K012632 · Medical Instruments Technology, Inc. · KCY · Oct 26, 2001 · General, Plastic Surgery
Device Facts
Record ID
K012632
Device Name
TOURNIQUET CUFF; PNEUMATIC TOURNIQUET
Applicant
Medical Instruments Technology, Inc.
Product Code
KCY · General, Plastic Surgery
Decision Date
Oct 26, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5910
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The intended use of a tourniquet cuff is to restrict blood flow to a patient's limb during surgery.
Device Story
Reprocessed pneumatic tourniquet cuffs consist of plastic or cloth covers over one or more plastic or rubber bladders. Plastic tubing connects the bladder to an external compressor. During surgery, the compressor inflates the bladder with air to restrict blood flow to the limb. The device is used in a surgical setting. Reprocessing involves washing with cleaning agents to remove bioburden, rinsing to remove residual cleansers, and ETO sterilization. The device is packaged in an environmentally controlled room to maintain cleanliness.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing confirmed the reprocessed devices maintain the ability to hold pressure equivalent to the predicate device.
Technological Characteristics
Pneumatic tourniquet cuff; plastic or cloth cover with plastic/rubber bladders; plastic tubing connector for external compressor. ETO sterilization. No changes to original manufacturer materials or functional design.
Indications for Use
Indicated for restricting blood flow to a patient's limb during surgery.
Regulatory Classification
Identification
A pneumatic tourniquet is an air-powered device consisting of a pressure-regulating unit, connecting tubing, and an inflatable cuff. The cuff is intended to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery.
K012614 — KENDALL, SEQUENTIAL COMPRESSION DEVICE; HUNTLEIGH, FLOWTRON; VENAFLOW; ACUFEX; ALPS; SHEEPSKIN · Medical Instruments Technology, Inc. · May 1, 2002
K123553 — ZIMMER A.T.S. 4000TS AUTOMATIC TOURNIQUET SYSTEM · Zimmer, Inc. · Aug 9, 2013
K060091 — ALLIANCE MEDICAL CORPORATION REPROCESSED COMPRESSION SLEEVES · Alliance Medical Corp. · Feb 17, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
### Medical Instruments Technology, Inc's Reprocessed Tourniquet Cuffs Premarket Notification
## Medical Instruments Technologyme
Quality Reprocessing and Surgical Cost Containment Systems
### Section 12: 510k Summary
#### l, Name of Submitter
Medical Instruments Technology, Inc. 385 North 3050 East Saint George, UT 84790 Tel: (435) 674-4010 Fax: (435) 674-9819
Contact persons: Tom Haueter, RA/QA Manager Crystal Batcabe, Assistant RA/QA Manager
### Summary Prepared August 10, 2001
#### Device name and Classification: II.
Common Name: Tourniquet cuff, Sleeve, Limb, Compressible Classification: Class II per 21 CFR 878.5910
#### III. Predicate Device:
MIT's reprocessed compressions cuffs are substantially equivalent to: Instrumed, Color Cuff (K890014)
#### IV. Description of Device
Tourniquet cuffs are composed of either a plastic or cloth cover, over one or more plastic or rubber bladders. There are one or more plastic tubes leading from the bladder to a connector that attaches to a compressor. All work on the same principal, a compressor inflates the bladder(s) with air to restrict the blood flow in the limb during surgery.
{1}------------------------------------------------
#### V. Intended Use:
The intended use of a tourniquet cuff is to restrict blood flow to a patient's limb during surgery.
#### VI. Technological Characteristics:
MIT's reprocessed tourniquet cuffs have the same technological characteristics as the Instrumed Color Cuffs (K890014). The materials used in the manufacture of the garments are not changed during the reprocessing process. Additionally, MIT does not change any part of the cuff that might affect its function.
Testing by MIT has shown that the reprocessed devices are substantially equivalent to the predicate in performance.
#### VII. Bio-Compatibility:
Because MIT does not change any of the materials in reprocessing, the biocompatibility of MIT's reprocessed tourniquet cuffs is equivalent to the original manufacturers'. MIT does wash the tourniquet cuffs to ensure that they are free of residual bioburden. The washing procedure utilizes cleaning agents, which have been chosen for their safety and effectiveness. MIT rinses all of the tourniquet cuffs after washing to ensure that no residual cleansers remain. Additionally, MIT packages the devices in an environmentally controlled room to ensure continued cleanliness of the devices.
MIT has validated the ETO sterilization system used. Additionally, MIT has performed residual ETO testing to ensure that residual ETO has been reduced to an acceptable level.
{2}------------------------------------------------
## Medical Instruments Technology, Inc's Reprocessed Tourniquet Cuffs Premarket Notification
#### Substantial Equivalence: VIII.
Physical Characteristics: Color, dimensions, damage
MIT compared the reprocessed tourniquet cuffs to the predicate device for the parameters above. In all cases the reprocessed devices were substantially equivalent to the new devices.
Performance Characteristics: Ability to hold pressure
MIT compared the reprocessed tourniquet cuffs to the predicate devices (for the parameters above) and in all cases the devices were substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the U.S. Department of Health & Human Services. The seal is circular, with the text "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol resembling an eagle or bird with three stylized wings or feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 6 2001
Mr. Jack Speer President Medical Instruments Technology, Inc. 385 North 3050 East St. George. Utah 84790
Re: K012632
Trade/Device Name: Reprocessed Pneumatic Tourniquet Cuffs Regulation Number: 878.5910 Regulation Name: Pneumatic tourniquet Regulatory Class: I Product Code: KCY Dated: August 10, 2001 Received: August 13, 2001
Dear Mr. Speer :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
### Page 2 - Mr. Jack Speer
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis leter will and in your e FDA finding of substantial equivalence of your device to a legally promated predicated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of Compliance at (301) 27 - 4 the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Outler general misticales, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susa Web
fa
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# KD12632 Medical Instruments Technology, Inc's Reprocessed Tourniquet Cuffs Premarket Notification Section 14:
Indications for Use Statement
The intended use of a tourniquet cuff is to restrict blood flow to a patient's limb during surgery.
I certify that, in my capacity as PRESIDENT of MEDICAL INSTRUMENTS TECHNOLOGY, INC., the indications for use stated above are accurate and that the device will not be marketed for any other indications for use.
Jack Jones
(Signature)
Jack Speer, President MIT, Inc. (Typed Name)
8-10-01
(Date)
Date)
.
Division Sign Off
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K012632
23
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.