K160994 · C2 Therapeutics, Inc. · GEH · May 19, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160994
Device Name
C2 CryoBalloon Focal Pear Catheter
Applicant
C2 Therapeutics, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
May 19, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The C2 CryoBalloon™ Focal Ablation System is intended to be used as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications.
Device Story
The C2 CryoBalloon Focal Pear Catheter is a component of a cryosurgical system used for tissue ablation. The system includes a sterile, single-use catheter, a reusable controller, and a single-use nitrous oxide cartridge. The device is inserted through the working channel of an endoscope (min 3.7 mm ID). The physician visualizes the target tissue through the catheter's distal balloon, which is inflated with nitrous oxide. The controller, operated by the physician, triggers the release of nitrous oxide to cool the balloon, freezing and destroying the unwanted tissue. The nitrous oxide is exhausted through the controller. The device is used in clinical settings for endoscopic procedures to treat unwanted tissue, providing a minimally invasive method for cryoablation.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included design verification for physical, reliability, and safety specifications, specifically dimensional verification, balloon integrity, balloon fatigue, and bond/connector tensile strength.
Technological Characteristics
Cryosurgical unit using nitrous oxide-cooled balloon. System components: single-use catheter, reusable controller, single-use cartridge. Compatible with endoscopes (min 3.7 mm working channel). Features user-controlled trigger for cryogen release. Battery-operated controller with software. Materials include modified tubing. Sterilization: sterile catheter.
Indications for Use
Indicated for destruction of unwanted tissue in general surgery via endoscopic application.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines extending from the heads.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2016
C2 Therapeutics, Inc. Ms. Theresa Brandner-Allen VP of Regulatory Affairs and Quality Assurance 303 Convention Way, Suite 1 Redwood City, California 94063
Re: K160994
Trade/Device Name: C2 Cryoballoon Focal Pear Catheter Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: May 10, 2016 Received: May 11, 2016
Dear Ms. Brandner-Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely yours,
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160994
Device Name
C2 CryoBalloon™ Focal Ablation System
### Indications for Use (Describe)
The C2 CryoBalloon™ Focal Ablation System is intended to be used as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg fill="none" height="16" style="vertical-align: middle;" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.2929 3.29289C13.6834 2.90237 14.3166 2.90237 14.7071 3.29289C15.0976 3.68342 15.0976 4.31658 14.7071 4.70711L5 14.4142C4.60948 14.8047 3.97631 14.8047 3.58579 14.4142L1.29289 12.1213C0.902369 11.7308 0.902369 11.0976 1.29289 10.7071C1.68342 10.3166 2.31658 10.3166 2.70711 10.7071L4.29289 12.2929L13.2929 3.29289Z" fill="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" style="vertical-align: middle;" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="2.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5. 510(k) Summary
## K160994
## Special 510(k)
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
#### I. SUBMITTER
C2 Therapeutics, Inc. 303 Convention Way, Suite 1 Redwood City, CA 94063
Phone: 650-521-5921 Fax: 650-556-1145
| Contact Person: | Theresa Brandner-Allen<br>VP of Regulatory Affairs and Quality Assurance |
|-----------------|--------------------------------------------------------------------------|
| Date Prepared: | May 9, 2016 |
#### II. DEVICE
Name of Device: C2 CryoBalloon™ Focal Pear Catheter Common Name: Cryosurgical Unit, Cryogenic Surgical Device Classification Name: Cryosurgical Unit, Cryogenic Surgical Device 21 CFR §878.4350(a)(2) Regulatory Class: Class II
### Product Code: GEH
#### III. PREDICATE DEVICE
CryoBalloon Focal Ablation System, K131523
This predicate device has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The subject device is the Catheter used as part of a cryosurgical unit with a nitrous oxide cooled balloon that is compatible with commercially available endoscopes with a minimum working channel inner diameter (ID) of 3.7 mm and maximum length of 100 cm. The subject Catheter is part of a cryosurgical system comprised of three components including a Catheter (sterile, single use), Controller (non-sterile, reusable), and Cartridge (non-sterile, single use).
{4}------------------------------------------------
The subject device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter is comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the pre-inflated balloon, and the treatment site is selected is by adjusting the endoscope and Controller. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.
#### V. INDICATIONS FOR USE
The C2 CryoBalloon™ Focal Ablation System is intended to be used as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery. specifically for endoscopic applications.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Cryoablation is the fundamental technological principle for both the subject C2 CryoBalloon™ Focal Ablation System and the predicate CryoBalloon Focal Ablation System. Both the subject device and predicate device are based on the same endoscopic instrumentation for removing unwanted tissue using extreme cold.
The subject C2 CryoBalloon™ Focal Ablation System has similar technological characteristics to the legally marketed predicate. The subject device and predicate device are based on the following same technological elements:
- Inserted through an endoscope to access the treatment site
- Apply a cryogen to ablate (freeze) the unwanted tissue ●
- Use of a compliant balloon to position the treatment diffuser and to contain ● and exhaust the cryogen
- . User-controlled (trigger) activates release of the cryogen
- Battery-operated Controller and software
The only modifications that were made include:
- The Catheter was modified to a different shaped balloon.
- The distal Catheter tip was modified to reduce the length. ●
- . The proximal Catheter connector cryogen inflow tubing was modified to change the tubing material.
- . The product name was changed slightly.
#### PERFORMANCE DATA VII.
Performance data were provided in support of the substantial equivalence determination. Design verification testing, based on results of risk analysis, was performed on the C2 CryoBalloon™ Focal Ablation System to evaluate the physical, reliability, and safety specifications. The testing performed includes dimensional verification, balloon integrity, balloon fatigue, and bond and connector tensile strength and integrity.
{5}------------------------------------------------
#### CONCLUSION VIII.
The subject C2 CryoBalloon™ Focal Ablation System has the same clinical attributes, technological characteristics, and expected performance as the legally marketed predicate, CryoBalloon Focal Ablation System (K131523). The performance data results demonstrate that the subject C2 CryoBalloon™ Focal Ablation System should perform as intended in the specified use conditions and should perform comparably to the legally marketed predicate that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.