Freezen Clear Skin Clinic Warts & Tags
K211099 · Cryoconcepts LP · GEH · Jun 17, 2022 · General, Plastic Surgery
Device Facts
| Record ID | K211099 |
| Device Name | Freezen Clear Skin Clinic Warts & Tags |
| Applicant | Cryoconcepts LP |
| Product Code | GEH · General, Plastic Surgery |
| Decision Date | Jun 17, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4350 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Freeze 'n Clear Skin Clinic for Warts and Skin Tags product is intended for the OTC treatment of common warts, plantar warts, and skin tags.
Device Story
Device utilizes cryogen gas (dimethyl ether, propane, isobutane) to remove warts and skin tags via freezing. User dispenses cryogen into foam applicator; applicator cools to approximately -55°C; user applies to target tissue for 40 seconds. Extreme cold destroys target tissue; tissue falls away over time; new epidermis regenerates. Intended for OTC consumer self-use. Treatment may require 1-4 applications at two-week intervals. Benefits include non-invasive removal of skin lesions without professional intervention.
Clinical Evidence
Clinical study conducted at 3 US dermatology sites with >300 subjects. Study included label comprehension for self-diagnosis/self-selection and an actual use component. Human factors usability testing evaluated label comprehension and simulated use. Results demonstrated the majority of subjects could correctly self-diagnose, select the product, and follow directions. Historical data from non-US markets showed low complaint/adverse event rates. Bench testing confirmed equivalence in cooling temperatures and cell destruction capability compared to Histofreezer Professional.
Technological Characteristics
Cryosurgical unit consisting of a canister containing a cryogen mixture (dimethyl ether, propane, isobutane) and foam applicators. Operates via rapid evaporation of cryogen to reach -55°C. Materials meet ISO 10993 biocompatibility requirements (cytotoxicity, sensitivity, irritation). Disposable unit; no electronics or software.
Indications for Use
Indicated for OTC treatment of common warts, plantar warts, and skin tags in consumers.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Predicate Devices
- OTC Wart Removal System (K023487)
- Histofreezer Professional (K933327)
- Claritag (K190747)
Related Devices
- K251493 — Skin Clinic Freeze Point for Warts and Skin Tags · Cryoconcepts LP · May 22, 2025
- K253000 — CryoFreeze Wart and Skin Tag Remover · Cryosurgery, Inc. · Oct 16, 2025
- K240226 — Pixie Skin Tag · Oystershell NV · Oct 18, 2024
- K243487 — CryoFreeze Wart and Skin Tag Remover · Cryosurgery, Inc. · Mar 13, 2025
- K242932 — Skin Clinic Nitro Clear Wart Remover · Cryoconcepts LP · Apr 15, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 17, 2022
CryoConcepts LP Sam Niedbala PhD CEO 205 Webster St Bethlehem, Pennsylvania 18015
Re: K211099
Trade/Device Name: Freeze'n Clear Skin Clinic Warts & Tags Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: April 7, 2021 Received: May 16, 2021
Dear Sam Niedbala PhD:
We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211099
#### Device Name
Freeze 'n Clear Skin Clinic for Warts and Skin Tags
#### Indications for Use (Describe)
The Freeze 'n Clear Skin Clinic for Warts and Skin Tags product is intended for the OTC treatment of common warts, plantar warts, and skin tags.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K211099 510(K) SUMMARY
## I. SUBMITTER
CryoConcepts LP, 205 Webster St. Bethlehem PA 18015 Phone: 855-355-2796 Contact: Sam Niedbala, Ph.D. June 16, 2022
## II. DEVICE
Name of Device: Freeze 'n Clear Skin Clinic for Warts and Skin Tags
Usual Name - Cryosurgical unit and accessories
Classification Name - General & Plastic Surgery
Regulatory Class: II
Code of Federal Regulation: 878.4350
Product Code: GEH
## III. PREDICATE DEVICES
OTC Wart Removal System K023487
Histofreezer Professional K933327
Claritag K190747
#### IV. DEVICE DESCRIPTION
The Freeze 'n Clear Skin Clinic for Warts and Skin Tags product utilizes extreme cold to facilitate the removal of warts and skin tags by freezing. Each kit contains a container of cryogen gas, foam applicators, tweezers and instructions for use. The device is for OTC use and utilizes a combination of dimethyl ether, propane and isobutane delivered from the canister into a
Page 1 of 5
{4}------------------------------------------------
foam applicator which acts as a reservoir for the cryogen gas. The gas rapidly evaporates and cools the applicator to approximately -55°C. The applicator is then placed against the wart or skin tag for 40 seconds which freezes the targeted tissue. The frozen skin tag or wart falls away over time and new epidermis grows in its place. One to four treatments with intervals of two weeks may be required.
# V. INDICATIONS FOR USE
The Freeze 'n Clear Skin Clinic for Warts and Skin Tags product is intended for the OTC treatment of common warts, plantar warts, and skin tags.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A summary of the technological characteristics for the Freeze 'n Clear Skin Clinic for Warts vs Freeze 'n Clear Skin Clinic for Warts and Skin Tags as well as the Histofreezer professional and Claritag predicates are provided in the following table. Each of these products utilize the same technological characteristics and cryogens as shown in the table below.
| Technological<br>Characteristics | Freeze 'n Clear<br>Skin Clinic for<br>Warts (K023487) | Histofreezer<br>Professional<br>(K933327) | Claritag<br>(K190747) | Freeze 'n Clear<br>Skin Clinic for<br>Warts and Skin<br>Tags (Subject<br>Device) |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| -Intended Use | The Freeze 'n Clear<br>Skin Clinic<br>Advanced Wart<br>product is intended<br>for consumer use<br>for the removal of<br>common warts and<br>plantar warts | For the<br>cryosurgical<br>treatment of lesions<br>including warts and<br>skin tags. Rx Only. | Claritag is<br>indicated for use in<br>the treatment of<br>acrochordons (skin<br>tags) | The Freeze 'n Clear<br>Skin Clinic for<br>Warts and Skin<br>Tags product is<br>intended for<br>consumer use for<br>the treatment of<br>common warts,<br>plantar warts and<br>skin tags |
| -Cryogen | Mixture of DMEP | Mixture of DMEP | Mixture of DMEP | Mixture of DMEP |
| Technological<br>Characteristics | Freeze 'n Clear<br>Skin Clinic for<br>Warts (K023487) | Histofreezer<br>Professional<br>(K933327) | Claritag<br>(K190747) | Freeze 'n Clear<br>Skin Clinic for<br>Warts and Skin<br>Tags (Subject<br>Device) |
| -Materials | -canister containing<br>cryogen<br>-Foam tipped<br>applicators | -canister containing<br>cryogen<br>-Foam tipped<br>applicators | -canister containing<br>cryogen and<br>disposable foam<br>pads, Tweezers | -canister containing<br>cryogen<br>-Foam tipped<br>applicators and<br>-Tweezers |
| -Mode of Use | Cryogen dispensed<br>into foam<br>applicator which is<br>then applied to the<br>lesion | Cryogen dispensed<br>into foam<br>applicator which is<br>then applied to the<br>lesion | Cryogen dispensed<br>into foam pads<br>which are then<br>applied to the<br>lesion | Cryogen dispensed<br>into foam<br>applicator which is<br>then applied to the<br>lesion |
| -Mechanism of<br>action | Extreme cold<br>destroys the target<br>tissue | Extreme cold<br>destroys the target<br>tissue | Extreme cold<br>destroys the target<br>tissue | Extreme cold<br>destroys the target<br>tissue |
| -Storage &<br>Safety<br>Conditions | -Keep away from<br>fire or flame<br>-Do not smoke<br>while using the<br>product<br>-Do not puncture or<br>incinerate canister | -Keep away from<br>fire or flame<br>-Do not smoke<br>while using the<br>product<br>-Do not puncture or<br>incinerate canister | -Highly<br>Flammable<br>-Store at room<br>temperature<br>between 68°-77°F<br>and away from<br>heat. | -Keep away from<br>fire or flame<br>-Do not smoke<br>while using the<br>product<br>-Do not puncture or<br>incinerate canister |
| | -Do not expose to<br>heat or store at<br>temperatures above<br>120°F.<br>-Store at room<br>temperature away<br>from heat | -Do not expose to<br>heat or store at<br>temperatures above<br>120°F.<br>-Store at room<br>temperature away<br>from heat | -Protect from<br>sunlight and do not<br>expose to<br>temperature above<br>120°F<br>-Contents under<br>pressure.<br>-Do not puncture or<br>incinerate | -Do not expose to<br>heat or store at<br>temperatures above<br>120°F.<br>-Store at room<br>temperature away<br>from heat |
| Technological<br>Characteristics | Freeze 'n Clear<br>Skin Clinic for<br>Warts (K023487) | Histofreezer<br>Professional<br>(K933327) | Claritag<br>(K190747) | Freeze 'n Clear<br>Skin Clinic for<br>Warts and Skin<br>Tags (Subject<br>Device) |
| | | | container, even if<br>empty | |
| -Treatment<br>Procedure | Spray the cryogen<br>into the applicator<br>to saturate it and<br>then place it<br>directly onto the<br>lesion for a<br>specified number of<br>seconds | Spray the cryogen<br>into the applicator<br>to saturate it and<br>then place it<br>directly onto the<br>lesion for a<br>specified number<br>of seconds | Spray the cryogen<br>into the foam pads<br>to saturate them<br>and then place<br>them directly onto<br>the lesion for a<br>specified number<br>of seconds | Spray the cryogen<br>into the applicator<br>to saturate it and<br>then place it<br>directly onto the<br>lesion for a<br>specified number<br>of seconds |
| -Disposal | Entire unit is<br>disposable after<br>emptied of<br>cryogen. | Entire unit is<br>disposable after<br>emptied of<br>cryogen. | Entire unit is<br>disposable after<br>emptied of<br>cryogen. | Entire unit is<br>disposable after<br>emptied of<br>cryogen. |
| -Defined<br>Operators | OTC for consumer<br>use | Rx Use Only | Rx Use Only | OTC for consumer<br>use |
| -Service /<br>Repair | None | None | None | None |
{5}------------------------------------------------
{6}------------------------------------------------
## VII. PERFORMANCE DATA
This submission added the indication to the Freeze 'n Clear Skin Clinic product to treat skin tags. As part of the data provided, the product was tested for biocompatibility including cytotoxicity, sensitivity, irritation according to FDA's biocompatibility 2020 guidance and meets the requirements of ISO 10993.
Bench testing compared the temperatures attained by the Histofreezer Professional predicate device and Freeze 'n Clear Skin Clinic product to demonstrate that they were equivalent. Additional bench testing using an in vitro model also demonstrated the Histofreezer Professional and Freeze 'n Clear Skin Clinic products were equivalent in their ability to freeze and destroy target cells.
A clinical study and human factors usability study was performed in support of the over-thecounter indication for skin tags. The clinical study was performed at 3 US, dermatology offices with over 300 subjects. It consisted of two parts, a label comprehension part for self-diagnosis and self-selection and an actual use part. Human factors usability testing also evaluated label
{7}------------------------------------------------
comprehension and simulated use. The studies demonstrated the majority of subjects were able to self-diagnosis, self-select for the product, understand the directions, and appropriately use the product. Additionally, historical data was presented showing the low number of complaints and adverse events seen outside the US with the Freeze 'n Clear product used either for warts or skin tags
#### VIII. CONCLUSIONS
The Freeze 'n Clear Skin Clinic K023487 predicate device is indicated for OTC treatment of common and plantar warts. This submission adds the indication for the treatment of skin tags to the Freeze 'n Clear Skin Clinic K211099 device. The combination of studies and performance data presented demonstrates the subject device is as safe and effective as the predicate device(s) as indicated for use.