K180460 · Theradome, Inc. · OAP · May 18, 2018 · Physical Medicine
Device Facts
Record ID
K180460
Device Name
Theradome LH40
Applicant
Theradome, Inc.
Product Code
OAP · Physical Medicine
Decision Date
May 18, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Theradome LH40 (Theragrow) is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of lla to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2; both with Fitzpatrick Skin Types I to IV.
Device Story
Theradome LH40 is a low-level laser therapy (LLT) helmet for home use. Device utilizes 40 red visible light diode lasers (678 nm) to deliver non-thermal energy to scalp hair follicles. Photobiostimulation promotes hair growth. User operates device via single button; audible timer automatically terminates 20-minute treatment session. Hands-free design allows patient self-use. Output is non-thermal laser energy; healthcare provider does not interact with device output. Benefit is hair growth stimulation for androgenetic alopecia.
Clinical Evidence
No clinical data required. Substantial equivalence established via comparison of technological characteristics and intended use to predicate devices. Bench testing, biocompatibility, mechanical, software, and EMC testing from predicate device remain applicable.
Technological Characteristics
Low-level laser therapy (LLT) device. 40 coherent light diodes (Class 3R), 678 nm wavelength, 5mW power. Helmet form factor (17.5x19.6x27.3 cm, 425g). Battery-operated. Single-button interface with audible timer. No software algorithm complexity; simple timer-based control.
Indications for Use
Indicated for males with Norwood-Hamilton IIa to V hair loss and females with Ludwig-Savin I-1 to I-4, II-1, II-2 hair loss; both with Fitzpatrick Skin Types I to IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
K113097 — LASER HELMET · Theradome, Inc. · Mar 2, 2012
K122950 — LASER HELMET · Theradome, Inc. · Jun 14, 2013
K171775 — Theradome LH80 PRO · Theradome, Inc. · Jan 9, 2018
K213025 — Laser hair growth helmet · Kam Yuen Plastic Products , Ltd. · May 12, 2022
K181253 — Bosley Revitalizer 272 Laser Cap, Bosley Revitalizer 164 Laser Cup · Pjm Worldwide, LLC D/B/A Phoenix Medical Technology, LLC · Aug 10, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 18, 2018
Theradome, Inc. % Jacqueline Hauge Regulatory Affairs Consultant Jacqueline A. Hauge 23300 Manning Trail N Scandia, Minnesota 55073
Re: K180460
Trade/Device Name: Theradome LH40 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: February 14, 2018 Received: February 20, 2018
Dear Jacqueline Hauge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -ਟਤੋ For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K180460
Device Name Theradome LH40 (Theragrow)
### Indications for Use (Describe)
The Theradome LH40 (Theragrow) is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2; both with Fitzpatrick Skin Types I to IV.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| ☐ Research Use (Part 21 CFR 821 Subpart B) | ☑ Same Tier Contracting (21 CFR 821 Subpart C) |
|--------------------------------------------|------------------------------------------------|
|--------------------------------------------|------------------------------------------------|
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "theradome" in a stylized font. The first part of the word, "thera," is in gray, while the second part, "dome," is in light blue. Above the word, there are three curved lines in light blue, adding a design element to the logo.
#### l. SUBMITTER
| Date Prepared: | February 14, 2018 |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Jacqueline A. Hauge<br>Regulatory Consultant<br>Phone: 763.742.8227<br>Email: jacqueline.hauge@gmail.com |
| Sponsor: | Theradome, Inc.<br>4900 Hopyard Road, Suite 100<br>Pleasanton, CA 94588<br>Phone: 510-714-0525<br>Fax: 925-374-1202<br>Establishment Registration# 3008295459 |
| | |
| Sponsor Contact: | Tamim Hamid, PhD, CEO<br>Phone: 510-714-0525 |
| II. DEVICE | |
#### Trade Name: Theradome LH40 (Theragrow) Common Name: Lamp, non-heating, for promotion of hair growth Classification Name: Infrared Lamp Product Code OAP Regulatory Class: Class II Regulation Number: 21 CFR 890.5500 Panel: General and Plastic Surgery
#### lll. PREDICATE DEVICES
| Primary: | K161046 Theradome LH40 Evo (Theradome, Inc.) |
|-------------|--------------------------------------------------------------|
| Additional: | K171775 Theradome LH80 Pro (Theradome, Inc.) |
| Reference: | K152473 HairMax LaserComb 41 (Lexington International, LLC.) |
#### IV. PURPOSE OF SUBMISSION
The purpose of this Premarket Notification is to obtain FDA clearance for the expanded indications for use statement for the Theradome LH40 device to include treatment of Androgenetic Alopecia and promotion of hair growth in males who have Norwood-Hamilton Classifications of IIa to V patterns of hair loss.
#### DEVICE DESCRIPTION V.
The Theradome LH40 (Theragrow) laser helmet is a low level laser therapy (LLT) device containing red, visible light diode lasers operating at 678 nanometers, designed to deliver non-thermal energy to the hair follicles to promote hair growth via photobiostimulation of the scalp. The lasers are contained inside a lightweight, one-size-fits-all helmet. The Theradome LH40 device utilizes 40 laser diodes to deliver laser stimulation to the entire scalp for hands-free operation during treatment. The device is one-button operated, and has an audible timer that automatically turns the lasers off after the 20 minute treatment is complete.
{4}------------------------------------------------
#### INDICATIONS FOR USE VI.
The Theradome LH40 (Theragrow) is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of lla to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2; both with Fitzpatrick Skin Types I to IV.
### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Low level laser therapy (LLT) is the principle technology for both the subject and predicate devices. At a high-level, the Theradome LH40 (Theragrow) device has the same technological characteristics as the predicate device, including:
- Number of Laser Diode ●
- Laser Class
- Laser Power ●
- Wavelength
- Laser Delivery Method
- Helmet Design
- Single Button Operation
- . Audible Timer
- Treatment Time and Duration ●
- Materials of Construction ●
- Risk Profile
- Battery Operation
The fundamental scientific technology of the Theradome LH40 device was not changed as part of this submission.
| Technological<br>Characteristic | Theradome<br>LH40 | Theradome<br>LH80 | HairMax<br>LaserComb 41 | Theradome<br>LH40 |
|------------------------------------------|--------------------------------|--------------------------------|------------------------------------------------------|--------------------------------|
| Regulatory Information | | | | |
| Manufacturer | Theradome, Inc. | Theradome, Inc. | Lexington Int'l, LLC | Theradome, Inc. |
| 510(k) Number | K161046 | K171775 | K142573 | K180460 |
| Product Code | OAP | OAP | OAP | OAP |
| Device Type | OTC Only | OTC Only | OTC Only | OTC Only |
| Technological Characteristics Comparison | | | | |
| # of Coherent Light Diodes | 40 | 80 | 41 | 40 |
| Coherent Light Class | Class 3R | Class 3R | Class 3R | Class 3R |
| Wavelength (nm) | 678 (+7) | 678 (+7) | 655 (+10) | 678 (+7) |
| Laser Power for Classification | 5mW | 5mW | <5mW | 5mW |
| Power Flux (mW/cm²) | 0.476 | 0.952 | Not Reported | 0.476 |
| Radiant Energy per Treatment (J) | 432 | 216 | Not Reported | 432 |
| Fluence (J/cm²) | 0.52 | 1.03 | 0.0 | 0.52 |
| Treatment Schedule | 4 times/week | 2 times/week | 3-4 times/week<br>(every other day) | 4 times/week |
| Treatment Time | 1200 sec<br>(20 min) | 1200 secs<br>(20 min) | 30 seconds at each of<br>6 positions on the<br>scalp | 1200 sec<br>(20 min) |
| Total Treatment Time | 4800 sec/week<br>(80 min/week) | 2400 sec/week<br>(40 min/week) | 720 seconds<br>(9-12 min/week) | 4800 sec/week<br>(80 min/week) |
| Device Configuration | | | | |
| Device Configuration | Helmet | Helmet | Band | Helmet |
| Device Dimensions (cm) | 17.5x19.6x27.3 | 17.5x19.5x26.8 | 20x5x13 | 17.5x19.6x27.3 |
| Device Weight (g) | 425 | 425 | 156 | 425 |
| Nominal Scalp Coverage (cm²) | 420 | 420 | Not Reported | 420 |
# Comparison of Theradome LH40 to Predicate Devices
{5}------------------------------------------------
#### PERFORMANCE DATA VIII.
Performance testing was not required to support a substantial equivalence determination for the Theradome LH40 device. The existing performance testing; including biocompatibility, mechanical, software, and electrical and electromagnetic compatibility (EMC), are unchanged from the predicate device. Animal testing was not required to support a substantial equivalence determination for the Theradome LH40 device. Clinical testing was not required to support a substantial equivalence determination for the Theradome LH40 device.
#### CONCLUSIONS IX.
The Theradome LH40 device has been demonstrated to be substantially equivalent to the predicate device based on the equivalence of technological characteristics and intended use. Because there are no new or changed technological characteristics presented in this submission, the existing risk profile for the device remains unchanged. Theradome has demonstrated that the Theradome device is safe and effective for the expanded indications for use.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.