COMFEEL SEASORB DRESSING

K983519 · Coloplast Corp. · KMF · Dec 30, 1998 · General Hospital

Device Facts

Record IDK983519
Device NameCOMFEEL SEASORB DRESSING
ApplicantColoplast Corp.
Product CodeKMF · General Hospital
Decision DateDec 30, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5090
Device ClassClass 1

Indications for Use

The Comfeel® Seasorb Dressing is indicated (under the guidance of a health care professional) for use in management of moderate to heavily exudating wounds, partial and full thickness wounds, including leg ulcers and pressure sores.

Device Story

Comfeel® Seasorb Dressing is a sterile, single-use topical wound dressing for moderate to heavily exudating wounds. It consists of a calcium sodium alginate and sodium carboxymethylcellulose xerogel matrix cast onto a high-density polyethylene net. The dressing is applied by a healthcare professional to the wound site to manage exudate. It is not a treatment or cure for wounds, nor is it an artificial skin. The device is contraindicated for dry wounds and third-degree burns. It must be removed prior to radiation, X-ray, ultrasonic, diathermy, or microwave treatments. The device benefits patients by providing an absorbent barrier for wound management.

Clinical Evidence

Bench testing only. Testing demonstrated no cumulative irritation, no evidence of delayed hypersensitivity, and non-cytotoxicity (cytotoxicity grade ≤ 2). No clinical data provided.

Technological Characteristics

Materials: Calcium sodium alginate, sodium carboxymethylcellulose, high-density polyethylene net. Form factor: 4x6 cm, 10x10 cm, 15x15 cm. Sterilization: Ethylene oxide. Single-use. Non-active device.

Indications for Use

Indicated for management of moderate to heavily exudating wounds, partial and full thickness wounds, including leg ulcers and pressure sores. Not for use on dry wounds or third-degree burns.

Regulatory Classification

Identification

A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 310 1998 ### 510(k) Summary Comfeel® Seasorb Dressing Coloplast Corporation 1955 West Oak Circle Marietta. Georgia 30062-2249 Date: 10/05/98 ### 1. Contact Person Ms. Sydney Lilly. (770) 281-8260 ### 2. Name of the Medical Device | Proposed Classification name: | Hydrophylic Wound and Burn<br>Dressing | |-------------------------------|----------------------------------------| | Common/usual name: | Topical Wound Dressing | | Proprietary name: | Comfeel® Seasorb Dressing | ### 3. Device Classification The Comfeel® Seasorb Dressing has been classified by the FDA under the proposed heading of Hydrophylic Wound and Burn Dressing as an unclassified device. ### 4. Statement of Substantial Equivalence The revised Comfeel® Seasorb Dressing is equivalent to the original Comfeel® Alginate (Seasorb) Dressing, manufactured by Coloplast A/S and approved for market under 510(k) K953497, but with minor alterations. #### 5. Intended Use The revised Comfeel® Seasorb Dressing is indicated (under the guidance of a health care professional) for use in management of moderate to heavily exudating wounds, partial and full thickness wounds, including leg ulcers and pressure sores. ### 6. Description of Device The revised Comfeel® Seasorb Dressing consists of an absorbent calcium sodium alginate/sodium carboxymethy/ceilulose xerogel cast onto a high density polyethylene net. The dressing is supplied in three sizes: 4 x 6 cm (3705), and 10 x 10 cm (3710) and 15 x 15 cm (3715). The dressings are packaged in individual pouches and sterilized by ethylene oxlde. K 983519 {1}------------------------------------------------ K983519 ### 510(k) Summary Comfeel® Seasorb Dressing Page 2 of 3 The revised Comfeel® Seasorb Dressing is a modification of Comfeel® alginate (seasorb) dressing (K953497) which consists of 100% alginate. Through adjustment of the formula to 30% carboxymethylcullulose sodium and 70% alginate, the performance of the dressing has been improved measured on softness and absorption capacity. The revised Comfeel® Seasorb Dressing has been tested and shown no accumulative irritation effects, no evidence of delayed hypersensitivity and is non-cytotoxic (cytotoxicity grade ≤ 2). A comparison Matrix for the revised Comfeel® Alginate Dressing versus the original Comfeel® Alginate Dressing (K953497) is presented in Section 6. ### 7. Predicate Device | | Comfeel® Seasorb Dressing | Comfeel® Alginate (Seasorb)<br>Dressing | |-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Statement of<br>Identy | Seasorb Dressing is a sterile<br>Wound Dressing | Comfeel® Alginate (Seasorb) Dressing<br>is a sterile Wound dressing | | Device description | The Seasorb Dressing is a highly<br>absorbent wound dressing<br>consisting of an calcium sodium<br>alginate/sodium<br>carboxymethylcellulose Xerogel<br>cast into a high density<br>polyethylene net. | The Comfeel® Alginate (Seasorb)<br>Dressing is a highly absorbent material<br>composed of a Xerogel of calcium<br>sodium alginate cast into a high density<br>polyethylene net. | | Sizes | 4x6 cm, 10x10 cm, 15x15 cm | 4x6 cm, 10x10 cm, 15x15 cm | | Use (single,<br>reusable<br>disposable) | Single | Single | | Intended use | For management of moderate to<br>heavily exudating Stage I-IV<br>wounds, partial and full thickness<br>wounds, including leg ulcers,<br>pressure sores and burns etc. | For management of (under the<br>guidance of a health care professional)<br>moderate to heavily exudating wounds,<br>including leg ulcers and pressure<br>sores, etc. | Comfeel®: Alginate (Seasorb) Dressing (Colopiast, Inc. K953497) {2}------------------------------------------------ K983519 # 510(k) Summary Comfeel® Seasorb Dressing Page 2 of 3 | | Comfeel® Seasorb Dressing | Comfeel® Alginate (Seasorb)<br>Dressing | |---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Precautions | May be used on patients with<br>systemic infections when the<br>infectious conditions are under<br>medical control. | Wounds which are solely or mainly<br>caused by arterial insufficiency or<br>complicated diabetic wounds (primarily<br>lower leg and foot) should be inspected<br>by a physician or nurse regularly. | | | Seasorb Dressing must be<br>removed prior to radiation<br>treatment of long duration (X-rays,<br>ultrasonic treatment, diathermy<br>and micro waves). | A physician should be consulted before<br>using this product on wounds with a<br>high risk of infection, or on lesions<br>caused by syphilis, tuberculosis,<br>leprosy or cancer. | | | Not recommended for use on dry<br>wounds. | Comfeel®Seasorb (Seasorb) Dressing<br>must be removed prior to the following<br>treatments: radiation, X-rays, ultrasonic<br>treatment, diathermy and micro waves. | | | Do not use on patients with known<br>hypersensitivity to any of the<br>ingredients. | Wounds with signs of clinical infection,<br>fever and local symptoms such as pain<br>erythema (redness) or pus should have<br>a bacterial swab examination. Use of<br>this product may be continued at the<br>discretion of a physician. Current<br>systemic antibiotic treatment may be<br>given if indicated. | | | | Not recommended for use on dry<br>wounds or third degree burns. | | | | Do not use on patients with known<br>hypersensitivity to any of the<br>ingredients. | | Sterilization | Sterile | Sterile | | Packaging | Polyester pouches laminated<br>with peelable polyethylene prior<br>to sterilization | Polyester pouches laminated with<br>peelable polyethylene prior to<br>sterilization | {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three heads, depicted in black and white. ## DEC 30 1998 M. Sidney Lilly Coloplast Corporation 1955 West Oak Circle Marietta, Georgia 30062 Re: K983519 Trade Name: Comfeel Seasorb Dressing Regulatory Class: Unclassified Product Code: KMF Dated: October 5, 1998 Received: October 8, 1998 Dear Ms. Lilly: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act). You may, therefore, market your device subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations: - This device may not be labeled for use on third degree burns. 1. - This device may not be labeled as having any accelerating effect on the rate of 2. wound healing or epithelization. - 3. This device may not be labeled as a long-term, permanent, or no-change dressing, or as an artificial (synthetic) skin. - 4. This device may not be labeled as a treatment or a cure for any type of wound. The labeling claims listed above would be considered a major modification in the intended use of the device and would require a premarket notification submission (21 CFR 807.81). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 {4}------------------------------------------------ #### Page 2 - Ms. M. Sidney Lilly This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597 or at its internet address http://www.fda.gov/cdrh/dsmamain.html. Sincerely yours, Mark N Melburn Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ DEC. -18' 98(FRI) 13:51 P. 002 510(k) Number (if known): K983519 Device Name: Comfeel® Seasorb Dressing Indications for Use: The Comfeel® Seasorb Dressing is indicated for use (under the guidance of a health care professional) in management of moderate to heavily exudating wounds, partial and full thickness wounds, including leg ulcers and pressure sores. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark n Millenn (Division Sign of General Res 510(k) Number | Prescription Use<br>(Per 21 CFR 801.109 | | |-----------------------------------------|---| | OR Over-The-Counter Use | X | (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...