← Product Code [GEH](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH) · K972662 # CRYOGEN CRYOSURGICAL SYSTEM (K972662) _Cryogen, Inc. · GEH · Oct 1, 1997 · General, Plastic Surgery · SESE_ **Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K972662 ## Device Facts - **Applicant:** Cryogen, Inc. - **Product Code:** [GEH](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH.md) - **Decision Date:** Oct 1, 1997 - **Decision:** SESE - **Submission Type:** Traditional - **Regulation:** 21 CFR 878.4350 - **Device Class:** Class 2 - **Review Panel:** General, Plastic Surgery - **Attributes:** Therapeutic ## Intended Use The CryoGen Cryosurgical System is intended for use in the surgical ablation of tissue by the application of extreme cold in the fields of dermatology, general surgery, neurosurgery, thoracic surgery, E.N.T., gynecology, oncology, proctology and urology. ## Device Story Cryosurgical system for tissue destruction via extreme cold; consists of disposable control unit, cryoprobe, and console with compressor, microprocessor, and user interface. Operates on Joule-Thomson principle: pressurized coolants expand through orifice to generate tip temperatures of -100 to -120 °C. Closed system prevents coolant contact with patient/physician and atmospheric exhaust. Used by clinicians in various surgical specialties; provides localized tissue ablation. Benefits include precise, controlled tissue destruction without shaft/handle cooling. ## Clinical Evidence No clinical data provided. Substantial equivalence is based on design, materials, and functional comparison to predicate devices. ## Technological Characteristics Closed-loop cryosurgical system; Joule-Thomson cooling principle; microprocessor-controlled; disposable control unit and cryoprobe; console-based compressor. Materials and design specifications are equivalent to predicate devices. ## Regulatory Identification (1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications. ## Predicate Devices - Frigitronics CE-4 & CE-4G (Pre-Amendment) - Frigitronics CCS 100 ([K811390](/device/K811390.md)) - CMS AccuProbe ([K904421](/device/K904421.md)) - CMS AccuProbe 550/530 ([K953637](/device/K953637.md)) - CryoGen Cryosurgical System 1 ([K964971](/device/K964971.md)) ## Related Devices - [K021211](/device/K021211.md) — MODIFICATION TO SANARUS VISICA TREATMENT SYSTEM · Sanarus Medical, Inc. · Jun 26, 2002 - [K231661](/device/K231661.md) — CryoThin™ Surgical System and Accessories · Medinux (Tianjin) Technologies Co., Ltd. · Nov 27, 2023 - [K980668](/device/K980668.md) — CMS BLIZZARD 700 SERIES MODELS 730, 740,750,760,770,780 & 790 · Cryomedical Sciences, Inc. · Jul 17, 1998 - [K240892](/device/K240892.md) — XSense Cryoablation System with Cryoprobes · IceCure Medical , Ltd. · Jun 28, 2024 - [K052861](/device/K052861.md) — MODIFICATION TO:SANARUS VISICA TREATMENT SYSTEM , MODELS VS2000 · Sanarus Medical, Inc. · Nov 15, 2005 ## Submission Summary (Full Text) {0}------------------------------------------------ 1007 K972662 CryoGen, Inc. San Diego, CA Premarket Notification July 2, 1997 ## X. 510(k) Summary ## Name of Device Trade name: Common name: Classification name: CrvoGen Crvosurgery System Cryosurgical Unit and Accessories Cryosurgical Unit and Accessories (21 CFR 878.4350) #### Predicate devices | Device | Premarket Notification | |-------------------------------|------------------------| | Frigitronics CE-4 & CE-4G | Pre-<br>Amendment | | Frigitronics CCS 100 | K811390 | | CMS AccuProbe | K904421 | | CMS AccuProbe 550/530 | K953637 | | CryoGen Cryosurgical System 1 | K964971 | #### Device description & Principle of Operation The CryoGen Cryosurgery System consists of three components: the disposable Control Unit, the Cryoprobe and the Console, which contains the compressor system, microprocessor and user interface. The CryoGen Cryosurgery System is a cryosurgical device incorporating a gas cooled cryoprobe. Operation of the System is based on the Joule-Thomson principle in which pressurized coolants are expanded through a small orifice to produce cooling. The device is intended to destroy tissue by the application of extreme cold. Temperatures of -100 to -120 ℃ are developed at the tip of the cryoprobe. These temperatures are within the range of the predicate devices and is sufficient to achieve the desired tissue effect. None of the coolant comes into contact with the patient or physician. In addition, none of the coolant gases are exhausted into the atmosphere, the system is closed. There is no coeling along the shaft of the probe nor at the handle that is held by the user during treatment. # 000019 {1}------------------------------------------------ CryoGen, Inc. San Diego, CA Premarket Notification July 2, 1997 #### Intended use The CryoGen Cryosurgical System is intended for use in the surgical ablation of tissue by the application of extreme cold in the fields of dermatology, general surgery, neurosurgery, thoracic surgery, E.N.T., gynecology, oncology, proctology and urology. This intended use is identical to the intended use for the predicate crvosurgical devices. ### Technological characteristics The technological characteristics of the CryoGen Cryosurgical System 2 are the same as those of the CryoGen Cryosurgical System 1, and the other predicates listed elsewhere in this premarket notification. These devices are substantially equivalent in terms of design, materials, principle of operation, product specifications and sterilization. ### Summarv By virtue of design, materials, function and intended use, the CryoGen Cryosurgical System 2 is substantially equivalent to the CryoGen Cryosurgical System 1 which is cleared under K964971. It is also equivalent to the predicate devices, both pre-amendment and cleared via the Premarket Notification process, which have been included in this submission. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features the department's seal, which includes an emblem of a stylized human figure with three faces in profile, representing health, hope, and service. The seal is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Ms. Cheryl L. Shea. . Vice President Regulatory Affairs/Quality Assurance CryoGen, Inc. 6199 Cornerstone Court East Suite 106 San Diego, California 92121 OCT - 1 1997 Re: K972662 > Trade Name: CryoGen Cryosurgery System Regulatory Class: II Product Code: GEH Dated: July 2, 1997 Received: July 3, 1997 Dear Ms. Shea: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. ´ Please be advised that this substantial equivalence determination does not include an indication for cryoablation of the endometrium. The use of cryosurgery for endometrial ablation raises new types of safety and effectiveness questions when compared to currently identified predicate devices used for this purpose and therefore will require approval of a premarket approval application (PMA) for this indication. Since, no data has been developed to establish the safety and effectiveness of this cryosurgical device for endometrial ablation, you may not market or promote such use until you have submitted such data and received clearance or approval for this claim. {3}------------------------------------------------ #### Page 2 - Ms. Shea If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that. through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. --------Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, folly Celia M. Witten, M.D., Ph.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ CryoGen, Inc. San Diego, California Premarket Notification July 2, 1997 # Indications Statement Device Name: CryoGen Cryosurgical System 510(k) Number: ## Indications for use: છે. The CryoGen Cryosurgical System is indicated for the ablation of tissue by the application of extreme cold in the areas of dermatology, general surgery, neurosurgery, thoracic surgery, E.N.T., gynecology, oncology, proctology and urology. | | | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | | |--|--|----------------------------------------------------------------------------------------------------------------------|--| | | | станные мести при седения с высокие при седения при при пристемент при пристемента при пристемента с мести с седания | | | | | | | #### Concurrence of CDRH, Office of Device Evaluation (ODE) t.co/e (Division Sign-Off) (Division of General Restorative Devices 510(k) Number X Prescription Use (Per 21 CFR 801.109) OR Over-the-Counter Use --- **Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K972662](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K972662) **Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact). **Cite as:** FDA Device Explorer (Innolitics), https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K972662
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...