LIQUISHIELD-S (0.7G), MODEL LS 0010, LIQUISHIELD-S (2.0G), MODEL LS 0020
K023163 · Medlogic Global Limited · KMF · Jan 13, 2003 · General Hospital
Device Facts
Record ID
K023163
Device Name
LIQUISHIELD-S (0.7G), MODEL LS 0010, LIQUISHIELD-S (2.0G), MODEL LS 0020
Applicant
Medlogic Global Limited
Product Code
KMF · General Hospital
Decision Date
Jan 13, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5090
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
LiquiShield™-S is intended to protect intact or damaged skin from the effects of moisture, friction, (rubbing) or shear (tearing). LiquiShield™-S helps protect skin exposed to irritation from moisture such as urine, faeces, digestive juices and wound drainage. LiquiShield™-S can also be used in areas that are exposed to friction and shear such as occurs when items, such as bedding, clothing, or shoes, rub against the skin. LiquiShield™-S helps protect the skin against irritation caused by adhesive products.
Device Story
LiquiShield®-S is a cyanoacrylate-based liquid barrier film; applied topically to skin; dries within 45 seconds to form a transparent, flexible, waterproof film; wears off naturally as skin regenerates. Used to protect skin from moisture, friction, shear, and adhesive irritation. Applicator and packaging are sterilized. No complex electronics or software involved.
Clinical Evidence
Biocompatibility testing performed per ISO 10993-1. Mechanical and performance testing conducted to demonstrate equivalence to predicate. Clinical evaluations conducted to demonstrate device meets intended use.
Technological Characteristics
Cyanoacrylate-based liquid barrier film; non-cytotoxic; forms transparent flexible film upon drying; sterilized applicator and packaging. Biocompatibility testing per ISO 10993-1.
Indications for Use
Indicated for protection of intact or damaged skin from moisture (urine, feces, digestive juices, perspiration, wound drainage), friction, shear, and irritation from adhesive products. Contraindicated for deep, open, or bleeding wounds, chronic wounds, second or third-degree burns, and infected areas.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
{0}------------------------------------------------
JAN 1 3 2003
# 510 (k) Notification - LiquiShield®-S
| Submitted by: | MedLogic Global Limited<br>Western Wood Way<br>Langage Science Park<br>Plymouth, Devon. PL7 5BG UK<br>+44 1752 209955 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact Name: | Howard Beaumont |
| Date prepared: | 13th September 2002 |
| Device Trade Name: | LiquiShield®-S |
| Common Name: | Liquid Barrier Film |
| Classification name: | Liquid Bandage: CFR 880.5090 |
| Classification | Class 1 |
| Predicate Device: | 3M No Sting Barrier Film<br>3M Medical Products Group<br>K955103 |
510(k) Summary
#### Intended Use:
LiquiShield™-S is intended to protect intact or damaged skin from the effects of moisture, friction, (rubbing) or shear (tearing).
LiquiShield™-S helps protect skin exposed to irritation from moisture such as urine, faeces, digestive juices and wound drainage. LiquiShield™-S can also be used in areas that are exposed to friction and shear such as occurs when items, such as bedding, clothing, or shoes, rub against the skin.
LiquiShield™-S helps protect the skin against irritation caused by adhesive products.
## Contraindications:
Do not apply directly to deep, open, or bleeding wounds. Do not apply to chronic wounds Do apply to second or third degree burns. Do not apply to infected areas
MedLogic Global Limited - LiquiShield -- S Liquid Barrier Film
{1}------------------------------------------------
# 510(k) Summary
#### Description:
LiquiShield®-S is a non-cytotoxic (not harmful to the bodv) cyanoacrylate based, rapid drying, liquid barrier film for the protection of the skin. It is applied as a liquid and dries, within 45 seconds, adhering to the contours of the skin to form a transparent flexible film. LiquiShield®-S will wear off, naturally, as the skin regenerates. The applicator and packaging are sterilized.
### Substantial Equivalence:
LiquiShield®-S Liquid Barrier film is substantially equivalent to the following predicate device:
No Sting Barrier Film - 3M™ Medical Products Group K955103
LiquiShield®-S is applied as a liquid solution, which, upon contact with the skin, dries to form a barrier film, which is substantially equivalent to the predicate device. Substantial equivalence is also based upon intended use, application, product performance, including MVTR rates, waterproof properties, friction reduction and barrier film duration.
### Testing Summary:
LiquiShield®-S has been subjected to the appropriate biocompatibility testing in accordance with ISO 10993-1, the results of which confirm that product is safe for its intended use. LiquiShield®-S has also been subjected to mechanical and performance testing to demonstrate equivalence to the predicate device, with clinical evaluations conducted to demonstrate LiquiShield®-S meets its intended use.
MedLogic Global Limited – LiquiShield –S Liquid Barrier Film
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, often associated with healthcare.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 3 2003
Mr. Howard Beaumont Director of Quality Assurance and Regulatory Affairs MedLogic Global Limited Western Wood Way Langage Science Park Plympton, Plymouth Devon PL7 5BG England
Re: K023163
Trade/Device Name: LiquiShield® -S Regulation Number: 21 CFR 880.5090 Regulation Name: Liquid Bandage Regulatory Class: I Product Code: KMF Dated: December 23, 2002 Received: December 26, 2002
Dear Mr. Beaumont:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Mr. Howard Beaumont
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
tal Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510 (k) Notification - LiquiShield®-S
| Applicant: | MedLogic Global Limited |
|---------------------------|-------------------------|
| 510(k) Number (if known): | |
| Device Name: | LiquiShield®-S |
Indications for Use:
LiquiShield®-S is intended to protect intact or damaged skin from the effects of moisture, friction (rubbing), or shear (tearing).
LiquiShield®-S helps protect skin exposed to irritation from moisture such as urine, faeces, digestive juices, perspiration and wound drainage.
LiquiShield®-S can also be used in areas that are exposed to friction and shear such as occurs when items, such as bedding, clothing, or shoes, rub against the skin.
LiquiShield®-S helps protect the skin against irritation caused by adhesive products.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K023163
Prescription Use (per 21 CFR 801.109) OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
MedLogic Global Limited – LiquiShield –S Liquid Barrier Film
Revised on the 23 Dec 2002
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.