Cavilon Advanced High Endurance Skin Protectant

K153571 · 3M Healthcare · KMF · Aug 23, 2016 · General Hospital

Device Facts

Record IDK153571
Device NameCavilon Advanced High Endurance Skin Protectant
Applicant3M Healthcare
Product CodeKMF · General Hospital
Decision DateAug 23, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5090
Device ClassClass 1
AttributesTherapeutic

Indications for Use

Cavilon™ Advanced Skin Protectant forms a film barrier intended to protect intact or damaged skin. It is effective in conditions where skin is frequently or continuously exposed to moisture and caustic irritants such as feces, digestive fluids, wound drainage and urine. Cavilon™ Advanced Skin Protectant also can be used in areas exposed to friction and shear from bedding, clothing, shoes or any other material that would rub against the skin.

Device Story

Polymeric-cyanoacrylate solution; applied as liquid to intact or damaged skin; dries rapidly to form transparent, elastomeric, waterproof, breathable, durable film barrier. Protects against moisture, caustic irritants, friction, and shear. Used in clinical or home settings; applied by healthcare providers or patients. Film adheres to dry, moist, or wet skin; remains intact during exposure; wears off naturally without requiring removal. Benefits include skin protection and reduced irritation from external sources.

Clinical Evidence

Clinical evidence includes a controlled, randomized open-label durability study on healthy subjects demonstrating barrier duration up to 72 hours, and a prospective, open-label wash-off resistance study on healthy subjects showing performance comparable to Medline® Marathon® Liquid Skin Protectant.

Technological Characteristics

Polymeric-cyanoacrylate solution in non-stinging solvent. Liquid form; polymerizes upon contact to form elastomeric, transparent, breathable film. Surface-contacting device for breached/compromised skin. Biocompatibility testing per ISO-10993 (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity).

Indications for Use

Indicated for protection of intact or damaged skin in patients exposed to moisture, caustic irritants (feces, digestive fluids, wound drainage, urine), or friction/shear from external materials.

Regulatory Classification

Identification

A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized caduceus or a family of people. The symbol is composed of three figures in profile, one behind the other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 23, 2016 3M Healthcare Ms. Maria Ruiz Regulatory Affairs Specialist 3M Center 2510 Conway Avenue, Bldg. 275-5W-06 St. Paul. MN 55144 Re: K153571 Trade/Device Name: Cavilon Advanced Skin Protectant Regulation Number: 21 CFR 880.5090 Regulation Name: Liquid bandage Regulatory Class: Class I Product Code: KMF Dated: July 25, 2016 Received: July 27, 2016 Dear Ms. Ruiz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153571 ### Device Name 3M™ Cavilon™ Advanced Skin Protectant ### Indications for Use (Describe) Cavilon™ Advanced Skin Protectant forms a film barrier intended to protect intact or damaged skin. It is effective in conditions where skin is frequently or continuously exposed to moisture and caustic irritants such as feces, digestive fluids, wound drainage and urine. Cavilon™ Advanced Skin Protectant also can be used in areas exposed to friction and shear from bedding, clothing, shoes or any other material that would rub against the skin. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510K SUMMARY #### SUBMITTER I. 3M Company 3M Health Care 3M Center 2510 Conway Ave., Bldg 275-5W-06 St. Paul, MN 55144 | Phone: | 651-736-2733 | |-----------------|----------------| | Fax: | 651-737-5320 | | E-Mail: | meruiz@mmm.com | | Contact Person: | Maria Ruiz | | Contact Person: | Maria Ruiz | |-----------------|-------------------------------| | | Regulatory Affairs Specialist | Date Prepared: December 4, 2015 #### II. DEVICE | Name of Device: | 3MTM Cavilon™ Advanced Skin Protectant | |-----------------------|----------------------------------------| | Common or Usual Name: | Barrier Film | | Classification Name: | Liquid Bandage | | Regulatory Class: | Class 1 | | Product Code: | KMF per 21 CFR 880.5090 | #### III. PREDICATE DEVICE 3MTM Cavilon™ No Sting Barrier Film (K955103) Medline® Marathon® Liquid Skin Protectant (K133443) #### IV. DEVICE DESCRIPTION 3M™ Cavilon™ Advanced Skin Protectant is a polymeric-cyanoacrylate solution intended for the protection of intact or damaged skin. Upon application to skin, the liquid dries rapidly to form a primary long-lasting waterproof, highly durable film barrier. It is elastomeric, adhering to the contours of the skin and providing a uniform film. The film is transparent and possesses good oxygen and moisture vapor permeability. {4}------------------------------------------------ The polymer-cyanoacylate is dispersed in a non-stinging solvent. The film is colorless and non-cytotoxic. The film adheres to dry, moist or wet skin surfaces and remains intact during conditions of continuous or repeated exposure to moisture or caustic irritants. It will wear off the skin and does not require removal. #### V. INDICATIONS FOR USE Cavilon™ Advanced Skin Protectant forms a film barrier intended to protect intact or damaged skin. It is effective in conditions where skin is frequently or continuously exposed to moisture and caustic irritants such as feces, digestive fluids, wound drainage and urine. Cavilon™ Advanced Skin Protectant also can be used in areas exposed to friction and shear from bedding, clothing, shoes or any other material that would rub against the skin. The Indications for Use statement for Cavilon™ Advanced Skin Protectant is similar to the predicate devices. Both the Subject Device and predicate devices have the same intended use to form a liquid barrier film when applied to the skin, by providing a barrier against irritants and caustic substances (e.g., urine, fecal matter, digestive juices) and to reduce friction and shear from materials rubbing against the skin (e.g., bedsheets, clothing, shoes). ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The technological principle for the Subject Device and the predicate devices are that they are all liquid bandages. Just like its predicate devices, Cavilon™ Advanced Skin Protectant forms a transparent and durable barrier film to protect intact or damaged skin from the caustic effects of moisture, urine or feces and friction and shear. It is applied in liguid form to the area of interest on the skin and quickly polymerizes upon contact to form the barrier film. The Subject Device and predicate devices all have similar technological elements which include: - Operational principle ● - Device is transparent - Device provided in a liquid form - Device forms a protective barrier - Device is elastomeric - Device is a durable barrier ● - Device protects from friction . - Device is breathable ● {5}------------------------------------------------ The minor differences do not raise different questions of safety and effectiveness. The following technological differences exist between the subject device and one or both of the predicate devices: - Liquid technological components - Monomer and/or polymer composition ● - Device dries rapidly - Device cures upon contact - Delivery system ● - Sterilization method #### VII. PERFORMANCE DATA The following performance data is provided in support of the substantial equivalence determination. # Biocompatibility Testing The biocompatibility evaluation for Cavilon™ Advanced Skin Protectant was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. Cavilon™ Advanced Skin Protectant is characterized as a surface contacting device for use on breached or compromised skin for periods of time greater than 30 days. The battery of testing included the following tests: - Cytotoxicity ● - Sensitization - Irritation - Systemic Toxicity ● - Genotoxicity ● # Bench Testing Bench testing for Cavilon™ Advanced Skin Protectant was conducted in support of substantial equivalence to the predicate devices. The bench testing included the following tests: - Elongation ● - Cure and/or Dry Time ● - Friction ● - Moisture Vapor Transmission Rate {6}------------------------------------------------ ## Clinical Studies Clinical testing of Cavilon™ Advanced Skin Protectant include: - Controlled, randomized open label durability study on healthy subjects that . demonstrated that Cavilon™ Advanced Skin Protectant is a durable barrier film that lasts up to 72 hours. - . Prospective, open label wash off resistance study on healthy subjects showed that the Cavilon™ Advanced Skin Protectant is similarly to Marathon® Liquid Skin Protectant with respect to the 'active' wash-off resistance and its ability to protect skin. # VIII. CONCLUSIONS 3M™ Cavilon™ Advanced Skin Protectant is substantially equivalent to the predicate devices, 3MTM Cavilon™ Barrier Film (K955103) and Medline Marathon® Liquid Skin Protectant (K133443). Cavilon™ Advanced Skin Protectant is composed of similar components and similar performance, intended use and indications for use as the predicate devices. Based on the similarities to the predicate devices, the minor differences do not present or raise any new safety or effectiveness issues and the device is substantially equivalent to the predicates.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...