K080212 · Medela AG · BTA · Jul 23, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080212
Device Name
MEDELA THOPAZ, MODEL 200 0288
Applicant
Medela AG
Product Code
BTA · General, Plastic Surgery
Decision Date
Jul 23, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medela® THOPAZ Suction Pump is indicated for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials from a patient's respiratory support system (after surgery). Generally the Medela® THOPAZ is intended to be used in all situations where chest drains are applied - especially for thoracic drainage and in situations such as pneumothorax, after surgery (post operative), thorax injury, pleura effusion, pleuryempyem or other related conditions.
Device Story
Medela THOPAZ is an AC/DC powered, maintenance-free secretion aspirator for thoracic drainage. Device utilizes a DC-motor with membrane aggregate to provide suction (max -10 kPa / -75 mmHg; 5 L/min flow). System is dry, requiring no fluids for operation. Features an electronic measuring/monitoring system with MMI (man-machine interface) for settings, alarm handling, and data transfer. Status displayed digitally and graphically on-device; therapy data transferable to PC. Used in clinical settings to assist patient mobilization during chest drainage. Healthcare providers use output to monitor therapy progress and adjust suction parameters. Benefits include quiet operation, patient mobility, and integrated safety features.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on identical technological characteristics and performance data compared to predicate devices.
Technological Characteristics
AC/DC powered, maintenance-free suction pump. Features DC-motor with membrane aggregate. Dry system (no fluids required). Max vacuum -10 kPa (-75 mmHg); flow rate 5 L/min. Includes electronic measuring/monitoring system with optical/acoustic status display. MMI interface for user control. Data transfer capability to PC.
Indications for Use
Indicated for post-surgical patients requiring aspiration/removal of surgical fluids, tissue, gases, or infectious materials from respiratory support systems. Applicable for thoracic drainage in conditions including pneumothorax, thorax injury, pleural effusion, and empyema.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
K080212
# Section E - 510(k) Summary
This 510(k) summary for the Medela® THOPAZ Suction Pump meets the requirements of 21 CFR 807.92.
JUL 2 3 2008
# Sponsor's Name, Address and Contact Person 1
Sponsor: Medela AG Medical Equipment Laettichstrasse 4b 6341 Baar Switzerland +41 41 769 5151 ext. 247 Ph: +41 41 769 5100 Fax:
Contact Person Bruno Gretler Manager Regulatory Affairs
Date Summary Prepared: January 24, 2008
#### 2 Name of Device
| Trade Name: | Medela® THOPAZ<br>Secretion & Surgical Aspirator |
|----------------------|--------------------------------------------------------------------------------------------|
| Common Name: | Powered Suction Pump |
| Classification Name: | PUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)<br>Classified Class II, per 21 CFR 878.4780 |
| Product Code: | BTA |
## 3 Name of the predicate Device(s)
- Medela® Vario 8/18/ci Suction Pumps, by Medela Inc. . K061205
- Medela® Dominant 35 c/i Suction Pump, by Medela AG . K043544
E_510k_Summary_THOPAZ.doc
2008-01-24
Section E - 1 of 3
{1}------------------------------------------------
#### ব Device Description
The Medela® THOPAZ pump is an innovative secretion aspirator intended to be used for thoracic drainage. Its well-proven membrane system guarantees maximum suction performance and quiet, dependable operation. Additional advantages of the Medela® THO-PAZ are: user friendliness, patient mobility, simple cleaning and integrated safety features. A comprehensive range of accessories makes the Medela® THOPAZ ideally suited for thoracic drainage while mobilizing the patient.
The Medela® THOPAZ suction pump is an AC/DC powered, maintenance-free aspirator which incorporates a DC-motor with membrane aggregate power actuation in its housing. A user friendly MMI (man machine interface) guides the user through first installation, change of settings, use, data transfer and alarm handling.
The Medela® THOPAZ suction pump has an electronic measuring and monitoring system with optical and acoustic status display. The device is a dry system, which means that no fluids are necessary for operation. Important information about the course of therapy is displayed digitally and as graphics in the display. These data can be transferred to a PC upon completion of the therapy.
The Medela® THOPAZ suction pump has a suction capacity of 5 liters per minute and a maximum vacuum up to -10 kPa (-75 mmHg). The pump is marked "low flow - low vacuum".
A variety of reusable and disposable accessories for thoracic drainage are available.
#### 5 Indications for use
The Medela® THOPAZ Suction Pump is indicated for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials from a patient's respiratory support system (after surgery).
Generally the Medela® THOPAZ is intended to be used in all situations where chest drains are applied - especially for thoracic drainage and in situations such as pneumothorax, after surgery (post operative), thorax injury, pleura effusion, pleuryempyem or other related conditions.
## Summary of Technological Characteristics 6
The Medela® THOPAZ has the same performance characteristics as the predicate devices. These differences regarding type of aggregate and display are marginal and state of the art today. Other FDA approved devices using the same technology (aggregate and display).
The Medela® THOPAZ Suction Pump is equipped with the identical technology like other marketed devices. These technological features do not affect safety and effectiveness of the device or the application (thoracic drainage).
E_510k_Summary_THOPAZ.doc
2008-01-24
{2}------------------------------------------------
#### 7 Conclusion
There are no differences in performance or technology which significantly affect the safety and effectiveness of the device or the application (thoracic drainage). All conclusions are made by the decision making process according to the recommendations in the "510(k) SE Decision Making Process" document.
The Medela® THOPAZ suction pump has the identical intended uses and, where applicable, the identical technological characteristics and performance data as the predicate devices.
Based upon the information presented in this submission, it is proven that the proposed Medela® THOPAZ powered suction pump is substantially equivalent, safe and effective for the intended use.
E_510k_Summary_THOPAZ.doc
2008-01-24
Section E - 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" arranged in a circular fashion around the eagle. The text is in all capital letters and is relatively small compared to the eagle symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL. 2 3 2008
Medela AG % Mr. Bruno Gretler Manager Regulatory Affairs Laettichstrasse 4b 6341 Baar Switzerland
Re: K080212
Trade/Device Name: Medela THOPAZ Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: II Product Code: BTA Dated: July 10, 2008 Received: July 10, 2008
Dear Mr. Gretler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Bruno Gretler
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Millerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation . Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): Kc 80212
Device Name:
Medela THOPAZ
The Medela THOPAZ Suction Pump is indicated for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials from a patient's respiratory support system (after surgery).
Generally the Medela THOPAZ is intended to be used in all situations where chest drains are applied - especially for thoracic drainage and in situations such as pneumothorax, after surgery (post operative), thorax injury, pleura effusion, pleuryempyem or other related conditions.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
i
Page 1 of 1
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.