← Product Code [GZB](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZB) · K150924

# A.I.M. Lead Loop Suturing Device (1), A.I.M. Tension Assist/Suture Cutter Device, A.I.M. Lead Loop Suturing Device (1) plus Tension Assist/Suture Cutter Device, A.I.M. Lead Loop Suturing Device (2), A.I.M. Lead Loop Suturing Device plus Tension Assist/Suture Cutter Device (K150924)

_Anchor Innovation Medical (A.I.M.) · GZB · Jul 27, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94neurological-therapeutic-devices/GZB/K150924

## Device Facts

- **Applicant:** Anchor Innovation Medical (A.I.M.)
- **Product Code:** [GZB](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZB.md)
- **Decision Date:** Jul 27, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5880
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The A.I.M. Lead Loop Suturing Device is intended for use in securing Spinal Cord Stimulation (SCS) leads and catheters to the fascia or intra-spinous/supra-spinous ligament.

## Device Story

Device is a singular anchor threaded with an adjustable loop of non-absorbable size 0 suture; used to secure SCS leads and catheters to fascia or ligaments. Anchor composed of carbon fiber reinforced PEEK; pre-loaded in a needle-tipped, disposable delivery device for tissue insertion. Ancillary instrument assists with tensioning and trimming suture. Used by physicians during SCS implantation procedures. Provides mechanical fixation of leads to prevent migration; benefits patient by maintaining lead position for effective stimulation therapy.

## Clinical Evidence

Bench testing only. Comparative performance testing against predicate demonstrated equivalent or superior performance in cyclic loading and ultimate construct failure. Biocompatibility confirmed per ISO 10993.

## Technological Characteristics

Anchor material: Carbon fiber reinforced polyetheretherketone (PEEK OPTIMA). Suture material: Ultra-High Molecular Weight Polyethylene or Polyester braid (size 0). Delivery: Needle-tipped, disposable delivery device. Sterilization: Provided sterile. Mechanical fixation principle.

## Regulatory Identification

An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.

## Predicate Devices

- Boston Scientific Fixate Suturing Device ([K113805](/device/K113805.md))

## Reference Devices

- A.I.M. Knotless Meniscal Repair Device ([K133770](/device/K133770.md))
- A.I.M. Suture Anchor ([K132461](/device/K132461.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 27, 2015

Anchor Innovation Medical c/o Karen M. Becker, Ph.D. 5410 Edson Lane, Suite 308 Rockville, MD 20852

Re: K150924

Trade/Device Name: A.I.M. Lead Loop Suturing Device Regulation Number: 21 CFR 882.5880 Regulation Name: Stimulator, Spinal Cord, Implanted (Pain Relief) Regulatory Class: Class II Product Code: GZB and GAT Dated: April 30, 2015 Received: May 1, 2015

Dear Dr. Becker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -5-D/Δ

Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150924

### Device Name

A.I.M. Lead Loop Suturing Device

Indications for Use (Describe)

The A.I.M. Lead Loop Suturing Device is intended for use in securing Spinal Cord Stimulation (SCS) leads and catheters to the fascia or intra-spinous/supra-spinous ligament.

| Type of Use (Select one or both, as applicable)                                                |                                               |
|------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

| X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 2.0 510(K) SUMMARY: A.I.M. LEAD LOOP SUTURING DEVICE

July 23, 2015 Submission Date:

## Submitter Information:

| Company:        | Anchor Innovation Medical (A.I.M.)<br>5410 Edson Lane<br>Suite 308<br>Rockville, MD 20852 |
|-----------------|-------------------------------------------------------------------------------------------|
| Contact Person: | Cory Stolneaker                                                                           |

### Contact Person: Cary Stalnecker Chief Executive Officer

## Correspondent Information:

Karen M. Becker, Ph.D. 5410 Edson Lane, Suite 308 Rockville, MD 20852

Phone: 301-770-7871 Fax: 301-770-7874 kbecker@anivmed.com

# Device Information:

| Trade Name:                         | A.I.M. Lead Loop Suturing Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Name:                    | Stimulator, Spinal Cord, Implanted (Pain Relief)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification:                     | 21 CFR 882.5880                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Codes:                      | GZB/GAT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Class:                       | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Devices:                  | Boston Scientific Fixate Suturing Device (K113805)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Intended Use:                       | The A.I.M. Lead Loop Suturing Device is intended for use in<br>securing Spinal Cord Stimulation (SCS) leads and catheters to the<br>fascia or intra-spinous/supra-spinous ligament.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Description:                 | The A.I.M. Lead Loop Suturing Device is a singular anchor<br>threaded with an adjustable loop of non-absorbable size 0 suture.<br>The suture material is commercially available Ultra-High<br>Molecular Weight Polyethylene or Polyester braid. The anchor is<br>composed of carbon fiber reinforced polyetheretherketone (PEEK<br>OPTIMA®) supplied by Invibio®. The device is provided sterile<br>and pre-loaded in a needle-tipped, disposable-delivery device to<br>facilitate insertion into tissue. A sterile, disposable ancillary<br>instrument is available separately, or included with the Lead Loop<br>Suturing Device, to assist with tensioning the construct and<br>trimming the suture once the appropriate tension is achieved. |
| Substantial Equivalence<br>Summary: | The A.I.M. Lead Loop Suturing Device is substantially equivalent<br>to the cited predicate, Boston Scientific's Fixate Suturing Device,<br>having the same, or similar, intended use, indications for use, and<br>fundamental scientific technology.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                     | The A.I.M Lead Loop Suturing Device and predicate differ in<br>suture anchor count, the means of tensioning and capturing the<br>SCS lead anchor or catheter, and retaining tension.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                     | Comparative performance testing with the predicate demonstrated<br>equivalent performance and did not raise any new questions of<br>safety or effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Performance Testing:                | Bench testing was conducted to confirm that the A.I.M. Lead Loop<br>Suturing Device performs as well or better than the predicate in<br>cyclic loading and ultimate device/construct failure.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Safety Testing:                     | The A.I.M. Lead Loop Suturing Device was determined to be<br>biocompatible based on conformity with ISO 10993. Component<br>materials and assembly processes are identical or similar to those<br>utilized within the A.I.M Knotless Meniscal Repair Device<br>(K133770) and Suture Anchor (K132461).                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Conclusion:                         | Based on the similarity in intended use, materials, fundamental<br>scientific technology and performance testing, the A.I.M. Lead<br>Loop Suturing Device is substantially equivalent to the cited<br>predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94neurological-therapeutic-devices/GZB/K150924](https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94neurological-therapeutic-devices/GZB/K150924)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
