Tevdek II, Silky II Polydek, 'Cottony' II, NextStitch
K192490 · Teleflex Medical · GAT · Mar 30, 2020 · General, Plastic Surgery
Device Facts
Record ID
K192490
Device Name
Tevdek II, Silky II Polydek, 'Cottony' II, NextStitch
Applicant
Teleflex Medical
Product Code
GAT · General, Plastic Surgery
Decision Date
Mar 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Polyester Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, orthopedic and neurological procedures. Polyester Surgical Suture Tapes are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, orthopedic and neurological procedures. TEVDEK® II NextStitch® and 'silky' II POLYDEK® NextStitch® cardiovascular valve suture are indicated for use for soft tissue approximation and/or ligation in cardiovascular valve replacement procedures.
Device Story
Non-absorbable surgical sutures composed of twisted or braided Polyethylene Terephthalate (PET) fibers; available undyed or dyed green; offered with or without stainless steel needles and PTFE pledgets. Configurations include uncoated 'cottony' II, lightly PTFE-impregnated 'silky' II/Polydek, and heavily PTFE-impregnated Tevdek II. NextStitch variant provides continuous chain of linked sutures for cardiovascular valve replacement. Used by surgeons in OR settings for tissue approximation/ligation. Device provides mechanical closure of surgical sites; benefits patient through secure tissue approximation. Subject device represents a change in source of PET resin and PTFE coating materials compared to predicates.
Clinical Evidence
Bench testing only. Performance verified per USP 40-NF 35 <861> (Diameter), <871> (Needle Attachment), and <881> (Tensile Strength). Biocompatibility assessed per ISO 10993-1. Stability and MR safety testing performed.
Technological Characteristics
Non-absorbable braided/twisted PET fibers. PTFE coating (varying density). 300/400 series stainless steel needles. USP sizes 7-0 to 9. Sterilization via Ethylene Oxide (EO) to SAL 10^-6. Compliant with ISO 10993-1 for biocompatibility.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in cardiovascular, ophthalmic, orthopedic, and neurological procedures; cardiovascular valve replacement procedures for NextStitch variants. No contraindications known.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
silky II POLYDEK, 'cottony' II, TEVDEK II, NextStitch, DEKLENE II, DEKLENE MAXX Gabbay-Frater, NYLON, SILK, STAINLESS STEEL (K153076)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
March 30, 2020
Teleflex Medical Robin Haden Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, North Carolina 27560
Re: K192490
Trade/Device Name: Tevdek II, "Silky" II Polydek, 'Cottony' II, NextStitch Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: February 21, 2020 Received: February 24, 2020
Dear Ms. Haden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192490
Device Name
"silky" II POLYDEK®, 'cottony' II, TEVDEK® II Polyester Surgical Suture
Indications for Use (Describe)
Polyester Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, orthopedic and neurological procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192490
Device Name
'cottony' II Polyester Surgical Suture Tape
Indications for Use (Describe)
Polyester Surgical Suture Tapes are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, orthopedic and neurological procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192490
Device Name
TEVDEK® II NextStitch® and "silky" II POLYDEK® NextStitch® Cardiovascular Valve Suture
Indications for Use (Describe)
TEVDEK® II NextStitch® and 'silky' II POLYDEK® NextStitch® cardiovascular valve suture are indicated for use for soft tissue approximation and/or ligation in cardiovascular valve replacement procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) SUMMARY
## Deknatel® Polyester Surgical Suture
## A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Inc. 3015 Carrington Mill Blvd Morrisville, NC 27560
Phone: 919-544-8000 919-433-3914 Fax:
# B. Contact Person
Robin Haden Surgical Business Unit Tel – (919) 433-8046 Fax - (919) 433-3914 Email – robin.haden@teleflex.com
# C. Date Prepared
March 18, 2020
### D. Device Name
| Trade Name(s) | Deknatel® Polyester Surgical Sutures, including: | | |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | – "silky" II POLYDEK® Polyester Surgical Suture<br>– 'cottony' II Polyester Surgical Suture and Tape<br>– TEVDEK® II Polyester Surgical Suture<br>– TEVDEK® II NextStitch® Cardiovascular Valve Suture<br>– "silky" II POLYDEK® NextStitch® Cardiovascular<br>Valve Suture | | |
| Product Code | GAT (Suture, Nonabsorbable, Synthetic, Polyethylene) | | |
| Regulation | 21 CFR 878.5000 | | |
| Classification Name | Nonabsorbable poly(ethylene terephthalate) surgical suture | | |
{6}------------------------------------------------
# E. Device Description
The Deknatel® Polyester Surgical Suture is a non-absorbable surgical suture composed of twisted or braided Polyethylene Terephthalate fibers. Deknatel® Polyester Surgical Suture is available undyed (white) or dyed (green) and is offered in a variety of cut lengths, with or without needles, and with or without pledgets. Product labeling carries the following specific information on the suture: material (including color, number and length of sutures in the package); USP size (diameter, including any difference in diameter); the number, profile and size of any attached needles; and the number, size and density of pre-attached pledgets.
The Deknatel® Polyester Surgical Suture meets all nonabsorbable surgical suture requirements established by the United States Pharmacopeia (USP) for tensile strength, diameter and needle attachment, except for suture size 7-0 which differs from USP in diameter only.
Deknatel® Polyester Surgical Suture is available in USP sizes 7-0 to 9 in the following configurations:
- . 'cottony' II: Uncoated (Not PTFE impregnated)
- "silky II" or Polydek®: Lightly coated (PTFE impregnated) .
- Tevdek® II: Heavily coated (more PTFE impregnated than "silky")) .
The Deknatel® Polyester Surgical Sutures are also available in the following productspecific configurations with or without needles, and with or without pledgets:
- Surgical Suture Tape as uncoated 'cottony' II in flat braid sizes of 1/8", 1mm, 2 mm, ● 3mm, 4mm and 5mm.
- . NextStitch® Cardiovascular Valve Suture is comprised of "silky" II POLYDEK and/or TEVDEK II polyester suture. NextStitch is a continuous chain of linked sutures designed to provide an alternative suturing technique for cardiovascular valve replacement procedures. NextStitch Sutures are available in USP Sizes 4-0, 3-0, 2-0 and 0 (metric sizes 1.5 through 3.5).
- Readi-Cut TEVDEK® II Bulk Non-Sterile Polyester Surgical Suture is bulk, non-● sterile Tevdek® II polyester suture available in USP Size 2.
# F. Indications for Use and Contraindications
### Sterile Deknatel® Polyester Suture, previously cleared under K153076 and K153089. carries the following indications and contraindications:
# "silky" II POLYDEK®, 'cottony' II, TEVDEK® II Polyester Surgical Suture and Tape
#### Indications:
Polyester Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, orthopedic and neurological procedures.
{7}------------------------------------------------
Contraindications: None known.
# TEVDEK® II NextStitch® and "silky" II POLYDEK® NextStitch® Cardiovascular Valve Suture
#### Indications:
TEVDEK II NextStitch and "silky" II POLYDEK NextStitch cardiovascular valve suture are indicated for use for soft tissue approximation and/or ligation in cardiovascular valve replacement procedures.
Contraindications: None known.
Non-sterile Deknatel® Polyester Suture, previously cleared under K930739, carries the following indications and contraindications:
#### READI-CUT TEVDEK® II Non-Sterile Polyester Surgical Suture
Indications:
Deknatel Polyester Surgical Sutures (i.e. Dacron®, Polydek, and Tevdek) are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurological procedures.
Contraindications: None known.
#### G. Substantial Equivalence
The proposed Deknatel Polyester Surgical Suture is substantially equivalent to the predicate devices cleared below:
| Predicate Device | Manufacturer | 510(k) | Clearance<br>Date |
|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------|---------|-------------------|
| "silky" II POLYDEK, 'cottony' II,<br>TEVDEK II, NextStitch, DEKLENE<br>II, DEKLENE MAXX Gabbay-Frater,<br>NYLON, SILK, STAINLESS STEEL | Teleflex Medical,<br>Inc. | K153076 | June 16, 2016 |
| 'cottony' II Polyester Suture Tape | Teleflex Medical,<br>Inc. | K153089 | July 15, 2016 |
{8}------------------------------------------------
The following reference device information is included to support the inclusion of the nonsterile Deknatel® Polyester Surgical Suture. The proposed non-sterile device is substantially equivalent to the reference non-sterile Deknatel® Polyester Surgical Suture cleared below:
| Reference Device | Manufacturer | 510(k) | Clearance Date |
|--------------------------------------------------|---------------------------|---------|----------------|
| SUTURES POLYPROP.,<br>POLYETHYLENE, NYLON & SILK | Teleflex Medical,<br>Inc. | K930739 | July 26, 1994 |
# H. Comparison To Predicate Devices
The proposed sterile Deknatel® Polyester Surgical Suture is substantially equivalent in intended use and fundamental scientific technology to the Deknatel® Polyester Surgical Suture predicate devices cleared under K153076 on June 16, 2016 and K153089 on July 15, 2016. The proposed non-sterile Deknatel® Polyester Surgical Suture is substantially equivalent in intended use and fundamental scientific technology to the non-sterile Deknatel® Polyester Surgical Suture devices cleared under reference K930739 on July 26, 1994. Nonclinical testing on functionality, stability, biocompatibility, and MR safety was conducted to confirm that the supplier changes did not negatively impact device performance. The testing results concluded that the only difference between the subject and predicate sutures are the sources for the PET resin and the PTFE impregnation/dispersion coating materials.
The following tables contain the technological characteristics of the proposed devices in comparison to the predicate and reference devices.
| Comparative<br>Characteristics | Predicate Device | Proposed Device | Explanation<br>of Differences |
|---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|---------------------------------------------|
| Manufacturer | Teleflex Medical | | Identical |
| 510(k) Number | K153076, K153089 | K192490 | N/A |
| Product Code | GAT | | Identical |
| Regulation | 878.5000 | | Identical |
| Device Class | 2 | | Identical |
| Indications for<br>Use<br>('cottony' II,<br>"silky" II<br>POLYDEK,<br>TEVDEK II<br>Sutures) | Polyester Surgical Sutures are indicated for use in general soft<br>tissue approximation and/or ligation, including use in<br>cardiovascular, ophthalmic, orthopedic and neurological<br>procedures. | | Identical |
| Comparative<br>Characteristics | Predicate Device | Proposed Device | Explanation<br>of Differences |
| Indications for<br>Use<br>('cottony' II<br>Suture Tapes) | Polyester Surgical Suture Tapes are indicated for use in general<br>soft tissue approximation and/or ligation, including use in<br>cardiovascular, ophthalmic, orthopedic and neurological<br>procedures. | | Identical |
| Indications for<br>Use<br>(NextStitch) | TEVDEK II NextStitch and "silky" II POLYDEK NextStitch<br>cardiovascular valve suture are indicated for use for soft tissue<br>approximation and/or ligation in cardiovascular valve<br>replacement procedures. | | Identical |
| Contraindications | | None known. | Identical |
| Trade Names | "silky" II POLYDEK®<br>'cottony' II, TEVDEK® II, NextStitch® | | Identical |
| USP Size<br>('cottony' II,<br>"silky" II<br>POLYDEK,<br>TEVDEK II) | 9-0 to 9 | 7-0 to 9 | Equivalent,<br>within cleared<br>size range |
| USP Size<br>(NextStitch) | 4-0 to 0 | | Equivalent |
| Size<br>'cottony II'<br>Suture Tape | 1/8", 1mm, 2mm, 3mm, 4mm, 5mm | | Equivalent |
| Suture Diameter | All sizes meet USP <861> requirements for nonabsorbable<br>surgical suture except tape which does not conform to USP<br>Diameter | | Identical |
| Suture Tensile<br>Strength | All sizes meet USP <881> requirements for nonabsorbable<br>surgical suture | | Identical |
| Needle<br>Attachment<br>Strength | All sizes meet USP <871> requirements for nonabsorbable<br>surgical suture | | Identical |
| Material | Polyester<br>[Poly(ethylene terephthalate)] | | Identical |
| Suture<br>Construction | Braided | | Identical |
| Impregnated/Coat<br>ed | Yes ("silky" II POLYDEK or TEVDEK II) with<br>Polytetrafluoroethylene | | Identical |
| Uncoated | Yes ('cottony' II) | | Identical |
| Undyed White | Yes | | Identical |
| D&C Green No. 6 | Yes | | Identical |
| Co-braid<br>(White/Green) | Yes | | Identical |
| Impregnated | Yes | | Identical |
| Comparative<br>Characteristics | Predicate Device | Proposed Device | Explanation<br>of Differences |
| PTFE Pledget | | | |
| Pledget Material | polytetrafluoroethylene (PTFE) | | Identical |
| Needle Material | 300 or 400 Series Stainless Steel | | Identical |
| Package Materials | Solid Sulfite Board (SBS)/ Uncoated Tyvek Pouch | | Identical |
| Sterility Method | Ethylene Oxide (EO);<br>SAL 10-6 | | Identical |
| Biocompatibility | Per ISO 10993-1 | | Identical |
| Shelf Life | 5 Years | | Identical |
#### Substantial Equivalence Summary Table (Sterile Product)
{9}------------------------------------------------
{10}------------------------------------------------
# Substantial Equivalence Summary Table (Non-Sterile Product)
| Comparative<br>Characteristics | Reference Device | Proposed Device | Explanation<br>of Differences |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|-------------------------------|
| Manufacturer | Teleflex Medical | Teleflex Medical | Identical |
| 510(k) Number | K930709 | K192490 | N/A |
| Product Code | GAT | GAT | Identical |
| Regulation | 878.5000 | 878.5000 | Identical |
| Device Class | 2 | 2 | Identical |
| Indications for<br>Use<br>TEVDEK® II | Polyester Surgical Sutures are indicated for use in general soft<br>tissue approximation and/or ligation, including use in<br>cardiovascular, ophthalmic, and neurological procedures. | | Identical |
| Contraindications | None known. | | Identical |
| Trade Names | TEVDEK® II | | Identical |
| USP Size<br>TEVDEK® | 2 | | Identical |
| Suture Diameter | All sizes meet USP <861> requirements for nonabsorbable<br>surgical suture except tape which does not conform to USP<br>Diameter | | Identical |
| Suture Tensile<br>Strength | All sizes meet USP <881> requirements for nonabsorbable<br>surgical suture | | Identical |
| Needle<br>Attachment<br>Strength | All sizes meet USP <871> requirements for nonabsorbable<br>surgical suture | | Identical |
| Material | Polyester<br>[Poly(ethylene terephthalate)] | | Identical |
| Suture<br>Construction | Braided | | Identical |
| Impregnated/Coat<br>ed | Polytetrafluoroethylene | | Identical |
| D&C Green No. 6 | Yes | | Identical |
| Package Materials | Solid Sulfite Board (SBS)/ Uncoated Tyvek Pouch | | Identical |
| Sterility Method | N/A- Sold Non-Sterile | | Identical |
{11}------------------------------------------------
| Comparative<br>Characteristics | Reference Device | Proposed Device | Explanation<br>of Differences |
|--------------------------------|-----------------------|-----------------|-------------------------------|
| Biocompatibility | Per ISO 10993-1 | | Identical |
| Shelf Life | N/A- Sold Non-Sterile | | Identical |
{12}------------------------------------------------
# I. Materials
All patient contacting materials are in compliance with ISO10993-1. The proposed modification is replacing the source of the polyester resin and polytetrafluoroethylene (PTFE) coating used to manufacture the Deknatel® Polyester Surgical Sutures.
### J. Technological Characteristics
A comparison of the technological characteristics of the proposed Deknatel® Polyester Surgical Suture and the predicate has been performed. The results of this comparison demonstrate that the Deknatel Polyester Surgical Sutures are equivalent to the marketed predicate devices in performance, stability, biocompatibility and MR safety characteristics.
### K. Performance Data
Non-clinical testing has been conducted in accordance with USP (United States Pharmacopeia) 40-NF 35 <861> Sutures – Diameter, <871> Sutures- Needle Attachment, and <881>Tensile Strength in order to verify the new polyester suture resin and polytetrafluoroethylene (PTFE) coating of the proposed Deknatel Polyester Surgical Sutures are substantially equivalent to the predicate devices in performance Additional non-clinical testing was conducted to confirm substantial equivalence with regards to stability, MR safety and biocompatibility (per ISO 10993-1).
#### L. Conclusion
Based upon the comparative test results, the proposed Deknatel Polyester Surgical Sutures are substantially equivalent in performance, stability, biocompatibility and MR safety to the predicate devices cleared to market via 510(k)s K153076, K153089, and reference 510(k) K930739. The proposed new sources of polyester resin and PTFE coating are the only changes to the product and do not introduce any new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.