K193530 · Teleflex Medical · GAW · Mar 19, 2020 · General, Plastic Surgery
Device Facts
Record ID
K193530
Device Name
Deklene MAXX
Applicant
Teleflex Medical
Product Code
GAW · General, Plastic Surgery
Decision Date
Mar 19, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Deknatel® Deklene® Maxx™ Polypropylene Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular and neurological procedures, but not for use in ophthalmic procedures.
Device Story
Nonabsorbable, sterile, monofilament polypropylene surgical suture; used for soft tissue approximation and ligation in cardiovascular and neurological procedures. Available in USP sizes 8-0 through 0; supplied with or without 300/400 series stainless steel needles and PTFE felt pledgets. Used by surgeons in clinical settings. Provides mechanical tissue closure; maintains tensile strength over time. MR safe (needles removed).
Clinical Evidence
No clinical trials were conducted. Substantial equivalence is supported by bench testing (USP non-absorbable suture requirements) and biocompatibility evaluation per ISO 10993-1:2018.
Technological Characteristics
Monofilament polypropylene (synthetic linear polyolefin). USP sizes 8-0 to 0. 300/400 series stainless steel needles. PTFE felt pledgets. Sterilized via Ethylene Oxide (SAL 10^-6). Compliant with USP <861>, <871>, <881> and ISO 10993-1. MR safe (needles removed).
Indications for Use
Indicated for general soft tissue approximation and/or ligation, including cardiovascular and neurological procedures. Contraindicated for ophthalmic procedures.
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the words "Convert to PDF" in a large, bold font. The text is centered and appears to be the main focus of the image. The background is a light gray color, which provides a neutral backdrop for the text.
Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 19, 2020
Teleflex Medical Robin Haden Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, North Carolina 27560
Re: K193530
Trade/Device Name: Deklene MAXX Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable Polypropylene Surgical Suture Regulatory Class: Class II Product Code: GAW Dated: December 19, 2019 Received: December 20, 2019
Dear Robin Haden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193530
#### Device Name
Deknatel® Deklene® MAXX™ Polypropylene Surgical Suture
Indications for Use (Describe)
Deknatel® Deklene® MAXX™ Polypropylene Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, and neurological procedures, but not for use in ophthalmic procedures.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
| Submission Date:<br>K193530 | 12/20/2019 |
|-----------------------------|----------------------------------------------------------------|
| SUBMITTER INFORMATION: | |
| Company Name: | Teleflex Medical, Inc. |
| Company Address: | 3015 Carrington Mill Blvd, Morrisville, NC 27560 |
| Contact Person: | Robin Haden<br>Phone: 919-544-8000<br>robin.haden@teleflex.com |
| Device Trade Name: | Deknatel® Deklene® Maxx™ Polypropylene Surgical Suture |
| Device Common Name: | Polypropylene nonabsorbable surgical suture |
| Class: | Class II |
| Classification: | 21 CFR 878.5010 |
| Product Code: | GAW |
# Predicate Devices:
The predicate is Deknatel® Deklene® Maxx™ Polypropylene Suture devices cleared under K930738 (July 26, 1994) and K153076 (June 16, 2016).
# Device Description:
Deknatel® Deklene® Maxx™ Polypropylene Suture is a nonabsorbable, sterile surgical suture composed of a strand of polypropylene, a synthetic linear polyolefin. Deknatel® Deklene® MAXX™ Polypropylene Surgical Suture is available dyed (blue) or undyed (colorless).
Deknatel® Deklene® Maxx™ Polypropylene Surgical Suture meets all nonabsorbable surgical suture requirements established by the United States Pharmacopeia (USP) except for suture sizes 7-0 and 8-0 which differ from USP in diameter only.
Deknatel® Deklene® Maxx™ Polypropylene Surgical Suture is available in USP sizes 8-0 through 0. The suture is provided in a variety of lengths, with and without 300 or 400 series stainless steel needles, with and without PTFE felt pledgets, and may be supplied in a variety of cut lengths. Finished suture may be packaged in cartons as single packs, multipacks or procedure packs.
{5}------------------------------------------------
# Intended Use:
Deknatel® Deklene® Maxx™ Polypropylene Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular and neurological procedures, but not for use in ophthalmic procedures.
{6}------------------------------------------------
# Technological Characteristics
The proposed Deknatel® Deklene® Maxx™ Polypropylene Suture is substantially equivalent to its predicate Deknatel® Deklene® Maxx™ Polypropylene Suture devices because there are no differences in technological or performance characteristics between the proposed and predicate devices. The difference between the proposed device and the predicate device is the supplier from which the resin used for extrusion is acquired. The resin from the new supplier is equivalent to that from the current supplier and does not introduce any new issues of safety and effectiveness. The results of the comparison between the technological characteristics of Deknatel® Deklene® Maxx™ Polypropylene Suture manufactured with the proposed and predicate material demonstrate that the proposed Deknatel® Deklene® Maxx™ Polypropylene Sutures are equivalent to the marketed predicate devices in performance characteristics.
# Materials
All patient contacting materials are in compliance with the requirements of ISO 10993-1. The proposed modification is replacing the source of the polypropylene resin used to manufacture the Deknatel® Deklene® Maxx™ Polypropylene Suture.
# Summary of Testing
Deknatel® Deklene® Maxx™ Polypropylene Suture is tested in accordance with the requirements of USP – non-absorbable surgical sutures for suture diameter, , tensile strength and needle attachment, and meet the requirements of Class II Special Controls Guidance: Surgical Sutures; Guidance for Industry and FDA; June 3, 2003.
All materials used in the fabrication of the Deknatel® Maxx™ Polypropylene Suture were evaluated through biological qualification safety tests as outlined in ISO 10993-1:2018 – Biological Evaluation of Medical Devices - Part I: Evaluation and Testing.
Deknatel® Deklene® Maxx™ Polypropylene Suture is tested to demonstrate it is "MR Safe" and poses no hazards in an MR environment with any needles removed.
{7}------------------------------------------------
# Substantial Equivalence:
The proposed Deknatel® Deklene® Maxx™ Polypropylene Surgical Suture is substantially equivalent in intended use and fundamental scientific technology to the Deknatel® Deklene® Maxx™ Polypropylene Suture devices cleared under K930738 (July 26, 1994) and K153076 (June 16, 2016). The substantial equivalence table for the comparison of the previously cleared devices and the proposed devices is included in the Table below.
| Comparative<br>Characteristics | Predicate Device<br>(K153076, K930738) | Proposed Device | Explanation<br>of Differences |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|--------------------------------------|
| Manufacturer | Teleflex Medical | | Identical |
| Product Code | GAW | | Identical |
| Regulation | 878.5010 | | Identical |
| Device Class | 2 | | Identical |
| Indications for Use | Polypropylene Surgical Sutures are indicated for use in<br>general soft tissue approximation and/or ligation, including<br>use in cardiovascular, and neurological procedures, but not for<br>use in ophthalmic procedures. | | Identical |
| Contraindications | None known. | | Identical |
| Trade Names | Deklene® Maxx™ | | Identical |
| Deklene®<br>Maxx™ USP<br>Size | 8-0 to 2 | 8-0 to 0 | Equivalent<br>within stated<br>range |
| Suture Diameter | All sizes except 7-0 and 8-0 meet USP <861> requirements<br>for nonabsorbable surgical suture | | Identical |
| Suture Tensile<br>Strength | All sizes meet USP <881> requirements for nonabsorbable<br>surgical suture | | Identical |
| Needle Attachment<br>Strength | All sizes meet USP <871> requirements for nonabsorbable<br>surgical suture | | Identical |
| Material | Polypropylene (synthetic linear polyolefin) | | Identical |
| Suture Construction | Monofilament | | Identical |
| Coated | No | | Identical |
| Uncoated | Yes | | Identical |
| Undyed White | Yes | | Identical |
| Dyed Blue with<br>[phthalocyaninato(2<br>-)] copper | Yes | | Identical |
| Impregnated PTFE<br>Pledget | Yes | | Identical |
| Pledget Material | polytetrafluoroethylene (PTFE) | | Identical |
| Needle Material | | 300 or 400 Series Stainless Steel | Identical |
| Package Materials | Solid Sulfite Board (SBS)/ Uncoated Tyvek Pouch | | Identical |
| Sterility Method | Ethylene Oxide (EO);<br>SAL 10-6 | | Identical |
| Biocompatibility | Per ISO 10993-1 | | Identical |
| Shelf Life | | 5 Years | Identical |
{8}------------------------------------------------
# Clinical Tests Performed:
# No clinical trials were conducted
# Conclusion:
The proposed Deknatel® Deklene® Maxx™ Polypropylene Suture is substantially equivalent to its predicate Deknatel® Deklene® Maxx™ Polypropylene Suture devices because there are no differences in technological or performance characteristics between the proposed and predicate devices.
The proposed modification is replacing the source of the polypropylene resin used to manufacture the Deknatel® Deklene® Maxx™ Polypropylene Suture. It also has the same design being a sterile, flexible, monofilament nonabsorbable thread meeting the requirements of the United States Pharmacopeia.
The biocompatibility data and the results of performance testing presented, demonstrate the substantial equivalence of Deklene® Maxx™ Polypropylene Suture to that of the predicate devices. It further demonstrates conformance with the USP, ISO 10993 and FDA Guidance for Surgical Suture 510(k).
The new device Deknatel® Deklene® MAXX™ Polypropylene Surgical Suture is substantially equivalent in intended use and fundamental scientific technology to the predicate devices Deknatel® Deklene® Maxx™ Polypropylene Suture cleared under K930738 and K153076.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.