K191068 · Wuhan Lotuxs Technology Co., Ltd. · PKT · Jul 17, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191068
Device Name
Powersculp laser lipolysis system
Applicant
Wuhan Lotuxs Technology Co., Ltd.
Product Code
PKT · General, Plastic Surgery
Decision Date
Jul 17, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Powersculp laser lipolysis system is intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.
Device Story
Diode laser system (1060 nm) for non-invasive lipolysis; delivers laser energy to subcutaneous tissue layers to disrupt adipocyte cells. System comprises console and four applicators; applicators attached to patient via belt. Used in clinical settings; operated by healthcare professionals. Provides heat-assisted lipolysis; contact cooling integrated for patient comfort. Output affects appearance of visible fat bulges in abdomen and flanks; intended to provide aesthetic benefit to patients.
Clinical Evidence
No clinical data. Bench testing only. Includes software verification/validation, power output verification, electromagnetic compatibility (IEC 60601-1-2), electrical safety (ANSI AAMI ES60601-1, IEC 60601-2-22, IEC 60825-1), and biocompatibility (ISO 10993-1).
Technological Characteristics
Diode laser system; 1060 nm wavelength; 50W peak power per applicator; 4x8 cm² spot size; contact cooling; belt-based attachment. Electrical safety/EMC per ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1. Biocompatible materials per ISO 10993-1.
Indications for Use
Indicated for non-invasive lipolysis of flank and abdomen in individuals with BMI ≤ 30 to achieve disruption of adipocyte cells for aesthetic purposes.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in a sans-serif font, with "FDA" in a blue square and the rest of the text in blue.
July 17, 2019
Wuhan Lotuxs Technology Co., Ltd. % Jinghua Zhou Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 215, Huaming Building, Chebei Road Guangzhou, China 511660
Re: K191068
Trade/Device Name: Powersculp laser lipolysis system Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System for Aesthetic Use Regulatory Class: Class II Product Code: PKT Dated: April 22, 2019 Received: April 22, 2019
Dear Jinghua Zhou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
for
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil Ogden Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191068
Device Name
Powersculp laser lipolysis system
Indications for Use (Describe)
The Powersculp laser lipolysis system is intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive a desired a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|× Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Section 5 - 510(k) Summary
K191068
Date of Summary Preparation: April 15, 2019
### 1. Submitter's Identifications
Submitter's Name: Wuhan Lotuxs Technology Co., Ltd. Address: 5F, E2 Building, NO 999 Gaoxin Avenue, Future City, East lake High-Tech Development Zone, Wuhan 430206, China Contact Person: Na Wu Contact Title: QA Manager Contact Email Address: na.wu@lotuxs.com Telephone: +86-27-87619668 Fax: +86-27-87515058
### 2. Correspondent's Identifications
Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 215, Huaming Building, Chebei Road, Guangzhou, P.R. China ZIP Code: 511660 Contact Person: Jinghua Zhou Contact Title: Regulation Control Manager Contact E-mail Address: admanzhou@126.com Telephone: +86-20-82329549 Fax: +86-20-82329549
### 3. Name of the Device
Device Classification Name: Laser for disruption of adipocyte cells for aesthetic use Product Name: Low level laser system for aesthetic use Trade Name: Powersculp laser lipolysis system Model: PSP050 Classification Panel: General & Plastic Surgery Product Code: PKT Regulation Number: 21 CFR 878.5400 Device Classification: Class II
### 4. The Predicate Devices
Primary Predicate device: K160470 SculpSure Secondary predicate device: K150230 Sculpsure
#### 5. Device Description
The Powersculp laser lipolysis system is a diode laser system. Electrically efficient semiconductors generate optical radiation (1060 mm) which is used to deliver laser
{5}------------------------------------------------
## Wuhan Lotuxs Technology Co., Ltd.
energy to subcutaneous tissue layers. The Powersculp laser lipolysis system is capable of peak powers of 50W. The main components of Powersculp laser lipolysis system are a console and four applicators.
### 6. Intended Use of Device
The Powersculp laser lipolysis system is intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.
{6}------------------------------------------------
## Wuhan Lotuxs Technology Co., Ltd.
### 7. Summary of Substantial Equivalence
Table 1
| | Proposed device | Primary predicate device | Secondary predicate device | Comparison |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number | ------ | K160470 | K150230 | ------ |
| Product Code | PKT | PKT | PKT | Same |
| Proprietary Name | Powersculp laser lipolysis system | Sculpsure | Sculpsure | ------ |
| Model | PSP050 | / | / | ------ |
| Manufacturer | Wuhan Lotuxs Technology Co., Ltd. | Cynosure, Inc. | Cynosure, Inc. | ------ |
| Indications for use | The Powersculp laser lipolysis system is intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less. | The Cynosure SculpSure is a diode laser system intended for noninvasive lipolysis of the abdomen and flanks in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen and flanks. | The Cynosure SculpSure is intended for non-invasive lipolysis of the flanks to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. Intended for individuals with a Body Mass Index (BMI) of 30 or less. | Same<br>The indications for use of the proposed device are same as those of predicate device K160470 and K150230. |
| Operating theory | The Powersculp laser lipolysis system is a diode laser system. Electrically efficient | The Cynosure SculpSure is a diode laser system. The main components of | The Cynosure SculpSure is a diode laser system. Electrically efficient | Same<br>Although there are differences in expression |
| | | | | |
| | semiconductors generate optical radiation (1060 nm) which is used to deliver laser energy to subcutaneous tissue layers. The Powersculp laser lipolysis system is capable of peak powers of 50W. The main components of Powersculp laser lipolysis system are a console and four applicators. | SculpSure are a console and four applicators that deliver the laser energy to the patient. Electrically efficient semiconductors generate opticalradiation (1060 nm) which is used to deliver laser energy tosubcutaneous tissue layers. | semiconductors generate optical radiation (1064 nm) which is used to directly irradiate the skin's surface. The Sculpsure is intended for noninvasive lipolysis of the flanks to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. | among the three devices, the technical principles are essentially the same. |
| Structure and main components | The main components of Powersculp laser lipolysis system are a console and four applicators. | The main components of SculpSure are a console and four applicators. | The main components of SculpSure are a console and applicator. | Same<br>The main components of the proposed device are same as those of the primary predicate device K160470. |
| Laser type | diode laser | diode laser | diode laser | Same |
| Lipolysis method | Heat-assisted | Heat-assisted | Heat-assisted | Same |
| Wavelength | 1060nm±20 nm (infrared) | 1060 ±20 nm (infrared) | 1064nm | Same |
| Spot size | 4 × 8 cm² (A single applicator of four applicators) | 4 × 6 cm² on each of the applicator heads (4X) | 4 × 6 cm² (3X) | Similar<br>The amount of applicator is same between the primary predicate device K160470 and the proposed device. |
| Pulse Width (laser<br>ON time) | CW | CW | CW | The spot size of applicator<br>does not affect the safety<br>and effectiveness. |
| Power density | Up to 0.7-1.7W/ cm² | Up to 1.4 W/cm² | 1.7W/ cm² | Similar<br>The power density of the<br>proposed device is<br>customizable, and the<br>maximum power density is<br>same as the secondary<br>predicate device K150230,<br>so this defference is not<br>affect safety and<br>effectiveness. |
| Power supply | AC100-240V, 50/60Hz, 15A | 200-240V~, Single Phase,<br>20A | 120V, 20A | Similar<br>The power supply is<br>different, not affect safety<br>and effectiveness. |
| Peak power | 50W (per applicator) | 30W (per applicator) | 40W (per applicator) | Similar<br>The peak power is different,<br>which related to power<br>density and spot size. The<br>difference does not affect<br>safety and effectiveness. |
| Cooling | Contact cooling | Contact cooling | Contact cooling | Same |
| Attachment to<br>patient | belt | belt | belt | Same |
| Software control | Yes | Yes | Yes | Same |
| Electromagnetic | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| compatibility and | ANSI AAMI ES60601-1 | ANSI AAMI ES60601-1 | ANSI AAMI ES60601-1 | |
| electrical safefy | IEC 60825-1 | IEC 60825-1 | IEC 60825-1 | |
| compliance | IEC 60601-2-22 | IEC 60601-2-22 | IEC 60601-2-22 | |
| Discussion for<br>Substantially<br>Equivalent (SE) | The proposed device PSP050 has the same as the predicate device: indications for use, operating theory, structure and<br>main components, laser type, lipolysis method, wavelength, power width, cooling method, attachment to patient. The<br>differences only exist in such contents: spot size, power density, power supply, peak power that both can be controlled in<br>range of application. These minor differences between proposed device and predicate device raise no new issue of safety<br>and effectiveness. According to the non-clinical test results, the proposed devices are as safe, as effective and perform as<br>well as the predicate device.<br>So the proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. | | | |
{7}------------------------------------------------
# Wuhan Lotuxs Technology Co., Ltd.
{8}------------------------------------------------
# Wuhan Lotuxs Technology Co., Ltd.
{9}------------------------------------------------
# Wuhan Lotuxs Technology Co., Ltd.
{10}------------------------------------------------
## Wuhan Lotuxs Technology Co., Ltd.
#### 8. Non-Clinical Tests Submitted:
Software verification and validation was performed, and it was demonstrated that the software performs as intended according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Testing confirmed that the power output meets specification. Electromagnetic compatibility and electrical safety testing was performed per standards ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22 and IEC 60825-1. Results confirmed the device meets the standards. All patient contacting materials were assessed as per ISO 10993-1 and found to be biocompatible.
#### 9. Clinical Tests Submitted:
According to "8. Clinical Testing" of "Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use", and operating instruction and "7. Summary of Substantial Equivalence" submitted of the proposed device, the proposed device PSP050 is Substantially Equivalent (SE) to the predicate device, includes: indication for use, design (such as operating theory, structure and main component, software control), technology (such as laser type, lipolysis method, wavelength, power width, cooling method, attachment to patient), the non-clinical tests complied with the requirements of relevant recognized standards, and passed the bench tests (such as performance tests, storage condition tests), so Lotuxs belives that the proposed device Powersculp laser lipolysis system PSP050 does not need to carry out clinical tests, and the clinical study data that have been the legally marketed device can be used for reference.
#### 10. Conclusions drawn from clinical and non-clinical tests submitted:
Lotuxs believes that Powersculp laser lipolysis system PSP050 is substantially equivalent to its predicate devices with same indications for use, similar technological characteristics. The non-clinical data for Powersculp laser lipolysis system PSP050 supports the safety of the device and the biocompatibility, hardware and software verification and validation demonstrate that the Powersculp laser lipolysis system PSP050 should perform as intended in the specified use conditions.
--- End of this section ---
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.