AMID STAPLER

K093253 · Safestitch, LLC · GDW · Nov 12, 2009 · General, Plastic Surgery

Device Facts

Record IDK093253
Device NameAMID STAPLER
ApplicantSafestitch, LLC
Product CodeGDW · General, Plastic Surgery
Decision DateNov 12, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SafeStitch LLC AMID Stapler & Non-Absorbable Staples has applications in general surgery procedures for fixation of mesh, in the repair of hernia defects and in other surgical specialties for the approximation of tissue(s), including skin.

Device Story

The AMID Stapler is a sterile, single-use, disposable manual surgical stapler containing 17 titanium staples. It is used by surgeons in general surgery and other specialties to approximate tissue or fixate mesh during hernia repair. The device operates manually; the surgeon triggers the mechanism to deploy staples into the target site. The device features an angled, non-swiveling firing end to facilitate access. By providing a mechanical means of tissue approximation and mesh fixation, the device assists in surgical closure and repair, potentially improving procedural efficiency compared to manual suturing.

Clinical Evidence

Bench testing only. Testing was performed to verify performance against internal specifications and to demonstrate substantial equivalence to the predicate devices.

Technological Characteristics

Sterile, single-use, disposable manual stapler. Materials include titanium staples. Features an angled, non-swiveling firing end. Mechanical actuation.

Indications for Use

Indicated for general surgery procedures requiring mesh fixation, hernia repair, and approximation of tissues including skin.

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K093253 រូបរិ l i : i । | Company: | SafeStitch LLC | NOV 12 2009 | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | Contact: | Stewart B. Davis M.D.<br>Chief Operating Officer<br>SafeStitch LLC and SafeStitch Medical Inc.<br>4400 Biscayne Boulevard<br>Suite A-100<br>Miami, FL 33137<br>Phone 305.575.4145<br>Fax 305.575.4130 | | | Trade Name: | AMID Stapler | | | Device Type: | Surgical Stapler | | | Classification Regulation: | 878.4750 | | | Class: | II | | | Panel: | General and Plastic Surgery | | | Product Code: | GDW | | | Predicate Devices: | AutoSuture Surgical Stapling Instrument, United States<br>Surgical Corporation, a division of Tyco Healthcare<br>(K771177)<br>AutoSuture Disposable Stapling Instrument, United<br>States Surgical Corporation, a division of Tyco<br>Healthcare (K780695)<br>AutoSuture Titanium Surgical Staples, United States<br>Surgical Corporation, a division of Tyco Healthcare<br>(K855047) | | | Device Description: | The AMID Stapler is a sterile, single use disposable<br>stapler. The AMID Stapler consists of a manual stapler<br>and 17 titanium staplers. It is designed for the stapling of<br>tissue and mesh, specifically for hernia repairs | | | Indications for use: | The SafeStitch AMID Stapler has application in general<br>surgery procedures for fixation of mesh, in the repair of<br>hernia defects and in other surgical specialties for the<br>approximation of tissue(s), including skin. | | | Technological Characteristics: | The SafeStitch AMID Stapler is similar to the predicate devices in design and operation. The primary difference are the firing end does not swivel and the tip is angled. | | | Performance Data: | Bench testing was performed to verify the AMID Stapler's performance to internal specifications. In addition, bench testing was also performed to demonstrate that the AMID Stapler is substantially equivalent to the predicate device(s). | | ## SafeStitch AMID Stapler 510(k) Summary .j ( ! ・ ; {1}------------------------------------------------ ・ . *, .. . CONFIDENTIAL 10 Samment of Children Sammer Sample Career } - ( : l {2}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 Safestitch LLC % Stewart B. Davis, M.D. Chief Operating Officer 4400 Biscayne Boulevard, Suite A-100 Miami. Florida 33137 JAN - 4 2010 Re: K093253 Trade/Device Name: Safestitch LLC AMID Stapler & Non-Absorbable Staples Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, GAG Dated: November 12, 2009 Received: November 12, 2009 Dear Dr. Davis: This letter corrects our substantially equivalent letter of November 12, 2009. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other {3}------------------------------------------------ Page 2 - Stewart B. Davis, M.D. requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark N. Malkerson Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K093253 Device Name: SafeStitch LLC AMID Stapler & Non-Absorbable Staples The SafeStitch LLC AMID Stapler & Non-Absorbable Staples has applications in general surgery procedures for fixation of mesh, in the repair of hernia defects and in other surgical specialties for the approximation of tissue(s), including skin. | Prescription Use | X | |-----------------------------|---| | (Part 21 CFR 801 Subpart D) | | | AND/OR | | | Over-The-Counter Use | | | (21 CFR 807 Subpart C) | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices | 510(k) Number | 1093253 | |---------------|---------| |---------------|---------| Page 1 of ___
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...