DYNA-FLEX MULTI-LAYER COMPRESSION SYSTEM

K970721 · Johnson & Johnson Medical, Inc. · MCY · Mar 21, 1997 · General Hospital

Device Facts

Record IDK970721
Device NameDYNA-FLEX MULTI-LAYER COMPRESSION SYSTEM
ApplicantJohnson & Johnson Medical, Inc.
Product CodeMCY · General Hospital
Decision DateMar 21, 1997
DecisionSN
Submission TypeTraditional
Regulation21 CFR 880.5075
Device ClassClass 2
AttributesTherapeutic

Indications for Use

DYNA-FLEX* Multi-Layer Compression System is indicated for the management of venous leg ulcers and related conditions. The compression system may be used on patients with an ankle circumference of 18 cm or larger (padded). Compression data indicates that the DYNA-FLEX* Multi-Layer Compression System provides sustained compression for up to seven days at both the ankle and below the knee, when used according to package insert directions. The DYNA-FLEX Multi-Layer Compression System should not be used on diabetic patients with advanced small vessel disease, patients with lymphadema as a result of cancer, or on patients with an ankle:brachial index (ABI) of less than 0.8. Doppler ultrasound is recommended to rule out arterial disease.

Device Story

Multi-layer compression kit for venous leg ulcer management; applied by healthcare professionals. Components: ADAPTIC non-adhering dressing (petrolatum-impregnated), polyurethane foam/cotton padding layer, elastomeric/nylon compression bandage, and cohesive spandex/latex compression bandage. Layer two features visual indicator (rectangle-to-square knit pattern) to guide correct compression application. Provides sustained compression for up to seven days. Benefits include ulcer management and edema control.

Clinical Evidence

Bench testing only. Biocompatibility testing conducted per ISO 10993 Part-1 (non-sensitizing, non-cytotoxic, non-irritating). Performance testing confirmed sustained compression for up to seven days at ankle and below knee.

Technological Characteristics

Kit includes: ADAPTIC dressing (petrolatum emulsion), polyurethane foam/cotton wrap, elastomeric/nylon compression bandage, and nonwoven spandex/latex cohesive bandage. Mechanical compression principle. No energy source. Non-sterile components (except dressing).

Indications for Use

Indicated for management of venous leg ulcers and related conditions in patients with ankle circumference ≥18 cm. Contraindicated in diabetic patients with advanced small vessel disease, patients with cancer-related lymphedema, or patients with ankle:brachial index (ABI) <0.8.

Regulatory Classification

Identification

An elastic bandage is a device consisting of either a long flat strip or a tube of elasticized material that is used to support and compress a part of a patient's body.

Predicate Devices

Submission Summary (Full Text)

{1} K970721 510(k) - DYNA-FLEX* Multi-Layer Compression System MAR 21 1997 # 510(k) SUMMARY 1. DATE PREPARED February 27, 1997 2. SUBMITTER Johnson &amp; Johnson Medical, Inc. 2500 Arbrook Blvd. P.O. Box 90130 Arlington, TX 76004-3130 3. CONTACT PERSON Wm. P. Robb Senior Regulatory Affairs Project Manager Phone: 817-784-5210 Fax: 817-784-5462 4. NAME OF THE MEDICAL DEVICE Classification Name: Bandage, Compression Common/Usual Name: Compression Bandaging System Proprietary Name: DYNA-FLEX* Multi-Layer Compression System 5. DEVICE CLASSIFICATION Regulatory Class: Unclassified Product Code: MCY 6. STATEMENT OF SUBSTANTIAL EQUIVALENCE The DYNA-FLEX* Multi Layer Compression System is substantially equivalent and identical in function to the Wound Care Compression System (WCCS) marketed by Suncoast Medical Products. * Trademark 000062 {2} 510(k) - DYNA-FLEX* Multi-Layer Compression System ## 7. INDICATIONS FOR USE DYNA-FLEX* Multi-Layer Compression System is indicated for the management of venous leg ulcers and related conditions. The compression system may be used on patients with an ankle circumference of 18 cm or larger (padded). Compression data indicates that the DYNA-FLEX* Multi-Layer Compression System provides sustained compression for up to seven days at both the ankle and below the knee, when used according to package insert directions. The DYNA-FLEX Multi-Layer Compression System should not be used on diabetic patients with advanced small vessel disease, patients with lymphadema as a result of cancer, or on patients with an ankle:brachial index (ABI) of less than 0.8. Doppler ultrasound is recommended to rule out arterial disease. ## 8. PHYSICAL DESCRIPTION The DYNA-FLEX* Multi-Layer Compression System is a kit that is comprised of one wound dressing component and three bandage layers for the management of venous leg ulcers and related conditions and is packaged in a polyethylene resealable bag. The description of each of these components/bandage layers is as follows: - The wound dressing component is the ADAPTIC* Non-Adhering Dressing (Sterile). This non-adherent 5" x 9" pad is impregnated with a petrolatum emulsion and is packaged in a polyester polyethylene film that is heat sealed prior to sterilization. - The first bandage layer is the DYNA-FLEX* Layer One (Padding/Absorption Layer). This polyurethane stretch foam/cotton wrap (non-sterile) is packaged in a polyurethane shrink wrap. Layer One is supplied in a 4" x 156" roll. (unstretched). - The second bandage layer is the DYNA-FLEX* Layer Two (Compression Bandage). This non-sterile compression wrap is comprised of absorbent yarns knitted with elastomeric and nylon yarns and is packaged in a polyurethane shrink wrap. Layer Two is supplied in a 4" x 90" roll (unstretched). To assist the healthcare professional in wrapping, this bandage layer is knitted with a rectangle design, such that as the bandage is extended it will change to a square design thus indicating a satisfactory compression range. - The third bandage layer is the DYNA-FLEX* Layer Three (Cohesive Compression Bandage). This non-sterile cohesive compression bandage is comprised of a nonwoven fabric with spandex elastic yarns bonded with a natural rubber cohesive latex based system. Layer Three is supplied in a 4" x 216" roll (stretched). ## 9. BIOCOMPATIBILITY Safety tests conducted in accordance with ISO 10993 Part-1 "Biological Evaluation Of Medical Devices" demonstrate the DYNA-FLEX* Multi-Layer Compression System is non-sensitizing, non-cytotoxic, non-systemic toxic, non-irritating and is suitable for its intended use. In addition, testing was done to demonstrate that sustained compression for up to seven days at both the ankle and below the knee was maintained, when used according to package insert directions. * Trademark 000063
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...