K210278 · Standard Bariatrics · GDW · Apr 28, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210278
Device Name
Titan SGS
Applicant
Standard Bariatrics
Product Code
GDW · General, Plastic Surgery
Decision Date
Apr 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended for transection and resection of gastric tissue.
Device Story
Titan SGS is a single-use, sterile linear cutter for longitudinal transection of gastric tissue during sleeve gastrectomy. Device consists of a stapler (handle, shaft, end effector with 342 preloaded titanium staples) and a cable connecting to a reusable, wall-powered Power Supply Unit. Operated by surgeons in an OR setting; device is inserted laparoscopically via trocar. Surgeon uses trigger to open/close jaws and fire; mechanical linkage drives knife to transect tissue while staples are formed into 'B' shapes in 6 staggered rows. Output is a stapled, transected gastric pouch. Device facilitates sleeve gastrectomy by enabling consistent tissue resection and hemostasis. Benefits include standardized staple line creation and effective tissue management.
Clinical Evidence
Evidence includes a comparative study on 36 excised human stomachs and a multisite clinical study (IDE G200085) enrolling 62 subjects. Results demonstrated that the Titan SGS performs comparably to the predicate device in longitudinal gastric resection, providing reasonable assurance of safety and effectiveness.
Technological Characteristics
Single-use linear cutter; 230 mm end effector; 342 titanium staples in 6 staggered rows; closed staple height 1.2-2.2 mm. Powered by external wall-outlet DC power supply. Mechanical linkage for jaw actuation and firing. Sterilized via gamma radiation (10-6 SAL). Complies with ISO 10993-1 (biocompatibility), ISO 14971 (risk management), IEC 60601-1 (electrical safety), and ISO 11737-2 (sterility).
Indications for Use
Indicated for longitudinal transection of gastric tissue for sleeve gastrectomy pouch creation.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 28, 2021
Standard Bariatrics Alison Sathe Regulatory Affairs 4362 Glendale Milford Rd. Cincinnati, Ohio 45242
# Re: K210278
Trade/Device Name: Titan SGS Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW, GAG Dated: January 28, 2021 Received: February 1, 2021
# Dear Ms. Sathe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210278
Device Name Titan SGS
Indications for Use (Describe)
The Titan SGS linear cutter is intended for longitudinal transection of gastric tissue for sleeve gastrectory pouch creation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Standard Bariatrics. The logo consists of a stylized red letter "S" on the left and the words "STANDARD BARIATRICS" in black on the right. The letter "S" is made up of three parallel lines that curve around to form the shape of the letter. The word "STANDARD" is above the word "BARIATRICS", and there is a small trademark symbol to the right of the word "BARIATRICS".
# 510(k) Summary
#### I. Submitter's Information
| Company Name: | Standard Bariatrics, Inc. |
|-----------------|----------------------------------------------------|
| Address: | 4362 Glendale Milford Road<br>Cincinnati, OH 45242 |
| Phone Number: | 513-658-0328 |
| Fax Number: | 513-436-0201 |
| Contact Person: | Alison Sathe |
| Phone Number: | 513-304-7971 |
| Email Address: | alison@regulatorymark.com |
| Date Prepared: | January 28, 2021 |
#### II. Device Information
| Device Name: | Titan SGS |
|-------------------|---------------------|
| Common Name: | Staple, Implantable |
| Regulatory Class: | Class II |
| Regulation: | 21 CFR 878.4750 |
| Product Code: | GDW |
#### III. Predicate Device:
Echelon Flex Power Plus, K140560, 21CFR 878.4750, Class II, Product Code GDW Ethicon Endo-Surgery, LLC
#### IV. Device Description
The Titan SGS with implantable staples (Titan Stapler) is a single patient use, sterile instrument used for cutting and stapling gastric tissue for sleeve gastrectomy pouch creation. The Titan Stapler is supplied preloaded with staples, fires once and cannot be reloaded.
The Titan Stapler is comprised of three main sections:
- . Stapler: located in the sterile field, it has three main sections
- the handle, which does not contact the patient and is handled by user within the sterile O field,
- the device shaft and end effector which are surgically invasive components, O
- the staples, which are housed in the end effector until they are applied to the tissue O where they are permanently implanted.
- . Cable: permanently attached to the stapler, it is passed from the sterile field to the nonsterile area to connect with the reusable Power Supply Unit, and
{4}------------------------------------------------
- . Power Supply Unit: which is stored and used in the nonsterile area of the O.R. and powers the stapler opening, closure, and firing
The stapler end effector is 230 mm long and contains 342 staples that are organized in 6 staggered rows, 3 on each side of the cut line. The staples range in closed staple height from 2.2 mm to 1.2 mm. Staples are formed into a traditional 'B' shape: similar to existing devices. As with other powered staylers, the Titan Stapler opens and closes through a simple mechanical linkage housed in the shaft and effector of the device.
The device of the single-use Titan Stapler and a reusable power source which is supplied separately. There are no accessories supplied with the instrument. The Titan Stapler Power Source is designed to supply energy to the Titan Stapler and has a unique receptable port specific to the Titan Stapler.
#### V. Intended Use
Intended for transection and resection of gastric tissue.
#### VI. Indications for Use
The Titan SGS linear cutter is intended for longitudinal transection of gastric tissue for sleeve gastrectomy pouch creation.
#### VII. Technological Characteristics
The technological specifications of Titan Stapler and its predicate have been evaluated to determine equivalence. As detailed on Section 012 - Substantial equivalence of this 510(k) submission, upon reviewing and comparing intended use, design, materials, principle of operation and overall technological characteristics, the Titan Stapler is determined by Standard Bariatrics to be substantially equivalent to existing legally marketed devices (Table 1).
| | Standard Bariatrics'<br>Device | Predicate Device | Determination |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Product Name | Titan SGS | Echelon Flex Powered Plus | N/A |
| 510(k) Holder | Standard Bariatrics | Ethicon Endo-Surgery, LLC | N/A |
| Regulatory Information | | | |
| 510(k) Number | TBD | K140560 | N/A |
| Product Code | GDW | GDW | Same |
| Regulation | 21 CFR 878.4750<br>Implantable Staple | 21 CFR 878.4750<br>Implantable Staple | Same |
| Classification | II | II | Same |
| Design Information | | | |
| Principle of operation | Tissue is placed between<br>jaws of stapler, jaws are<br>closed, stapler is fired by<br>depressing trigger, knife<br>transects tissue as staples<br>are formed in tissue | Tissue is placed between jaws of<br>stapler, jaws are closed, stapler<br>is fired by depressing trigger,<br>knife transects tissue as staples<br>are formed in tissue | Same |
| Table 1: Overview of Substantial Equivalence | | | |
|----------------------------------------------|--|--|--|
|----------------------------------------------|--|--|--|
{5}------------------------------------------------
| Method of Insertion | Used in laparoscopic<br>applications by insertion<br>through a trocar | Used in laparoscopic<br>applications by insertion<br>through a trocar | Same |
|----------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|-------------|
| Power Source | DC powered - wall outlet | DC powered - battery | Equivalent |
| Staple Shape | B | B | Same |
| Staple Material | Titanium | Titanium Alloy | Equivalent. |
| Closed Staple Height | 1.2 - 2.2 | 1.0-2.3mm | Equivalent |
| Staple Arrangement | 3 staggered rows on each<br>side of cut line | 3 staggered rows on each side of<br>cut line | Same |
| Staple Line Extension<br>Past Cut Line | 1.5 staples (6.0 mm) | 1.5 staples (6.0 mm) | Same |
| Method of Resection | Blade | Blade | Same |
| Method of Activation | Trigger to open/close<br>device jaws and fire | Trigger to open/close device<br>jaws and fire | Same |
| Sterilization | Provided sterile via gamma.<br>10-6 SAL | Provided sterile via gamma. 10-6<br>SAL | Same |
The technological differences between the predicate and subject device do not impact the safety and effectiveness of the subject device as described in Section 012 - Substantial Equivalence.
#### VIII. Performance Data
### Non-Clinical
Performance Testing to evaluate and compare the technological and performance characteristics included bench, animal, and clinical studies.
Pre-determined performance specifications were tested, and validation activites were conducted to demonstrate that the Titan Stapler met the defined criteria. Testing on the subject device included biocompatibility, MR compatibility, usability, Electrical Safety and EMC testing. Animal studies evaluating hemostasis as compared to the predicate device and a survival study were conducted.
Performance evaluation of the Titan Stapler during the design validation and verification was completed by applying methods of internationally recognized standards such as, EN ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and Testing within a risk management process, EN ISO 14971:2012, Medical Devices - Application of Risk Management to Medical Devices, IEC 60601-1:2005/(R)2012, Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance, ISO 11737-2:2009, Sterilization of Medical Devices - Microbiological Methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process, among others.
The Titan Stapler met acceptance criteria and demonstrated comparable performance to the predicate device for the equivalent indications for use.
### Clinical
A comparative clinical study including 36 subjects at one site was conducted between the Titan Stapler and the predicate device, the Echelon Flex Powered Plus GST System on excised human stomach. Results demonstrated substantially equivalent performance between the two devices for the equivalent indications for use.
{6}------------------------------------------------
IDE G200085 with protocol title Multisite Study of Titan SGS Stapler in Longitudinal Gastric Resection was developed based on FDA's recommendations (Q200176). The study enrolled 62 subjects at three sites. Results demonstrated that the Titan SGS does not raise any new types of questions and the performance data provided reasonable assurance of safety and effectiveness to demonstrate substantial equivalence.
#### IX. Conclusion
The Titan Stapler has the same intended use as the Echelon Flex Powered Plus GST System. The conclusion drawn from the nonclinical and clinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device. The design/technological differences do not raise any new types of questions and the performance data provided reasonable assurance of safety and effectiveness to demonstrate substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.