K182501 · Pulsaderm, LLC · OAP · Oct 26, 2018 · Physical Medicine
Device Facts
Record ID
K182501
Device Name
Hair Laser Headband
Applicant
Pulsaderm, LLC
Product Code
OAP · Physical Medicine
Decision Date
Oct 26, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hair Laser Headband is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood Hamilton Classifications of Ila to V patterns of hair loss and treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal, both with Fitzpatrick Skin Types I to IV.
Device Story
Low-level laser device; portable headband form factor; hands-free operation; rechargeable battery. Emits 655 +/- 5 nm laser light via 40 laser diodes (5mW each). User-operated at home; device moved across three scalp sections to ensure coverage. Intended to promote hair growth in patients with Androgenetic Alopecia. Bench testing confirms output parameters and dose rates equivalent to predicate helmet device.
Clinical Evidence
Bench testing only. Device performance verified against IEC 60601-1, IEC 60601-1-2, and IEC 60825-1 standards. Risk management performed per ISO 14971:2012. Testing confirmed output parameters and dose rates are equivalent to the predicate device.
Technological Characteristics
Low-level laser therapy (LLLT) device; 40 laser diodes; 655 +/- 5 nm wavelength; 5mW output per diode; Class 3R laser. Rechargeable battery power. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, ISO 14971. Form factor: headband.
Indications for Use
Indicated for males with Norwood Hamilton Classifications IIa-V and females with Ludwig (Savin) Scale I-4, II-1, II-2, or frontal patterns of Androgenetic Alopecia; Fitzpatrick Skin Types I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Pulsaderm LLC Gloria Avendano Regulatory Affairs Manager 12801 Commonwealth Dr. Units 2-6 Fort Myers, Florida 33913
## Re: K182501
Trade/Device Name: Hair Laser Headband Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: September 12, 2018 Received: September 12, 2018
October 26, 2018
Dear Gloria Avendano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Neil R Ógden -S 2018.10.26 10:23:41 -04'00' For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Hair Laser Headband
#### Indications for Use (Describe)
The Laser Headband is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood Hamilton Classifications of 11a to V patterns of hair loss and treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale 1-4, 11-1, or frontal, both with Fitzpatrick Skin Types I to IV.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# FDA 510(k) Summary of Safety and Effectiveness for The Hair Laser Headband
# General Information
Submitter: Pulsaderm LLC 12801 Commonwealth Dr. Units 2-7 Fort Myers, Florida 33913
# Contact Person:
Gloria Avendano Regulatory Affairs Manager Pulsaderm LLC 12801 Commonwealth Dr. Units 2-7 Fort Myers, Florida 33913 Gloria@pulsaderm.com
### Summary Preparation Date:
Device
| Device Name: | Hair Laser Headband |
|----------------------|-----------------------------------------------------------------------------------------|
| Classification Name: | Laser Surgical Instrument, for use in General and Plastic<br>Surgery and in Dermatology |
| Regulation Numbers: | 21 CFR 890.5550 |
| Regulatory Class: | II |
| Product Codes: | OAP |
### 3. Predicate Devices:
The Laser Headband is substantially equivalent to the following predicate devices;
| Device | 510(k) Number | Manufacturer |
|-----------------------------|---------------|---------------|
| NutraStim Hair Laser Helmet | K160728 | Pulsaderm LLC |
{5}------------------------------------------------
# Device Description
Similar to the Hair Laser Helmet, the Hair Laser Headband is a low-level laser device and is intended/indicated for over- the-counter use. The device emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp at 655 +/- 5 nanometers with an individual laser output of 5mW per laser diode. The lasers are configured inside a headband, designed for hands-free operation during treatment. The laser headband is for portable use with rechargeable battery.
# Intended Use and Indications:
The Hair Laser Headband is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood Hamilton Classifications of Ila to V patterns of hair loss and treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal, both with Fitzpatrick Skin Types I to IV.
## Technological Characteristics
The Hair Laser Headband shares the exact same fundamental scientific technology characteristics to the Hair Laser Helmet. The only modifications are the following:
- Size and dimensions change
- . Reduce number of diodes
The main technological characteristics in the Hair Laser Headband remain the same, such as wavelength and output of laser diodes, treatment regiment regimen, materials, delivery method, indications for use, helmet-like design, single on/off push button, and both devices use a rechargeable battery.
The modifications mentioned above have no effect in the safety or efficacy of the Hair Laser Headband as users are instructed to move the device in three different sections of the head to achieve the same surface treatment as the Hair Laser Helmet.
### Performance Data:
Similar to the Hair Laser Helmet, the Hair Laser Headband complies with the following test standards:
IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
IEC 60825-1 Safety of laser products - Part 1: Equipment classification and requirements.
{6}------------------------------------------------
Based on the nature of modifications implemented, the Hair Laser Headband underwent bench testing and results indicate that the Hair Laser Headband shares the same output parameters and dose rate characteristics as the Hair Laser Helmet.
Design verification and design validation testing were performed according to FDA's Design control activities per 21 CFR 820.30. Additionally, Risk Management activities were performed according to ISO 14971:2012 to evaluate new risks from design changes and all residual risks were found to be acceptable.
In all instances, the Hair Laser Headband performed as intended.
# Substantial Equivalence
The Hair Laser Headband shares the same technological characteristics as the Hair Laser Helmet K160728:
| Device | Hair Laser Headband | Hair Laser Helmet |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510 (k) Number | pending | K160728 |
| Product Code | OAP | OAP |
| Regulation<br>Number | 21 CFR 890.5500 | 21 CFR 890.5500 |
| Legal<br>Manufacturer | Pulsaderm LLC | Pulsaderm LLC |
| Device Type | Portable headband sits on patient's head | Portable helmet sits on patient's head, |
| Indications for<br>Use | Indicated to treat Androgenetic Alopecia and<br>promote hair growth in Males who have<br>Norwood Hamilton Classifications of IIa to V<br>patterns of hair loss and treat Androgenetic<br>Alopecia and promote hair growth in females<br>who have Ludwig (Savin) Scale I-4, II-1, II-2, or<br>frontal, both with Fitzpatrick Skin Types I to IV. | Indicated to treat Androgenetic Alopecia<br>and promote hair growth in Males who have<br>Norwood Hamilton Classifications of IIa to<br>V patterns of hair loss and treat<br>Androgenetic Alopecia and promote hair<br>growth in females who have Ludwig<br>(Savin) Scale I-4, II-1, II-2, or frontal, both<br>with Fitzpatrick Skin Types I to IV. |
| Type of Use | Over the Counter | Over the Counter |
| Number of Laser<br>Diodes | 40 | 82 |
| Wavelength | 655 +/- 5 nanometers | 655 +/- 5 nanometers |
| Laser<br>Classification | Class 3R | Class 3R |
| Treatment<br>Protocol | 20 Minutes- 3 times a week | 20 Minutes- 3 times a week |
| Standards<br>Applied | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60825-1<br>ISO 14971 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60825-1<br>ISO 14971 |
| Device Class | Class II | Class II |
| Sterilization | Basic cleaning, instructions provided. | Basic cleaning, instructions provided. |
| Common Name | Lamp, Non-Heating | Lamp, Non-Heating |
{7}------------------------------------------------
The minor modifications in dimensions and number of diodes have no effect in the safety or efficacy of device, as confirmed with performance testing and risk analysis activities.
# Conclusion:
Based upon the indications for use and data provided in this pre-market notification, all functional modes of the Hair Laser Headband have been shown to be substantially equivalent to the Hair Laser Helmet with respect to intended use, technological characteristics and safety characteristics. Pulsaderm LLC believes that no significant differences exist between the device and the predicate. Therefore, Substantial equivalency is requested.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.