Echelon Endoscopic Linear Cutter Reloads, White, Blue, Gold, Green and Black
K183435 · Ethicon Endo-Surgery, LLC · GDW · Feb 6, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183435
Device Name
Echelon Endoscopic Linear Cutter Reloads, White, Blue, Gold, Green and Black
Applicant
Ethicon Endo-Surgery, LLC
Product Code
GDW · General, Plastic Surgery
Decision Date
Feb 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX™ families of Endoscopic Linear Cutters and Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
Device Story
Sterile, single-patient-use endoscopic linear cutters; simultaneously cut and staple tissue using six staggered rows of staples. Reloads feature Gripping Surface Technology; color-coded for specific tissue thicknesses. Operated by surgeons in OR settings; shaft rotates and articulates for lateral access. Instruments require manual loading of reloads and installation of primary lithium battery pack. Lock-out feature prevents firing of used or improperly installed reloads. Output is a stapled, transected tissue line; assists in surgical resection and anastomosis. Benefits include consistent staple line formation and tissue management during complex procedures.
Clinical Evidence
No clinical data was collected. Substantial equivalence demonstrated via bench testing only.
Technological Characteristics
Sterile, single-use endoscopic linear cutter reloads. Stapling principle; six staggered rows of staples. Materials cleared under K140560. Sterilization via Cobalt 60 irradiation (min 25 kGy) per ISO 11137-1:2006 and ISO 11137-2:2012. No electronics or software.
Indications for Use
Indicated for patients undergoing open or minimally invasive general, gynecologic, urologic, thoracic, or pediatric surgical procedures requiring tissue transection, resection, or anastomosis creation.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Echelon Endoscopic Linear Cutter Reload, White (+ Gripping Surface Technology) (K140560)
Echelon Endoscopic Linear Cutter Reload, Blue (+ Gripping Surface Technology) (K140560)
Echelon Endoscopic Linear Cutter Reload, Green (+ Gripping Surface Technology) (K140560)
Echelon Endoscopic Linear Cutter Reload, Black (+ Gripping Surface Technology) (K140560)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
February 6, 2019
Ethicon Endo-Surgery, LLC % Brian Ruble Associate Director, Regulatory Affairs Ethicon Endo-Surgery, Inc 4545 Creek Road Cincinnati, Ohio 45245
Re: K183435
Trade/Device Name: Echelon Endoscopic Linear Cutter Reloads, White, Blue, Gold, Green and Black Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: December 7, 2018 Received: December 11, 2018
## Dear Mr. Ruble:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by David David Krause -S Krause -S Krause -S
Date: 2019.02.06 12:56:59 -05'00'
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183435
Device Name
The ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX™ families of Endoscopic Linear Cutters and Reloads
Indications for Use (Describe)
The ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX™ families of Endoscopic Linear Cutters and Reloads are intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Company | Ethicon Endo-Surgery, LLC<br>475 Calle C<br>Guaynabo, PR 00969 |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Brian Ruble<br>Associate Director, Regulatory Affairs<br>Ethicon Endo-Surgery, Inc.<br>Telephone: (513) 337-1593<br>Email: BRuble1@its.jnj.com |
| Date Prepared: | December 07, 2018 |
| Common or Usual Name: | Surgical Stapler with Implantable Staples |
| Classification Name: | Staple, implantable (21 CFR 878.4750) |
| Regulatory Class: | Class II |
| Product Code: | GDW |
### Subject Device Trade Name
Echelon Endoscopic Linear Cutter Reload, White (+ Gripping Surface Technology) Echelon Endoscopic Linear Cutter Reload, Blue (+ Gripping Surface Technology) Echelon Endoscopic Linear Cutter Reload, Gold (+ Gripping Surface Technology) Echelon Endoscopic Linear Cutter Reload, Green (+ Gripping Surface Technology) Echelon Endoscopic Linear Cutter Reload, Black (+ Gripping Surface Technology)
### Predicate Devices Trade Name (K140560)
Echelon Endoscopic Linear Cutter Reload, White (+ Gripping Surface Technology) Echelon Endoscopic Linear Cutter Reload, Blue (+ Gripping Surface Technology) Echelon Endoscopic Linear Cutter Reload, Gold (+ Gripping Surface Technology) Echelon Endoscopic Linear Cutter Reload, Green (+ Gripping Surface Technology) Echelon Endoscopic Linear Cutter Reload, Black (+ Gripping Surface Technology)
#### Device Description
The ENDOPATH ECHELON™ and ECHELON FLEX™ families of endoscopic linear cutters are sterile, single patient use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. The ECHELON 60 instruments have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. The shaft can rotate freely in both directions and an articulation mechanism on articulating instruments enables bending the distal portion of the shaft to facilitate lateral access of the operative site.
{4}------------------------------------------------
The ECHELON ENDOPATH™ Echelon Endoscopic Linear Cutter Reloads come in five colors corresponding to different closed staple heights, with the intention of being used in different tissue thickness.
The instruments are packaged with a primary lithium battery pack that must be installed prior to use.
The instruments are packaged without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. The instruments' lock-out feature is designed to prevent a used or improperly installed reload from being re-fired or an instrument from being used in different tissue thickness.
The ECHELON ENDOPATH™ Echelon Endoscopic Linear Cutter Reloads (+ Gripping Surface Technology) are designed for use with the Echelon Endoscopic Linear Cutter devices (Echelon, Echelon Flex, Echelon Flex Powered and Echelon Flex Powered Plus) in a 60mm configuration.
#### Indications for Use
The ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX™ families of Endoscopic Linear Cutters and Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
#### Technological Characteristics
The design and performance of the subject device is based on the currently marketed Echelon Endoscopic Linear Cutter Reloads. The changes described in the submission do not affect the intended use of the devices or alter fundamental scientific technology of the device. Magnetic Resonance studies are not required because the dimensional changes in the cartridge sub assembly will not impact the shape and form of the staples. Stapling is the technological principle for both the subjects and predicate devices. It is based on the use of endoscopic instrumentation for transection, resection and/or creation of anastomoses.
#### Performance Data
Bench testing was performed to demonstrate that the changes introduced in the updated devices support substantial equivalence to the predicate device.
{5}------------------------------------------------
## Labelling
Update to IFU. Tyveks and Cartons to bring them up to date with current global labelling requirements.
# Sterilization
The subject devices will be sterilized by Cobalt 60 irradiation. The device will be validated to a minimum sterilization (radiation) dose of 25 kGy to achieve a 10 € sterility assurance level (SAL).
The sterilization process will be validated, and the sterilization dose will be established per the requirements of the following FDA recognized standards:
ISO 11137-1:2006 - Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices (FDA Recognition Number 14-328).
ISO 11137-2:2012 - Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose (FDA Recognition Number 14-364).
The subject devices are intended for single patient use and are not intended to be reused or resterilized.
ISO11737-1:2006, Sterilization of health care product-Microbiological methods- Part 1: Determination of the population of microorganisms on product.
ISO11737-2:2009, Sterilization of Medical Devices-Part 2: Tests of sterility performed in the validation of a sterilization process.
### Biocompatibility Testing
Biocompatibility testing was not required for this submission as no new materials were introduced on this device. All materials are already cleared under the predicate device submission (K140560).
### Electrical Safety and Electromagnetic Compatibility
Electrical Safety and Electromagnetic Compatibility section does not apply because the Echelon Endoscopic Linear Cutter Reloads (+ Gripping Surface Technology) do not contain any electronics.
### Bench Testing
Force to Close, Staple Form Ouality, Formed Staple Height and Staple Line Integrity were evaluated for Echelon Endoscopic Linear Cutter Reloads (+ Gripping Surface Technology) to support substantial equivalence to the predicate device.
### Clinical Testing
Clinical Testing section is not applicable. No clinical data was collected to demonstrate equivalency.
{6}------------------------------------------------
# Conclusion
Four outputs from the staple-line testing were evaluated for equality with the primary goal of improving the cartridge pan retention at the proximal end of the reload to predicates. In all cases, the subject reloads demonstrated equivalent performance relative to the predicates. The overall results from the bench testing are therefore acceptable since the subject reloads demonstrated substantial equivalency to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.