K121515 · Solta Medical, Inc. · ONF · Jun 5, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121515
Device Name
ISOLAZ 2 INTENSE PULSED LIGHT SYSTEM
Applicant
Solta Medical, Inc.
Product Code
ONF · General, Plastic Surgery
Decision Date
Jun 5, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Isolaz® 2 System is intended for treatment of mild to moderate acne, including pustular acne, comedonal acne and mild to moderate inflammatory acne (acne vulgaris) in all skin types (Fitzpatrick I-VI).
Device Story
Isolaz 2 Intense Pulsed Light System delivers non-coherent 400-1200 nm and 550-1200 nm light via photopneumatic handpiece; system comprises main console and treatment handpiece. Operator sets power and delivery time. Modifications from predicate include simplified device settings and hardware/circuitry updates. Used in clinical settings for acne treatment; light energy targets acne lesions. Benefits include reduction of pustular, comedonal, and inflammatory acne. Healthcare provider operates device to deliver light therapy to patient skin.
Clinical Evidence
Bench testing only. Verification and validation data demonstrate device meets all product specifications and functions as intended. Compliance with electrical safety and biocompatibility standards confirmed.
Technological Characteristics
Intense pulsed light system; 400-1200 nm and 550-1200 nm wavelengths; photopneumatic delivery handpiece. Compliant with IEC 60601-1, IEC 60601-1-4, IEC 60601-1-2, and ISO 10993-1. Hardware/circuitry modifications for simplified settings.
Indications for Use
Indicated for treatment of mild to moderate acne, including pustular, comedonal, and inflammatory acne (acne vulgaris) in patients of all skin types (Fitzpatrick I-VI).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K121515
## SECTION 5: 510(k) SUMMARY *****
JUN - 5 2012
This 510(k) summary is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990, 21 CFR 807.92
#### 1. General Information
Date of Submission: May 17, 2012
Submitted By:
Solta Medical, Inc. 25881 Industrial Blvd Hayward, CA 94545
Contact Person:
Kristine Foss V.P., Regulatory, Clinical & Quality 510-780-4657 (Direct Phone) 510-780-4857 Fax kfoss@solta.com
#### 2. Trade/Proprietary Name of Device:
Isolaz® 2 Intense Pulsed Light System Trade Name: Powered Light Non-laser Surgical Instrument Common Name: C Regulation Number 878.4810 Product Code: ONF General Surgery/Restorative Devices Device Panel: Device Classification: Class II
- 3. Legally Marketed Predicate Device for Claimed Equivalence: Name: Isolaz® Intense Pulsed Light System 510(k) #: K083730
#### 4. Device Description
The Isolaz® 2 System delivers non-coherent 400-1200 nm and 550-1200 nm intense pulsed light via a delivery handpiece utilizing photopneumatic technology. The system consists of a main console and a treatment handpiece. The desired power and delivery time are set by the operator. The modifications made to the device were to reduce and simplify device settings.
### 5. Intended Use:
The Isolaz® 2 System is intended for treatment of mild to moderate acne, including pustular acne, comedonal acne and mild to moderate inflammatory acne (acne vulgaris) in all skin types (Fitzpatrick I-VI).
{1}------------------------------------------------
#### 6. Technological Characteristics:
The Isolaz® 2 System is similar to the predicate device in design specification, output energy, and delivery system. They are both intense pulsed light instruments designed to produce light energy for treatment of mild to moderate acne including pustular acne, comedonal acne and inflammatory acne (acne vulgaris) in all skin types (Fitzpatrick I-VI). The modifications to the system console and handpiece hardware and circuitry do not significantly affect the safety or effectiveness of the device. The system was evaluated and found compliant with IEC 60601-1 and 60601-1-4 for electrical safety, IEC 60601-1-2 for EMI/EMC, and ISO 10993-1 for biocompatibility of the treatment tips. Verification and validation data show that the device meets all product specifications.
#### 7. Performance Data:
Laboratory and performance tests were executed to ensure that the device functioned as intended and met design specifications. Sufficient data were obtained to show that the device is substantially equivalent to the predicate device and meets safety and effectiveness criteria.
#### 8. Conclusion:
By virtue of the design, materials, function and intended use, the Isolaz® 2 Intense Pulsed Light System is as safe, as effective and performs as well as or better than the predicate device. In establishing substantial equivalence to the predicate device. Solta Medical evaluated the indications for use, product specifications, and energy requirements of the device.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Solta Medical % Ms. Kristine Foss Vice President, Regulatory , Clinical & Quality 25881 Industrial Boulevard Hayward, California, 94545
JUN - 5 2012
Re: K121515
Trade/Device Name: Isolaz 2 Intense Pulsed Light System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II Product Code: ONF Dated: May 17, 2012 Received: May 22, 2012
Dear Ms. Foss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading:
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it . may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
for. Rpt. De curson
Mark N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### SECTION 4:
# Indications for Use
510(k) Number (if known): K121515
Device Name: Isolaz® 2 Intense Pulsed Light System
Indications For Use:
Treatment of mild to moderate acne, including pustular acne, comedonal acne and mild to moderate inflammatory acne (acne vulgaris) in all skin types (Fitzpatrick I-VI).
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K121515
510(k) Number
ຂອງເບືອນໄປສາມາດສອງ bus ,ວibəququQ ,ഭാഭൂഢമ 10 поiгiviQ
of 8.8.8.8.j.ril
Isolaz II Special 510k 25 ર્વ ભ સ્ટ
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.