IPL Hair Removal Device (KCA511/KCA516/KCA522)

K250253 · Dongguan Boyuan Intelligent Technology Co.,Ltd · OHT · May 6, 2025 · General, Plastic Surgery

Device Facts

Record IDK250253
Device NameIPL Hair Removal Device (KCA511/KCA516/KCA522)
ApplicantDongguan Boyuan Intelligent Technology Co.,Ltd
Product CodeOHT · General, Plastic Surgery
Decision DateMay 6, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Device Story

Over-the-counter, home-use IPL device for hair reduction; utilizes Intense Pulsed Light (IPL) to target hair follicles below skin surface. Device powered by external adapter; activated via finger switch. Features include 3.3cm² treatment window, integrated cooling system, and skin contact sensor preventing pulse emission unless in full contact with skin. User operates device manually on body areas (underarms, bikini line, arms, legs, facial hair below chin). Output consists of light pulses; cooling function provides comfort. Device design aims to reduce hair growth without cutting or pulling. Safety features include contact-based activation and photobiological safety compliance.

Clinical Evidence

Bench testing only. Evidence includes biocompatibility (ISO 10993-5, -10, -23), electrical safety/EMC (IEC 60601-1, -1-2, -1-11, -2-83), photobiological eye safety (IEC 62471), and software verification/validation. No clinical trial data presented.

Technological Characteristics

IPL-based hair removal; 3.3cm² treatment window; 610-1200nm wavelength; 2.18-7.27J/cm² energy density. Powered by external 24V DC adapter. Features skin contact sensor and active cooling. Biocompatibility per ISO 10993. Electrical safety per IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-83. Software control for pulse delivery.

Indications for Use

Indicated for removal of unwanted hair and permanent reduction in hair regrowth in adults with Fitzpatrick Skin Types I-V. Suitable for underarm, bikini line, arms, facial hair below the chin line, back, and legs.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 6, 2025 Dongguan Boyuan Intelligent Technology Co., Ltd % Reanny Wang General Manager Shenzhen Reanny Medical Devices Management Consulting Co.Ltd Room 1509, Jingting Building, Dongzhou Community, Guangming Street, Guangming Shenzhen, Guangdong 818000 China Re: K250253 Trade/Device Name: IPL Hair Removal Device (KCA511/KCA516/KCA522) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: January 26, 2025 Received: January 28, 2025 Dear Reanny Wang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} 2K250253 - Reanny Wang Page Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} 3K250253 - Reanny Wang Page assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TANISHA L. HITHE Digitally signed by TANISHA L. HITHE Date: 2025.05.06 17:23:28 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250253 | | | Device Name IPL Hair Removal Device (KCA511/KCA516/KCA522) | | | Indications for Use (Describe) IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K250253 - 510(k) Summary I. SUBMITTER Dongguan Boyuan Intelligent Technology Co.,Ltd. Room 801, Building 3, No.3 Kuiqiao Road, Puxin Lake , Tangxia Town ,Dongguan , Guangdong, China. Phone: 86-13422889315 Fax: 978953150@qq.com Contact Person: Li Jinlong Date Prepared: Sep 29, 2024 II. DEVICE Name of Device: IPL Hair Removal Device Model(s): KCA511,KCA516,KCA522. Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: II Product Code: OHT Regulation Number:21 CFR 878.4810 III. PREDICATE DEVICE Predicate device: | Manufacturer | Predicate Device | 510(k) Number | Approval Date | | --- | --- | --- | --- | | Dongguan Boyuan Intelligent Technology Co.,Ltd. | IPL Hair Removal Device | K240282 | April 3, 2024 | This predicate has not been subject to a design-related recall. Reference device: | Manufacturer | Predicate Device | 510(k) Number | Approval Date | | --- | --- | --- | --- | | Shenzhen Ulike Smart Electronics Co.,Ltd. | IPL Hair Removal Device | K230122 | April 10, 2023 | {5} This predicate has not been subject to a design-related recall. ## IV. DEVICE DESCRIPTION IPL Hair Removal Device (Model: KCA511, KCA516, KCA522), is an over-the-counter, home-use and personal device for hair reduction by using Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts irreplaceable treatment window and the spot size is $3.3\mathrm{cm}^2$ that is suitable for multiple hair removal areas. It contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the device has the cooling function that will be activated throughout the whole hair removal process to provide users with a more comfortable experience. IPL Hair Removal Device, model: KCA511, KCA516 and KCA522 have the same indication for use, performance, structure design and operation, the only deference is their appearance size and weight. IPL Hair Removal is used for removing excess facial hair and body hair and can also reduce hair regrowth. The products can be used on body parts such as underarm, bikini line, arms, facial hair below the chin line, back and legs. It is designed so that the flash window requires full contact with the skin to work (flash), it will no works as long as the flash window does not contact the skin. ## V. Indications for Use IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. ## VI. Comparison of Technological Characteristics with the Predicate Device The IPL Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate device and reference device. Any minor differences between the subject device and the listed predicate device and reference {6} device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device and reference device for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate device and reference device. IPL Hair Removal Device is compared with the following Predicate Device and Reference Device in terms of intended use, design, specifications and performance: | Comparison Elements | Subject device | Predicate device | Reference device | Remark | | --- | --- | --- | --- | --- | | 510(k) Number | Pending | K240282 | K230122 | / | | Trade name | IPL Hair Removal Device | IPL Hair Removal Device | IPL Hair Removal Device | / | | Manufacturer | Dongguan Boyuan Intelligent Technology Co.,Ltd. | Dongguan Boyuan Intelligent Technology Co.,Ltd. | Shenzhen Ulike Smart Electronics Co.,Ltd. | / | | Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same | | Product code | OHT | OHT | OHT | Same | | Device classification | Class II | Class II | Class II | Same | | Indication for use/ Intended use | IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when | IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when | IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when | Same | {7} | | measured at 6, 9 and 12 months after the completion of a treatment regime. | measured at 6, 9 and 12 months after the completion of a treatment regime. | when measured at 6, 9 and 12 months after the completion of a treatment regime. | | | --- | --- | --- | --- | --- | | Prescription or OTC | OTC | OTC | OTC | Same | | Applicable skin | Fitzpatrick Skin Types I-V | Fitzpatrick Skin Types I-V | Fitzpatrick Skin Types I-V | Same | | Treatment area | Body parts such as underarm, bikini line, arms, facial hair below the chin line, back and legs. | Body parts such as underarm, bikini line, arms, facial hair below the chin line, back and legs. | large areas (e.g. arms, legs, chest) and small areas (e.g. lip) | Same | | Device Design | | | | | | Source energy | Supplied by Power Adapter | Supplied by Power Adapter | Supplied by external adapter | | | Power source | External power supply 1) Model: FX65U-2403 00C Input: 100-240V AC 50/60Hz 2A Output: 24.0V DC 3.0A 2) Model: DZ072CHL2 40300U Input: 100-240V AC 50/60Hz 1.5A Output: 24.0V DC 3.0A | External power supply: KCA450, KCA446, KCA504, KCA505, KCA506, KCA507, KCA508, KCA509: 1) Model: FX48U-120400C Input: 100-240V AC; 50/60Hz; 1.0A max Output: 12,0 V d.c.; 4,0 A 2) Model: DZ048CHL 120400U Input: 100-240V AC; 50/60Hz; 1.5A max | External power supply: 100-240 VAC, 50/60Hz | Different | {8} | | | Output: 12,0 V d.c.; 4,0 A | | | | --- | --- | --- | --- | --- | | | | An external power supply: KCA501, KCA423, KCA448, KCA449, KCA502, KCA521: 1) Model: FX48U-120300C Input: 100-240V AC; 50/60Hz; 1.0A max Output: 12,0 V d.c.; 3,0 A 2) Model: DZ048CHL 120300U Input: 100-240V AC; 50/60Hz; 1.5A max Output: 12,0 V d.c.; 3,0 A | | | | Sterilization | Not required | Not required | Not required | Same | | Dimensions | KCA511 : 178.2 *61.9 *36mm KCA516 : 177.3*65.4*37.4mm KCA522 : 177*70.2*43.1mm | KCA450: 170.8 *80.4 *41.3mm KCA446: 170.0 *80.5 *41.3mm KCA504: 172.1 *69.0 *37.8mm KCA505: 172.1 *69.0 *37.8mm KCA506: 171.0 *74.8 *41.9mm KCA507: 171.5 *74.5 *37.3mm | 60mm x 38mm x 170mm | Different | {9} | | | KCA508: 171.0 *74.8 *41.9mm KCA509: 171.5 *74.5 *37.3mm | | | | --- | --- | --- | --- | --- | | | | KCA501: 163.5*79.1*42.8mm KCA423: 164.8*73.7*41.1mm KCA448: 162.0*76.4*42.2mm KCA449: 162.2*78.0*42.7mm KCA502: 164.2*75.8*38.3mm KCA521: 162.5*80.8*39.1mm | | | | Output Specification | | | | | | Light source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same | | Wavelength range | (610nm-1200nm) ±15nm | (600nm~1200nm) ±15nm | 560-1200nm | Similar | | Energy density | 2.18~7.27J/cm² | 1.6~6J/cm² | 2.4~7.2J/cm2 | Similar | | Output energy | (9J~20J) ± 20% | (7.11J~17.41J) ± 20% | 9.9~19.8J | Similar | | Spot size | 3.3cm² | 3.5 cm² | 3.3cm² | Same | | Pulse duration | 0.45ms±0.2ms (0.67-4.5)ms ±2ms | 8~12ms | 1.15-6.2ms | Different | | Output intensity level | 5 | 9 | 5 | Same | | Software/Firmware/Microprocess | Software control | Software control | Software control | Same | {10} | or Control? | | | | | | --- | --- | --- | --- | --- | | Additional Features | | | | | | Electrical safety | ANSI AAMI ES60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 | ANSI AAMI ES60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 | ANSI AAMI ES60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 60601-2-83 | Same | | Eye safety | IEC 62471 | IEC 62471 | IEC 62471 | Same | | Biocompatibility | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-5 ISO 10993-10 | Same | ## VII. Performance Data The following performance data were provided in support of the substantial equivalence determination: 1) Biocompatibility testing The biocompatibility evaluation for the IPL Hair Removal Device was conducted in accordance with the "Use of International Standard ISO-10993-1, 'Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process'" as recognized by FDA. The battery of testing included the following tests: - ISO 10993-5:2009, Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity - ISO 10993-10:2010, Biological Evaluation of Medical Devices –Part 10: Tests for Irritation and Skin Sensitization - ISO 10993-23: 2021, Biological evaluation of medical devices --Part 23: Tests for irritation The IPL Hair Removal Device is considered tissue contacting for a duration of less than 24 hours. 2) Electrical Safety and EMC Electrical safety and EMC testing was performed to, and passed, as per the following standards: {11} IEC 60601-1:2005/AMD1:2012/AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. IEC 60601-1-11:2015/AMD1:2020 Medical Electrical Equipment – Part 1-11: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment. IEC 60601-2-83:2019/AMD1:2022 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment 3) Eye Safety IEC 62471:2006 Photobiological safety of lamps and lamp systems 4) Software Verification and Validation Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. ## Summary Based on the above performance as documented in this application, IPL Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device and reference device. ## VIII. Conclusions Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...