K133554 · Microtek Medical, Inc. an Ecolab Company · FTA · Apr 8, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133554
Device Name
SURGICAL LIGHT HANDLE COVER
Applicant
Microtek Medical, Inc. an Ecolab Company
Product Code
FTA · General, Plastic Surgery
Decision Date
Apr 8, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The Surgical Light Handle Cover is intended to be used as a disposable barrier for a surgical lighting system and which allows the surgical team member to manually adjust the lighting system. This is a single-use device.
Device Story
Surgical Light Handle Cover is a polycarbonate, injection-molded, single-use sterile barrier; designed to fit over surgical light handles with release buttons. Device features apertures that interface with light handle release buttons for secure attachment. Distal end is transparent to permit use of light-mounted recording devices without providing image enhancement or magnification. Used in surgical settings by surgical team members to manually adjust light positioning during procedures. Provides sterile interface for light manipulation; prevents contamination of surgical field.
Clinical Evidence
Bench testing only. Performance testing included simulated distribution (ASTM D4169), dye migration (ASTM F1929), seal peel (ASTM F88), visual inspection, sterility (ISO 11135-1), EO/ECH residuals (ISO 10993-7), cytotoxicity (ISO Elution), material leachables (USP physical-chemical), and functional installation/removal and light manipulation testing.
Technological Characteristics
Polycarbonate injection-molded handle cover; transparent distal end for recording device compatibility. Sterile, single-use. Sterilization: EO gas (SAL 10^-6). Standards: ASTM D4169, ASTM F1929, ASTM F88, ISO 11135-1, ISO 10993-7, ISO Elution.
Indications for Use
Indicated for use as a disposable barrier for surgical lighting systems to allow surgical team members to manually adjust the light position during procedures.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Ecolab Inc. The text "Ecolab Inc." is at the top of the image. Below that is the Ecolab logo, with the word "ECOLAB" in bold black letters and a symbol in the middle of the O. There is a registered trademark symbol to the right of the B.
:
# APR 0 8 2014 K133554
#### Section 5: 510(k) Summary
## Surgical Light Handle Cover
As required by 21 CFR 807.92.
| Date: | November 11, 2013 | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------|--|
| Administrative Information | | |
| Submitter: | Microtek Medical Inc., an Ecolab Company | |
| Establishment<br>Registration Number: | 1043582. | |
| Contact Person: | Andy Roller<br>370 Wabasha Street North<br>St. Paul, MN 55102-1390<br>Sr. Regulatory Specialist<br>651.293.2080 | |
| Device Identification | | |
| Device Name: | Surgical Light Handle Cover | |
| Common Name: | Surgical Light Accessory | |
| Device Classification Name: | Light, Surgical, Accessories | |
| Device Classification: | Class II | |
| Panel: | General and Plastic Surgery | |
| Classification Regulation: | 878.4580 | |
| Product Code: | FTA | |
Performance Standards: No Recognized Consensus Standards
{1}------------------------------------------------
Ecolab Inc.
Surgical Light Handle Cover
| Predicate Device: | Surgical Lamp Handle and Cover, cleared on 3/20/1999 via K901154 |
|-------------------|------------------------------------------------------------------|
| Trade Name: | Skytron Disposable Light Handle Cover |
Device Description
The Surgical Light Handle Cover is a polycarbonate injection molded device shaped to fit a surgical light that has release buttons protruding from the surgical light handle. The Surgical Light Handle Cover has apertures located at the end of the device, proximal to the surgical light, which correspond to the release buttons on the surgical light handle. When the Surgical Light Handle Cover is attached to the light, the release buttons will interface with the apertures to secure the Surgical Light Handle Cover to the light handle.
The Surgical Light Handle Cover is hollow in the center, with transparent polycarbonate located near the distal end. The transparent end permits the use of a recording device, which may be mounted onto the surgical light. The transparent end allows light to pass through for the purpose of recording images, it does not provide image enhancement or magnification.
#### Statement of Intended Use
The Surgical Light Handle Cover is intended to be used as a disposable barrier for a surgical lighting system and which allows the surgical team member to manually adjust the lighting system. This is a single-use device.
### Substantial Equivalence Discussion
The predicate device and design basis for the Surgical Light Handle Cover is the Microtek Medical Surgical Lamp Handle and Cover (K901154, cleared 3/20/1990). The fundamental scientific technology of the device remains unchanged in that it is a disposable sterile light cover that allows the user to adjust position of the light during surgical procedures. The following table illustrates the similarities and differences in the product designs.
| Property or<br>Characteristic | Proposed Device -<br>Surgical Light Handle Cover | Predicate Device -<br>Surgical Lamp Handle and<br>Cover (K901154) |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/Indications for<br>Use | A device intended to be used as a<br>disposable barrier for a surgical<br>lighting system and which allows a<br>surgical team member to manually | A device intended to be used as<br>a disposable barrier for a<br>surgical lighting system and<br>which allows a surgical team |
| | | Table 1: Substantial Equivalence | | |
|--|--|----------------------------------|--|--|
|--|--|----------------------------------|--|--|
{2}------------------------------------------------
ද
| | adjust the lighting system. This is a<br>single-use device | member to manually adjust the<br>lighting system. This is a<br>single-use device |
|----------------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------|
| Conditions of Use | Single Use, Disposable | Single Use, Disposable |
| Materials | Polycarbonate Handle Cover | Polypropylene Handle Cover |
| Property or<br>Characteristic | Proposed Device -<br>Surgical Light Handle Cover | Predicate Device -<br>Surgical Lamp Handle and<br>Cover (K901154) |
| | Color | Clear - Frosted |
| Sterility | Provided sterile | Provided sterile |
| Sterility Assurance<br>Level | 10-6 via EO Gas | 10-3 via Gamma Radiation |
| Principle of<br>Operation | Sterile cover to allow manipulation<br>of a surgical light | Sterile cover to allow<br>manipulation of a surgical light |
| Interface with<br>Surgical Light | Fully detachable | Lamp Handle - Not detachable<br>Handle Cover - Fully<br>detachable |
| | Transparent Distal<br>End | Yes |
Performance Data Summary
Table 2: Performance Data Summary of the Surgical Light Handle Cover
| | Requirement | Specification | Method | Result |
|-----------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|--------|--------|
| Packaging | Simulated Distribution-<br>Test | ASTM D4169 | Pass | |
| | Dye Migration Test | ASTM F1929 | Pass | |
| | Seal Peel Test | ASTM F88 | Pass | |
| | Inspection for Shipping<br>Damage | Visual Inspection | Pass | |
| Sterility | SAL 10-6 | ISO 11135-1 | Pass | |
| | EO/ECH Residuals | ISO 10993-7 | Pass | |
| Material<br>Compatibility<br>Evaluation | Cytotoxicity | ISO Elution | Pass | |
| | Material Leachables | USP Physical-Chemical<br>Analysis | Pass | |
| Functional<br>Requirements | Product must be easily<br>installed and removed | Attached/Detached Surgical<br>Light Handle Cover to surgical<br>light | Pass | |
| | Product must not interfere<br>with the functionality of<br>surgical light | With Surgical Light Handle<br>Cover attached: rotate light,<br>move light in transverse<br>direction, compare visual clarity | Pass | |
{3}------------------------------------------------
#### Substantial Equivalence Conclusion
The differences between the Surgical Light Handle Cover and the predicate device do not constitute a new intended use, and the differences in technological characteristics do not raise different questions of safety and effectiveness. Furthermore, the changes to the device design do not impact the fundamental scientific technology or principle of operation, which is to allow a user to manipulate a surgical light during surgical procedures using a disposable sterile cover.
The Surgical Light Handle Cover is substantially equivalent to the predicate device cleared under K901154.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 8, 2014
Microtek Medical, Incorporated An Ecolab Company Mr. Andy Roller Senior Regulatory Specialist 370 Wabasha Street North St. Paul. MN 55102
Re: K133554
Trade/Device Name: Surgical Light Handle Cover Regulation Number: 21 CFR 878.4580 Regulation Name: Light, Surgical, Accessories Regulatory Class: II Product Code: FTA Dated: January 30, 2014 Received: February 12, 2014
Dear Mr. Roller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Mr. Roller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fdagov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tejashri
Clinical D
DAGRID
Teiashri Purohit-Sheth, M.D. Clinical Deputy Director
FOR
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Statement of Indications for Use Section 4:
510(k) Number (if known): _ K133554
Device Name: Surgical Light Handle Cover
Model Number: B1-715-65
Indications For Use: A device intended to be used as a disposable barrier for a surgical lighting system and which allows a surgical team member to manually adjust the lighting system. This is a single-use device
Image /page/6/Picture/7 description: The image shows the phrase "Prescription Use (Part 21 CFR 801 Subpart D)" in bold font. There is a line underneath the phrase. Above the phrase is a large X mark.
AND/OR
Over-The-Counter Use (2) CFR 801 Subparl C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (CDDE) Elizabeth F. Claverie -S 2014.04.07 22:55:00 -04'00'
4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.