K150042 · Clearflow, Inc. · OTK · Jun 2, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150042
Device Name
PleuraFlow System
Applicant
Clearflow, Inc.
Product Code
OTK · General, Plastic Surgery
Decision Date
Jun 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The PleuraFlow System is indicated for use during cardiothoracic surgical procedures and chest trauma. Its active clearance technology proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.
Device Story
PleuraFlow System is a wound drain catheter system designed to maintain chest tube patency. It consists of a silicone chest tube (20FR-32FR) and a clearance apparatus. The apparatus includes a guide tube and a PTFE-coated clearance wire with a 105-degree angled distal loop. A magnetic shuttle advances the clearance wire/loop within the chest tube to mechanically break up or prevent clot formation and tube occlusion. Used in clinical settings by healthcare providers to ensure continuous drainage of blood and fluid post-surgery. By preventing chest tube clogging, the device reduces retained blood, which is associated with complications like post-operative atrial fibrillation (POAF).
Clinical Evidence
No new clinical trials were required for this submission. Evidence relies on previously cleared design validation and verification (K093565) and four years of post-market performance data. Literature cited indicates that active clearance of chest drainage catheters reduces retained blood and post-operative atrial fibrillation (POAF) interventions.
Technological Characteristics
System comprises a silicone chest tube (20FR, 24FR, 28FR, 32FR) with barium stripe and a clearance apparatus (PTFE-coated wire with distal loop). Operates via manual mechanical advancement of the clearance wire using a magnetic shuttle. No electronic energy source or software algorithms involved.
Indications for Use
Indicated for adult and pediatric patients (infant, preadolescent, adolescent) undergoing cardiothoracic surgery or experiencing chest trauma requiring evacuation of blood and fluid from the operative site.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
K203394 — PleuraFlow System with FlowGlide · Clearflow, Inc. · Mar 2, 2021
K153681 — PleuraFlow System · Clearflow, Inc. · Jan 20, 2016
K093565 — PLEURAFLOW CATHETER SYSTEM · Clear Catheter Systems · Dec 3, 2010
K182067 — PleuraFlow System with FlowGlide Extra Drainage Length · Clearflow, Inc. · Aug 14, 2018
K163139 — PleuraFlow System with FlowGlide · Clearflow, Inc. · Nov 30, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with a stylized eagle or bird-like element above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2015
ClearFlow Incorporated Dr. Dov Gal, DVM Vice President Regulatory Affairs, Quality Assurance and Clinical Affairs 1630 South Sunkist Street, Suite E Anaheim, California 92806
Re: K150042
Trade/Device Name: PleuraFlow® System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OTK, GBX Dated: April 14, 2015 Received: April 20, 2015
Dear Dr. Gal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) 150042
Device Name PleuraFlow® System
Indications for Use (Describe)
The PleuraFlow System is indicated for use during cardical procedures and chest trauma. Its active clearance technology proactively removes clots formed inside to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and patients including infant, preadolescent and adolescent patients under clinical settings.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for PleuraFlow® System 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
| Date of Submission: | April 14, 2015 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Clear Flow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA<br>92806 |
| Primary Contact Person: | Dov Gal, DVM<br>ClearFlow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA<br>92806<br>Phone: 714-916-5014<br>Fax: 714-916-5019<br>Email: dgal@clearflow.com |
| CEO/President: | Paul Molloy<br>ClearFlow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA<br>92806<br>Phone: 714-905-5271<br>Fax: 714-916-5019<br>Email: pmolloy@clearflow.com |
| Device Proprietary Name: | PleuraFlow® System |
| Device Common Name: | Introduction/drainage; wound drain catheter system. |
{4}------------------------------------------------
### Regulatory Class and Name: Class II, Powered Suction Pump
#### Product Codes: OTK and GBX
- Indication For Use: The PleuraFlow System is indicated for use during cardiothoracic surgical procedures and chest trauma. Its active clearance technology proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.
- Predicate Device: Predicate device is the PleuraFlow Catheter System (K093565) by ClearFlow, Inc., AKA Clear Systems, Inc.
- The PleuraFlow System is comprised of a silicone Chest Tube Device Description: and a Clearance Apparatus. The PleuraFlow Chest Tube is available in four (4) standard sizes (20FR, 24FR, 28FR and 32FR). Each Chest Tube has a Cut Length of 19 inches (48.3 cm) with graduated measurements in centimeters from the distal eyelet. Each Chest Tube has a barium stripe to facilitate visualization. The Chest Tube is connected to a Clearance Apparatus, which is connected to the tubing from the drainage canister. The Clearance Apparatus consists of a Guide Tube and a PTFE-coated Clearance Wire with a Loop set on its distal end, bent at a 105-degree angle. The Clearance Apparatus is advanced into the PleuraFlow Chest Tube using a magnetic Shuttle. When indicated, the Clearance Wire and Loop is advanced and retracted within the PleuraFlow Chest Tube to proactively prevent or break up and clear any tube obstructions or clogging to keep the tube patent.
{5}------------------------------------------------
- The PleuraFlow System is indicated for use during Indications for Use: cardiothoracic surgical procedures and chest trauma. Its active clearance technology proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.
- The safety and effectiveness of the PleuraFlow System has Performance Data: been previously demonstrated through design validation and verification that were cleared under 510(k) (K093565). Performance has been further demonstrated through post-market data.
Use of the device over the last four (4) years has shown that the product has significantly reduced the complications for patients who have received treatment with PleuraFlow versus other products during and after surgery.
Studies have linked both chest tube clogging and retained pericardial blood with POAF.2, 10 There is a substantial body of literature illustrating that shunting blood through pericardial windows to divert the blood to the pleural spaces can reduce POAF.1% demonstrates by adhering to a protocol developed to maximize chest tube patency, POAF can be reduced. Consistent with prior studies, this link appears to be related to reducing RBS. A manuscript titled "Reduction in Interventions for Post Operative Effusions and Atrial Fibrillation with Active Clearance of Chest Drainage Catheters" was submitted to the Journal of Thoracic and Cardiovascular Surgery. Authors are Sirch J.1, Ledwon M.1, Püski T.1, Grossmann I.1, Boyle EM.2, Pfeiffer S.1,Fischlein T.1 From the following institutions:
{6}------------------------------------------------
| 1. Cardiac Surgery, Heart Center, Paracelsus Medical | |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| University, Nurenberg, Germany, | |
| 2. St. Charles Medical Center, Bend, Oregon, United States. | |
| Conclusion: | The evaluation of the PleuraFlow System does not raise any<br>additional concerns regarding safety and effectiveness and may<br>therefore be considered substantially equivalent to the<br>predicate device. |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.