OLYMPUS PSD-20 ELECTROSURGICAL SYSTEM & ITS ASSOCIATED ACCESSORIES
K021852 · Olympus Optical Co., Ltd. · GEI · Feb 11, 2003 · General, Plastic Surgery
Device Facts
Record ID
K021852
Device Name
OLYMPUS PSD-20 ELECTROSURGICAL SYSTEM & ITS ASSOCIATED ACCESSORIES
Applicant
Olympus Optical Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 11, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This instrument has been designed to be used with Olympus recommended Electrosurgical accessories, Endoscope, Light Source and other ancillary equipment for bronchial endoscopic treatment (cutting and coagulating) including: Hemostasis of superficial bleeding - Treatment of benign tumors like papillomatosis, granulomas, polyps, lipomas and hemangiomas in the trachea and bronchi - Recanalization of malignant stenoses - Treatment of Cicatricial stenoses of the respiratory tract
Device Story
The Olympus PSD-20 Electrosurgical System is an electrosurgical unit used in bronchial endoscopic procedures. It delivers high-frequency electrical current to compatible accessories (e.g., snares, hot biopsy forceps) to perform tissue cutting and coagulation. The system provides five monopolar cut modes (PURE, BLEND 1/2/3/4) and one coagulation mode (NORMAL COAGULATION). It is operated by a physician in a clinical setting. The device includes safety features such as voltage output level settings and a monitor circuit to detect irregularities or improper connections. The output allows clinicians to manage bleeding, remove benign tumors, and treat stenoses in the respiratory tract, potentially improving patient outcomes by enabling minimally invasive endoscopic interventions.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, technological characteristics, and adherence to voluntary safety standards (IEC 60601 series).
Technological Characteristics
Electrosurgical unit providing high-frequency electrical current for cutting and coagulation. Features monopolar output with 5 cut modes and 1 coagulation mode. Includes voltage output monitoring and connection detection circuits. Complies with IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18, and IEC 60601-1-2 (EMC). No patient-contacting materials in the unit itself.
Indications for Use
Indicated for bronchial endoscopic treatment (cutting and coagulating) in patients requiring hemostasis of superficial bleeding, treatment of benign tumors (papillomatosis, granulomas, polyps, lipomas, hemangiomas) in the trachea and bronchi, recanalization of malignant stenoses, or treatment of cicatricial stenoses of the respiratory tract.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Olympus PSD-20 Electrosurgical Unit and its associated accessories (For GI application) (K970797)
Submission Summary (Full Text)
{1}------------------------------------------------
#### 510(k) Summary of Safety and Effectiveness SMDA
# KO21852
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807 subpart E, Section 807.92.
#### A. Sponsor /Manufacturer Name and Address
- 1. Applicant Olympus Optical Co., Ltd. 2-3-1 Shinjuku Monolis Nishi-shinjuku Shinjuku-ku, Tokyo, Japan, 163-0914 Establishment registration number : 8010047
- Initial Importer 2. Olympus America Inc. Two Corporate Center Drive Melville, NY 11747-3157 E stablishment registration number : 2429304
- 3. Contact Person Laura Storms-Tyler Director, Regulatory Affairs Olympus America Inc. Two Corporate Center Drive Melville, NY 11747-3157
- Tel 631-844-5688 Fax 631-844-5416
#### B. Device Name, Common Name
- 1. Common/Usual Name
Electrosurgical Unit and its ancillary equipment.
- 2. Device Name
- Olympus PSD-20 Electrosurgical System and its associated accessories
- 3. Classification
21 CFR 876.4300 Endoscopic electrosurgical unit and accessories, Class II
#### C. Identification of the predicate or legally marketed device
The following devices are substantially equivalent devices in consideration to its characteristics or specifications.
| Model Name | 510(k) | Applicant |
|------------------------------------------------------------------------------------------------|---------|---------------------------|
| Olympus PSD-20 Electrosurgical<br>Unit and its associated accessories<br>(For GI application ) | K970797 | Olympus Optical Co., Ltd. |
#### D. Device Description
- 1. Summary
{2}------------------------------------------------
This instrument has been designed to be used with Olympus recommended Electrosurgical accessories, Endoscope, Light Source and other ancillary equipment for bronchial endoscopic treatment (cutting and coagulating.
Recommended endoscopes for this instrument are Olympus's series BF series, and Olympus's electrosurgical instruments such as electrosurgical snare, hot biopsy forceps. Other Olympus electrosurgical devices or instruments can be used with this subject device. Be sure to check compatibility for each device in each instrument's instruction manual. The PSD-20 unit offers Monopolar output modes; 5 Cut modes (PURE, BLEND 1/2/3/4), 1 Coagulation mode. (NORMAL COAGULATION), It offers several features to ensure the safe operation of the unit. For example, the voltage output level setting and a monitor circuit detect irreqularity or improper connections.
### 2. Design
This device has been designed to be comply with the following voluntary standards.
- · IEC 60601-1
- · IEC 60601-2-2
- · IEC 60601-2-18
- · IEC 60601-1-2 (EMC)
#### 3. Materials
There aren't any patient contacting material in PSD-20 itself. Some ancillary equipment have patient contact materials, however there are no new patient contacting material in those devices.
### 4. Technology
This device does not have any special technology or characteristic i.e. this device is intended to provide electrical power to accessories (such as biopsy) via the use of high frequency electrical current waveform passing into the tissue during endoscopic bronchial treatment.
#### E. Indication for Use
This instrument has been designed to be used with Olympus recommended Electrosurgical accessories, Endoscope, Light Source and other ancillary equipment for bronchial endoscopic treatment (cutting and coagulating) including:
- Hemostasis of superficial bleeding
- Treatment of benign turnors like papillomatosis, granulomas, polyps,
lipomas and hemangiornas in the trachea and bronchi
- Recanalization of malignant stenoses
- Treatment of Cicatricial stenoses of the respiratory tract
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# FEB 1 1 2003
Olympus Optical Co., Ltd. c/o Ms. Laura Storms-Tyler Director, Regulatory Affairs Olympus America, Inc. Two Corporate Center Drive Melville, New York 11747-3157
Re: K021852
Trade/Device Name: Olympus PSD-20 Electrosurgical System and Its Associated Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: November 12, 2002 Received: November 13, 2002
Dear Ms. Storms-Tyler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (P.MA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - o Ms. Laura Storms-Tyler
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Muriam C. Provost
Felia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. The font is sans-serif and slightly distressed, giving it a textured appearance. Below the word, there is a line of smaller, illegible text that appears to be a description or tagline associated with the brand name.
## Indication for Use Statement
KO21852 510(k) Number(if known): Net assigned yet. Device Name: Olympus PSD-20 Electrosurgical System and its associated accessories
#### Indications for Use :
This instrument has been designed to be used with Olympus recommended Electrosurgical accessories, Endoscope, Light Source and other ancillary equipment for bronchial endoscopic treatment (cutting and coagulating) including:
- Hemostasis of superficial bleeding -
- Treatment of benign tumors like papillomatosis, granulomas, polyps, lipomas and hemangiomas in the trachea and bronchi
- Recanalization of malignant stenoses -
- Treatment of Cicatricial stenoses of the respiratory tract ﯩ
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-The-Counter Use (Prescription 21 CFR 801.109) (Optional Format 1-2-96) (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Numb
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.