K182562 · Wontech · OAP · Nov 16, 2018 · Physical Medicine
Device Facts
Record ID
K182562
Device Name
Hair Boom 69
Applicant
Wontech
Product Code
OAP · Physical Medicine
Decision Date
Nov 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hair Boom 69 is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I – II, males who have Norwood-Hamilton Classifications of IIa - V and for both, Fitzpatrick Classification of Skin Phototypes I to IV.
Device Story
Hair Boom 69 is a low-level laser therapy (LLLT) helmet device used for the promotion of hair growth. It utilizes 69 diode lasers (650 +/- 5 nm wavelength) configured within a helmet and protective inner liner to provide coverage to the upper third of the scalp. The device is intended for prescription use in a home setting. The patient wears the helmet for a 30-minute treatment session, performed every other day or three times per week for 16 weeks. The system provides audible feedback via a beep pattern to signal the completion of the therapy cycle. By delivering radiant energy to the scalp, the device aims to stimulate hair growth in patients with androgenetic alopecia. The device is hands-free and passive during operation.
Clinical Evidence
No clinical performance data was produced for this submission. Substantial equivalence is based on equivalent optical, electronic, and mechanical function to the predicate device. Bench testing was performed to verify compliance with IEC 60825-1 (laser safety), IEC 60601-1 (basic safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), IEC 62304 (software), ISO 14971 (risk management), and ISO 10993 (biocompatibility).
Technological Characteristics
LLLT device featuring 69 diode lasers (650 +/- 5 nm wavelength). Form factor is a helmet with protective inner liner. Class 3R laser system. Standards: IEC 60825-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62304, ISO 14971, ISO 10993-1/10. Operates on a 30-minute timer with audible beep notification for cycle completion.
Indications for Use
Indicated for females with androgenetic alopecia (Ludwig-Savin I-II) and males with androgenetic alopecia (Norwood-Hamilton IIa-V) with Fitzpatrick Skin Phototypes I-IV to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
WONTECH % Raymond Blanche Consultant NST Consultants. Inc. 641 Shunpike Road, Suite 311 Chatham, New Jersey 07928
November 16, 2018
## Re: K182562
Trade/Device Name: Hair Boom 69 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: September 4, 2018 Received: September 18, 2018
Dear Raymond Blanche:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Neil R.P.
Ogden
Digitally signed by Neil
R.P. Ogden
Date: 2018.11.16
14:44:27 -05'00'
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182562
Device Name Hair Boom 69
Indications for Use (Describe)
The Hair Boom 69 is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications ofl - II, males who have Norwood-Hamilton Classifications of Ila- V and for both, Fitzpatrick Classification of Skin Phototypes I to IV.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
1ZI Prescription Use (Part 21 CFR 801 Subpart D)
D Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."
{3}------------------------------------------------
## 510(k) Summary –Revised 10-25-2018
# WONTECH
## Submitter's Contact Information
| Name: | Raymond R. Blanche |
|------------|----------------------------------------------------------------------------|
| Address | NST Consultants, Inc.<br>641 Shunpike Road, Suite 311<br>Chatham, NJ 07928 |
| Telephone: | (973)539-7444 |
| Facsimile: | (973)539-7445 |
## Name of Device and Name/Address of Sponsor
| Trade Name: | Hair Boom 69 |
|---------------------------------|---------------------------------------------------------------------------------------------------------|
| Sponsor Contact<br>Information: | Ms. Gyuri (Erin) Park<br>WONTECH.<br>64 Techno 8-ro, Yuseong-Gu<br>Daejeon<br>Republic of Korea 305-500 |
| Common or Usual Name: | Lamp, non-heating, for promotion of hair growth |
|-----------------------|-------------------------------------------------|
| Classification Name: | Infrared lamp per 21 CFR 890.5500 |
| Classification Code: | OAP (Laser, comb, hair) |
## Predicate Devices:
| Device Trade Name | Manufacturer |
|--------------------------|------------------------------|
| HairMax Laser 80 K180885 | Lexington International,LLC. |
## Reference Devices:
{4}------------------------------------------------
Date Prepared: September 4, 2018
## Intended Use / Indications for Use
The Hair Boom 69 is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I – II, males who have Norwood-Hamilton Classifications of IIa - V and for both, Fitzpatrick Classification of Skin Phototypes I to IV.
## Technological Characteristics
The Hair Boom 69 diode laser is configured within an outer helmet and protective inner liner. The use of diode lasers provides for a full coverage of the upper 1/3 of the head; i.e., the area commonly covered with stylized hair. At the end of the therapy cycle, the system signals that therapy is complete and ready to be powered down, by emitting an audible beep pattern.
## Performance Data:
No clinical performance data was produced for this submission because the Hair Boom 69 is an equivalent device in optical, electronic and mechanical function as well as recommended clinical treatment regime, to the predicate device, the HairMax Laser 80. There is one ergonomic design distinction between the two systems. The Hair Boom 69 is a substantive helmet design versus a cap design for the HairMax Laser 80.
The Hair Boom 69 was tested to internationally recognized standards, consistent with the current recommendations adopted by the FDA.
- 1. IEC 60825-1 Edition 2.0 2007 03 Laser Safety & Classification
- 2. IEC 60601-1:2005 (currently called AAMI/ANSI standard) Basic Safety and Essential Performance
- 3. IEC 60601-1-2 Edition 1.0. 2010 -04 EMC. This replaces Edition 3.0 2007 03
- 4. IEC 60601-1-11 Edition 1.0 2014 06 Home Use. This replaces Edition 2010 04
- 5. IEC 62304 Edition 1.1 2015 06 Software and Life Cycle Processes
- 6. ISO 14971 Second edition 2007 03 01 Application of Risk Management to Medical Devices
- 7. ISO 10993-1 2009 (R) 2013 Biological Evaluation of Medical Devices Part 1 Evaluation and Testing within a Risk Management Process. ISO 10993-10:2010/(R) 2014 -Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
{5}------------------------------------------------
#### Substantial Equivalence
The Hair Boom 69 is equivalent to the device known as the HairMax Laser 80 cleared under 510(k) number, K180885. It is as safe and effective as the predicate device. Both systems, which use red light diode lasers which are class IIIa/3R laser systems by the IEC standard for allowable emission levels, which is a recognized standard by the FDA as well, and the adverse event profile is the same. The sponsor believes that the difference in the physical appearance or in the method of delivering the radiant energy of the two systems is of no consequence and does not effect the therapeutic value or the safety profile.
Finally, data summarized in the 510(k) notice confirms the safety and efficacy of the Hair Boom 69 for Prescription Use, according to Part 21 CFR 801 Subpart D. For these reasons, the Hair Boom 69 satisfies the FDA's standard for substantial equivalence with respect to intended use, technological and design characteristics.
#### Treatment Protocol
The Hair Boom 69 and the HairMax Laser 80 devices possess the same treatment regime of 30 minutes, every other day, on non-consecutive days, or three times per week for 16 weeks.
| Hair Boom 69 | HairMax Laser 80 |
|----------------------------------------------------|-------------------------------------------------|
| LLLT Device Type | LLLT Device Type |
| Prescription | OTC |
| Intended Use - Androgenetic Alopecia | Intended Use - Androgenetic Alopecia |
| Contain Laser Diodes- 69 Class 3R | Contain Laser Diodes- 80 Class 3R |
| Helmet Design | Cap Design |
| Wavelength 650 +/- 5 nms. | Wavelength 650 +/- 5 nms. |
| Marketing Clearance -Females & Males, Prescription | Marketing Clearance -Females & Males OTC |
| Passive Use-Hands Free | Passive Use-Hands Free |
| OAP Classification | OAP Classification |
| Classification Name -Infrared Lamp | Classification Name -Infrared Lamp |
| Common Usage Name - Lamp, Non-Heating | Common Usage Name - Lamp, Non-Heating |
| General & Plastic Surgery Committee | General & Plastic Surgery Committee |
| Skin Phototypes - I- IV | Skin Phototypes - I- IV |
| Hamilton-Norwood IIa-V Hair Loss Classification | Hamilton-Norwood IIa-V Hair Loss Classification |
| Ludwig-Savin I - II Hair Loss Classification | Ludwig-Savin I – II Hair Loss Classification |
| Device Class II | Device Class II |
The following Comparison Chart in support of substantial equivalence is provided:
{6}------------------------------------------------
## Conclusion
With the data presented in the Comparison Chart, the sponsor believes that this demonstrates the Hair Boom 69, is substantially equivalent to the HairMax Laser 80 and based upon the equivalent technological designs of the compared devices, the sponsor requests the FDA to clear the device via the 510(k) notice.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.