K192566 · Reach Surgical, Inc. · GDW · Oct 17, 2019 · General, Plastic Surgery
Device Facts
Record ID
K192566
Device Name
Reach and CQ'ENCE Circular Staplers
Applicant
Reach Surgical, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Oct 17, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CQ ENCE Circular Staplers have application throughout the alimentary tract for the creation of end-to-end, side-to-side and end-to-side anastomoses in both open and laparoscopic surgeries.
Device Story
Sterile, single-use, hand-held, non-powered circular stapler; used in general surgery for alimentary tract anastomoses. Device places double-staggered circular row of titanium staples while simultaneously resecting tissue. Surgeon manually activates device via trigger; controls tissue compression by varying closed staple height. Features include anvil, cartridge, instrument shaft, trigger, safety release, wing nut, staple height indicator, and ancillary trocar. Audible indicator clicks when device is fully open for safe removal. Used in OR by surgeons; provides mechanical tissue joining and resection. Benefits include standardized anastomosis creation and adjustable compression for varying tissue requirements.
Clinical Evidence
Bench testing only. Performance data demonstrated compliance with specifications for visual inspection, dimensional analysis, staple form, and safety mechanism performance, consistent with predicate K120179.
Technological Characteristics
Materials: ASTM F 67 unalloyed titanium staples; polycarbonate and stainless steel end effector; stainless steel shaft; polycarbonate handle. Lubricant: PTFE and silicone grease. Energy: Manual. Dimensions: 21mm, 25mm, 28mm, 31mm, 34mm head diameters; 27cm or 35cm shaft lengths. Sterilization: EtO (SAL 10^-6). Connectivity: None (mechanical).
Indications for Use
Indicated for patients requiring alimentary tract anastomoses (end-to-end, side-to-side, end-to-side) in open or laparoscopic surgeries.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 17, 2019
Reach Surgical, Inc. % Rich Grant Official Correspondent SeCQure Surgical 4480 Lake Forest Dr. Cincinnati. Ohio 45242
Re: K192566
Trade/Device Name: Reach and CQ'ENCE Circular Staplers Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: September 6, 2019 Received: September 18, 2019
Dear Rich Grant:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192566
Device Name CQ'ENCE TM Circular Staplers
Indications for Use (Describe)
The CQ ENCE Circular Staplers have application throughout the alimentary tract for the creation of end-to-end, side-toside and end-to-side anastomoses in both open and laparoscopic surgeries.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
#### I. SUBMITTER
Rach Surgical Inc. 120 Xinxing Road West Zone, TEDA Tianjin, Tianjin CHINA 300462
Contact Person: Richard Grant Phone: 513-608-4017 Email: rgrant@secquresurgical.com
Date Prepared: September 6, 2019
#### II. DEVICE
Name of Device: Reach™ and CQ ENCE™ Circular Stapler Common or Usual Name: Circular Stapler Classification Name: Staple, implantable (21 CFR 878.4750) Regulatory Class: II Product Code: GDW
III. PREDICATE DEVICE
REACHTM Circular Staplers, (K120179)
#### IV. DEVICE DESCRIPTION
The REACH™ and SeCQure CO'ENCE™ Circular Staplers (CO'ENCE™ Circular Staplers) are sterile, single-use, hand-held, non-powered stapling devices used during general surgical procedures for the creation of end-to-end, side-to-side and end-to-side anastomoses in both open and laparoscopic surgeries.
The Circular Staplers place a double-staggered, circular row of titanium staples while simultaneously resecting tissue when the device is manually activated. The device is provided in five head diameter sizes: 34mm, 31mm, 28mm, 25mm, and 21mm (product codes CS34, CS31, CS28, CS25, and CS21 respectively). Each size device can be purchased with a standard 27cm length shaft. The CS21 and SC25 can also be purchased with a longer, 35cm shaft (product codes CS21L, CS25L) to facilitate surgeon access to desired tissues.
The CQ 'ENCE™ Circular Stapler allows the surgeon to control tissue compression by varying the closed staple height. The CQ ENCE Circular Stapler forms traditional B-shaped staples that are comprised of titanium. Safe removal of the device is facilitated by opening the instrument one full the Audible Indicator makes a clicking sound.
#### V. INDICATIONS FOR USE
{4}------------------------------------------------
The CQ ENCE Circular Staplers have application throughout the alimentary tract for the creation of endto-end, side-to-side and end-to-side anastomoses in both open and laparoscopic surgeries.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Reach™ and CQ 'ENCE™ Circular Stapler has the same technological characteristics as the predicate device with the exception of an additional device size and minor technological characteristics as described below
| Product | Reach TM and SeCQure CQ'ENCETM<br>Circular Stapler (Subject) | Predicate ReachTM Circular<br>Stapler<br>(K120179) | Comparison |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Anatomical Site Used | Various soft tissues accessible during<br>laparoscopic procedures | Various soft tissues accessible<br>during laparoscopic procedures | Same |
| Product Models Provided | SeCQure CQ'ENCETM 1<br>CS34, CS31, CS28, CS25, CS25L,<br>CS21, CS21L<br>ReachTM 1<br>ACS34D, ACS31D, ACS28D,<br>ACS25D, ACS25DL, ACS21D<br>ACS21DL | RCS34D, RCS31G, RCS28G,<br>RCS25G, RCS23G,<br>RCS17G | Equivalent. The subject<br>device is provided in a<br>21mm size which is<br>within the previously<br>cleared range and was<br>cleared in K071023. The<br>17 mm size is<br>discontinued. The models<br>are available in two brand<br>types, ReachTM or<br>SeCQure CQ'ENCETM,<br>which are the same with<br>the exception of branding. |
| Cutting Mechanism | Circular Knife | Circular Knife | Same |
| Method of Activation | Single Trigger manually activated | Single Trigger manually activated | Same |
| Disposable/Reusable | Disposable | Disposable | Same |
| Safety Mechanism | Firing Safety with Out-of-Range<br>Lockout | Firing Safety with Out-of-Range<br>Lockout | Same |
| Main Device<br>Features/Components | Anvil<br>Cartridge | Anvil<br>Cartridge | Same |
| | Instrument Shaft | Instrument Shaft | |
| | Trigger | Trigger | |
| | Safety Release | Safety Release | |
| | Wing Nut | Wing Nut | |
| | Staple Height Indicator | Staple Height Indicator | |
| | Ancillary Trocar | Ancillary Trocar | |
| | Shaft | Shaft | |
| | Staple | Staple | |
| Materials | Staples: ASTM F 67 Unalloyed<br>Titanium | Staples: ASTM F 67 Unalloyed<br>Titanium | Equivalent. A<br>biocompatible lubricant<br>was added to the<br>manufacturing process in<br>the current configuration. |
| | End Effector: polycarbonate and<br>stainless steel | End Effector: polycarbonate and<br>stainless steel | |
| | Shaft: stainless steel | Shaft: stainless steel | |
| | Handle: polycarbonate | Handle: polycarbonate | |
| | Lubricant: PTFE & Silicone Grease | Lubricant: Silicone Grease | |
| Staple Shape | Standard "B" shaped staple | Standard "B" shaped staple | Same |
| Staple Height (open) | 3.5 and 4.8 mm | 3.5 and 4.8 mm | Same |
| Staple Height (closed) | 1.0 to 1.8 for 3.5 mm Staples<br>1.2 to 2.2 mm for 4.8 mm staples | 1.5 and 2.2 mm | Same staple height. The<br>Staple Height labeling has<br>been amended to align<br>with industry standard to<br>describe the range of<br>staple heights to be<br>selected by the user. The<br>staple height remains<br>adjustable in the control<br>of the user. |
| End Effector Diameters | 34mm, 31mm, 28mm, 25mm, and<br>21mm | 34mm, 31mm, 28mm, 25mm,<br>23mm and 17 mm | Equivalent. The subject<br>device is provided in a<br>21mm size which is<br>within the previously<br>cleared range. The 17 mm<br>size is being discontinued. |
| Shaft Length | Standard (27cm), Long (35 cm) | Standard (27cm), Long (35 cm) | Same |
| Markings on Handle | Printed instructions for opening | None | Equivalent. Printed instructions were added for ease of use. |
| Anvil Tilt | Anvil Tilt Feature not offered | Anvil Tilt Feature offered | Equivalent. The Anvil Tilt feature is no longer offered as non-tilting is a standard for circular staplers. |
| Opening Method | Turn the Wing Nut counterclockwise one half to a full turn until the Audible Indicator clicks | Turn the Wing Nut counterclockwise one half to a full turn | Equivalent. Opening is the same. An audible click has been added to indicate to the user when the device is fully open. |
| Biocompatibility | Stapler is biocompatible for blood/bone/tissue contact for limited duration<br>Staples are biocompatible for permanent implantation | Stapler is biocompatible for blood/bone/tissue contact for limited duration<br>Staples are biocompatible for permanent implantation | Same |
| Endotoxin Limit | 0.5 EU/mL (20 EU/Device) | 0.5 EU/mL (20 EU/Device) | Same |
| Sterilization Method | EtO, SAL 10-6 | EtO. SAL 10-6 | Same |
| Packaging | Tyvek covered tray in cardboard box | Tyvek covered tray in cardboard box | Same |
| Shelf Life | 5 years | 5 years | Same |
Table 1: Technological Characteristics
<sup>4</sup> Note: The Reach and SeCQure CQ ENCE™ products are the same products with different branding i.e. the CS34 is the 34mm size SeCQure branded CQ ENCE™ Stapler and the ACS34D is the 34mm size Reach™ branded CQ ENCE™ Stapler. The devices are the same other than the branding on the labels and pad printed branding on the handles.
{5}------------------------------------------------
{6}------------------------------------------------
### PERFORMANCE DATA VII.
Reach™ Circular Stapler performance bench testing was provided in K120179. Testing for the Reach™ and CQ 'ENCE™ Circular Stapler is provided in Section 018 - Performance Testing - Bench of this submission to the same specifications as the Reach™ Circular Stapler previously provided in K120179. The data demonstrates that the Reach™ and CQ ENCE™ Circular Stapler meets the same specifications for visual inspection, dimensional analysis, staple form and safety mechanism performance. The materials used in the device, packaging, and sterilization method remain unchanged.
### VIII. CONCLUSIONS
The CQ 'ENCE™ Circular Stapler has the same intended use and principles of operation as its predicate device. The minor differences in technological characteristics do not change the device use or application and have been tested to the same standards as the currently marketed devices. There are no new issues of safety or effectiveness. The CQ ENCE™ Circular Stapler is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.