------|
| CelonProSleep: | Submucosal tissue shrinkage and tissue coagulation in the uvula/soft palate for the treatment of snoring |
| CelonProSleep PLUS: | Submucosal tissue shrinkage and tissue coagulation in the uvula/soft palate for the treatment of snoring |
The system is intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
{3}------------------------------------------------
# K032838
#### Comparison with Legally Marketed Device:
It is the opinion of Celon AG medical instruments that the Celon ENT System is of a type and substantially equivalent to legally marketed devices with respect to intended use and technological characteristics. It will comply with the safety requirements of IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, ANSI/AAMI HF-18 and applicable collateral and particular standards. Furthermore, the CelonProBreath, CelonProSleep, and CelonProSleep PLUS bipolar coagulation electrodes, which have direct contact with the patient, are biocompatible according to ISO 10993 or USP 23 Class VI. In addition, performance validation testing has been done to validate the performance of the device. The comparison results presented in this 510(k) notification show that the device is substantially equivalent to legally marketed devices and is safe and effective in its intended use.
### Conclusions
The Celon ENT System does not result in any new potential safety risks and performs as well as or better than devices legally on the market. Celon AG medical instruments consider the Celon ENT System to be equivalent to other marketed devices with the same indications for use and meeting similar standards.
#### Legal Notice
CelonProBreath® and CelonProSleep® are registered trademarks of Celon AG medical instruments.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with its wings spread, composed of three thick, curved lines.
Public Health Service
SEP 2 6 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Celon AG Medical Instruments c/o Ms. Pamela Gwynn Underwriters Laboratories, Inc. 12 Laboratory Drive P.O. Box 13995 Research Triangle Park, North Carolina 27709-3995
Re: K032838
Trade/Device Name: Celon ENT System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 18, 2003 Received: September 11, 2003
Dear Ms. Gwynn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Pamela Gwynn
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Statement of Intended Use 6
| 510(k) Number: | pending KC32838 |
|----------------|-----------------|
|----------------|-----------------|
Device Name:
Celon ENT System
Indications For Use:
This application system, comprising the CelonLab ENT power control unit and the CelonProBreath, CelonProSleep, and CelonProSleep PLUS bipolar coagulation electrodes, is indicated for ablation and coagulation of soft tissue in otorhinolaryngology (ENT) surgery including:
| CelonProBreath: | Submucosal tissue shrinkage for nasal airway obstruction by reduction of hypertrophic nasal turbinates |
|---------------------|----------------------------------------------------------------------------------------------------------|
| CelonProSleep: | Submucosal tissue shrinkage and tissue coagulation in the uvula/soft palate for the treatment of snoring |
| CelonProSleep PLUS: | Submucosal tissue shrinkage and tissue coagulation in the uvula/soft palate for the treatment of snoring |
The system is intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X |
|-------------------------|---|
| OR Over-The-Counter Use | |
(Per 21 CFR 801.109)
Miriam C. Provost
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
| 510(k) Number | K032838 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.