Bosley Revitalizer 272 Laser Cap, Bosley Revitalizer 164 Laser Cup
K181253 · Pjm Worldwide, LLC D/B/A Phoenix Medical Technology, LLC · OAP · Aug 10, 2018 · Physical Medicine
Device Facts
Record ID
K181253
Device Name
Bosley Revitalizer 272 Laser Cap, Bosley Revitalizer 164 Laser Cup
Applicant
Pjm Worldwide, LLC D/B/A Phoenix Medical Technology, LLC
Product Code
OAP · Physical Medicine
Decision Date
Aug 10, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BOSLEY Revitalizer 272 and 164 Laser Caps are indicated to treat Androgenic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-3, or frontal; both with Fitzpatrick Skin Types I to IV.
Device Story
Low-level laser therapy (LLLT) device; promotes hair growth via photobiostimulation. Input: 272 or 164 visible red laser diodes (650-nm, 5mW each) housed in lightweight cap. Operation: battery-powered; 30-minute automatic shut-off; detects presence on user's head. Used in home setting by patient (OTC). Output: red light exposure to scalp. Clinical decision-making: patient self-selects based on hair loss classification and skin type; device provides hands-free treatment. Benefit: stimulation of hair follicles for androgenic alopecia.
Clinical Evidence
No clinical efficacy data. Provided OTC usability study (n=60; 30 male, 30 female) to assess self-selection and labeling comprehension. Results: 100% of females and 90% of males correctly identified candidacy based on hair loss and skin type; 95% overall accuracy. Bench testing confirmed compliance with IEC 60825-1 (Class 3R laser), IEC 60601-1, and biocompatibility (ISO 10993).
Technological Characteristics
LLLT device; 272 or 164 laser diodes (650-nm, 5mW). Powered by Lithium-ion battery (IEC 62133) with AC adapter (IEC 60950). Form factor: cap. Biocompatibility per ISO 10993-5/10. Electrical safety per IEC 60601-1, 60601-1-2, 60601-1-11. Laser classification: 3R (IEC 60825-1).
Indications for Use
Indicated for males with Norwood-Hamilton IIa-V and females with Ludwig (Savin) I-1 to I-4, II-1, II-3, or frontal androgenic alopecia; Fitzpatrick Skin Types I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 10, 2018
PJM Worldwide, LLC d/b/a Phoenix Medical Technology, LLC % Ronald Berglund Member/Manager Grace Consulting, LLC 6615 Lake Shore Drive, Suite 806 Minneapolis, Minnesota 55423
Re: K181253
Trade/Device Name: Bosley Revitalizer 272 Laser Cap, Bosley Revitalizer 164 Laser Cup Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: May 8, 2018 Received: May 30, 2018
Dear Ronald Berglund:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/ofpmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181253
Device Name
BOSLEY Revitalizer 272 Laser Cap; BOSLEY Revitalizer 164 Laser Cap
Indications for Use (Describe)
The BOSLEY Revitalizer 272 and 164 Laser Caps are indicated to treat Androgenic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-3, or frontal; both with Fitzpatrick Skin Types I to IV.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## Premarket Submission Applicant Name/510k Holder: Phoenix Medical Technology, LLC
Address: 1499 Northwest 79th Avenue, Miami, Florida 33126 (305)477-2515 Phone number: Email Address: Ira@piminc.com
### Official Correspondent: Ronald Berglund, GRACE Consulting, LLC Regulatory Consultant to Phoenix Medical Technology, LLC 6615 Lake Shore Drive, Suite 806 Minneapolis, MN 55423 (952)220-3014 Email address: Ronald @ronaldberglundlaw.com
Date of Original Submission: May 22, 2018
# Device Name(s) and Classification
Product Name(s) and Models: BOSLEY Revitalizer 272 Laser Cap BOSLEY Revitalizer 164 Laser Cap
Product Code: OAP Regulation Number: 21 CFR 890.5500 Panel: General and Plastic Surgery Class: II (performance standards)
### Predicate Devices:
Illuniflow Laser Cap (K162071)-Cleared for OTC use; Grivamax Hair Growth System (K171895)-Cleared for OTC use; Theradome LH80 PRO (K122950)-Cleared for OTC use; Capillus 272 Pro (K143199) - original clearance for use in females; K160285 adds treatments for males to the indications for use of all Capillus models)
### Reference Devices:
iGrow-II Hair Growth System (K122248; K140931, additional indication for use in females; K141567, original clearance for use in males)
{5}------------------------------------------------
HairMax LaserComb 82 (now sold as the HairMax LaserBand) (K142573, hands-free version of the HairMax LaserComb for use in males and females)
# Substantial Equivalence:
The BOSLEY Revitalizer 272 Laser Cap and BOSLEY Revitalizer164 Laser Cap are substantially equivalent to the referenced predicate devices based on indications for use, manufacturing materials, physical specifications, and performance specifications. The BOSLEY Revitalizer 272 and 164 Laser Caps raise no safety or efficacy concerns when compared to the predicate devices.
# Description of the Device:
The BOSLEY Revitalizer 272 and 164 Laser Caps are low level laser therapy (LLLT) devices designed to promote hair growth in women and men by exposing the entire scalp to the photobiostimulation of 272 or 164 visible red light-emitting diodes at 650-nm and 5mW each. The lasers are contained inside a lightweight cap. The device is powered by an included battery pack and automatically turns off after 30 minutes.
# Intended Use/Indications for Use:
The BOSLEY Revitalizer 272 and 164 Laser Caps are indicated to treat Androgenic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-3, or frontal; both with Fitzpatrick Skin Types I to IV.
# Technological Characteristics:
The BOSLEY Revitalizer 272 and 164 Laser Caps and the predicate devices have the following similar and substantially equivalent technological characteristics:
- · Design/Operating Principle: Both the BOSLEY Revitalizer 272 and 164 Laser Caps and the predicate devices are LLLT/photobiostimulation devices (with either 272, 202, 82 or 80 laser diodes) in cap form for the hands-free delivery of 650-nm visible red light to the scalp in order to treat androgenic alopecia and stimulate hair growth.
- Energy Source: The BOSLEY Revitalizer 272 and 164 Laser Caps and the . predicate devices are powered by Lithium-ion battery packs. The batteries can be recharged with the provided standard AC adapters.
{6}------------------------------------------------
- Performance: The BOSLEY Revitalizer 272 and 164 Laser Caps and the . predicate devices provide timed 30-minute LLLT treatments.
The predicate Illumiflow Laser Cap, Grivamax Hair Growth System, Capillus 272 Pro and Theradome LH80 Pro are cleared as OTC devices.
Other minor differences between the subject and predicate devices are summarized below:
- The BOSLEY Revitalizer 272 and 164 Laser Caps do not contain a . 'safety interlock' which automatically pauses therapy if the subject's head is in a less-than-optimal position.
- The BOSLEY Revitalizer 272 and 164 Laser Caps can detect whether or not the o device is on the user's head.
No new questions of safety and effectiveness have been raised as a result of these differences. From the evidence presented in the Premarket Notification, the subject device can be considered substantially equivalent.
### Biocompatibility:
Biocompatibility tests of the patient-contacting materials were tested by Mid-Link Technology Testing Co., Ltd., Tianjin, China in accordance with FDA Good Laboratory Practice Regulations (21 CFR Part 58) and International Organization for Standardization 10993-10, Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization and 10993-5:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, ISO 14971:2007 - Medical devices-Application of risk management to medical devices, FDA General Program Memorandum #G95-1, and FDA's Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.(Part 5: Tests for In Vitro Cytotoxicity). Biocompatibility tests were selected appropriate for the duration and type of contact to the device design of the BOSLEY Revitalizer 272 and 164 Laser Caps.
Cytotoxicity tests performed on the BOSLEY Revitalizer 272 and 164 Laser Caps demonstrated that the material used in the manufacturing of the caps is non-cytotoxic. The caps were considered by the testing facility to have met the requirements of ISO 10993-1:2009, ISO 14971:2007, and FDA General Program Memorandum #G95-1, for surface devices with limited contact (<24 hours) with the skin and was considered safe for use. Skin sensitization tests performed on the BOSLEY Revitalizer 272 and 164 Laser Caps further demonstrated that the material used in the manufacturing of the caps were not considered a sensitizer in the tests performed.
### Non-clinical Performance Testing:
{7}------------------------------------------------
Performance tests were conducted in support of the design verification of the BOSLEY Revitalizer 272 and 164 Laser Caps to confirm compliance to design specifications; all functions were verified to operate as designed, the BOSLEY Revitalizer 272 and 164 Laser Caps met all acceptance criteria in the performance testing.
{8}------------------------------------------------
Testing to IEC 60825-1 certifies the laser system to classification 3R, which is the same as the predicate devices. The BOSLEY Revitalizer 272 and 164 Laser Caps were also tested and complies with Electrical Safety and EMC testing, which include the requirements of IEC 60601-1:2012 3rd Edition "Medical Electrical Equipment Part 1- General Requirements for Safety" IEC 60601-1-2 "Medical Electrical Equipment Part 1-2, General Requirements for Safety -Collateral Standard Electromagnetic Compatibility Requirements and Tests IEC and 60601-1-11 Home healthcare medical equipment. The charger conforms to IEC 60950 and the battery to IEC 62133.
### Over-the-Counter/Usability Testing:
A single study was conducted in February 2018 in 30 female participants and 30 male participants to judge self-selection, comprehension of labeling and instructions, and usability of the Bosley Revitalizer Laser Cap. Each participant was provided with a Survey and Consent Form and a document entitled "Am I a Candidate for Usage of the Bosley Laser Cap?" These documents described the purpose of the study, the basic procedure steps, a consent form, permission to use their photos and non-disclosure terms.
Each participant was provided the proposed retail package for the Bosley Revitalizer Laser Cap, instructions for use, and the survey form containing 13 questions. Each participant was given as much time as he or she needed to read and understand the packaging information (an understanding of indications, contraindications, warnings and precautions), and assess whether they were within any contraindicated group. Assistance was available from the interviewer, but none of the study participants asked for assistance. The subjects were required to answer all questions. The interviewer verified that the participant completed the questionnaire in his or her presence. One male was excluded from the study as he wore a hair system (toupee) and thus his answers could not be verified. The total male population in the analysis is 29. A series of global photos using superior (male and female) and vertex (male only) were taken by the interviewer.
The survey forms and photos were securely stored electronically is a single folder for each participant. After completion of the data acquisition an independent reviewer verified completeness of the survey forms and the acceptable quality of the photos. Then, a licensed physician skilled in hair loss and skin clinical studies rated the participants pictures for hair loss and skin type.
An independent blinded statistician was provided a chart of hair loss rating and skin type rating, gender, and participant identifier. Consistent with the requirements cited in the comprehension and usability studies for the other laser cap devices granted over-the-counter clearance by the FDA, an 80% success rate was required.
{9}------------------------------------------------
All participants in the Bosley study accurately categorized their skin type as suitable candidates for treatment or not a suitable candidate when compared with physician ratings (i.e. 100%). Zero females and three males incorrectly identified themselves as treatment candidates for treatment based on hair loss. This means 100% of women and 90% of men correctly identified their candidacy for treatment or incorrectly judged themselves unsuitable for treatment based on hair loss self-rating. Overall, 95% percent of participants judged themselves correctly on hair loss.
Based on the data submitted and reviewed in the Statistical Analysis of the Comprehension and Usability Study of Revitalizer Laser Cap. Phoenix Technology LLC believes the Bosley Revitalizer Laser Cap for males and females should be granted the OTC intended use as requested.
### Conclusion:
Based on the non-clinical verification performance testing and as described above, it can be concluded that the BOSLEY Revitalizer 272 and 164 Laser Caps are substantially equivalent to the predicate Illumiflow, Theradome, Grivamax and Capillus Pro 272 Laser Caps (K162071, K122950, K171895 and K143199)
with respect to intended use, principles of operation and technological characteristics. No new questions of safety and effectiveness have been raised as a result of the minor differences between the devices. All nonclinical verification performance testing was done using the BOSLEY Revitalizer 272 Laser Cap. The only difference between the BOSLEY Revitalizer 272 and 164 Laser Caps is the number of laser diodes contained in the product. Phoenix Technology LLC believes the results of the non-clinical performance testing based on the use of 272 laser diodes would have been virtually identical had the 164 laser diode model been tested instead. In addition, the predicate Theradome LH80 PRO (K122950)—also cleared for OTC use-utilized only 80 laser diodes and was found to be substantially equivalent to the other predicates cited. Based on the data submitted and reviewed in the Statistical Analysis of the Comprehension and Usability Study of Revitalizer Laser Cap, Phoenix Technology LLC believes the Bosley Revitalizer Laser Cap for males and females should also be granted the OTC intended use as requested.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.