K111020 · AtriCure, Inc. · FTD · Jun 9, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111020
Device Name
ATRICURE DISSECTOR
Applicant
AtriCure, Inc.
Product Code
FTD · General, Plastic Surgery
Decision Date
Jun 9, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4580
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AtriCure Dissector™ is intended to dissect soft tissue during general, ENT, thoracic, urological and gynecological surgical procedures. The Dissector's battery-powered light source is used to navigate soft tissue for identification of anatomic structures.
Device Story
AtriCure Dissector is a hand-held, single-use, surgical articulated dissector featuring an integral battery-powered light source. Operated by qualified surgeons, the device allows manual articulation of the distal tip via a rotation knob to facilitate tissue dissection. The integrated light source illuminates the surgical field to assist in the identification of anatomic structures. Available in multiple shaft lengths to accommodate varying patient habitus and surgeon preference. The light source remains illuminated for a minimum of four hours. The device is used in general, ENT, thoracic, urological, and gynecological surgical procedures to aid in navigation and tissue separation.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1) for limited patient contact (<24 hours), risk analysis, and process verification to confirm conformance to product specifications.
Technological Characteristics
Hand-held, single-use, articulated surgical dissector. Materials include medical grade silicone, stainless steel, medical grade adhesive, and acrylic epoxy resin. Features an integral battery-powered light source. Biocompatibility tested per ISO 10993-1. No electronic software or complex algorithms; mechanical articulation controlled by manual rotation knob.
Indications for Use
Indicated for soft tissue dissection during general, ENT, thoracic, urological, and gynecological surgical procedures. Intended for use by qualified surgeons to navigate and identify anatomic structures.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
K///.020
# ATRICURE DISSECTOR™ 510(k) SUMMARY
### General Information
| Date Compiled | April 11, 2011 |
|----------------|------------------------------------------------------------------------------------|
| Classification | Class II (Surgical Lamp)<br>21 CFR 878.4580 |
| Product Code | FTD, GDI |
| Trade Name | AtriCure DissectorTM |
| Manufacturer | AtriCure, Inc<br>6217 Centre Park Drive<br>West Chester, OH 45069 |
| Contact | James L. Lucky<br>VP of Quality Assurance and Regulatory Affairs<br>(513) 755-5754 |
### Indications for Use
The AtriCure Dissector™ is intended to dissect soft tissue during general, ENT, thoracic, urological and gynecological surgical procedures. The Dissector's battery-powered light source is used to navigate soft tissue for identification of anatomic structures.
#### Cleared Device
The device proposed for modification in this submission is the AtriCure Dissector™ (K041681).
#### Device Description
The AtriCure Dissector™ is a hand held, single use, surgical articulated dissector with intearal light source intended to for use by qualified surgeons only. The surgeon is able to articulate the distal member of the device by means of hand actuated rotation knob. The surgeon directly controls the amount of articulation and tissue dissection. A small light source is attached to the distal tip of the articulation member, which remains illuminated for a minimum of four (4) hours. Multiple models of the Dissector™ are available with varving shaft lengths to accommodate surgeon preference and differing patient habitus.
#### Materials
All materials used in the manufacture of the AtriCure Dissector™ are suitable for their intended use and have been used in numerous previously cleared products. Materials used include, but are not limited to, medical grade silicone, stainless steel, medical grade adhesive, and acrylic epoxy resin. Testing was conducted in Accordance with ISO 10993-1 to ensure appropriate biocompatibility of all materials.
#### Testing
Appropriate product testing was conducted according to the Design Control requirements as defined in 21 CFR 820.30 to evaluate conformance to product specification and substantial equivalence to the legally marketed device. These activities included Risk
1/2
{1}------------------------------------------------
Analysis, Process Verification, and Biocompatibility Testing. Biocompatibility testing was conducted according to ISO 10993-1 to confirm that the modified Dissector™ is safe for limited use (<24 hrs) as a patient contacting device. .
Summary of Substantial Eguivalence
substantially equivalent to the Dissector™ (MID1) device cleared via K041681. The modification does not affect the technological characteristics, scientific principles, or performance specifications. The indications for use, basic overall function, and materials used are substantially equivalent.
\$\frac{2}{2}\$
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN - 9 2011
AtriCure, Inc. % Mr. James L. Lucky VP, Quality Assurance and Regulatory Affairs 6217 Centre Park Drive West Chester, Ohio 45069
Re: K111020
Trade/Device Name: AtriCure Dissector"™ Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FTD, GDI Dated: May 19, 2011 Received: May 23, 2011
Dear Mr. Lucky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Mr. James L. Lucky
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fdai.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
> Sincerely yours. Mark N. Melkerson
> Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
KII/020 510(k) Number (if known):
Device Name: AtriCure Dissector™
Indications For Use:
The AtriCure Dissector™ is intended to dissect soft tissue during general, ENT, thoracic, urological and gynecological surgical procedures. The Dissector's battery-powered light source is used to navigate soft tissue for identification of anatomic structures.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Ogden for mxm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.