K082676 · Genadyne Biotechnologies, Inc. · OMP · Dec 18, 2008 · General, Plastic Surgery
Device Facts
Record ID
K082676
Device Name
GENADYNE A4 WOUND VACUUM SYSTEM
Applicant
Genadyne Biotechnologies, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Dec 18, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Genadyne A4 Wound Vacuum System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material and tissue debris.
Device Story
Portable suction pump system; provides negative pressure wound therapy (NPWT). Device consists of suction pump, collection canister, and wound sealing kit (gauze, film, drain). Operated by clinician; used to remove exudates, infectious material, and tissue debris from wounds. System generates vacuum; collects fluids into disposable 800ml canister with hydrophobic shut-off filter. Powered by 24 VDC (Ni-MH battery). Intended to promote wound healing. Differences from predicate include lighter weight and longer battery life. Used in clinical settings.
Clinical Evidence
Bench testing only. Performance testing conducted on dressing kits and system specifications to demonstrate safety and effectiveness compared to predicate.
Technological Characteristics
Powered suction pump; 24 VDC, 1A power; Ni-MH battery. Dimensions: 200 x 180 x 80 mm; weight: 1.36 kg. Suction capacity: 5 L/min; max vacuum: 230 mmHg. Includes 800 ml disposable canister with hydrophobic shut-off filter. Non-sterile. Complies with UL 60601-1, CAN/CSA C22.2 No. 601-1-M90, IEC 60601-1-2, FCC Part 15 Class B, EN 55011, IEC 61000-4-2, and IEC 61000-4-3.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to promote healing via removal of excess exudates, infectious material, and tissue debris. Contraindicated for necrotic tissue, untreated osteomyelitis, malignancy (except to enhance quality of life), untreated malnutrition, and exposed arteries, veins, or organs.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
K082676 page 1/8
### 510k Summary
### General Information
- 1. Applicant:
Genadyne Biotechnologies Inc. 65 Watermill Lane, Great Neck, NY 11021 (t) 516.487.8787 (f) 516.487.7878 www.genadyne.com
DEC 1 8 2008
- Mr. Chien-Ming GOH (Andrew) 2. Contact Person: Vice President Genadyne Biotechnologies Inc. 65 Watermill Lane, Great Neck, NY 11021 Tel: 516-487-8787 Fax: 516-487-7878 Andrew@gcnadyne.com
# 3. Trade/Proprietary Name Including Model Number of Device:
Genadyne A4 Wound Vacuum System
# 4. Common Name or Classification Name (21 CFR Part 807.87) of Device:
Powered Suction Pump (21 CFR 878.4780, Product Code BTA)
### 5. Class in which Device has been placed:
Class II
### 6. Reason for Premarket Notification:
Introduction of a new device that is substantially equivalent to a legally marketed device.
# 7. Identification of Legally Marketed Device Which We Claim Substantial Equivalence (Predicate Device):
BlueSky VISTA"M Wound Vacuum System, K061367 (Aug 10, 2006)
{1}------------------------------------------------
K082676 Page 2/8
# 8. Compliance with Requirements of the Federal FD&C Act:
The General and Rostorative Device Panel (DGRD) has classified this device as Class II, 21 CFR 878.4780
Product Code: BTA
### 9. Kit Certification and Information:
This device kit has been tested and is substantially equivalent to the kit of the predicate device.
Please refer to Attachment E for the performance testing on the dressing kit for both the Genadyne A4 Wound Vacuum System and the predicate device. Please do refer to Attachment F for the kit information and material data safety sheets for each component of the dressing kit.
### 10. Description of the Device
The product is a portable suction device that may promote wound healing when used with accessory wound sealing kits.
### 11. Intended use of the Device
The Genadyne A4 Wound Vacuum System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material and tissue debris.
#### 12. Substantial Equivalence
In establishing substantial equivalence to the predicate device, Genadyne Biotechnologies evaluated the indications for use, material, technology, product snecifications, and energy requirements of the system. Performance testing has been completed to demonstrate the safe and cffective use of the Genadyne A4 Wound Vacuum System for the intended use.
#### 13. Summary of Safety and Effectiveness
Performance testing and device comparison demonstrates that the subject device is substantially equivalent to the predicate device, and is safe and effective for the intended use.
{2}------------------------------------------------
K082676 Page 3/5
# 14. Comparison to Predicate Device
# Table of Comparison to Legally Marketed Device:
| Comparative Information | | |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | New Device | Predicate Device |
| Company | Genadyne Biotechnologies | BlueSky Medical Group, Inc. |
| | | BlueSky VISTATM Wound Vacuum<br>System |
| Device Name | A4 Wound Vacuum System | System |
| 510 (K) Number | K082676 | K061367 |
| | | |
| Technical Data | | |
| Suction Capacity | 5 liters per minute | 8 liters per minute |
| Max Vacuum | 230 mmHg | 200 mmHg |
| Power Requirements | 24 VDC, 1A | 100-240 V AC, 60 Hz |
| Battery Type | Ni-MH | Ni-MH |
| Dimensions and<br>Weight | 200 x 180 x 80 mm / 1.36 Kg | 260 x 250 x 106 mm / 1.9 Kg |
| | | |
| Accessories | | |
| Canisters | 800 ml disposable canister with<br>a build-in hydrophobic shut off<br>filter for overflow protection | 250 ml or 800 ml disposable<br>canister |
| Reusable | Yes | Yes |
| | | |
| Sterile | Non Sterile | Non Sterile |
| | | |
| Accessory Kit | | |
| | A4 Wound Sealing Kit which is<br>based on the teachings of Dr.<br>Mark Chariker and Dr.<br>Katherine Jeter1, which includes: | BlueSky Medical Wound Sealing<br>Kits, which includes: |
| | - Non-adherent gauze | - Non-adherent gauze |
| | - Anti-Microbial gauze | - Anti-Microbial gauze |
| | - Transparent film dressing | - Transparent film dressing |
| | - Silicone flat drain | - Silicone flat drain |
| Indications for Use | | |
| | The Genadyne A4 Wound<br>Vacuum System is indicated<br>for use in patients who<br>would benefit from negative<br>pressure wound therapy<br>particularly as the device<br>may promote wound healing<br>by the removal of excess<br>exudates, infectious material<br>and tissue debris. | The BlueSky VISTATM Wound<br>Vacuum System is indicated for<br>patients who would benefit from a<br>suction device particularly as the<br>device may promote wound healing. |
| Contraindications | | |
| | The Genadyne A4 is<br>contraindicated in the presence<br>of : | The VISTATM is contraindicated in<br>the presence of: |
| | Necrotic tissue | Necrotic tissue |
| - | Untreated osteomyelitis | Untreated osteomyelitis |
| - | Malignancy (with exception to<br>enhance quality of life) | Malignancy (with exception to<br>enhance quality of life) |
| | Untreated malnutrition | Untreated malnutrition |
| | Exposed arteries, veins, or<br>organs | Exposed arteries, veins, or organs |
| Precautions | | |
| | Precautions should be taken for | Precautions should be taken for |
| | patients who are or may be: | patients who are or may be: |
| | Receiving anticoagulant<br>therapy | Receiving anticoagulant therapy |
| | Suffering from difficult<br>hemostasis | Suffering from difficult hemostasis |
| | Untreated for malnutrition | Untreated for malnutrition |
| | Non-complaint or combative | Non-complaint or combative |
| Compliance | | |
| | UL 60601-1 | UL 60601-1 |
| | CAN/CSA C22.2 No. 601-1-<br>M90 | IEC 0601-1-2 |
| | | CAN/CSA C22.2 No. 601.1 |
| Storage / Transport | | |
| | -18°C to +43°C (0°F to 110°F) | -30 to +50° C (-22 to 122° F) |
| | Relative Humidity 10% to 95 % | 5 to 90% Humidity, non condensing |
| | | |
| | pressure | pressure |
| Operation | 18°C to 34°C (65°F to 94°F) | +5 to +35° C (41 to 95° F) |
| | Relative Humidity 10% to 95 % | 20 to 80 % Humidity, non condensing |
| | 700 - 1060 mbar Atmospheric<br>pressure | 700 - 1060 mbar Atmospheric<br>pressure |
| Testing | | |
| | IEC 60601-1-2 | - |
| | FCC part 15 Class B | - |
| | EN 55011 | - |
| | IEC 61000-4-2 | - |
| | IEC 61000-4-3 | - |
<sup>1</sup> Chariker ME, Jeter KF, Tintle TE. Effective management of incisional and cutaneous fistulae with closed suction wound drainage. Contemporary Surg. 1989;(34):59-63.
{3}------------------------------------------------
K082676 page 4/8
:
{4}------------------------------------------------
K082676 Page 5/6
# 15. Comparative Performance Evaluation:
The FDA Decision Tree for substantial equivalence was followed and the steps involved have been considered. The rationale for each step is discussed below. For reviewer convenience, the numbering system used by FDA on the decision tree has been followed by Gcnadyne in their process for the substantial equivalence determination rationale.
{5}------------------------------------------------
K082676 page
6/45
510(k) “SUBSTANTIAL EQUIVALENCE”
#### DECISION-MAKING PROCESS New Device is Compared to Marketed Device * 2 l Descriptive Information Does New Device Have Same NO Do the Differences Alter the Intended Not Substantially about New or Marketed Indication Statement? Therapeutic/Diagnostic/etc. Effect YES Equivalent Determination Device Requested as Needed (in Deciding, May Consider Impact on AES Safety and Effectiveness)?** New Device Has Same Intended NO Use and May be "Substantially Equivalent" O New Device Has New Intended {Jse ﺮ ﻟﻠ 4 Does New Device Have Same ර Technological Characteristics, NO Could the New Do the New Characteristics e.g. Design, Materials, etc. ? Characteristics YES Affect Safety or · Raise New Types of Safety YES Effectiveness? or Effectiveness Questions? 5 NO Are the Descriptive NO Characteristics Precise Enough NO to Ensure Equivalence? -8 7 NO Are Performance Data Do Accepted Scientific Available to Assess Equivalence? YES Methods Exist for NO Assessing Effects of the New Characteristics? YES YES ହୁ Performance Are Performance Data Available NO Data Required To Assess Effects of New Characteristics? *** YES 9 0 Performance Data Demonstrate Performance Data Demonstrate Equivalence? Equivalence? YES YES NO NO "Substantially Equivalent" へ To べ Determination To
* 510(k) Submissions compare new devices to marketed devices. FDA requests additional information if the relationship between marketed and "predicate" (pre-Amendments or reclassified post-Amendments) devices is unclear.
本本 This decision is normally based on descriptive information alone, but limited testing information is sometimes required.
*** Data maybe in the 510(k), other 510(k)s, the Center's classification files, or the literature
{6}------------------------------------------------
K0552676 Page 7/8
### 16. Discussion of Substantial Equivalence:
Note 1: Does new device have same indication statement as the predicate device(s)?
Yes. The Genadyne A4 Wound Vacuum System and the BlueSky VISTATM Wound Vacuum System are both indicated for patients who would benefit from a suction device particularly as the device may promote wound healing.
Note 3: Does the device have the same technological characteristics, e.g., design, materials, etc. ?
Yes. The Genadyne A4 Wound Vacuum System is design with a suction pump, can be externally and internally powered and has a collection canister, same as the BlueSky VISTA™ Wound Vacuum System, which also features a suction pump, a collection canister and can be powered externally and internally.
Note 5: Are the descriptive characteristics precise enough to ensure equivalence?
Yes. The characteristics of the Genadyne A4 Wound Vacuum System and the BlueSky VISTATM Wound Vacuum System are precise enough to ensure equivalence, based on the table in item 14 in the 510K Summary.
### 17. Discussion of Similarities and Differences
### Device Similarities
#### Indication for use
The indication for use is identical for the Genadyne A4 Wound Vacuum System and the predicate device.
### Configuration
All devices are sold non-sterile and are intended to be reusable. Devices are compatible with off-the-shelf accessories, such as disposable 800ml canisters.
#### Basic Product Function
The Genadyne A4 Wound Vacuum System and the predicate device have the same product function of generating a vacuum to provide general use suction and collection of liquids into an off-the-shelf canister reservoir which may promote wound healing for patient who would benefit from it.
{7}------------------------------------------------
K082676 Pages/s
# Device Differences
In comparison to the predicate devices, the Genadyne A4 Powered Wound Vacuum System has several differences which do not affect the device safety and effectiveness of the Genadyne A4 Wound Vacuum System. These differences between the Genadyne A4 Wound Vacuum System and the predicate device are described in further detail below.
# BlueSky VISTA134 Wound Vacuum System
The differences between the Genadyne A4 Wound Vacuum System and the BlueSky VISTA™ Wound Vacuum System are that the Genadyne A4 Wound Vacuum System is lighter in weight and has a longer battery life. In all other aspects, the Genadyne A4 Wound Vacuum System and the predicate device is substantially equivalent.
### 18. Conclusions:
Genadyne believes the Genadyne A4 Wound Vacuum System is substantially equivalent to the predicate device.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The eagle is positioned to the right of a circular emblem containing the words "DEPARTMENT OF HEALTH + HUMAN SERVICES USA" arranged around the perimeter.
Public Health Service
APR - 7 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Genadyne Biotechnologies, Inc. % Mr. Chien-Ming Goh 65 Watermill Lane Great Neck, New York 11021
Re: K082676
Trade/Device Name: Genadyne A4 Wound Vacuum System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: II Product Code: OMP Dated: November 12, 2008 Received: November 14, 2008
Dear Mr. Goh:
This letter corrects our substantially equivalent letter of December 18, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not
{9}------------------------------------------------
limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
K 062676
### Indications for Use
510(k) Number (if known):
Device Name: Genadyne A4 Wound Vacuum System
Indications For Use:
The Genadyne A4 Wound Vacuum System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material and tissue debris.
Prescription Use X (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** k082676
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.