K132493 · Covidien, LLC · GDW · Oct 23, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132493
Device Name
ENDO GIA RADIAL RELOAD WITH TRI-STAPLE TECHNOLOGY
Applicant
Covidien, LLC
Product Code
GDW · General, Plastic Surgery
Decision Date
Oct 23, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Endo GIA™ Radial Reload with Tri-Staple™ Technology has application in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
Device Story
Endo GIA™ Radial Reload with Tri-Staple™ Technology is a single-use, disposable surgical stapling component; compatible with Endo GIA™ Ultra Universal, Endo GIA™ Universal, GIA™ Universal, and iDrive™ Ultra handles. Device places three rows of curved titanium staples (60mm length) on each side of a cut line while simultaneously dividing tissue. Used in open or minimally invasive abdominal, gynecologic, pediatric, and thoracic surgeries; operated by surgeons. Provides curved transection capability for deep pelvic access and organ resection (liver, pancreas). Benefits include precise tissue management and anastomosis creation in restricted surgical spaces.
Clinical Evidence
Bench testing only. In vitro testing included firing force, retraction force, and staple formation. In vivo testing included free bleed evaluation, air leak test, burst evaluation, and tissue grasping/trauma assessment. Biocompatibility testing conducted per ISO 10993-1.
Technological Characteristics
Implantable titanium staples; 3-row curved cartridge geometry; 60mm staple line length. Materials compliant with ISO 10993-1. Sterilized via ethylene oxide (ETO) to SAL 10^-6. Single-use, disposable. Mechanical actuation via compatible stapler handles.
Indications for Use
Indicated for patients undergoing open or minimally invasive general abdominal, gynecologic, pediatric, and thoracic surgery requiring tissue resection, transection, and anastomosis, including deep pelvic procedures (e.g., low anterior resection) and procedures involving liver, hepatic vasculature, biliary structures, and pancreas.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Endo GIA™ Radial Reload with Tri-Staple™ Technology (K131705)
Endo GIA™ Single Use Black Reload with Tri-Staple™ Technology (K093410)
Endo GIA™ Single Use Tan Reload with Tri-Staple™ Technology (K101444)
Submission Summary (Full Text)
{0}------------------------------------------------
- -
・・・
## 510(k) Summary
| SUBMITTER: | Covidien<br>60 Middletown Avenue<br>North Haven, CT 06473<br>(203) 492-5299 (T)<br>(203) 492-5029 (F) |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Katherine Kassim<br>Senior Specialist, Regulatory Affairs |
| DATE PREPARED: | August 19, 2013 |
| TRADE/PROPRIETRY NAME: | Endo GIA™ Radial Reload with Tri-Staple™ Technology |
| | COMMON/USUAL NAME: |
| CLASSIFICATION NAME: | Staples, Implantable |
| FDA PANEL NUMBER: | 79 |
| PRODUCT CODE: | GDW |
| CLASS CODE: | Pursuant to 21 CFR 878.4750, an implantable staple is a Class II device. |
| PREDICATE DEVICE(S): | Endo GIA™ Radial Reload with Tri-Staple™ Technology (K131705);<br>Endo GIA™ Single Use Black Reload with Tri-Staple™ Technology<br>(K093410); Endo GIA™ Single Use Tan Reload with Tri-Staple™<br>Technology (K101444) |
| DEVICE DESCRIPTION: | The single use Endo GIA™ Radial Reload with Tri-Staple™ Technology<br>places three radial (curved) staple rows 60mm in length on each side of<br>a cut line and simultaneously divides the tissue between the third and<br>fourth lines, creating a 60mm curved transection. The radial reload<br>places height progressive titanium staple rows. The reloads may be<br>inserted through an access device such as a hand access device or<br>comparable access port. The Endo GIA™ Radial Reload with Tri-<br>Stable™ Technology may be used with the Endo GIA™ Ultra Universal,<br>Endo GIA™ Universal, GIA™ Universal and iDrive™ Ultra stapler<br>handles. |
| INTENDED USE: | The Endo GIA™ Radial Reload with Tri-Staple™ Technology has<br>application in open or minimally invasive general abdominal,<br>gynecologic, pediatric and thoracic surgery for resection and transection<br>of tissue and creation of anastomosis, as well as application deep in the<br>pelvis, i.e. low anterior resection. It may be used for transection and<br>resection of liver substance, hepatic vasculature and biliary structures<br>and for transection and resection of pancreas. |
------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
·
---
- -
. .
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
· · ·- ·-
ﺳﻌﺮ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍ
. . . . . . . . . . . . . .
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{1}------------------------------------------------
TECHNICAL CHARACTERISTICS:
The Endo GIA™ Radial Reload with Tri-Staple™ Technology black reload and tan reload are substantially equivalent and have not altered the fundamental scientific technology to the predicate devices with regards to stapling technologies. The following summary tables show the similarities and differences between the subject device and the predicate devices.
| Substantial | Endo GIA™ Radial | Endo GIA™ Radial | Endo GIA™ Single |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Equivalence Chart | Reload with Tri- | Reload with | Use Black Reload with |
| Black Reload | Staple™ Technology | Tri-Staple™ | Tri-Staple Technology |
| | Black Reload | Technology Purple | (predicate - K093410) |
| | (subject devices) | Reload | |
| | | (predicate - K131705 ) | |
| Intended Use | The Endo GIA™ Radial<br>Reload with Tri-Staple™<br>Technology has<br>application in open and<br>minimally invasive<br>general abdominal,<br>gynecologic, pediatric<br>and thoracic surgery for<br>resection and<br>transection of tissue and<br>creation of anastomosis,<br>as well as application<br>deep in the pelvis, i.e.<br>low anterior resection. It<br>may be used for<br>transection and<br>resection of liver<br>substances, hepatic<br>vasculature and biliary<br>structures and for<br>transection and | Same | The AutoSuture™ Endo<br>GIA™ Staplers have<br>application in<br>abdominal, gynecologic,<br>pediatric and thoracic<br>surgery for resection,<br>transection and creation<br>of anastomosis. They<br>may be used for<br>transection and<br>resection of liver<br>substance, hepatic<br>vasculature and biliary<br>structures. |
| | resection of pancreas. | | |
| Staple Line Length | 60mm | 60mm | 45mm or 60mm |
| Cartridge Color | Black | Purple | Same |
| Staple Sizes | 4.0mm, 4.5mm, 5.0mm | 3.0mm, 3.5mm, 4.0mm | Same |
| Cartridge Geometry | Curved | Same | Straight |
| Staple Rows | 3 rows | Same | Same |
| Staple Material | Titanium | Same | Same |
| Single Use | Yes | Same | Same |
| Disposable | Yes | Same | Same |
| Sterilization | The subject devices<br>shall be sterilized via a<br>validated ethylene oxide<br>(ETO) cycle to a SAL of<br>106 | Same | Same |
| Packaging | The subject devices<br>shall be packaged in a<br>sterile tray with Tyvek®<br>lid. Each tray is<br>packaged in a display | Same | Same |
| Substantial<br>Equivalence Chart for<br>Tan Reload | Endo GIA™ Radial<br>Reload with Tri-<br>Staple™ Technology<br>Tan Reload<br>(subject device) | Endo GIA™ Radial<br>Reload with Tri-<br>Staple™ Technology<br>Purple Reload<br>(predicate K131705) | Endo GIA™ Single<br>Use Tan Reload with<br>Tri-Staple Technology<br>(predicate-K101444) |
| Intended Use | The Endo GIA™ Radial<br>Reload with Tri-Staple™<br>Technology has<br>application in open and<br>minimally invasive<br>general abdominal,<br>gynecologic, pediatric<br>and thoracic surgery for<br>resection and<br>transection of tissue and<br>creation of anastomosis,<br>as well as application<br>deep in the pelvis, i.e.<br>low anterior resection. It<br>may be used for<br>transection and<br>resection of liver<br>substances, hepatic<br>vasculature and biliary<br>structures and for<br>transection and<br>resection of pancreas. | Same | The AutoSuture™ Endo<br>GIA™ Staplers have<br>application in<br>abdominal, gynecologic,<br>pediatric and thoracic<br>surgery for resection,<br>transection and creation<br>of anastomoses. They<br>may be used for<br>transection and<br>resection of liver<br>substance, heptic<br>vasculature and biliary<br>structures. |
| Staple Line Length | 60mm | 60mm | 45mm or 60mm |
| Cartridge Color | Tan | Purple | Same |
| Staple Sizes | 2.0mm, 2.5mm, 3.0mm | 3.0mm, 3.5mm, 4.0mm | Same |
| Cartridge Geometry | Curved | Same | Straight |
| Staple Rows | 3 rows | Same | Same |
| Staple Material | Titanium | Same | Same |
| Single Use | Yes | Same | Same |
| Disposable | Yes | Same | Same |
| Sterilization | The subject devices<br>shall be sterilized via a<br>validated ethylene oxide<br>(ETO) cycle to a SAL of<br>10-6 | Same | Same |
| Packaging | The subject devices<br>shall be packaged in a<br>sterile tray with Tyvek®<br>lid. Each tray is<br>packaged in a display<br>box. | Same | Same |
:
{2}------------------------------------------------
.
.
:
{3}------------------------------------------------
All components of the Endo GIA™ Radial Reload with Tri-Staple™ MATERIALS: Technology are similar to the predicate Endo GIA™ Radial Reload with Tri-Staple™ Technology. All materials are similar and are in accordance with ISO Standard 10993-1.
Design verification and pre-clinical validation studies were conducted to PERFORMANCE DATA: demonstrate that the proposed devices Endo GIA™ Radial Reload with Tri-Staple™ Technology black reload and tan reload are safe and effective and perform as intended. In vitro and in vivo testing to support the intended use of this device includes:
- . In Vitro
- Firing Force o
- Retraction Force o
- Staple Formation O
- In Vivo .
・
- Free Bleed Evaluation o
- o Air Leak Test
- o Burst Evaluation
- Tissue Grasping and Trauma o
- . Biocompatibility
The result of these tests demonstrates that the proposed devices Endo GIA™ Radial Reload with Tri-Staple™ Technology black reload and tan reload are substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 23, 2013
Covidien, LLC Ms. Katherine Kassim Senior Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K132493
Trade/Device Name: Endo GIA" Radial Reload with Tri-Staple" Technology Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: September 23, 2013 Received: September 24, 2013
Dear Ms. Kassim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Katherine Kassim
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Endo GIA™ Radial Reload with Tri-Staple™ Technology
### Indications for Use
K132493 510(k) Number (if known):
Device Name: Endo GIA™ Radial Reload with Tri-Staple™ Technology
.. .
Indications for Use:
The Endo GIA™ Radial Reload with Tri-Staple™ Technology has application In open or minimally invasive general abdominal, gynecologic, pediatre and thorado surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the perior, it e. low antentiment which and may be used for transection and resection of liver substance, hepatic vasculature and billiary stuctures and for transection and resection of pancreas.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
# David Krause -S
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K132493
Covidien Supplamental Information Радв 18
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.