K222081 · Freedom Laser Therapy, Inc. · OAP · Oct 13, 2022 · Physical Medicine
Device Facts
Record ID
K222081
Device Name
ID-510 iRestore Elite
Applicant
Freedom Laser Therapy, Inc.
Product Code
OAP · Physical Medicine
Decision Date
Oct 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ID-510 iRestore Elite is indicated to promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V, and in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, and in both, Fitzpatrick Classification of Skin Phototypes I to IV.
Device Story
Low-level laser/light therapy (LLLT) helmet for home use; treats androgenetic alopecia. Device inputs: line voltage (100/240V). Device transforms electrical energy into fixed-level light emission via 300 laser diodes (680nm) and 200 super luminescent diodes (625nm, 655nm). Delivers 2500mW total output per 12-minute session. Operator (patient) wears helmet; system operation is fixed and cannot be altered. Output: red light exposure to scalp. Benefit: promotes hair growth. Construction: ABS plastic helmet with silicone pad fixation ring.
Clinical Evidence
Bench testing only. Testing included cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), electrical safety (ANSI AAMI 60601-1), EMC (IEC 60601-1-2), software V&V, lifetime testing (5 years), and transportation testing (ASTM D4169).
Technological Characteristics
ABS plastic helmet with silicone fixation ring. Light source: 300 laser diodes (680±10nm) and 200 super luminescent diodes (625±10nm, 655±10nm). Laser Class 3R. Power: 2500mW total. Line voltage 100/240V. Standards: ISO 10993-5/10, ANSI AAMI 60601-1, IEC 60601-1-2, ASTM D4169.
Indications for Use
Indicated for promotion of hair growth in males (Norwood-Hamilton IIa-V) and females (Ludwig-Savin I-II) with androgenetic alopecia, Fitzpatrick Skin Phototypes I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 13, 2022
Freedom Laser Therapy, Inc. % Sharon Chen Regulatory Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K222081
Trade/Device Name: ID-510 iRestore Elite Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: July 12, 2022 Received: July 15, 2022
Dear Sharon Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K222081
Device Name
# ID-510 iRestore Elite
Indications for Use (Describe)
The ID-510 iRestore Elite is indicated to promote hair growth in males who have Norwood-Hamilton Classifications of Ila-V, and in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, and in both, Fitzpatrick Classification of Skin Phototypes I to IV.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{4}------------------------------------------------
#### 5. 510(k) Summary K222081
### 5.1 Submission Sponsor
Freedom Laser Therapy Inc. 16782 Von Karman Ave, Unit 15 Irvine, CA, 92606 USA Contact: Wei-Chih (Kevin) Chen Title: President
## 5.2 Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Email: LST.AUS.ProjectManagement(@ul.com Contact: Sharon Chen Title: Regulatory Consultant
#### 5.3 Date Prepared
July 12, 2022
#### 5.4 Device Identification
| Trade Name: | ID-510 iRestore Elite |
|----------------------|-----------------------|
| Classification Name: | Physical Medicine |
| Regulation Name: | Infrared Lamp |
| Regulation Number: | 21 CFR 890.5500 |
| Product Code: | OAP |
| Class: | Class II |
#### 5.5 Legally Marketed Predicate Device
| Trade Name: | ID-500 iRestore Hair Growth System |
|----------------------|------------------------------------|
| 510(k) Number: | K213094 |
| Classification Name: | Physical Medicine |
| Regulation Name: | Infrared Lamp |
| Regulation Number: | 21 CFR 890.5500 |
| Product Code: | OAP |
| Class: | Class II |
#### 5.6 Reference Devices
iRestore Professional 282 (K183417), manufactured by Remax Medi-Tech (Shenzhen) Corporation
REVIAN RED (K173729), manufactured by PhotonMD, Inc.
Theradome LH40 (K180460), manufactured by Theradome, Inc.
{5}------------------------------------------------
## 5.7 Indications for Use Statement
The ID-510 iRestore Elite is indicated to promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V, and in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, and in both, Fitzpatrick Classification of Skin Phototypes I to IV.
### 5.8 Device Description
The ID-510 iRestore Elite is a low-level laser/light system operating at 680 ± 10, 655 ± 10, and 625± 10 nanometers. The physical configuration is that of a helmet containing an inner and outer liner, stabilized with a silicone pad lined fixation ring. The device helmet is constructed of an ABS type plastic. The system operates on line voltage at 100 or 240 volts. The helmet's inner liner permits full adjustment to any heads of non-toxic silicone pads. The helmet contains 300 units of five-millwatt-diode lasers and 200 units of fivemilliwatt, super luminescent diodes, that emit red light. This system delivers fixed laser emission levels, measured to be 2500 total millwatts per 12 minute treatment session, which cannot be altered by the operator.
#### 5.9 Substantial Equivalence Discussion
The following table compares the ID-510 iRestore Elite to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.
| Attribute | ID-510<br>iRestore Elite | ID-500<br>iRestore Hair<br>Growth<br>System | iRestore<br>Professional<br>282 | REVIAN<br>RED | Theradome<br>LH40 | Capillus 112,<br>Capillus 244 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Category | Subject<br>Device | Predicate<br>Device | Reference<br>Device | Reference<br>Device | Reference Device | Reference<br>Device |
| 510(k)<br>Number | Not applicable | K213094 | K183417 | K173729 | K180460 | K192012 |
| Manufacturer | Remax Medi-<br>Tech<br>(Shenzhen)<br>Corporation | Remax Medi-<br>Tech<br>(Shenzhen)<br>Corporation | Remax Medi-<br>Tech<br>(Shenzhen)<br>Corporation | PhotonMD,<br>Inc. | Theradome, Inc. | Capillus |
| Product Code | OAP | OAP | OAP | OAP | OAP | OAP |
| Regulation<br>Number | 21 CFR<br>890.5500 | 21 CFR<br>890.5500 | 21 CFR<br>890.5500 | 21 CFR<br>890.5500 | 21 CFR 890.5500 | 21 CFR<br>890.5500 |
| Indications<br>for Use | The ID-510<br>iRestore Elite<br>is indicated to<br>promote hair<br>growth in<br>males who<br>have<br>Norwood-<br>Hamilton<br>Classifications | The predicate<br>Is indicated to<br>treat<br>androgenetic<br>alopecia in<br>men<br>It is<br>designed to<br>promote | The iRestore<br>Professional<br>282<br>is indicated<br>for promote<br>hair and<br>growth in<br>females with<br>to androgenetic<br>hair alopecia | REVIAN RED<br>is indicated to<br>treat<br>Androgenetic<br>hair Alopecia<br>in promote<br>growth<br>males<br>who have<br>Norwood-Hamilton | The Theradome<br>LH40<br>(Theragrow) is an<br>over the counter<br>and (OTC)<br>hair therapeutic<br>in device<br>to treat<br>Androgenetic<br>Alopecia | The CapillusX<br>and CapillusX+<br>laser domes are<br>intended to treat<br>Androgenetic<br>Alopecia and<br>promotion of<br>hair growth in<br>males who have<br>and Norwood |
| | of IIa-V, and males<br>in females have<br>with androgenetic<br>alopecia who<br>have Ludwig-<br>Savin<br>Classifications<br>of I-II, and in<br>both,<br>Fitzpatrick<br>Classification<br>of<br>Phototypes<br>to IV. | | who Savin<br>Classifications<br>of I-II, males<br>who have loss and<br>Norwood-<br>Hamilton<br>Classifications<br>of IIa to V and<br>in Ludwig-<br>Savin<br>Classifications<br>of I to II. All of<br>Skin users should<br>I also<br>Skin Types I<br>to IV. | Classifications<br>promote<br>treat<br>Androgenetic<br>Alopecia<br>for both, growth<br>Fitzpatrick<br>Classification<br>Skin Savin Scale I-1<br>I to I-4, II-1, II-2<br>or frontal<br>patterns of hair | | hair Hamilton<br>Classifications<br>of IIa - V growth in<br>have of IIA to V<br>loss and treating<br>Classifications of<br>and IIa to V patterns<br>hair of hair loss and to<br>in treat<br>females who<br>and have<br>hair (Savin) Scale 1-<br>in 4. II-1, II-2, or<br>who frontal; both<br>Skin Types I to |
| OTC or Rx Only | OTC | OTC | OTC | OTC | OTC | OTC |
| Wavelength<br>of LED | 625 ± 10 nm<br>and 655 ± 10 nm | 655 ± 10 nm | 655 ± 10 nm | 620 – 660 nm | No LED | No LED |
| Wavelength<br>of Laser | 680 ± 10 nm | 655 ± 10 nm | 655 ± 10 nm | No laser | 620 - 660 nm | 678 ± 7 nm |
| Laser Power<br>for<br>Classification | Laser Class 3R | Laser Class 3R | Laser Class 3R | Not applicable | Laser Class 3R | Laser Class 3R |
| Time and<br>Frequency of<br>Treatment | 12 minutes per<br>treatment, to be<br>used daily. | 25 minutes per<br>treatment, to be<br>used every day. | 25 minutes per<br>treatment, to be<br>used other every<br>day. | 10 minutes per<br>treatment, to be<br>used daily. | 20 minutes per<br>treatment, to be<br>used four times<br>per week. | Six minutes per<br>treatment, to be<br>used daily. |
Table 5.1: Comparison of Characteristics
{6}------------------------------------------------
#### 5.10 Non-Clinical Performance Data
To demonstrate safety and effectiveness of ID-510 iRestore Elite and to show substantial equivalence to the predicate device, Remax Medi-Tech (Shenzhen) Corporation completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The ID-510 iRestore Elite passed the testing in accordance with internal requirements, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:
- · Cytotoxicity testing per ISO 10993-5 Passed
- · Sensitization testing per ISO 10993-10 Passed
- · Electrical safety testing per ANSI AMMI 60601-1 Passed
- · Electromagnetic Disturbance (EMC) testing per IEC 60601-1-2 Passed
- · Software verification and validation per FDA Guidance Compliant
- · Lifetime Testing Supports lifetime of five years
{7}------------------------------------------------
- · Transportation Testing per ASTM D4169 Demonstrates package integrity maintained
#### 5.11 Statement of Substantial Equivalence
The ID-510 iRestore Elite has the same indications for use as the ID-500 iRestore Hair Growth System. Any minor differences in the technological characteristics of the subject device when compared to the predicate device have been successfully evaluated through appropriate safety and performance testing which demonstrates that the subject device, when compared to the predicate device, does not raise any new questions of safety and effectiveness. Therefore, ID-510 iRestore Elite has been determined to be substantially equivalent to ID-500 iRestore Hair Growth System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.