ODONTIT AUTOGENOUS BONE COLLECTION DEVICE

K971036 · Interpore Intl. · GCY · Jun 3, 1997 · General, Plastic Surgery

Device Facts

Record IDK971036
Device NameODONTIT AUTOGENOUS BONE COLLECTION DEVICE
ApplicantInterpore Intl.
Product CodeGCY · General, Plastic Surgery
Decision DateJun 3, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4680
Device ClassClass 1
AttributesTherapeutic

Intended Use

The Odontit Autogenous Bone Collection Device is indicated for use as an aspirator for the removal and collection of blood, autogenous bone chips, or other extraneous fluids from a surgical site. Using a vacuum source, the Odontitit aspirator will remove materials (including blood, bone chips and other fluids) from the surgical site and pass them through a stainless steel screen which is designed to capture and collect autogenous bone. At the option of the surgeon, this bone may be removed from the aspirator, washed and cleansed as necessary, and returned to the same patient as an autogenous bone graft. The capacity for bone collection in the screen is approximately 16 cc.

Device Story

Portable suction device; connects to external vacuum source. Components: body assembly, aspirator tip, screen filter, flex/rigid tubing, plunger. Operates by aspirating surgical site materials; stainless steel screen captures bone chips (16 cc capacity). Used by surgeons in clinical settings. Surgeon separates device body, removes filter, cleanses collected bone chips, and uses plunger to compact material for grafting. Benefits patient by enabling autogenous bone collection during standard aspiration procedures.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Portable suction apparatus; stainless steel screen filter; plastic body/tubing components. Manual operation via external vacuum source. Single-patient use; non-sterile (aseptic technique required).

Indications for Use

Indicated for patients requiring autogenous bone graft harvesting during surgery. Used by surgeons to aspirate and collect bone chips, blood, and fluids from surgical sites for potential autograft reuse.

Regulatory Classification

Identification

A nonpowered, single patient, portable suction apparatus is a device that consists of a manually operated plastic, disposable evacuation system intended to provide a vacuum for suction drainage of surgical wounds.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) Premarket Notification Odontit Autogenous Bone Collection Device ODONTIT S.A. - Confidential - 510(k) SUMMARY K971036 ## SUBMITTED BY Mario Gerzberg President ODONTIT S.A. Av. Rivadavia 5032 1*7 1424 Buenos Aires ARGENTINA Telephone: 011-541-902-9500 Facsimile: 011-541-903-9330 Date Submitted: March 20, 1997 ## CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME Classification Name: Non-powered, Single Patitent, Portable Suction Apparatus Common/Usual Name: Disposable Aspirator Proprietary Name: Odontit Autogenous Bone Collection Device ## PREDICATE DEVICE KAM Super Sucker™ [reference 510(k) K822065 cleared 8/16/82] ## DEVICE DESCRIPTION The Odontit Autogenous Bone Collection Device is a portable suction device intended for use as an aspirator for the removal and collection of blood, autogenous bone, or other extraneous fluids from a surgical site. The device consists of the following components: Body Assembly, Aspirator Tip, Screen Filter, Flex tubing, Rigid tubing and a Plunger. ## PRODUCT USE 1. The package is opened and the contents removed. 2. The rigid tubing connector is attached to a vacuum source. 3. The aspirator tip is attached to the flexible tubing. 4. The Autogenous Bone Collection Device is used in accordance with standard surgical procedures and as required to remove blood, autogenous bone chips or other extraneous fluids from the surgical site. 5. The lower body of the device is then aseptically separated from the upper body by twisting and carefully separating the two components. 6. The Screen Filter assembly is then carefully removed using aseptic technique. 7. Wash or rinse the collected bone chips per standard procedure if deemed necessary. JUN - 3 1997 15 {1} 510(k) Premarket Notification Odontit S.A. - Confidential - Odontit Autogenous Bone Collection Device 8. Use the plunger to compact the bone chips as required. 9. Remove the Filter Standoff to harvest the bone chips for delivery to the surgical site. ## INDICATIONS FOR USE: The Odontit Autogenous Bone Collection Device is indicated for use as an aspirator for the removal and collection of blood, autogenous bone chips, or other extraneous fluids from a surgical site. Using a vacuum source, the Odontititit aspirator will remove materials (including blood, bone chips and other fluids) from the surgical site and pass them through a stainless steel screen which is designed to capture and collect autogenous bone. At the option of the surgeon, this bone may be removed from the aspirator, washed and cleansed as necessary, and returned to the same patient as an autogenous bone graft. The capacity for bone collection in the screen is approximately 16 cc. ## WARNINGS AND PRECAUTIONS Aseptic Technique Required Warning: For one time use only - Do not resterilize. Caution: Federal (USA) law restricts the use of this device to sale by or on the order of a licensed physician. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS The product design, material of construction, and function as an aspirator is substantially equivalent to the FDA cleared predicate device. 16 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN - 3 1997 Mr. Mario Gersberg President Interpore International 181 Technology Drive Irvine, California 92618 Re: K971036 Trade Name: Odontit Autogenous Bone Collection Device Regulatory Class: I Product Code: GCY Dated: March 20, 1997 Received: March 20, 1997 Dear Mr. Gersberg: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3} Page 2 - Mr. Mario Gersberg This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} 510(k) Premarket Notification Odontit Autogenous Bone Collection Device ODONTIT S.A. - Confidential - 510(k) Number (if known): Not Known K971036 Device Name: Odontit Autogenous Bone Collection Device ## Indications for Use: The Odontit Autogenous Bone Collection Device is indicated for use as an aspirator for the removal and collection of blood, autogenous bone chips, or other extraneous fluids from a surgical site. Using a vacuum source, the Odontit aspirator will remove materials (including blood, bone chips and other fluids) from the surgical site and pass them through a stainless steel screen which is designed to capture and collect autogenous bone. At the option of the surgeon, this bone may be removed from the aspirator, washed and cleansed as necessary, and returned to the same patient as an autogenous bone graft. The capacity for bone collection in the screen is approximately 16 cc. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-1.jpeg](img-1.jpeg) Prescription Use ☑ (PER 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...