K101378 · Suzhou Frankenman Medical Equipment Co., Ltd. · GDW · Jun 29, 2010 · General, Plastic Surgery
Device Facts
Record ID
K101378
Device Name
FRANKENMAN SURGICAL STAPLERS
Applicant
Suzhou Frankenman Medical Equipment Co., Ltd.
Product Code
GDW · General, Plastic Surgery
Decision Date
Jun 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Frankenman Disposable Alimentary Canal Stapler is used throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic techniques. The Frankenman Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids is a Circular Stapler product, with accessories, that has application for general surgical treatment of haemorrhoids and anorectal wall defects by means of transanal stapling and resection of mucosal and musculo-mucosal tissue resulting in occlusion of haemorrhoidal inflow, restoring the haemorrhoidal tissue to its normal physiological position. The Frankenman Disposable Reloadable Linear Stapler (and Reloads) is used in the resection or transaction of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures. The Frankenman Disposable Reloadable Linear Cutter Stapler (and Reloads) has application in abdominal, gynecological, thoracic and pediatric surgery transaction, resection, and the creation of anastomoses.
Device Story
Frankenman Surgical Staplers are manual, single-use, disposable instruments designed for mechanical tissue suturing and resection. The family includes alimentary canal staplers, circular staplers for rectal prolapse/hemorrhoids, reloadable linear staplers, and reloadable linear cutter staplers. Operation involves manual activation by squeezing the handle, which drives titanium staples through tissue in staggered rows; circular and linear cutter models incorporate a knife blade to resect excess tissue simultaneously. Used in OR settings by surgeons for open or laparoscopic procedures. The devices facilitate tissue closure and anastomosis, potentially reducing operative time and improving post-operative outcomes compared to manual suturing.
Clinical Evidence
Clinical testing of each stapler line demonstrated superior results in post-operative healing, pain management, anastomotic leakage, and bleeding compared to manual suturing and competitor devices. Bench testing validated physical characteristics (appearance, dimensions, cartridge compatibility) and performance (strength, closure) per ISO 11737 and EN 552 standards.
Technological Characteristics
Manual surgical staplers; titanium staples; mechanical actuation via handle squeeze. Available in various staple line lengths (30mm-101mm) and diameters (21mm-34mm). Disposable, single-use. Sterilization per ISO 11737 and EN 552. No software or electronic components.
Indications for Use
Indicated for patients undergoing alimentary tract, abdominal, gynecological, pediatric, or thoracic surgical procedures requiring tissue resection, transaction, or anastomosis, and for patients requiring surgical treatment of hemorrhoids and anorectal wall defects.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Chex™ Single Use Curved Intraluminal Circular Stapler (K090821)
Endopath ILS Endoscopic Circular Stapler (K920752)
Chex™ Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids (K090821)
Chex™ Single Use Reloadable Linear Stapler and Reloads (K090821)
Chex™ Linear Cutter Disposable Reloadable Stapler (K090821)
Submission Summary (Full Text)
{0}------------------------------------------------
K101378
## 510(k) Summary
(per 21 CFR 807.92(c))
### 1. Applicant
# JUN 2 9 2010
Suzhou Frankenman Medical Equipment Co., Ltd. 88 Jinfeng Road High-new District Suzhou 215011 · · · China
Contact Person: Zhou Tiejun, Office Manager. Tel: 86 512 6878 6125 Fax: 86 512 6808 0025 E-mail: manager@frankenman.com
Date Prepared: April 12, 2010
- 2. Device Name
| Trade Name: | Frankenman Family of Surgical Staplers |
|----------------------|----------------------------------------|
| Common/ Usual Name: | Staple, Implantable |
| Classification Name: | Implantable Staple |
| Regulation Number: | 878.4750 |
| Product Code: | GDW |
| Classification: | II |
| Panel: | General & Plastic Surgery |
### 3. Predicate Devices
The Frankenman Su: gical Staplers which include:
- Disposable Alimentary Canal Staplers; .
- . Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids;
- . Disposable Reloadable Linear Stapler and Reloads; and
- Disposable Reloadable Linear Cutter Stapler and Reloads; .
are substantially equivalent to the devices listed in Table 1. In fact, the Frankenman Staplers are identical to the Chex Staplers with the exception of the manufacturer and product names as listed on the labeling.
{1}------------------------------------------------
# 510(k) Summary
(per 21 CFR 807.92(c))
### 1. Applicant
Suzhou Frankenman Medical Equipment Co., Ltd. 88 Jinfeng Road High-new District Suzhou 215011 China
Contact Person: Zhou Tiejun, Office Manager Tel: 86 512 6878 6125 Fax: 86 512 6808 0025 E-mail: manager@frankenman.com
Date Prepared: April 12, 2010
### 2. Device Name
| Trade Name: | Frankenman Family of Surgical Staplers |
|----------------------|----------------------------------------|
| Common/ Usual Name: | Staple, Implantable |
| Classification Name: | Implantable Staple |
| Regulation Number: | 878.4750 |
| Product Code: | GDW |
| Classification: | II |
| Panel: | General & Plastic Surgery |
### 3. Predicate Devices
The Frankenman Surgical Staplers which include:
- Disposable Alimentary Canal Staplers; .
- Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids; .
- Disposable Reloadable Linear Stapler and Reloads; and .
- Disposable Reloadable Linear Cutter Stapler and Reloads; .
are substantially equivalent to the devices listed in Table 1. In fact, the Frankenman Staplers are identical to the Chex Staplers with the exception of the manufacturer and product names as listed on the labeling.
{2}------------------------------------------------
| | Predicate Device | | |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------|--------------------------------------|---------------|
| Subject Device | Name | Company | 510(k) Number |
| Disposable Alimentary<br>Canal Staplers | Chex™ Single Use Curved<br>Intraluminal Circular Stapler | Frankenman™<br>International Limited | K090821 |
| Disposable Alimentary<br>Canal Staplers | Endopath ILS Endoscopic<br>Circular Stapler | Ethicon | K920752 |
| Single Use Circular<br>Stapler for Rectal<br>Prolapse and<br>Hemorrhoids | Chex™ Single Use Circular<br>Stapler for Rectal Prolapse and<br>Hemorrhoids | Frankenman™<br>International Limited | K090821 |
| Disposable Reloadable<br>Linear Stapler and<br>Reloads | Chex™ Single Use Reloadable<br>Linear Stapler and Reloads | Frankenman™<br>International Limited | K090821 |
| | Disposable Linear Stapler (LSF) | Ethicon | K822345 |
| Disposable Reloadable<br>Linear Cutter Stapler<br>and Reloads | Chex™ Linear Cutter Disposable<br>Reloadable Stapler | Frankenman™<br>International Limited | K090821 |
### 4. Indications for Use
The indications for use for the Frankenman Family of Staplers include:
#### Disposable Alimentary Canal Stapler �
The Frankenman Disposable Alimentary Canal Stapler is used throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic techniques.
#### . Single Use Circular Stapler for Rectal Prolapse and Hemorrhoid
The Frankenman Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids is a Circular Stapler product, with accessories, that has application for general surgical treatment of haemorrhoids and anorectal wall defects by means of transanal stapling and resection of mucosal and musculo-mucosal tissue resulting in occlusion of haemorrhoidal inflow, restoring the haemorrhoidal tissue to its normal physiological position.
#### . Disposable Reloadable Linear Stapler and Reloads
The Frankenman Disposable Reloadable Linear Stapler is used in the resection or transaction of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures.
#### Disposable Reloadable Linear Cutter Stapler and Reloads .
The Frankenman Disposable Reloadable Linear Cutter Stapler has application in abdominal, gynecological, thoracic and pediatric surgery transaction, resection, and the creation of anastomoses.
{3}------------------------------------------------
| | Predicate Device | | |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------|--------------------------------------|---------------|
| Subject Device | Name | Company | 510(k) Number |
| Disposable Alimentary<br>Canal Staplers | Chex™ Single Use Curved<br>Intraluminal Circular Stapler | Frankenman™<br>International Limited | K090821 |
| | Endopath ILS Endoscopic<br>Circular Stapler | Ethicon | K920752 |
| Single Use Circular<br>Stapler for Rectal<br>Prolapse and<br>Hemorrhoids | Chex™ Single Use Circular<br>Stapler for Rectal Prolapse and<br>Hemorrhoids | Frankenman™<br>International Limited | K090821 |
| Disposable Reloadable<br>Linear Stapler and<br>Reloads | Chex™ Single Use Reloadable<br>Linear Stapler and Reloads | Frankenman™<br>International Limited | K090821 |
| | Disposable Linear Stapler (LSF) | Ethicon | K822345 |
| Disposable Reloadable<br>Linear Cutter Stapler<br>and Reloads | Chex™ Linear Cutter Disposable<br>Reloadable Stapler | Frankenman™<br>International Limited | K090821 |
### 4. Indications for Use
The indications for use for the Frankenman Family of Staplers include:
#### Disposable Alimentary Canal Stapler .
The Frankenman Disposable Alimentary Canal Stapler is used throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic techniques.
#### . Single Use Circular Stapler for Rectal Prolapse and Hemorrhoid
The Frankenman Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids is a Circular Stapler product, with accessories, that has application for general surgical treatment of haemorrhoids and anorectal wall defects by means of transanal stapling and resection of mucosal and musculo-mucosal tissue resulting in occlusion of haemorrhoidal inflow, restoring the haemorrhoidal tissue to its normal physiological position.
#### Disposable Reloadable Linear Stapler and Reloads .
The Frankenman Disposable Reloadable Linear Stapler is used in the resection or transaction of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures.
#### Disposable Reloadable Linear Cutter Stapler and Reloads .
The Frankenman Disposable Reloadable Linear Cutter Stapler has application in abdominal, gynecological, thoracic and pediatric surgery transaction, resection, and the creation of anastomoses.
{4}------------------------------------------------
#### Description of the Devices 5.
The Frankenman Staplers were designed in reference to the general principles of surgical staplers. Each stapler/ instrument is activated by squeezing the handle firmly as far as it will go. Specifics for each stapler include:
- The Frankenman Disposable Alimentary Canal Stapler places a circular, double staggered . row of titanium staples. Immediately after staple formation, the instrument's knife blade resects the excess tissue, creating a circular anastomosis. The diameter of the staple line is determined by the selection of the 21mm, 28mm, or 32mm stapler.
- . The Frankenman Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids places two circular peripheral lines of alternating and overlapping staples, thereby sealing off the rectal mucosa above the anal canal. The central circular cutting blade cuts the surplus tissue after the sealing to reconstruct the rectal mucosa. The diameter of the staple line is determined by the selection of the 32mm or 34mm stapler.
- The Frankenman Disposable Reloadable Linear Stapler places a double(or triple in the . case of the LS30-W) staggered row of titanium staples used for mechanical suturing and closure of tissue, prior to the removal of excess tissue. The Linear Stapler is available in 30mm, 45mm, 60mm, and 90mm line lengths for use in various applications. The instrument may be reloaded during a single procedure but cannot be reloaded more than seven times for a maximum of eight firings per instrument.
- . The Frankenman Disposable Reloadable Linear Cutter Stapler delivers two doubled staggered rows of titanium staples and anastomoses the internal tissues and incisions during surgical procedures. The Linear Cutter stapler is available in three staple line lengths (61mm, 81mm, or 101mm). The instrument may be reloaded during a single procedure but cannot be reloaded more than seven times for a maximum of eight firings per instrument. The use of the instrument with staple buttressing materials, or across a previous staple line, may reduce the number of firings.
### 6. Summary of Performance Data
Suzhou Frankenman tested each stapler to validate both physical (i.e., appearance, dimensions, stapler compatibility with cartridge, sterility per ISO 11737 and EN 552) and performance characteristics (strength, closure performance). In addition, a performance test was conducted comparing the Frankenman Disposable Alimentary Canal Stapler and the Frankenman Disposable Reloadable Linear Stapler to analogous Ethicon staplers. Based on this test, the design and construction of the Frankenman and Ethicon staplers were determined to be substantially equivalent. Finally, clinical testing of each stapler line produced superior results (in terms of post-operative healing, pain management, anastomotic leakage, and bleeding) as compared to manual suturing and competitors' devices.
### 7. Safety & Effectiveness
There are no substantial differences between the Frankenman Family of Staplers defined in this 510(k) submission and the predicate devices. They have the same or similar Indications for Use. In addition, the minor differences in technological characteristics do not raise issues of safety and effectiveness.
Page 5-3
{5}------------------------------------------------
#### Description of the Devices 5.
The Frankenman Staplers were designed in reference to the general principles of surgical staplers. Each stapler/ instrument is activated by squeezing the handle firmly as far as it will go. Specifics for each stapler include:
- The Frankenman Disposable Alimentary Canal Stapler places a circular, double staggered row of titanium staples. Immediately after staple formation, the instrument's knife blade resects the excess tissue, creating a circular anastomosis. The diameter of the staple line is determined by the selection of the 21mm, 28mm, or 32mm stapler.
- The Frankenman Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids places . two circular peripheral lines of alternating and overlapping staples, thereby sealing off the rectal mucosa above the anal canal. The central circular cutting blade cuts the surplus tissue after the sealing to reconstruct the rectal mucosa. The diameter of the staple line is determined by the selection of the 32mm or 34mm stapler.
- The Frankenman Disposable Reloadable Linear Stapler places a double(or triple in the . case of the LS30-W) staggered row of titanium staples used for mechanical suturing and closure of tissue, prior to the removal of excess tissue. The Linear Stapler is available in 30mm, 45mm, 60mm, and 90mm line lengths for use in various applications. The instrument may be reloaded during a single procedure but cannot be reloaded more than seven times for a maximum of eight firings per instrument.
- The Frankenman Disposable Reloadable Linear Cutter Stapler delivers two doubled . staggered rows of titanium staples and anastomoses the internal tissues and incisions during surgical procedures. The Linear Cutter stapler is available in three staple line lengths (61mm, 81mm, or 101mm). The instrument may be reloaded during a single procedure but cannot be reloaded more than seven times for a maximum of eight firings per instrument. The use of the instrument with staple buttressing materials, or across a previous staple line, may reduce the number of firings.
### 6. Summary of Performance Data
Suzhou Frankenman tested each stapler to validate both physical (i.e., appearance, dimensions, stapler compatibility with cartridge, sterility per ISO 11737 and EN 552) and performance characteristics (strength, closure performance). In addition, a performance test was conducted comparing the Frankenman Disposable Alimentary Canal Stapler and the Frankenman Disposable Reloadable Linear Stapler to analogous Ethicon staplers. Based on this test, the design and construction of the Frankenman and Ethicon staplers were determined to be substantially equivalent. Finally, clinical testing of each stapler line produced superior results (in terms of post-opera:ive healing, pain management, anastomotic leakage, and bleeding) as compared to manual suturing and competitors' devices.
### 7. Safety & Effectiveness
There are no substantial differences between the Frankenman Family of Staplers defined in this 510(k) submission and the predicate devices. They have the same or similar Indications for Use. In addition, the minor differences in technological characteristics do not raise issues of safety and effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle symbol with three curved lines representing the eagle's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Suzhou Frankenman Medical Equipment Co., Ltd % Emergo Group, Inc. Jean Asquith 1705 S. Capital of Texas Highway, Suite 500 Austin, Texas 78746
JUN 2 9 2010
Re: K101378
Trade/Device Name: Frankenman Surgical Staplers Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: April 12, 2010 Received: May 21, 2010
Dear Jean Asquith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the includines for use stated in the enclosure) to legally marketed predicate devices marked in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Fedral Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleding.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other maden wor not not hits incan or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{7}------------------------------------------------
## Page 2 - Jean Asquith
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Mark N. Millman
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ KIOI.378
Device Name: Frankenman Surgical Staplers
Indications for Use:
Frankenman Staplers are indicated as follows:
#### Disposable Alimentary Canal Stapler .
The Frankenman Disposable Alimentary Canal Stapler is used throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic techniques.
#### Single Use Circular Stapler for Rectal Prolapse and Hemorrhoid .
The Frankenman Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids is a Circular Stapler product, with accessories, that has application for general surgical treatment of haemorrhoids and anorectal wall defects by means of transanal stapling and resection of mucosal and musculo-mucosal tissue resulting in occlusion of haemorrhoidal inflow, restoring the haemorrhoidal tissue to its normal physiological position.
# Disposable Reloadable Linear Stapler and Reloads The Frankenman Disposable Reloadable Linear Stapler (and Reloads) is used in the resection or transaction of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures.
# Disposable Reloadable Linear Cutter Stapler and Reloads The Frankenman Disposable Reloadable Linear Cutter Stapler (and Reloads) has application in abdominal, gynecological, thoracic and pediatric surgery transaction, resection, and the creation of anastomoses.
Prescription Use Over-The-Counter Use X AND/OR {Part 21 CFR 801 Subpart D} (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Donald Kione for MXM
of Surgical. Onhopedic. storative Devices
510(k) Number: K101378
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.