K103540 · Psoria-Shield, Inc. · FTC · Feb 10, 2011 · General, Plastic Surgery
Device Facts
Record ID
K103540
Device Name
PSORIA-LIGHT
Applicant
Psoria-Shield, Inc.
Product Code
FTC · General, Plastic Surgery
Decision Date
Feb 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Psoria-Light Model PS1000 is indicated for use in targeted PUVA photochemistry and UVB phototherapy for the treatment of skin conditions including psoriasis, vitiligo, atopic dermatitis (eczema), and seborrheic dermatitis. In addition, the system UVB channel is indicated for the treatment of leukoderma.
Device Story
Psoria-Light Model PS1000 delivers targeted UVA (350-395nm) or UVB (300-320nm) light to small skin regions (2.88 cm) while sparing healthy tissue. Device utilizes LED technology, eliminating mercury/xenon/chlorine gas lamps. Handpiece includes non-contact sensor to detect skin surface; UV dosage trigger mechanism remains disabled until skin contact is confirmed. Used by clinicians for targeted phototherapy. Output allows precise treatment of dermatologic conditions. Benefits include localized delivery, reduced exposure to healthy skin, and avoidance of hazardous gas-based light sources.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on the device's similarity to previously cleared predicate devices for the same indications for use. Bench testing and certification to international standards (IEC 60601 series, UL 60601-1, FCC Part 15, EN55011) and risk management (ISO 14971) were performed.
Technological Characteristics
Targeted UV phototherapy system using UVA (350-395nm) and UVB (300-320nm) LEDs. Features non-contact skin sensor for safety interlock. Conforms to IEC 60601-1, IEC 60601-1-2, UL 60601-1, and 21 CFR 1040.10. Complies with FCC Part 15 and EN55011 for EMC. Risk management per ISO 14971.
Indications for Use
Indicated for patients with psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis, and leukoderma requiring targeted PUVA or UVB phototherapy.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
Predicate Devices
TheraLight Inc. DuaLight UV120-2 UVA/UVB Phototherapy System (K024020)
PhotoMedex, Inc. XTRAC Ultra2 Excimer Laser System, Model AL10000 (K073659)
Submission Summary (Full Text)
{0}------------------------------------------------
K103540
FEB 10 20il
# ENCLOSURE E
Non-Confidential 510(k) Summary of Safety and Effectiveness
Psoria-Shield Inc. 6408 W. Linebaugh Avenue Suite 104 Tampa, FL 33625
| Official Contact: | Scot Johnson<br>Chief Technology Officer |
|---------------------|--------------------------------------------------------------------------------------|
| Trade/Device Name: | Psoria-Light Model PS1000 |
| Device Common Name: | Targeted UVA / UVB Phototherapy System |
| Regulation Number: | 878.4630 |
| Regulation Name: | Ultraviolet lamp for dermatologic disorders |
| Regulatory Class: | Class II |
| Product Code: | FTC |
| Predicate Devices: | This product is substantially equivalent in design,<br>composition, and function to: |
| | TheraLight Inc.<br>DuaLight UV120-2 UVA/UVB Phototherapy System<br>K024020 |
| | PhotoMedex, Inc.<br>XTRAC Ultra2 Excimer Laser System, Model AL10000<br>K073659 |
{1}------------------------------------------------
K103540
pg 2 of 3
# ENCLOSURE E
## Non-Confidential 510(k) Summary of Safety and Effectiveness
## Device Description:
Psoria-Light™ Model PS1000 delivers targeted ultraviolet (UV) light to small regions of affected skin (2.88 cm), sparring healthy skin. The system is capable of delivering either UVB light (300 to 320nm) or UVA light (350 to 395nm). Psoria-Light™ Model PS1000 generates both UVA and UVB light via LED technology, avoiding environmental concerns associated with disposal and exchange of mercury gas lamps, Xenon, and Chlorine gases commonly associated with targeted UV phototherapy devices. Psoria-Light's™ UVB LEDs emit light between 300 and 320 nm, avoiding light emissions below 300nm.
Psoria-Light™ Model PS1000 incorporates a unique safety feature, a non-contact sensor located about the aperture of the handpiece that detects human skin. The trigger mechanism for UV dosage administration (located on the handpiece) is disabled until a suitable surface is detected by this sensor.
## Labeling and Instructions for Use:
See Proposed Labeling and Instructions for Use in sections J and Q respectively.
#### Intended Use:
The Psoria-Light Model PS1000 is indicated for use in targeted PUVA photochemistry and UVB phototherapy for the treatment of skin conditions including psoriasis, vitiligo, atopic dermatitis (eczema), and seborrheic dermatitis. In addition, the system UVB channel is indicated for the treatment of leukoderma.
## Substantial Equivalence:
The characteristics of the Psoria-Shield Psoria-Light Model PS1000 are substantially equivalent to those of the predicate devices. Operation of the predicate devices, including output wavelengths, power, and light sources have received 510(k) clearance previously for these type of treatments.
#### Clinical Performance Testing:
All clinical indications requested in this application have been previously cleared in the identified predicate devices. The Psoria-Light Model PS1000 does not introduce any new indications for use, and will perform in a substantially equivalent manner as the identified predicates. Therefore Psoria-Shield believes duplicative clinical data is not required as a condition of granting market clearance for the Psoria-Light Model PS1000.
{2}------------------------------------------------
K103540
pg 3 of 3
# ENCLOSURE E
## Non-Confidential 510(k) Summary of Safety and Effectiveness
### Product Performance Testing:
Testing and certification relevant to the Psoria-Light Model PS1000 includes conformance to current applicable international IEC 60601 series of standards, 21 CFR Part 1040.10, Performance Standard for Light-Emitting Products and certification to the UL 60601-1 Medical Electrical Equipment classification standard. Product will also comply with EMC (Electromagnetic Compatibility) requirements per FCC rules part 15, IEC 60601-1-2 and its particular Standards for susceptibility and emissions, as well as EN55011. The Psoria-Light Model PS1000 has been reviewed for risk management utilizing ISO 14971, Application of risk management to medical devices ensuring all aspects of the device are reviewed for potential hazards.
The Psoria-Light Model PS1000 shall have completed all testing to the standards and procedures identified in this submission prior to any product being released for sale. Products shall be produced and distributed by a facility that has been registered with the FDA to manufacture medical devices.
### Conclusions:
The comparison to the predicate devices demonstrates that the proposed device is safe and effective and is Substantially Equivalent to the predicate devices and it does not introduce any new issues of safety or efficacy. The Indications for Use and methods of operation are substantially equivalent to the predicates for the treatment of skin conditions including psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis, and of leukoderma.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Psoria-Shield. Inc. % Technireg, Inc. Mr. Wayne Glover 19404 Pine Valley Drive Odessa, Florida 33556
FEB 10 201
Re: K103540
Trade/Device Name: Psoria-Light Model PS 1000 Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet lamp for dermatologic disorders Regulatory Class: Class II Product Code: FTC Dated: November 30, 2010 Received: December 02, 2010
Dear Mr. Glover:
and the comments of the comments of the comments of
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Orug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Ice The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
a participal production in the lineral in
{4}------------------------------------------------
Page 2 – Mr. Wayne Glover
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
A.S. B. R. h
for
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# ENCLOSURE D
# INDICATIONS FOR USE
K103540
510(k) Number (if known):
Device Name:
Psoria-Light Model PS1000
Indications for Use:
The Psoria-Light Model PS1000 is indicated for use in targeted PUVA photochemistry and UVB phototherapy for the treatment of skin conditions including psoriasis, vitiligo, atopic dermatitis (eczema), and seborrheic dermatitis. In addition, the system UVB channel is indicated for the treatment of leukoderma.
Prescription Use X (Per 21 CFR 801 Subpart D) and/or
Over-the-counter use (Per 21 CFR 801 Subpart C)
PJ
Tofal
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nut ReQs for mxm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103540
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.