LS SERIES LINEAR STAPLER AND RELOADS, CLC SERIES CURVED LINEAR CUTTER AND RELOADS, LC SERIES LINEAR CUTTER AND RELOADS
K141367 · Touchstone International Medical Science Co., Ltd. · GDW · Jul 25, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141367
Device Name
LS SERIES LINEAR STAPLER AND RELOADS, CLC SERIES CURVED LINEAR CUTTER AND RELOADS, LC SERIES LINEAR CUTTER AND RELOADS
Applicant
Touchstone International Medical Science Co., Ltd.
Product Code
GDW · General, Plastic Surgery
Decision Date
Jul 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The LS Series Linear Stapler and Reloads have application in the resection or transection of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures. The CLC Series Curved Linear Cutter and Reloads is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general (gastrointestinal and skeletal muscle), gynecologic, urologic, and thoracic surgical procedures. The LC Series Linear Cutter and Reloads have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and the creation of anastomoses. The ELC Series Endoscopic Linear Cutter and Single Use Loading Unit have applications in general, abdominal, gynecologic, pediatric and thoracic surgery for resection, and creation of anastomoses. They may be used for transection of liver substance, hepatic vasculature and biliary structures.
Device Story
Sterile, disposable surgical staplers and reloads for tissue resection, transection, and anastomosis. Devices include LS (Linear), CLC (Curved Linear Cutter), LC (Linear Cutter), and ELC (Endoscopic Linear Cutter) series. Manual operation; places titanium staples in staggered rows (double, triple, or quadruple depending on model) while simultaneously cutting tissue between rows. Available in various lengths and staple sizes to accommodate tissue thickness. Used in open or minimally invasive procedures by surgeons. Provides mechanical tissue closure and hemostasis; benefits include reduced manual anastomosis difficulty and standardized staple line formation.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included physical specifications, closed staple height dimensions, pressure resistance evaluation, maximum tensile strength of staple line repair, force required to fire, endotoxin limits, package integrity, and shelf life.
Technological Characteristics
Manual surgical staplers; titanium staples (ASTM F 67-06); stainless steel and polycarbonate construction; irradiation sterilization (SAL 10^-6); various lengths (30-100mm) and staple heights (0.75-4.8mm); safety release mechanisms for mis-firing prevention.
Indications for Use
Indicated for patients undergoing abdominal, gynecological, pediatric, thoracic, gastrointestinal, urologic, or general surgery requiring tissue resection, transection, or anastomosis creation. Includes liver, hepatic vasculature, and biliary structure transection/resection for ELC series. Prescription use only.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Disposable Linear Cutter Staplers and Reloads (K131511)
Endoscopic Linear Cutting Staplers with Single Use Loading Units (K120179)
Submission Summary (Full Text)
{0}------------------------------------------------
Premarket Notification 510(k) Submission
Section 3 510(k) Summary
## Section 3 510(k) Summary
#### 2 5 2014
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: _
Submission Date: 05/08/2014 1.
Sponsor Identification 2.
> Touchstone International Medical Science Co., Ltd. 21 A Science Plaza, International Science Park, No.1355 Jinjihu Avenue, Suzhou, 215021 P.R.CHINA
Establishment Registration Number:
Contact Person: Jo.Qiao Position: Management Representative Tel: +86 512 62991985 Fax: +86 512 62991998 Email: tsmo@touchstone.hk
{1}------------------------------------------------
#### Page 2 of 10 K141367
## Premarket Notification 510(k) Submission
3. Proposed Device Identification
3.1 Proposed Device for LS Series
Proposed Device Name: LS Series Linear Stapler and Reloads 3.1.1
3.1.2 Proposed Device Common Name: Stapler
3.1.3 Regulatory Information Classification Name: staple, implantable Classification: II Product Code: GDW Regulation Number: 21 CFR 878.4750 Review Panel: General & Plastic Surgery
3.1.4 Intended Use Statement
The LS Series Linear Stapler and Reloads have application or transection of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures.
Device Description 3.1.5
The proposed device, LS Series Linear Stapler and Reloads is a sterilized and disposable surgical instrument intended to be used in the resection of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures.
It places a double staggered row of titanium staples. It is available in 30 mm, 45 mm, 60 mm and 90 mm staple line length for use in various applications. Three staple sizes (2.5 mm, 3.5 mm, and 4.8 mm) are available to accommodate various tissue thicknesses. Each stapler could be reloaded no more than 11 times for total 12 firings.
It has two configurations: (1) LS serials are staplers, and (2) LSC serials are Reloads. Each of them has various specifications.
{2}------------------------------------------------
1/
## Premarket Notification 510(k) Submission
## 3.2 Proposed Device for CLC Series
Proposed Device Name: CLC Series Curved Linear Cutter and Reloads 3.2.1
Proposed Device Common Name: Stapler 3.2.2
3.2.3 Regulatory Information Classification Name: staple, implantable Classification: II Product Code: GDW Regulation Number: 21 CFR 878.4750 Review Panel: General & Plastic Surgery
Intended Use Statement 3.2.4
The CLC Series Curved Linear Cutter and Reloads is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general (gastrointestinal and skeletal muscle), gynecologic, and thoracic surgical procedures.
3.2.5 Device Description
The proposed device, CLC Series Curved Linear Cutter and Reloads are sterilized and disposable surgical instrument, which has application in multiple open or minimally invasive general (gastrointestinal and skeletal muscle), gynecologic, urologic, and thoracic surgical procedures. It is mainly suitable for low rectal resection and anastomosis of tissues where manual anastomosis is difficult.
It places four staggered curved row of titanium staples on the tissue upon activation, and cut the tissue between staple lines. It is available in 37mm and 43mm lengths. Two staple sizes (3.8 mm and 4.8 mm) are available to accommodate various tissue thicknesses. Each stapler could be reloaded no more than 7 times for total 8 firings.
It has two configurations: (1) CLC serials are staplers, and (2) CLCC serials are reloads. Each of them has various specifications.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text 'K141367' in a simple, sans-serif font. The text is arranged horizontally, with each character clearly legible. The characters are uniformly sized and spaced, creating a balanced and easily readable sequence.
page 4 of 10
## Premarket Notification 510(k) Submission
Proposed Device for LC Series 3.3
Proposed Device Name: LC Series Linear Cutter and Reloads 3.3.1
Proposed Device Common Name: Stapler 3.3.2
3.3.3 Regulatory Information Classification Name: staple, implantable Classification: II Product Code: GDW Regulation Number: 21 CFR 878.4750 Review Panel: General & Plastic Surgery
3.3.4 Intended Use Statement
The LC Series Linear Cutter and Reloads have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and the creation of anastomoses.
Device Description 3.3.5
The proposed device, LC Series Linear Cutter Stapler and Reloads are sterilized and disposable surgical instrument, which has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and the creation of anastomoses.
It places two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. It is available in 60 mm, 80 mm and 100 mm lengths. Three staple sizes (3.8 mm, 4.2mm and 4.5 mm) are available to accommodate various tissue thicknesses. Each stapler could be reloaded no more than 11 times for total 12 firings.
It has two configurations: (1) LC serials are staplers, and (2) LCC serials are reloads. Each of them has various specifications.
{4}------------------------------------------------
## k141367
10 Paqe 5 of
### Section 3 510(k) Summary
Premarket Notification 510(k) Submission
Proposed Device for ELC Series 3.4
Proposed Device Name: ELC Series Endoscopic Linear Cutter and Single Use Loading Unit 3.4.1
Proposed Device Common Name: Stapler 3.4.2
3.4.3 Regulatory Information Classification Name: staple, implantable Classification: II Product Code: GDW Regulation Number: 21 CFR 878.4750 Review Panel: General & Plastic Surgery
Intended Use Statement 3.4.4
The ELC Series Endoscopic Linear Cutter and Single Use Loading Unit have applications in general, abdominal, gynecologic, pediatric and thoracic surgery for resection, and creation of anastomoses. They may be used for transection of liver substance, hepatic vasculature and biliary structures.
3.4.5 Device Description
The proposed device, ELC Series Endoscopic Linear Cutter and Single Use Loading Unit are sterilized and disposable-surgical instrument, which has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures.
It places two, triple-staggered rows of titanium staples and simultaneously divides the tissue from central line. Five staple sizes (2.0mm, 2.5mm, 4.0mm, and 4.8mm) are available to accommodate various tissue thicknesses. Each cartridge could be reloaded no more than 7 times for total 8 firings. It can be adapted for all of the reload sizes available.
It has two configurations: (1) ELC-G and ELC-T serials are staplers, and (2) MG and RG serials are cartridges. Each of them has various specifications.
{5}------------------------------------------------
## Premarket Notification 510(k) Submission
### Section 3 510(k) Summary
- Predicate Device Identification 4.
Predicate Device for Disposable Linear Staplers and Reloads 4.1
510(k) Number: K131511
Product Name: Disposable Linear Staplers and Reloads
Manufacturer: Changzhou Sinolinks Medical Innovation Co., Ltd.
Predicate Device for Disposable Curved Cutter Stapler 4.2
510(k) Number: K131511
Product Name: Disposable Curved Cutter Stapler
Manufacturer: Changzhou Sinolinks Medical Innovation Co., Ltd.
Predicate Device for Disposable Linear Cutter Staplers and Reloads 4.3
510(k) Number: K131511
Product Name: Disposable Linear Cutter Staplers and Reloads
Manufacturer: Changzhou Sinolinks Medical Innovation Co., Ltd.
Predicate Device for Endoscopic Linear Cutting Staplers with Single Use Loading Units 4.4 510(k) Number: K120179 Product Name: Endoscopic Linear Cutting Staplers with Single Use Loading Units
Manufacturer: Reach Surgical, Inc.
#### Non-Clinical Test Conclusion 5.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. These tests include:
Performance Testing, including Physical Specification, Closed Staple Height Dimensions, Pressure Resistance Evaluation, Maximum Tensile Strength of Staple Line Repair and Force Required to Fire Stapler.
Endotoxin Limit
Package Integrity, including dye penetration tests and seal strength test.
Shelf Life
#### Substantially Equivalent (SE) Conclusion રું.
The following table compares the device to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
{6}------------------------------------------------
page 7 of 10
Premarket Notification 510(k) Submission
Section 3 510(k) Summary
| Table 3-1 | LS Series Linear Stapler and Reloads | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Item | Proposed Device | Predicate<br>Device |
| Product Code | GDW | Same |
| Regulation No. | 21 CFR 878.4750 | Same |
| Class | II | Same |
| Intended Use | The LS Series Linear Stapler and Reloads have application in the<br>resection or transection of tissue for abdominal, gynecological,<br>pediatric and thoracic surgical procedures. | Same |
| Operation Principle | Manual | Same |
| Safety Mechanism | Safety Release for preventing from mis-firing | Same |
| Suture Line Length | 30, 45,50, 60, 90 mm | Similar |
| Number of Staples | 13, 19, 21, 23, 25,39 | Similar |
| Closed Staple Height | 1.0, 1.5, 2.0 mm | Similar |
| Closed Staple Form | Image: staple form | Same |
| Pressure Resistance after Suturing | ≥3.6kPa | Same |
| Staple Materiel | Unalloyed Titanium conforms to ASTM F 67-06 | Same |
| Stapler Materials | Stainless Steel, Polycarbonate | Same |
| Sterilization | Irradiation Sterilized, SAL: 10-6 | Same |
| Endotoxin Limit | 20 EU per Product | Same |
| Package | Tray with Tyvek Paper | Same |
| Labeling | Conforms to 21 CFR part 801 | Same |
Stanler and Reloads a creat T--L1
Difference in Suture Line Length, Number of Staples, Closed Staple Height. Between the proposed and the predicate device are discussed in the 510(k) submission documents, it is concluded that these differences will not affect the effectiveness and safety of the proposed device.
The proposed device, LS Series Linear Stapler and Reloads, is determined to be Substantially Equivalent (SE) to the predicate device, Disposable Linear Staplers and Reloads (K131511), in respect of safety and effectiveness.
3- 7
{7}------------------------------------------------
く
Premarket Notification 510(k) Submission
| Table 3-2 CLC Series Curved Linear Cutter and Reloads | | |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Item | Proposed Device | Predicate<br>Device |
| Product Code | GDW | Same |
| Regulation No. | 21 CFR 878.4750 | Same |
| Class | II | Same |
| Intended Use | The CLC Series Curved Linear Cutter and Reloads is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general (gastrointestinal and skeletal muscle), gynecologic, urologic, and thoracic surgical procedures. | Same |
| Operation Principle | Manual | Same |
| Cutting Mechanism | Curved Knife | Same |
| Safety Mechanism | Safety Release for preventing from mis-firing. | Same |
| Cutting Line | 37, 43 mm | Similar |
| Row Number of Staple | 4 | Same |
| Closed Staple Height | 1.5, 2.0 mm | Similar |
| Closed Staple Form | Image: staple form | Same |
| Hardness (Circular knife) | $\cong$ 380HV0.2 | Same |
| Pressure Resistance after Suturing | $\geq$ 3.6kPa | Same |
| Staple Materiel | Unalloyed Titanium conforms to ASTM F 67-06 | Same |
| Stapler Materials | Stainless Steel, Polycarbonate | Same |
| Sterilization | Irradiation Sterilized, SAL: 106 | Same |
| Endotoxin Limit | 20 EU per Product | Same |
| Package | Tray with Tyvek Paper | Same |
| Labeling | Conforms to 21 CFR part 801 | Same |
#### 2 CLC Series Curved Linear Cutter and Reloads Toble
Difference in Cutting Line, Closed Staples Height. Between the proposed and predicate device are discussed in the 510(k) submission documents, it is concluded that these differences will not affect the effectiveness and safety of the proposed device.
The proposed device, CLC Series Curved Linear Cutter and Reloads, is determined to be Substantially Equivalent (SE) to the predicate device, Disposable Curved Cutter Stapler(K131511), in respect of safety and effectiveness.
{8}------------------------------------------------
## Premarket Notification 510(k) Submission
## Section 3 510(k) Summary
| Table 3-3 LC Series Linear Cutter and Reloads | | |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Item | Proposed Device | Predicate<br>Device |
| Product Code | GDW | Same |
| Regulation No. | 21 CFR 878.4750 | Same |
| Class | II | Same |
| Intended Use | The LC Series Linear Cutter and Reloads have applications in<br>abdominal, gynecologic, pediatric and thoracic surgery for resection,<br>transection, and the creation of anastomoses. | Same |
| Operation Principle | Manual | Same |
| Cutting Mechanism | Linear Knife | Same |
| Safety Mechanism | Safety Release for preventing from mis-firing. | Same |
| Suture Length | 60, 80, 100 mm | Similar |
| Cutting Length | 54, 74, 94 mm | Similar |
| Number of Staple | 64, 84, 104 | Similar |
| Closed Staple Height | 1.5, 1.8, 2.0 mm | Similar |
| Closed Staple Form | Image: staple form | Same |
| Hardness (Circular knife) | $≥380HV_{0.2}$ | Same |
| Pressure Resistance after Suturing | $≥3.6kPa$ | Same |
| Staple Materiel | Unalloyed Titanium conforms to ASTM F 67-06 | Same |
| Stapler Materials | Stainless Steel, Polycarbonate | Same |
| Sterilization | Irradiation Sterilized, SAL: 10-6 | Same |
| Endotoxin Limit | 20 EU per Product | Same |
| Package | Tray with Tyvek Paper | Same |
| Labeling | Conforms to 21 CFR part 801 | Same |
I inear Cutter and Reloads 1 0 0 ----
Difference in Suture Length , Cutting Length , Number of Staple, Closed Staple Height. Between the proposed and predicate device are discussed in the 510(k) submission documents, it is concluded that these differences will not affect the effectiveness and safety of the proposed device.
The proposed device, LC Series Linear Cutter and Reloads, is determined to be Substantially Equivalent (SE) to the predicate device, Disposable Linear Cutter Staplers and Reloads(K131511), in respect of safety and effectiveness.
{9}------------------------------------------------
.
가
| Item | Proposed Device | Predicate<br>Device |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Product Code | GDW | Same |
| Regulation No. | 21 CFR 878.4750 | Same |
| Class | II | Same |
| Intended Use | The ELC Series Endoscopic Linear Cutter and Single Use Loading<br>Unit have applications in general, abdominal, gynecologic, pediatric<br>and thoracic surgery for resection, transection, and creation of<br>anastomoses. They may be used for transection and resection of liver<br>substance, hepatic vasculature and biliary structures. | Same |
| Operation Principle | Manual | Same |
| Cutting Mechanism | Linear Knife | Same |
| Safety Mechanism | Green button for preventing from mis-firing | Same |
| Suture Length | 30, 45, 60 mm | Same |
| Cutting Length | 26, 41, 56 mm | Same |
| Closed Staple Height | 0.75, 1.0, 1.5, 1.7, 2.0 mm | Same |
| Closed Staple Form | Image: staple form | Same |
| Hardness (Circular knife) | $\ge 380\text{HV}_{0.2}$ | Same |
| Pressure Resistance after Suturing | $\ge 3.6\text{kPa}$ | Same |
| Staple Materiel | Unalloyed Titanium conforms to ASTM F 67-06 | Same |
| Stapler Materials | Stainless Steel, Polycarbonate | Same |
| Sterilization | Irradiation Sterilized, SAL: 10-6 | Same |
| Endotoxin Limit | 20 EU per Product | Same |
| Package | Tray with Tyvek Paper | Same |
| Labeling | Conforms to 21 CFR part 801 | Same |
| Table 3-4 ELC Series Endoscopic Linear Cutter and Single Use Loading Unit | | |
|---------------------------------------------------------------------------|--|--|
| | | |
The proposed device, ELC Series Endoscopic Linear Cutter and Single Use Loading Unit, is determined to be Substantially Equivalent (SE) to the predicate device, Endoscopic Linear Cutting Staplers with Single Use Loading Units (K120179), in respect of safety and effectiveness.
3- 10
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure embracing a globe. The text is in all caps and reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 25, 2014
Touchstone International Medical Science Company, Ltd Mr. Jo Qiao Management Representative 21A Science Plaza, International Science Park No. 1355 Jinjihu Avenue Suzhou, 215021 P.R. CHINA
Re: K141367
Trade/Device Name: LS Series Linear Stapler and Reloads Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, GAG Dated: May 26, 2014 Received: May 28, 2014
Dear Mr. Qiao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{11}------------------------------------------------
Page 2 - Jo. Qiao
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
## 510(k) Number (if known)
KI413b1
Device Name LS Series Linear Stapler and Reloads
## Indications for Use (Describe)
The LS Series Linear Stapler and Reloads have application in transection of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED,
1 1 1 . 5. 5. 5. 5 5 5 5 3 . 3 . 3 . 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . *** FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/12/Picture/13 description: The image shows the name "Peter Hudson -S" in a simple, sans-serif font. The name is written in black ink on a white background. The letters are evenly spaced and easy to read. The word "Peter" is followed by "Hudson" and then "-S".
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this Information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{13}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
| 510(k) Number (if known) |
|--------------------------|
|--------------------------|
KI41367
Device Name
CLC Series Curved Linear Cutter and Reloads
Indications for Use (Describe)
The CLC Series Curved Linear Cutter and Reloads is intended for transection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general (gastrointestinal and skeletal muscle), gynecologic, urologic, and thoracic surgical procedures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
ることができる。 このため、 FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FO Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
$$\mathtt{Peter} = \mathtt{\mathtt{\mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}}} \mathtt{\mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}}} \mathtt{\mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt{\mathtt{\mathtt{\bullet}}}} \mathtt{\mathtt$$
This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suppestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{14}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
KJ41367
Device Name LC Series Linear Cutter and Reloads
Indications for Use (Describe)
The LC Series Linear Cutter and Reloads have applications in abdominal, genecologic, pediatric and thoracic surgery for resection, transection, and the creation of anastomoses.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY ----------------... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
PeterFDAson-S
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review Instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless It displays a currently valid OMB number."
{15}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
K14367
Device Name
ELC Series Endoscopic Linear Cutter and Single Use Loading Unit
## Indications for Use (Describe)
The ELC Series Endoscopic Linear Cutter and Single Use Loading Unit have applications in general, abdominal, gynecologic, pediatic and thoracic surgery for resection, and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY . ' . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Peter FDAudson-S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and roview the collection of information. Send comments regarding this burden estimate or any other aspect of this Information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.