LIGASURE 5MM LAPAROSCOPIC SEALER-DIVIDER, MODEL LS1500
K031011 · Valleylab, Inc. · GEI · May 29, 2003 · General, Plastic Surgery
Device Facts
Record ID
K031011
Device Name
LIGASURE 5MM LAPAROSCOPIC SEALER-DIVIDER, MODEL LS1500
Applicant
Valleylab, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
May 29, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LigaSure™ 5mm Vessel Sealer-Divider is a bipolar electrosurgical instrument intended for use with the LigaSure Generator in general and gynecologic laparoscopic surgical procedures where ligation and division of vessels is desired. The instrument creates a seal by the application of RF electrosurgical energy to vascular structures (vessels) interposed between the jaws of the instrument. A blade within the instrument is actuated to divide tissue. Indications for use include general laparoscopic surgical procedures including urologic, vascular, thoracic and thoracoscopic, and gynecologic laparoscopic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectomy, Nissen fundoplication, colectorny, adhesiolysis, oophorectomy, etc. The LigaSure 5mm Vessel Sealer-Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures. The LigaSure 5mm Vessel Sealer-Divider can be used on vessels up to and including 7mm diameter, and tissue bundles as large as will fit in the jaws of the instrument.
Device Story
Bipolar electrosurgical instrument for laparoscopic surgery; used with LigaSure Vessel Sealing Generator. Input: vascular structures/tissue bundles clamped between jaws. Operation: applies RF energy to seal vessels; surgeon-actuated blade divides tissue. Output: sealed and divided vessels/tissue. Used in OR by surgeons. Smart connector identifies instrument to generator. Benefits: enables vessel ligation and division in minimally invasive procedures. Single-use, sterile device.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by performance testing and pre-clinical studies ensuring device functions as intended and meets design specifications.
Technological Characteristics
Bipolar electrosurgical instrument; 5mm shaft diameter; 37cm working length. Features 'smart' connector for generator identification. Single-use, sterile. RF energy source. Manual surgeon-actuated blade for tissue division.
Indications for Use
Indicated for general, urologic, vascular, thoracic, thoracoscopic, and gynecologic laparoscopic procedures requiring vessel ligation and division (e.g., hysterectomy, fundoplication, colectomy, adhesiolysis, oophorectomy). Contraindicated for tubal sterilization or tubal coagulation. Suitable for vessels up to 7mm diameter and tissue bundles fitting within instrument jaws.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Valleylab LS1000 LigaSure™ LAP Laparoscopic Instrument (K981916)
{0}------------------------------------------------
# KO31011
## SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
Valleylab LigaSure™ 5mm Laparoscopic Sealer-Divider
#### Submitter Information 1.
Valleylab, Inc. 5920 Longbow Drive Boulder, CO 80301 Contact: Herbert Vinson Telephone: 303-530-6469
Date summary prepared: March 11, 2003
#### 2. Name of Device
Trade or Proprietary Name: LigaSure™ 5mm Laparoscopic Sealer-Divider
Common Name: Bipolar Laparoscopic Electrosurgical Instrument
Classification Name:
- Electrosurqical Cutting and Coagulation Device and Accessories, and ●
- Gynecologic Electrocautery and Accessories .
#### 3. Predicate Devices
The Valleylab LigaSure™ 5mm Laparoscopic Sealer-Divider is substantially equivalent to the Vallevlab LS1000 LigaSure™ LAP Laparoscopic Instrument (K981916), and the Valleylab LS1100 LigaSure ATLAS™ Laparoscopic Sealer-Divider Instrument (K010013). All three of these devices are used in laparoscopic surgery to seal vessels by the application of RF enerqy to the vessels and tissues interposed between the jaws of the instrument. The LigaSure™ 5mm Laparoscopic Sealer-Divider and the LS1100 LigaSure ATLAS™ Laparoscopic Sealer-Divider Instrument divide tissue using a surgeonactuated blade.
### 4. Device Description
The LigaSure 5mm Laparoscopic Sealer-Divider is a multi-functional electrosurgical instrument for use with the LigaSure Vessel Sealing Generator (K981916) when performing laparoscopic surgery. The instrument is capable of sealing vessels, dividing vessels and tissue clamped between its jaws, grasping tissue, and blunt dissection. The outer diameter of the instrument shaft is 5mm,
{1}------------------------------------------------
with a working length of 37 cm. Controls are located on the instrument handle. All controls can be operated with either the right or left hand.
The instrument attaches to the generator with a "smart" connector that identifies the instrument type to the LigaSure generator, and a ten (10) foot cable. The instrument is supplied sterile for single-use.
#### 5. Intended Use
The LigaSure™ 5mm Vessel Sealer-Divider is a bipolar electrosurgical instrument intended for use with the LigaSure Generator in general and gynecologic laparoscopic surgical procedures where ligation and division of vessels is desired. The instrument creates a seal by the application of RF electrosurgical energy to vascular structures (vessels) interposed between the jaws of the instrument. A blade within the instrument is actuated to divide tissue.
Indications for use include general laparoscopic surgical procedures including urologic, vascular, thoracic and thoracoscopic, and gynecologic laparoscopic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectomy, Nissen fundoplication, colectorny, adhesiolysis, oophorectomy, etc. The LigaSure 5mm Vessel Sealer-Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
The LigaSure 5mm Vessel Sealer-Divider can be used on vessels up to and including 7mm diameter, and tissue bundles as large as will fit in the jaws of the instrument.
#### 6. Summary of Technological Characteristics
The LigaSure™ 5mm Laparoscopic Sealer-Divider has the same basic technological characteristics as the predicate devices noted above.
#### 7. Performance Data
Performance testing and pre-clinical studies were performed to ensure that the LigaSure™ 5mm Laparoscopic Sealer-Divider functions as intended, and meets design specifications. Sufficient data were obtained to show that the device is substantially equivalent to the predicate devices, and meets safety and effectiveness criteria.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 9 2003
Mr. Herbert Vinson Senior Regulatory Associate Valleylab. Inc. 5920 Longbow Drive Boulder, Colorado 80301
Re: K031011
Trade/Device Name: LigaSure™ 5mm Laparoscopic Sealer-Divider Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: March 11, 2003 Received: March 31, 2003
Dear Mr. Vinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Herbert Vinson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Muriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known): KO31011
Device Name: LigaSure™ 5mm Laparoscopic Sealer-Divider
Indications For Use:
The LigaSure™ 5mm Vessel Sealer-Divider is a bipolar electrosurgical instrument intended for use with the LigaSure Generator in general and gynecologic laparoscopic surgical procedures where ligation and division of vessels is desired. The instrument creates a seal by application of RF electrosurgical energy to vascular structures (vessels) interposed between the jaws of the instrument. A blade within the instrument is surgeonactuated to divide tissue.
Indications for use include general laparoscopic procedures including urologic, vascular, thoracic and thoracoscopic, and gynecologic laparoscopic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectorny, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy, etc. The LigaSure Vessel Sealing System has not been shown to be effective for tubal sterilization or tubal coaqulation for sterilization procedures, and should not be used for these procedures.
The LigaSure 5mm Vessel Sealer-Divider can be used on vessels up to and including 7mm diameter, and tissue bundles as large as will fit in the jaws of the instrument.
> PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Musiam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
Prescription Use \ (Per 21 CFR 801.109)
510(k) Number K03/011
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.