K111714 · Lexington International, LLC · OAP · Aug 16, 2011 · Physical Medicine
Device Facts
Record ID
K111714
Device Name
HAIRMAX DUAL 12
Applicant
Lexington International, LLC
Product Code
OAP · Physical Medicine
Decision Date
Aug 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HairMax Dual 12 is indicated to Treat Androgenetic Alopecia and Promote Hair Growth in Females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal and Fitzpatrick Skin Types I to IV.
Device Story
Hand-held low-level laser device for home use by patients to promote hair growth. Device features comb teeth to part hair, ensuring laser light reaches scalp. Emits laser light via 12 modules (six 655nm, six 635nm). Includes audible/vibrating timer (every 4 seconds) to guide user to move device to new scalp sections. Output power identical to predicate. Clinical benefit demonstrated by increased terminal hair count after 16 and 26 weeks of treatment.
Clinical Evidence
Randomized, double-blind, multi-center clinical trial (n=3 sites). Population: females 25-60, androgenetic alopecia, Ludwig I-4, II-1, II-2, or frontal, Fitzpatrick I-IV. Results: 81% of subjects showed increased hair count at 16 weeks; 95% showed significant increases at 26 weeks (defined as minimum 5 new hairs). No serious adverse events reported.
Technological Characteristics
Hand-held low-level laser device. Emits laser light via 12 modules (six 655nm, six 635nm). Features comb teeth for scalp parting. Includes audible/vibrating timer. Safety testing per IEC 60601-1 and 60601-1-2.
Indications for Use
Indicated for females aged 25-60 with androgenetic alopecia, Ludwig (Savin) scale I-4, II-1, II-2, or frontal, and Fitzpatrick skin types I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
HairMax Dual 12
Image /page/0/Picture/2 description: The image shows the logo for Lexington International, LLC. The word "Lexington" is in a larger, bold font above the words "International, LLC". There is a curved line above the word "Lexington" and a straight line below it, which extends to the right and forms a star-like shape. The logo is simple and professional.
AUG 1 6 2011
## Section 5: 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for Lexington International, LLC HairMax Dual 12 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
k117
Sponsor: Lexington International, LLC 777 Yamato Rd. Suite 105 Boca Raton, FL 33431 (561) 417-0200 research@hairmax.com
Contact: Olsson Frank Weeda C/O Casper E Uldriks Esq. 1400 Sixteenth Street, NW Washington DC 20036
Date of Submission: June16, 2011
Proprietary Name: HairMax Dual 12
Common Name: Lamp, non-heating, for promotion of hair growth
Regulatory Class: II
Product Codes: OAP
Predicate Device(s): Lexington International, LLC - HairMax Lux 9 and HairMax Pro 12 (K103368) and Female Lux 9 (K110233)
#### · Device Description:
Substantially equivalent to the HairMax Pro 12 (K103368) and HairMax Lux 9 for females (K110233), the HairMax Dual 12 is a hand-held low-level laser device that emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure maximum laser light reaches the user's scalp. HairMax Dual 12 utilizes twelve laser modules, six with wavelengths of 655nm and six with wavelengths of 635nm with the same exact output power as the HairMax Pro 12.
{1}------------------------------------------------
#### Intended Use:
The HairMax Dual 12 is indicated to Treat Androgenetic Alopecia and Promote Hair Growth in Females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal and Fitzpatrick Skin Types I to IV.
#### Technological Characteristics
The HairMax Dual 12 consists of a hand-held low-level laser device that promotes hair growth. The device provides distributed laser light to the scalp while the device's comb teeth simultaneously part the user's hair to ensure maximum laser light reaches the user's scalp. When in use, the device emits a beep and vibration every four seconds to notify the user to move the device to a new section of the scalp.
#### Performance Testing
Testing to IEC 60601-1 and 60601-1-2 confirm the device's adherence to LVD . electrical and EMC safety requirements.
#### Clinical Testing
A randomized, double-blind, controlled, multi-center clinical trial was conducted at 3 sites, Cleveland Clinic - Wilma Bergfeld M.D., University of Minnesota - Maria Hordinsky M.D. and University of Miami - Lawrence Schachner M.D. Each site complied with Institutional Review Board approval and oversight and in accordance with applicable references defined by the Food and Drug Cosmetics Act and Title 21, Code of Federal Regulations. The clinical trials were listed on www.clinicaltrials.gov. The purpose of the clinical trial was to confirm the performance of the HairMax Dual 12 to treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal and Fitzpatrick Skin Types I to IV. After 16 weeks of treatment, 81% of the subjects using the HairMax Dual 12 experienced increases in hair count, (based on a minimum of 5 new hairs being observed at follow up). Benefits continued to improve after 26 weeks of treatment, 95% of the subjects using the HairMax Dual 12 experienced significant increases in hair count (based on a minimum of 5 new hairs being observed at follow up). No
{2}------------------------------------------------
### HairMax Dual 12 ______
subjects experienced any serious adverse event from the treatments.
The study population included females between the ages of 25 to 60 years with a diagnosis of androgenetic alopecia who had been experiencing active hair loss within the last 12 months. They were also required to have a Ludwig (Savin) classification of I-4, II-1, II-2 or Frontal, and have Skin Type I, II, III, or IV on the Fitzpatrick Skin Type Scale. Skin types were limited to the Fitzpatrick Skin types I-IV to facilitate the hair counting process, as it is difficult to count hairs on darker skin tones.
#### Substantial Equivalence
The HairMax Dual 12 is as safe and effective as the predicate devices. The HairMax Dual 12 has the same intended use of promoting hair growth as the predicate devices. The subject device has the same general indications, i.e., treating androgenetic alopecia and promoting hair growth.
The HairMax Dual 12 is identical in technological characteristics as the HairMax Pro 12 as cleared in K103368 and HairMax Lux 9 for females (K110233), except for the dual wavelength component. The HairMax Pro 12 contains twelve 655nm laser modules, the HairMax Dual 12 contains six 655nm laser modules and six 635nm laser modules. All other technical aspects, including its power output, its comb component, its instructions for use and its audible or vibrating timer remain identical. The modification to the HairMax Dual 12 does not change the intended use of the product nor does it affect the products fundamental scientific technology. Therefore this change does not raise new questions of safety or effectiveness. This was also demonstrated in a randomized, doubleblind, control clinical study evaluating changes in terminal hair-count in the evaluation zone, as well as usability studies to validate instructions for use, confirm that device modifications do not affect the safe and effective use of the devices when compared to the predicates.
For those reasons, HairMax Dual 12 satisfies FDA's substantial equivalence with respect to both the intended use and technological characteristics.
. . . . . .
. . . .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle symbol on the right and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" arranged in a circular fashion on the left. The eagle symbol is composed of three curved lines that suggest the shape of an eagle in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002
Lexington International. LLC % Olsson Frank Weeda Casper E. Uldriks, Esq. 1400 Sixteenth Street, NW Washington. District of Columbia 20036
Re: K111714
Trade/Device Name: HairMax Dual 12 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP Dated: June 16, 2011 Received: June 20, 2011
AUG 1 6 2011
Dear Mr. Uldriks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I'DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Casper E. Uldriks, Esq.
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Section 4: Indications for Use Statement
510(k) Number: k 111 7 / 4 Device Name: HairMax Dual 12 Indications for Use:
The HairMax Dual 12 is indicated to Treat Androgenetic Alopecia and Promote Hair Growth in Females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal and Fitzpatrick Skin Types I to IV.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use __ X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Milfredh for mton
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111714
Page 1 of __ 1_
8
Lexington International, LLC June 16, 2011
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.