K200929 · Xuzhou Kernel Medical Equipment Co., Ltd. · OAP · Jun 3, 2020 · Physical Medicine
Device Facts
Record ID
K200929
Device Name
Hair Growth System
Applicant
Xuzhou Kernel Medical Equipment Co., Ltd.
Product Code
OAP · Physical Medicine
Decision Date
Jun 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hair Growth System (Model: KN-8000B/KN-8000C/KN-8000D/KN-8000E) is intended for the promotion of hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenetic alopecia who have Norwood Hamilton Classifications Ila-V; and both genders having Fitzpatrick Classification of Skin Phototypes I-IV.
Device Story
Hands-free, portable, noninvasive low-level laser therapy (LLLT) device; utilizes red visible light diode lasers and/or LEDs (650nm/655nm) in a helmet form factor. Input: user-positioned head within the helmet. Operation: dot matrix laser arrangement targets hair follicles; automatic pause/resume function based on head position; audible voice prompts for treatment status (models KN-8000D/E). Power: internal rechargeable Li-ion battery or external AC adapter. Used in home/clinic settings by patients. Output: red light radiation to scalp. Clinical benefit: promotion of hair growth. Healthcare provider uses device output to support hair regrowth therapy.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including electrical safety (IEC 60601-1, IEC 60825-1, IEC 60601-2-57), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-5, ISO 10993-10), and usability (IEC 62366).
Technological Characteristics
LLLT helmet; red visible light diode lasers/LEDs (650-655nm); max output <5mW per diode; Class 3R laser classification. Powered by 7.2V Li-ion battery or 9V DC adapter. Patient-contacting materials biocompatible per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 62471, IEC 60601-2-57.
Indications for Use
Indicated for females with androgenic alopecia (Ludwig-Savin I-II) and males with androgenetic alopecia (Norwood Hamilton IIa-V), Fitzpatrick Skin Phototypes I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 3, 2020
Xuzhou Kernel Medical Equipment Co., Ltd. % Jet Li Regulation Manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong China
Re: K200929
Trade/Device Name: Hair Growth System Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: March 15, 2020 Received: April 7, 2020
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K200929
Device Name
Hair Growth System (Model: KN-8000B/KN-8000C/KN-8000D/KN-8000E)
#### Indications for Use (Describe)
The Hair Growth System (Model: KN-8000B/KN-8000C/KN-8000E) is intended for the promotion of hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenetic alopecia who have Norwood Hamilton Classifications Ila-V; and both genders having Fitzpatrick Classification of Skin Phototypes I-IV.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Date of the summary prepared: May 30, 2020
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. This is a traditional 510(K) submission with no previous application.
#### 1. Submitter's Information
#### Sponsor
- � Company Name: Xuzhou Kernel Medical Equipment Co., Ltd.
- Address: Kernel Mansion, Economic Development District, Xuzhou City, Jiangsu Province, � China
- � Phone: +86 1831 685 8036
- Fax: +86 0516-87732208 13776587162 �
- Contact Person (including title): Wang Jing (Management Representative) �
- E-mail: wikernel@126.com �
#### Application Correspondent:
- Company Name: Guangzhou KEDA Biological Tech Co., Ltd. �
- Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou City, China �
- � Contact Person: Jet Li
- Tel: +86-18588874857 �
- Email: med-jl@foxmail.com �
#### Subject Device Information 2.
- � Trade Name: Hair Growth System
- Model: KN-8000B/KN-8000C/KN-8000D/KN-8000E �
| Data<br>Model | LD Quantity | LED Quantity |
|---------------|-------------|--------------|
| KN-8000B | 204 | -- |
| KN-8000C | 80 | 30 |
| KN-8000D | 21 | 30 |
| KN-8000E | 81 | -- |
- Common Name: Lamp, non-heating, for promotion of hair growth �
- Classification name: Infrared lamp �
- � Review Panel: General & Plastic Surgery
{5}------------------------------------------------
- Product Code: OAP (Laser, comb, hair) �
- � Requlation Class: 2
- Regulation Number: 21 CFR 890.5500 �
# 3. Predicate Device Information
- � Predicate Devices for KN-8000B and KN-8000E
| | Predicate I | Predicate II |
|-------------------|----------------------|-----------------------------------------------------------------------------------------------------------|
| Sponsor | DermaScalp LLC | Capillus LLC |
| Device Name | DermaScalp Laser Cap | Capillus 82, Capillus 202, Capillus 272 Pro, 272 Office Pro, Capillus 302, Capillus 312, and Capillus 352 |
| 510(k) Number | K173846 | K163170 |
| Product Code | OAP | OAP |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 |
| Regulation Class | 2 | 2 |
#### Predicate Devices for KN-8000C and KN-8000D �
| | Predicate I | Predicate II |
|-------------------|------------------------------|-----------------------------|
| Sponsor | Freedom Laser Therapy, Inc. | Freedom Laser Therapy, Inc. |
| Device Name | iRestore Hair Grow th System | iRestore Professional 282 |
| 510(k) Number | K151662 | K183417 |
| Product Code | OAP | OAP |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 |
| Regulation Class | 2 | 2 |
#### 2. Device Description
The Hair Growth System (Model: KN-8000B/KN-8000D/KN-8000E) is hands-free, portable, noninvasive, low-level laser device, which consists of red visible light diode lasers and/or LEDs, to produce red light operating at 650nm wavelength (maximum output power of each is 5mW). The laser sources are arranged in a dot matrix arrangement in the inner of the helmet, which could take into account of every hair follicle and promote rapid hair growth. The device will automatically pause therapy if the user's head is moved outside of the zone of radiation and will resume therapy when the correct head position is reestablished.
The device could be powered by internal rechargeable li-on battery (Rated 7.2Vd.c. 2100mAh), and it can also be supplied by specified external adapter with rated Input 100-240Vac. 50/60Hz and rated Output 9Vdc 3A.
For KN-8000D and KN-8000E, the device emits an audible voice prompt when the treatment is going to start/pause/end, and the device is going to shut down.
{6}------------------------------------------------
# 3. Intended Use / Indications for Use
The Hair Growth System (Model: KN-8000B/KN-8000C/KN-8000E) is intended for the promotion of hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenetic alopecia who have Norwood Hamilton Classifications Ila-V; and both genders having Fitzpatrick Classification of Skin Phototypes I-IV.
# 4. Test Summary
The Hair Growth System has been evaluated for its safety and performance by lab bench testing as following:
- � Electrical safety and performance test according to IEC 60601-1, IEC 60825-1, IEC 62471 and IEC 60601-2-57 standard
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standard
- � Usability test according to IEC 62366
- � Software verification and validation test according to the requirements of the FDA "Guidance for Premarket Submissions and for Software Contained in Medical Devices"
#### 5. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, and intended use of Hair Growth System, model: KN-8000B/KN-8000D/KN-8000E is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
{7}------------------------------------------------
| Substantial Equivalence Comparison Table for<br>KN-8000B/KN-8000E | | | | | |
|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|--|
| Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict | |
| 510(k) Number | TBD | K173846 | K163170 | -- | |
| Device Name | Hair Grow th System<br>KN-8000B/ KN-8000E | DermaScalp Laser Cap | Capillus 82, Capillus 202, Capillus<br>272 Pro, 272 Office Pro, Capillus<br>302, Capillus 312, and Capillus 352 | -- | |
| Product Code | OAP | OAP | OAP | SE | |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | 21 CFR 890.5500 | SE | |
| Regulation Class | 2 | 2 | 2 | SE | |
| LLLT Device Type | LLLT | LLLT | LLLT | SE | |
| Prescription | OTC | OTC | OTC | SE | |
| Intended Use | The Hair Grow th System (Model: KN-<br>8000B/KN-8000C/KN-8000D/KN-8000E) is<br>intended for the promotion of hair grow th in<br>females with androgenic alopecia w ho<br>have Ludwig-Savin Classifications I-II, and<br>in males with androgenetic alopecia w ho<br>have Norwood Hamilton Classifications Ila-<br>V; and both genders having Fitzpatrick<br>Classification of Skin Phototypes I-IV. | The family of DermaScalp Laser Cap<br>devices are indicated to treat<br>Androgenetic Alpopecia and promote<br>hair grow th in Males w ho have<br>Norw ood Hamilton Classifications of<br>Ila to V patterns of hair loss and to<br>treat Androgenetic Alopecia and<br>promote hair grow th in Females w ho<br>have Ludwig (Savin) Scale I-1 to I-4,<br>Il-1, Il-2, or frontal; both with<br>Fitzpatrick Skin Types I to IV. | The Capillus laser domes 82, 202, 272<br>Pro, 272 OfficePro, 302, 312, and 352,<br>are intended to treat Androgenetic<br>Alopecia and promote hair grow th in<br>males w ho have Norw ood -Hamilton<br>Classifications of Ila to V patterns of<br>hair loss and to treat Androgenetic<br>Alopecia and promote hair grow th in<br>females who have Ludwig (Savin) Scale<br>I-1 to I-4, II-1, II-2, or frontal; both with<br>Fitzpatrick Skin Types I to IV. | SE<br>Minor<br>difference | |
| Waveform | Visible red laser | Visible red laser | Visible red laser | SE | |
| Wavelength | 655nm ±5nm | 650nm | 650nm | SE<br>Note 1 | |
| Amounts of laser diode | KN-8000B: 204<br>KN-8000E: 81 | 50, 80, 120, 148, 180, 202, 224, 272 | 82, 202, 272, 302, 312, 352 | SE<br>Note 2 | |
| Laser radiation output | <5mW | <5mW | ≤5mW | SE<br>Note 1 | |
| Classification according to<br>IEC60825-1 | Class 3R | Class 3R | Class 3R | SE | |
| Substantial Equivalence Comparison Table for<br>KN-8000B/KN-8000E | | | | | |
| Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict | |
| Treatment | Each Treatment: 25-35min,<br>Total Treatment: every other day | 17weeks, every other day (indefinite) | 17weeks, every other day (indefinite) | SE<br>Note 2 | |
| Total laser irradiance | KN-8000B: ≤1,020mW<br>KN-8000E: ≤405mW | Diode 50: <250mW,<br>Diode 80: <400mW,<br>Diode 120: < 600mW,<br>Diode 148: <740mW,<br>Diode 180: <900mW,<br>Diode 202: <1,010mW,<br>Diode 224: <1,120mW,<br>Diode 272: <1,360mW | Diode 82: <410mW,<br>Diode 202: <1,010mW,<br>Diode 272: <1,360mW,<br>Diode 302: <1,510mW,<br>Diode 312: <1,560mW,<br>Diode 352: <1,760mW | SE<br>Note 2 | |
| Appearance Design | Helmet cap | Cap | Cap | SE<br>Minor<br>difference | |
| Safety and Performance<br>Feature | Complied with IEC 60601-1, IEC 60601-1-<br>2, IEC 60825-1 | Complied with IEC 60601-1, IEC<br>60601-1-2, IEC 60825-1, IEC 60601-<br>1-11 | Complied with IEC 60601-1, IEC<br>60601-1-2, IEC 60825-1 | SE<br>Note 3 | |
| Biocompatibility | All patient contacting materials are<br>complied with ISO 10993-5, ISO 10993-10 | All patient contacting materials are<br>complied with ISO 10993-1, ISO<br>10993-5 | All patient contacting materials are<br>complied with ISO 10993-5, ISO 10993-<br>10 | SE<br>Note 3 | |
{8}------------------------------------------------
| Substantial Equivalence Comparison Table for<br>KN-8000C/KN-8000D | | | | |
|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict |
| 510(k) Number | TBD | K151662 | K183417 | -- |
| Device Name | Hair Grow th System | iRestore Hair Grow th System | iRestore Professional 282 | -- |
| Substantial Equivalence Comparison Table for<br>KN-8000C/KN-8000D | | | | |
| Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict |
| | KN-8000C/ KN-8000D | | | |
| Product Code | OAP | OAP | OAP | SE |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | 21 CFR 890.5500 | SE |
| Regulation Class | 2 | 2 | 2 | SE |
| LLLT Device Type | LLLT | LLLT | LLLT | SE |
| Prescription | OTC | OTC | OTC | SE |
| Intended Use | The Hair Growth System (Model: KN-<br>8000B/KN-8000C/KN-8000D/KN-8000E) is<br>intended for the promotion of hair growth in<br>females with androgenic alopecia who have<br>Ludwig-Savin Classifications I-II, and in<br>males with androgenetic alopecia who have<br>Norwood Hamilton Classifications IIa-V; and<br>both genders having Fitzpatrick Classification<br>of Skin Phototypes I-IV. | The iRestore Hair Growth System is<br>indicated to promote hair growth in<br>females with androgenetic alopecia<br>who have Ludwig-Savin<br>Classifications of I - II, males who<br>have Norwood-Hamilton<br>Classifications of IIa - V and for both,<br>Fitzpatrick Classification of Skin<br>Phototypes I to IV. | The iRestore Professional 282 is<br>indicated to promote hair growth in<br>females with androgenetic alopecia<br>who have Ludwig-Savin<br>Classifications of I-II, males who have<br>Norwood-Hamilton Classifications of<br>IIa-V and for both, Fitzpatrick<br>Classification of Skin Phototypes I to<br>IV. | SE |
| Waveform | Visible red laser and light | Visible red laser and light | Visible red laser and light | SE |
| Wavelength | LD: 655nm±5nm<br>LED: 655nm±20nm | 655nm | 650nm±10nm | SE<br>Note 1 |
| Amounts of LED | KN-8000C: 30<br>KN-8000D: 30 | 30 | 200 | SE<br>Note 2 |
| Amounts of laser diode | KN-8000C: 80<br>KN-8000D: 21 | 21 | 82 | SE<br>Note 2 |
| Classification according to<br>IEC60825-1 | Class 3R | Class 3R | Class 3R | SE |
| Treatment time | Each Treatment: 25-35min<br>Total Treatment: every other day | Each Treatment: 25min<br>Total Treatment: every other day, on<br>non-consecutive days, for 16weeks | Each Treatment: 25min<br>Total Treatment: every other day, on<br>non-consecutive days, for 16weeks | SE<br>Note 2 |
| Appearance Design | Helmet cap | Helmet | Helmet | SE |
| Substantial Equivalence Comparison Table for<br>KN-8000C/KN-8000D | | | | |
| Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict |
| Safety and Performance<br>Feature | Complied with IEC 60601-1, IEC 60601-1-2,<br>IEC 60825-1, IEC 62471, IEC 60601-2-57 | Complied with IEC 60601-1, IEC<br>60601-1-2, IEC 60825-1, IEC 60601-<br>1-11 | Complied with IEC 60601-1, IEC<br>60601-1-2, IEC 60825-1, IEC 60601-<br>1-11 | SE<br>Note 3 |
| Biocompatibility | All patient contacting materials are complied<br>with ISO 10993-5, ISO 10993-10 | All patient contacting materials are<br>complied with ISO 10993-5, ISO<br>10993-10 | All patient contacting materials are<br>complied with ISO 10993-5, ISO<br>10993-10 | SE |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
# Comparison in Detail(s):
### Note 1:
Although the Wavelength and Laser radiation output of subject device have a little difference to the predicate devices, these parameters are similar to the predicate devices, and these minor difference of wavelength will not affect the main function. So the differences will not raise any safety or effectiveness issue.
### Note 2:
Although the Amounts of LED and/or laser diode/Treatment Time/Total Irradiance of subject device and predicate devices are a little difference, the energy and power parameters' range of subject device can be covered by predicate device's several models' range; they are very similar. So these parameters' differences will not raise any safety or effectiveness issue.
### Note 3:
Although the accepted FDA recognized standards of subject device are a little different from the predicate devices, they all comply with IEC 60101-1/IEC 60601-1-2/IEC 60825, and the subject device also complies with IEC 62471 requirements, so these differences will not affect the critical functions or the normal use.
#### Summary for clinical test 6.
Clinical performance is not deemed necessary.
# 7. Conclusion
The subject device Hair Growth System (KN-8000B/KN-8000D/KN-8000E) has all features of the predicate devices for intended use. Thus, the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.