K030177 · Sri Orion Pharmaceuticals & Sutures · GAR · Jul 21, 2003 · General, Plastic Surgery
Device Facts
Record ID
K030177
Device Name
ORLON NONABSORABLE SURGICAL SUTURE USP
Applicant
Sri Orion Pharmaceuticals & Sutures
Product Code
GAR · General, Plastic Surgery
Decision Date
Jul 21, 2003
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.5020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ORLON® Polyamide Nonabsorbable Needle Suture USP (Polyamide filament) is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Device Story
ORLON® is a sterile, nonabsorbable, monofilament surgical suture composed of long-chain aliphatic polymers (Nylon 6 and Nylon 6,6). It is provided uncoated, with or without an attached needle. Used by surgeons in clinical settings for soft tissue approximation and ligation. The device functions as a mechanical support to hold tissue edges together during healing. It meets USP requirements for nonabsorbable sutures.
Clinical Evidence
Bench testing only. Sutures undergo physical testing for diameter, tensile strength, needle attachment, and sterility post-sterilization to ensure conformance with United States Pharmacopoeia standards.
Technological Characteristics
Material: Polyamide (Nylon 6 and Nylon 6,6). Form: Monofilament, nonabsorbable, sterile, flexible thread. Meets USP standards. Provided with or without needle.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in cardiovascular, ophthalmic, and neurological procedures.
Regulatory Classification
Identification
Nonabsorbable polyamide surgical suture is a nonabsorbable, sterile, flexible thread prepared from long-chain aliphatic polymers Nylon 6 and Nylon 6,6 and is indicated for use in soft tissue approximation. The polyamide surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. monograph for nonabsorbable surgical sutures; it may be monofilament or multifilament in form; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
# SRI ORION PHARMACEUTICALS & SUTURES
21/22, 1st Cross, Ganesh Block, Nandini Layout, BANGALORE 560096 ① : 349 8093 Fax : 91 80 359 3958 E mail : orion_suture@yahoo.com
Ko30177
### JUL 2 1 2003
## 3.0 SUMMARY OF SAFETY AND EFFECTIVENESS
#### SPONSOR IDENTIFICATION 3.1
S. Gururajarao Sri Orion Pharmaceuticals & Sutures 21/22, 2nd Floor, 1st Cross, Ganesh Block Nandini Layout Bangalore-560096 India Telephone: 91 80 3498093 Fax: 91 80 359 3958 Orion suture@yahoo.com
#### ESTABLISMENT REGISTRATION NUMBER: PENDING 3.2
#### OFFICIAL CONTACT PERSON AND AGENT 3.3
Norman F. Estrin, Ph.D., RAC President Estrin Consulting Group Inc. 9109 Copenhaver Drive Potomac, MD 20854 Tel: 301-279-2899 FAX: 301-294-0126 estrin@yourFDAconsultant.com
#### DATE OF PREPARATION: January 14, 2003 3.4
#### DEVICE NAMES 3.5
- Trade Name: ORLON® 3.5.1
- Common or Usual Name: Polyamide Surgical Suture 3.5.2
- Classification Name: Nonabsorbable Polyamide Surgical Suture 3.5.3 (21CFR 978.5020)
. .
- PROPOSED REGULATORY CLASS: Class II 3.6
- DEVICE PRODUCT CODE: 3.7 GAR
- CFR REFERENCE: 21CFR 878.5020 3.8
{1}------------------------------------------------
#### 3.9 PANEL : General & Plastic Surgery
#### DEVICE DESCRIPTION 3.10
The ORLON® NONABSORBABLE SURGICAL SUTURE USP is a polyamide suture. It is a nonabsorbable, sterile, flexible thread that is prepared from long-chain aliphatic polymers Nylon 6 and Nylon 6,6. It meets all the requirements established by the United States Pharmacopoeia for nonabsorbable sutures. It is offered in the monofilament form and is dyed with an appropriate FDA listed color additive. The suture is provided uncoated and with or without a standard needle attached.
### INDICATIONS FOR USE 3.11
The ORLON® Polyamide Nonabsorbable Needle Suture USP (Polyamide filament) is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
#### PREDICATE DEVICE COMPARISON 3.12
The ORLON® Polyamide Nonabsorbable Needle Suture USP is substantially equivalent to the NYLON Polyamide Surgical Suture (K993998) submitted by Trading Consultants and Distributors International Inc., 157
(ર) Wind Dance Drive, Chicago, IL 60046-6681 and the Surgilon Nonbsorbable Surgical Suture submitted by Sherwood-Davis & Geck.
Both the ORLON® Polyamide Nonabsorbable Needle Suture USP and the predicate devices are nonabsorbable polyamide sutures and are sold in the monofilament form and have the same indications for use. Both the
ORLON® product and the NYLON product meet the requirements of the United States Pharmacopoeia.
#### SUMMARY OF STUDIES 3.13
All the testing and the manufacturing process conform to the standards of the United States Pharmacopoeia. After sterilization, the sutures are subjected to physical tests of diameter, tensile strength, needle attachment and sterility. Once all these tests are conducted and the sutures pass these tests, they are released for distribution.
105
1 :
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 1 2003
Sri Orion Pharmaceuticals & Sutures c/o Norman F. Estrin, Ph.D., RAC Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, Maryland 20854
Re: K030177
Trade/Device Name: The ORLON® Polyamide Nonabsorbable Needle Suture USP Regulation Number: 21 CFR 878.5020 Regulation Name: Suture, nonabsorbable, synthetic, polyamide Regulatory Class: II Product Code: GAR Dated: June 15, 2003 Received: June 16, 2003
Dear Dr. Estrin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Norman F. Estrin, Ph.D., RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
030177
Page 1 of 1
510(k) Number (if known):
The ORLON® Polyamide Nonabsorbable Needle Suture USP Device Name:
Indications for Use:
The ORLON® Polyamide Nonabsorbable Needle Suture USP (Polyamide filament) is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109
OR
Over-the-Counter Use
(Optional Format 1-2-96)
ាន
ﺎ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
Muriam C. Provost
(Division Sign Off)
ral, Restorative d Neurological Devi
510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.