HAIRMAX LASERCOMB

K112524 · Lexington International, LLC · OAP · Sep 26, 2011 · Physical Medicine

Device Facts

Record IDK112524
Device NameHAIRMAX LASERCOMB
ApplicantLexington International, LLC
Product CodeOAP · Physical Medicine
Decision DateSep 26, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HairMax LaserComb Advanced 7 is indicated to treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal and Fitzpatrick Skin Types I to IV. The HairMax LaserComb Professional 12 is indicated to treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal and Fitzpatrick Skin Types I to IV.

Device Story

Hand-held low-level laser device; emits laser light to scalp to promote hair growth. Comb teeth part hair to ensure light reaches scalp. Advanced 7 model uses 7 laser modules; Professional 12 model uses 12 laser modules. Operated by user at home. Includes audible timer for treatment duration. Output is distributed laser light. Benefits patient by stimulating hair growth in androgenetic alopecia. Adjustments in treatment time compensate for module quantity differences.

Clinical Evidence

Bench testing only. Performance testing confirmed compliance with design specifications and verified all functions operate as designed. No clinical data presented.

Technological Characteristics

Hand-held low-level laser device. Emits laser light. Features hair-parting teeth mechanism. Includes audible timer. Advanced 7 contains 7 laser modules; Professional 12 contains 12 laser modules. Same laser modules and materials as predicate K110233.

Indications for Use

Indicated for females with androgenetic alopecia (Ludwig-Savin Scale I-4, II-1, II-2, or frontal) and Fitzpatrick Skin Types I-IV to promote hair growth.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 2 6 2011 # 6. 510(k) Summary The following information is provided as required by 21 CFR § 807.87 for Lexington International, LLC HairMax LaserComb Lux 9 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based. - Lexington International, LLC Sponsor: 777 Yamato Rd. Suite 105 Boca Raton, FL 33431 (561) 417-0200 research@hairmax.com Establishment Registration Number: 3006182775 - Olsson Frank Weeda Contact: C/O Casper E Uldriks Esq. 1400 Sixteenth Street, NW Washington DC 20036 August 30, 2011 Date: Proprietary Name: HairMax LaserComb Common or Usual Name: Lamp, nonheating, for promotion of hair growth. Product Code: OAP Classification Name: 21 CFR 890.5500 Infrared lamp Predicate Device: HairMax Lux 9 K110233 ## Device Description: Similar to the HairMax Lux 9 K110233, the modified HairMax Advanced 7 and Pro 12 consist of a hand-held low level laser device that emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. ### Intended Use / Indications for Use The HairMax LaserComb Advanced 7 is indicated to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-4, II-1, II-2 or frontal and Fitzpatrick Skin Types I to IV. {1}------------------------------------------------ The HairMax LaserComb Pro 12 is indicated to treat Androgenetic Alopecia and promote hair growth in females with who have Ludwig-Savin Scale 1-4, II-1, II-2 or frontal and Fitzpatrick Skin Types I to IV. #### Technological Characteristics The modifications to the HairMax Lux 9 since its previous clearance in K110233 do not alter the safety of the device. The predicate device contains 9 laser modules. The modified devices use the exact same laser modules and hair parting teeth mechanism. The difference in the predicate versus modified device is the quantity of the laser modules. The Advanced 7 contains 7 laser modules and the Pro 12 contains 12 laser modules. Adjustments in treatment time compensate for the different number of laser modules. ### Nonclinical Testing Based on the Risk Analysis, the verification and validation tests that were performed and the acceptance criteria applied for each are listed in Section 10. The HairMax Advanced 7 and Pro 12 were subject to the same preclinical requirements as the predicate device. Performance testing was conducted to confirm compliance to design specifications; all functions were verified to operate as designed. #### Substantial Equivalence The Advanced 7 and Pro 12 are as safe and effective as the predicate device, HairMax Lux 9. The subject devices have the same intended use of affecting hair growth as the predicate device. The subject devices have indications, i.e., treating androgenetic alopecia, and the same specific indication of promoting hair growth in females, who have Ludwig-Savin Scale I-4, II-1, II-2 or frontal and Fitzpatrick Skin Types I to IV, as the predicate device. The Advanced 7 and Pro 12 are identical in technological characteristics as the device cleared in K110233, including its laser power, wavelength, laser delivery method, its comb component, its instructions for use and its audible timer. The changes in the number of laser modules does not change the safety or effectiveness. {2}------------------------------------------------ ### Conclusion: The Advanced 7 and Pro 12 have the following similarities to the HairMax Lux 9: - has the same indicated use, ● - same identical laser modules . - . same hair parting teeth - uses the same operating principle . - incorporates the same basic device design and physical properties . - incorporates the same materials . Therefore the modification to the Lux 9 can be found substantially equivalent to the HairMax LaserComb cleared in K110233. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized emblem featuring three abstract shapes, possibly representing human figures or elements of health and well-being. The emblem is simple and modern in design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-Ci609 Silver Spring, MD 20993-0002 Lexington International, LLC % Olsson Frank Weeda Casper E Uldriks, Esq. 1400 16th Street Northwest. Suite 400 Washington, District of Columbia 20036 SEP 26. 2011 Re: K112524 Trade/Device Name: HairMax Advanced 7, HairMax Professional 12 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Il Product Code: OAP Dated: September 16, 2011 Received: September 19, 2011 Dear Mr. Uldriks: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 - Casper E Uldriks, Esq. CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. http://www.fda.gov/medicaldevices/ResourcesforYou/Industry/default.htm. Sincerely yours, for Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 5. Indications for Use 510(k) Number (if known): K (12524 Device Name: HairMax Advanced 7, HairMax Professional 12 Indications for Use: The HairMax LaserComb Advanced 7 is indicated to treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal and Fitzpatrick Skin Types I to IV. The HairMax LaserComb Professional 12 is indicated to treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal and Fitzpatrick Skin Types I to IV. Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use __X_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil RP Ogle for nuxen (Division Sign-C Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K112524
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...