ECHELON FLEX POWERED VASCULAR STAPLER WITH ADVANCED PLACEMENT TIP, ENDOPATH ECHELON VASCULAR WHITE RELOADS FOR ADVANCED
K141952 · Ethicon Endo-Surgery, Inc. · GDW · Nov 24, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141952
Device Name
ECHELON FLEX POWERED VASCULAR STAPLER WITH ADVANCED PLACEMENT TIP, ENDOPATH ECHELON VASCULAR WHITE RELOADS FOR ADVANCED
Applicant
Ethicon Endo-Surgery, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Nov 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip and its reloads are intended for transection and resection of tissue and vasculature. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures.
Device Story
Powered surgical stapler; performs simultaneous tissue/vasculature transection and resection. Device consists of powered handle, articulating shaft, and replaceable reload cartridge. Input: manual operation by surgeon; battery-powered motor drives staple firing and cutting. Output: four staggered rows of titanium staples with central cut line. Used in OR for open/minimally invasive procedures. Features include active articulation, narrow shaft, and curved anvil tip for access to confined spaces. Lockout mechanism prevents firing without reload or with used reload. Benefits: facilitates access to difficult-to-reach vessels; provides hemostasis via staple line.
Clinical Evidence
Bench testing and animal studies performed. Bench testing evaluated force-to-close, staple height, staple form, staple line integrity, electrical safety, electromagnetic compatibility, MR compatibility, and biocompatibility. Animal testing (acute and survival models) evaluated hemostasis performance. No human clinical data provided.
Technological Characteristics
Powered handle with primary lithium battery. Titanium alloy staples. 35mm staple line, 30mm cut line. 4-row staggered staple configuration. Features active articulation, narrow shaft, and curved anvil tip. Sterile, single-patient use. Electrical safety and EMC compliant.
Indications for Use
Indicated for transection and resection of tissue and vasculature in open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
ENDOPATH ETS Endoscopic Reload, White (K11111, K070887, K020779)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other, resembling a stylized bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 24, 2014
Ethicon Endo-Surgery, LLC. % Ms. Linda Hill Ethicon Endo-Surgery Incorporated 4545 Creek Road Cincinnati Ohio 45242
Re: K141952
Trade/Device Name: Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip, and Endopath Echelon™ Vascular White Reloads for Advanced Placement Tip Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: October 25, 2014 Received: October 27, 2014
Dear Ms. Hill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
Page 2 - Ms. Linda Hill
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
K141952
510(k) Number (if known) To be assigned
↑
Device Name
Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip, and Endopath Vascular White Reloads for Advanced Placement Tip
#### Indications for Use (Describe)
The Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip and its reloads are intended for transection and resection of tissue and vasculature. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Company | Ethicon Endo-Surgery, LLC<br>475 Calle C<br>Guaynabo, PR 00969 |
|---------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Linda Hill<br>Ethicon Endo-Surgery, Inc.<br>Portfolio Leader, Regulatory Affairs<br>Telephone (513) 337-7623<br>Fax (513) 337-2623<br>Email LHILL3@its.jnj.com |
Date Prepared July 17, 2014
#### Device Name
Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip, Trade Name: and ENDOPATH Echelon™ Vascular White Reloads for Advanced Placement Tip
Common or Usual Name: Surgical Stapler with Implantable Staples
Classification Name: Staple, Implantable
#### Predicate Devices
Echelon Flex Powered Articulating Endoscopic Linear Cutters, 45mm, 60mm (K110385, K081146)
ENDOPATH ETS Flex Articulating Linear Cutter. 35mm (K111111. K070887. K020779) Echelon Endoscopic Linear Cutter Reload, White (K121600, K081146, K070887) ENDOPATH ETS Endoscopic Reload, White (K11111, K 070887, K020779)
## Device Description
The Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip and ENDOPATH Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 Row) reloads are sterile, single patient use devices that simultaneously cut and staple tissue. There are four staggered rows of staples, two on either side of the cut line. The Echelon Flex Powered Vascular Stapler with Advanced Placement Tip and reloads have a staple line that is approximately 35 mm long and a cut line that is approximately 30 mm long. The shaft can rotate freely in both directions and an articulation mechanism enables the distal portion of the shaft to pivot to facilitate access to the operative site.
The instrument is packaged with a primary lithium battery pack that must be installed prior to use. The instrument is packaged without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping, transporting, and installing the reload. The instrument's lockout feature is designed to prevent a used or improperly installed reload from being refired or an instrument from being fired without a reload.
{4}------------------------------------------------
## Indications for Use
The Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip and its reloads are intended for transection and resection of tissue and vasculature. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures.
## Technological Characteristics
The subject Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip instrument is similar to the predicate Echelon Flex Powered Articulating Endoscopic Linear Cutters with respect to the powered handle, and closure and firing design for minimal tip movement. The subject instrument is similar to the predicate ENDOPATH ETS Flex 35mm Articulating Linear Cutter and Reloads with respect to the active articulation feature and to the tissue compression aspects of the end effector, with similar jaw length and staple line length. As compared with the predicate devices, the subject Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip instrument incorporates several design enhancements - including a narrower diameter shaft, narrower end effector tip, curved anvil tip, introducer-like cartridge body tip, and active articulation with higher maximum articulation angles - which facilitate access in surgical procedures with smaller surgical spaces or difficult-to-reach vessels.
The subject reload, the ENDOPATH Echelon™ Vascular White Reloads for Advanced Placement Tip, is similar to the predicate ENDOPATH ETS Endoscopic Reload, White, 35mm with regard to staple material (titanium alloy), reload color (white), intended tissue type (thin/vascular), and closed staple height (1.0mm). The principal difference between the subject reload and predicate reloads is that the subject reload, which features a narrower design, delivers 2 sides of double staggered staple lines, rather than 2 sides of triple staggered staple lines as in the predicate device.
## Performance Data
Testing conducted to demonstrate that the subject device will perform equivalently to the identified predicate devices for the intended use included: bench testing for force-to-close evaluation; staple height, staple form, and staple line integrity evaluation; animal testing for hemostasis evaluation (in acute and survival models); electrical safety and electromagnetic compatibility testing; MR compatibility testing; and aggregate device biocompatibility testing for patient contacting materials.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.