K121510 · Covidien, Formerly US Surgical A Divison of Tyco H · GDW · Jul 25, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121510
Device Name
IDRIVE ULTRA POWERED HANDLE ENDO GIA ADAPTER
Applicant
Covidien, Formerly US Surgical A Divison of Tyco H
Product Code
GDW · General, Plastic Surgery
Decision Date
Jul 25, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The iDrive ™ Ultra powered handle and Endo GIA ™ adapter, when used with Endo GIA ™ single use reloads have applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature, and biliary structures and for transection and resection of pancreas. The iDrive ™ Ultra powered handle and Endo GIA ™ adapter, when used with Endo GIA ™ curved tip single use reloads, can be used to blunt dissect or separate target tissue from other certain tissue. The iDrive ™ Ultra powered handle and Endo GIA ™ adapter, when used with the Endo GIA ™ Radial Reload with Tri-Staple ™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
Device Story
iDrive Ultra powered handle and Endo GIA adapter; electromechanical surgical stapler system. Input: manual activation via buttons on handle. Operation: powered delivery of triple-staggered rows of titanium staples; simultaneous tissue division via single-use reload. Usage: OR setting; operated by surgeons. Output: stapled and transected tissue. Benefit: provides powered assistance for stapling and cutting in open or minimally invasive procedures; facilitates anastomosis and resection in complex anatomical areas like pelvis, liver, and pancreas.
Clinical Evidence
Bench testing only. Testing included articulation and rotation verification, staple formation verification, visual indication intensity, aseptic battery transfer, grasping capability, tissue trauma evaluation, knife cutting performance, and lifecycle reliability. Electrical safety testing performed per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18.
Technological Characteristics
Powered handle and adapter system; delivers triple-staggered rows of titanium staples; simultaneous tissue division. Patient-contacting materials evaluated per ISO 10993-1:2009. Electrical safety per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18. Battery-powered operation.
Indications for Use
Indicated for patients undergoing abdominal, gynecological, pediatric, or thoracic surgery requiring tissue resection, transection, or anastomosis; includes liver, hepatic vasculature, biliary structures, and pancreas procedures; includes deep pelvic procedures (e.g., low anterior resection); allows for blunt tissue dissection.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
K121510
page 1/2
### 510(k) Summary of Safety and Effectiveness
SUBMITTER:
Covidien IIc (formerly registered as Tyco Healthcare, LP) 60 Middletown Avenue North Haven, CT 06473 Tel. No .: (203) 492-5000
JUL 25 2012
CONTACT PERSON:
Sarah Rizk Senior Product Specialist, Regulatory Affairs Covidien IIc Phone: (203) 492-5493 Fax: (203) 492- 5029
Surgical Stapler with Implantable Staples
Covidien, iDrive TM system (K102325)
DATE PREPARED:
May 18, 2011
TRADE/PROPRIETARY NAME:
y
iDrive™ Ultra powered handle Endo GIA™ adapter iDrive™ Battery Insertion Guide
COMMON/USUAL NAME:
Staple, Implantable CLASSIFICATION NAME:
PREDICATE DEVICE(S):
DEVICE DESCRIPTION:
INTENDED USE:
Surgical stapler with a powered handle, delivering implantable titanium staples.
Covidien, Endo GIA™ Ultra Universal Stapler (K111825)
The iDrive ™ Ultra powered handle and Endo GIA™ adapter, when used with Endo GIA™ single use reloads have applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature, and biliary structures and for transection and resection of pancreas.
The iDrive™ Ultra powered handle and Endo GIA™ adapter, when used with Endo GIA™ curved tip single use reloads, can be used to blunt dissect or separate target tissue from other certain tissue.
The iDrive™ Ultra powered handle and Endo GIA™ adapter, when used with the Endo GIA™ Radial Reload with Tri-Staple™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
Traditional 510(k)
Page 22 of 100
{1}------------------------------------------------
Covidien IIc (Formerly registered as Tyco Healthcare, LP)
K121510
page 2/2
TECHNOLOGICAL CHARACTERISTICS:
The iDrive™ Ultra powered handle with the Endo GIA™ adapter and Endo GIA™ reloads delivers two sets of triplestaggered rows of titanium staples and simultaneously divides the tissue between the two rows of staples via the single use reload, initiated by buttons on the powered handle.
#### MATERIALS:
All patient-contacting components of the iDrive™ Ultra powered handle and Endo GIA™ adapter are comprised of materials that have been evaluated in accordance with ISO 10993-1: 2009, Biological Evaluation of medical devices -- Part 1: Evaluation and Testing.
#### PERFORMANCE DATA:
In-vitro and in-vivo testing to support the intended use of this device includes:
- Articulation and Rotation Verification O
- Staple Formation Verification O
- Visual Indication Intensity O
- Aseptic Battery Transfer Test o
- Grasping Capability o
- Tissue Trauma Evaluation O
- Knife Cutting Performance Verification o
- Lifecycle Reliability Test O
Additional bench top testing has been performed and includes testing to the following electrical safety standards:
- IEC 60601-1: 1988 + A1 (1991) + A2 (1995) 0
- IEC 60601-1-2: 2007 0
- IEC 60601-2-18: 1996+ A1 (2000) 0
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
COVIDIEN. FORMERLY US SURGICAL A DIVISION OF TYCO % Ms. Sarah Rizk Regulatory Affairs Product Specialist 60 Middletown Avenue North Haven, Connecticut 06473
JUL 2 5 2012
Re: K121510
Trade/Device Name: iDrive Ultra Powered Handle and Endo GIA adapter Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: May 18, 2012 Received: May 21, 2012
Dear Ms. Rizk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
### Page 2- Ms. Sarah Rizk
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the . Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Eirnd Keith
~ Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Device Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
iDrive ™ Ultra powered handle and Endo GIA ™ adapter Device Name:
Indications For Use:
The iDrive ™ Ultra powered handle and En do GIA ™ adapter, when used with En do GIA ™ single use reloads have applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis. It may be used for transection and resection rof liver substance, hepatic vasculature, and biliary structures and for transection and resection of pancreas.
The iDrive ™ Ultra powered handle and Endo GIA TM adapter, when used with Endo GIA TM curved tip single use reloads, can be used to blunt dissect or separate target tissue from other certain tissue.
The iDrive ™ Ultra powered handle and En do GIA TM adapter, when used with the Endo GIA TM Radial Reload with Tri-Staple ™ Technology, has applications in open or minimally invasive raneral abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection. It may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
evice Evaluation (ODE)
David Kine for MM
(Division Sign-Off)
Division of Surgical. Orthopedic. and Restorative Devices
510(k) Number K/2/5/0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.